← Product Code [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH) · K183213

# IceSense 3, ProSense, MultiSense (K183213)

_IceCure Medical , Ltd. · GEH · Dec 20, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K183213

## Device Facts

- **Applicant:** IceCure Medical , Ltd.
- **Product Code:** [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH.md)
- **Decision Date:** Dec 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The IceCure Family (IceSense™3, ProSense™, and MultiSense™) cryoablation system is intended for cryogenic destruction of tissue during surgical procedures by the application of extreme cold temperatures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The system has the following specific indications: Urology (ablate prostate tissue in cases of prostate cancer and benign prostatic hyperplasia (BPH)). Oncology (ablation of cancerous or malignant tissue and benign tumors and palliative intervention). Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Palliation of turnors of the skin. Destruction of warts or lesions). Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia). ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth). General Surgery (ablation of leukoplakia of mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts actinic and sebortheic keratoses, cavernous hemangiomas, recurrent cancerous lesions. Palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions. Destruction of warts or lesions. Palliation of the oral cavity, rectum, and skin. Ablation of breast fibroadenomas). Thoracic Surgery (ablation of arrhythmic cardiac tissue and cancerous lesions). Proctology (ablation of benign or malignant growths of the anus and rectum and hemorrhoids). The system may be used with imaging device like ultrasound to provide real-time visualization of the cryosurgical procedure. MultiSense™ System is indicated for three probe configuration only.

## Device Story

Cryoablation system; delivers extreme cold via pressurized liquid nitrogen to disposable cryoprobes; destroys targeted tissue. Components: main chassis, controller, touch screen, foot pedal, cryohandle. Used in OR or clinic by physicians. MultiSense model supports up to three probes in parallel; IceSense 3/ProSense support single probe. System monitors pressure, flow, temperature, and microswitches; provides real-time status/error displays. Optional ultrasound integration for visualization. Benefits: precise tissue ablation for oncology, urology, and general surgery applications. Safety features: alarms, emergency stop, automatic procedure abortion on malfunction.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via electrical/EMC testing and bench studies comparing iceball size (40mm single, 94mm triple) and thermal profiles against predicates.

## Technological Characteristics

Closed-loop liquid nitrogen system. Probes made of 304/316 stainless steel. Active thawing via electrical heating element (up to 35°C). Touch screen interface. Connectivity: standalone. Sterilization: Ethylene Oxide (ISO 11135). Standards: ES60601-1, IEC 60601-1-2. Software-controlled valves, pump, and heaters.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- IceSense 3 System ([K102360](/device/K102360.md))
- Galil Medical Seednet family ([K052530](/device/K052530.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

December 20, 2019

IceCure Medical LTD. % Zvi Ladin Principal Boston MedTech Advisors Inc. 990 Washington Street Suite #204 Dedham, Massachusetts 02026

Re: K183213

Trade/Device Name: IceCure Family Cryoablation System (IceSense 3, ProSense, MultiSense) Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: November 20, 2019 Received: November 21, 2019

Dear Zvi Ladin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183213

#### Device Name IceCure Family Cryoablation System

#### Indications for Use (Describe)

The IceCure Family (IceSense™3, ProSense™, and MultiSense™) cryoablation system is intended for cryogenic destruction of tissue during surgical procedures by the application of extreme cold temperatures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The system has the following specific indications:

Urology (ablate prostate tissue in cases of prostate cancer and benign prostatic hyperplasia (BPH)).

Oncology (ablation of cancerous or malignant tissue and benign tumors and palliative intervention).

Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Palliation of turnors of the skin. Destruction of warts or lesions).

Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia).

ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth).

General Surgery (ablation of leukoplakia of mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts actinic and sebortheic keratoses, cavernous hemangiomas, recurrent cancerous lesions. Palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions. Destruction of warts or lesions. Palliation of the oral cavity, rectum, and skin. Ablation of breast fibroadenomas).

Thoracic Surgery (ablation of arrhythmic cardiac tissue and cancerous lesions).

Proctology (ablation of benign or malignant growths of the anus and rectum and hemorrhoids).

The system may be used with imaging device like ultrasound to provide real-time visualization of the cryosurgical procedure.

MultiSense™ System is indicated for three probe configuration only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

| CONTINUE ON A SEPARATE PAGE IF NEEDED. |
|----------------------------------------|
|----------------------------------------|

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 1. 510(K) SUMMARY

# lceCure's Family Cryoablation System

# Name and Address of Applicant:

lceCure Medical LTD. Haeshel 7, Caesarea I.P. 38900, Israel Elisabeth Sadka, VP RA QA clinical Telephone: 972-4-6230333 Fax: 972-4-6230222

# Contact Person and Phone Number:

Zvi Ladin, Ph.D., Principal, Boston MedTech Advisors Telephone: 781-407-0900 Fax: 781-407-0901 E-mail: zladin@bmtadvisors.com Date Prepared: December 19, 2019

## Name of Device

Trade/Proprietary Name: IceCure Family (IceSense™3, ProSense™ and MultiSense™) Cryotherapy Systems Common Name: Cryosurgical unit and accessories Classification Name: Cryosurgical unit and accessories (21 C.F.R. § 878.4350) Product Code: GEH

# Manufacturing Facility

IceCure Medical LTD. Haeshel 7, Caesarea I.P. 3079504, Israel

{4}------------------------------------------------

### Predicate Devices

The IceCure Family of Cryotherapy Systems is substantially equivalent to the cleared IceSense™3 System (K102360) and the cleared Galil Medical Seednet family (K052530).

### Intended Use / Indications for Use

The IceCure Family (IceSense™3, ProSense™, and MultiSense™) cryoablation system is intended for cryogenic destruction of tissue during surgical procedures by the application of extreme cold temperatures.

It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, proctology, and urology.

The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.

The system has the following specific indications:

Urology (ablate prostate tissue in cases of prostate cancer and benign prostatic hyperplasia (BPH)).

Oncology (ablation of cancerous or malignant tissue and benign tumors and palliative intervention).

Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Palliation of tumors of the skin. Destruction of warts or lesions).

Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia).

ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth).

General Surgery (ablation of leukoplakia of mouth, angjomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts actinic and seborrheic keratoses, cavernous hemanqiomas, recurrent cancerous lesions. Palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions. Destruction of warts or lesions. Palliation of tumors of the oral cavity, rectum, and skin. Ablation of breast fibroadenomas).

Thoracic Surgery (ablation of arrhythmic cardiac tissue and cancerous lesions).

Proctology (ablation of benign or malignant growths of the anus and rectum and hemorrhoids).

The system may be used with imaging device like ultrasound to provide real-time visualization of the cryosurgical procedure.

MultiSense™ System is indicated for three probe configuration only.

{5}------------------------------------------------

## Technological Characteristics

The IceCure Family of cryotherapy devices includes the IceSense™3 single-probe system (cleared under 510(k) #K102360 and rebranded as ProSense™), and MultiSense™ – a new system that accommodates up to three cryoprobes operating in parallel. The systems are used to destroy unwanted tissue by application of extreme cold to the selected sites. The devices deliver cold temperatures to targeted tissue by pressurized liquid nitrogen closed system and a disposable cryoprobe. Various cryoprobes are available.

The devices consist of a main chassis for the cooling system, a controller, a touch screen, a foot pedal and a cryohandle that controls the system and holds the probe.

Safety measures of the system include alarms, safety valve, emergency button and automatic abortion of the procedure in case of technical malfunction.

# Comparison with the Predicate Device

Cryoablation is the fundamental technological principle for both the subject IceCure Family of cryotherapy devices and the predicate lceSense™3 System (K102360) and Galil Medical Seednet family (K052530). The subject and marketed predicate devices are used to ablate unwanted tissue by application of extreme cold.

The subject and predicate devices are based on the same technological principles:

- . Delivery of cryogen from a dewar to a cryoprobe tip
- Application of cryogen to ablate (freeze) the unwanted tissue .
- Heat transfer at the tip of the cryoprobe for quick release of the probe
- User-controlled (trigger/foot pedal) to release cryogen .
- Software activated Controller .

ProSense™ is a mere rebranding of the cleared IceSense™3 single-probe system (K102360). Therefore, the ProSense™ System has the same hardware and software components as the lceSense™3 System. Hardware and software changes introduced since the device was originally cleared on November 29, 2010, were analyzed following FDA's Guidance Documents – "Deciding When to Submit a 510(k) for a Change to an Existing Device" (K97-1) dated January 10, 1997 and "Deciding When to Submit 510(k) for a Software Change to an Existing Device," dated October 25, 2017. All changes were determined not to require new 510(k) premarket notifications and were

{6}------------------------------------------------

therefore documented and released based on the company's standard operating procedures. This submission includes the current configuration of the system.

The MultiSense™ System has the same principle of operation and technological features as the single-probe IceSense™3 (K102360) and the multi-probe Galil Medical Seednet (K052530) systems. It can accommodate up to three cryoprobes operating in parallel.

The submission was focused on two key elements:

- IceCure Medical's multi-probe system design is similar to the Galil Medical Seednet . (K052530) predicate.
- . Streamline the Indications for Use of the IceCure Family cryoablation system in accordance with those of the predicate devices.

|                        | IceCure Family                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | IceSense3                                                                                                                                                                                                                                                                                                                                                                                                            | Seednet family                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number            | K183213                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K102360                                                                                                                                                                                                                                                                                                                                                                                                              | K052530                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Company                | IceCure Medical, Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | IceCure Medical, Ltd.                                                                                                                                                                                                                                                                                                                                                                                                | Galil Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device                 | Cryotherapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Cryotherapy system                                                                                                                                                                                                                                                                                                                                                                                                   | Cryotherapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Intended use           | The system is intended for<br>cryogenic destruction of<br>tissue during surgical<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                           | The IceSense3 System is<br>intended for cryogenic<br>destruction of tissue<br>during surgical<br>procedures.                                                                                                                                                                                                                                                                                                         | The system is intended<br>for cryogenic destruction<br>of tissue during surgical<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for<br>Use | IceCure Family<br>(IceSenseTM3, ProSenseTM,<br>and MultiSenseTM)<br>cryoablation system is<br>intended for cryogenic<br>destruction of tissue during<br>surgical procedures by the<br>application of extreme cold<br>temperatures.<br>It is indicated for use as a<br>cryosurgical tool in the fields<br>of general surgery,<br>dermatology, neurology<br>(including cryoanalgesia),<br>thoracic surgery, ENT,<br>gynecology, oncology,<br>proctology, and urology.<br>The system is designed to<br>destroy tissue by the | The IceSense3 System<br>is intended for cryogenic<br>destruction of tissue<br>during surgical<br>procedures. It is<br>indicated for use as a<br>cryosurgical tool in a<br>number of specific fields<br>which include the<br>following:<br>Urology<br>The system may be<br>used to ablate prostatic<br>tissue.<br>The system may be<br>used for the ablation of<br>prostate tissue in cases<br>of prostate cancer and | The system is intended<br>for cryogenic destruction<br>of tissue during surgical<br>procedures.<br>The system is indicated<br>for use as a cryosurgical<br>tool in the fields of<br>general surgery,<br>dermatology, neurology<br>(including cryoanalgesia),<br>thoracic surgery, ENT,<br>gynecology, oncology,<br>proctology, and urology.<br>The system is designed<br>to destroy tissue by the<br>application of extreme<br>cold temperatures<br>including prostate and<br>kidney tissue, liver |

{7}------------------------------------------------

| application of extreme cold<br>temperatures including<br>prostate and kidney tissue,<br>liver metastases, tumors,<br>skin lesions, and warts. | benign prostatic<br>hyperplasia<br>Oncology<br>The system may be                                    |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| The system has the<br>following specific<br>indications:                                                                                      | used for ablation of<br>cancerous or malignant<br>tissue.<br>The system may be                      |
| Urology (ablate prostate<br>tissue in cases of prostate<br>cancer and benign prostatic<br>hyperplasia (BPH)).                                 | used for ablation of<br>benign tumors.<br>The system may be<br>used for palliative<br>intervention. |
| Oncology (ablation of<br>cancerous or malignant<br>tissue and benign tumors<br>and palliative intervention).                                  | Dermatology<br>The system may be<br>used for the ablation or<br>freezing of skin cancers            |
| Dermatology (ablation or<br>freezing of skin cancers and<br>other cutaneous disorders.                                                        | and other cutaneous<br>disorders.                                                                   |
| Palliation of tumors of the<br>skin. Destruction of warts or<br>lesions).                                                                     | Gynecology<br>The system may be<br>used for the ablation of<br>malignant neoplasia or               |
| Gynecology (ablation of<br>malignant neoplasia or<br>benign dysplasia of the                                                                  | benign dysplasia of the<br>female genitalia.<br>General Surgery                                     |
| female genitalia).<br>ENT (Palliation of tumors of<br>the oral cavity and ablation                                                            | The system may be<br>used for the ablation of<br>leukoplakia of mouth,                              |
| of leukoplakia of the mouth).<br>General Surgery (ablation of                                                                                 | angiomas, sebaceous<br>hyperplasia, basal cell                                                      |
| leukoplakia of mouth,<br>angiomas, sebaceous<br>hyperplasia, basal cell                                                                       | tumors of the eyelid or<br>canthus area, ulcerated<br>basal cell tumors,                            |
| tumors of the eyelid or<br>canthus area, ulcerated<br>basal cell tumors,                                                                      | dermatofibromas, small<br>hemangiomas,<br>mucocele cysts, multiple                                  |
| dermatofibromas, small<br>hemangiomas, mucocele<br>cysts, multiple warts, plantar                                                             | warts, plantar warts,<br>hemorrhoids, anal<br>fissures, perianal                                    |
| warts, hemorrhoids, anal<br>fissures, perianal<br>condylomata, pilonidal cysts                                                                | condylomata, pilonidal<br>cysts, actinic and<br>seborrheic keratoses,                               |
| actinic and seborrheic<br>keratoses, cavernous<br>hemangiomas, recurrent                                                                      | cavernous<br>hemangiomas, recurrent<br>cancerous lesions,<br>ablation of breast                     |
| cancerous lesions.<br>Palliation of tumors of the<br>rectum, hemorrhoids, anal                                                                | fibroadenoma.<br>The system may be<br>used for the destruction                                      |
| fissures, pilonidal cysts, and<br>recurrent cancerous lesions.<br>Destruction of warts or                                                     | of warts or lesions.<br>The system may be                                                           |
| lesions.                                                                                                                                      | used for the palliation of<br>warts or lesions.                                                     |

metastases, tumors, skin lesions, and warts.

The system has the following specific indications:

Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")

Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)

Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin)

Gynecology (ablation of maliqnant neoplasia or benign dysplasia of the female genitalia)

General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas)

ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth).

{8}------------------------------------------------

| of the oral cavity, rectum,<br>and skin. Ablation of breast<br>fibroadenomas).<br>Thoracic Surgery (ablation<br>of arrhythmic cardiac tissue<br>and cancerous lesions).<br>Proctology (ablation of<br>benign or malignant growths<br>of the anus and rectum and<br>hemorrhoids).<br>The system may be used<br>with imaging device like<br>ultrasound to provide real-<br>time visualization of the<br>cryosurgical procedure.<br>MultiSense™ System is<br>indicated for three probe<br>configuration only. | tumors of the oral cavity,<br>rectum, and skin.<br>Thoracic Surgery<br>The system may be<br>used for the ablation of<br>arrhythmic cardiac<br>tissue.<br>The system may be<br>used for the ablation of<br>cancerous lesions<br>Proctology<br>The system may be<br>used for the ablation of<br>benign or malignant<br>growths of the anus and<br>rectum.<br>The systems may be<br>used for the ablation of<br>hemorrhoids.<br>The system may be<br>used with an ultrasound<br>device to provide real-<br>time visualization of the<br>cryosurgical procedure | Thoracic surgery<br>(ablation of arrhythmic<br>cardiac tissue cancerous<br>lesions)<br>Proctology (ablation of<br>benign or malignant<br>growths of the anus or<br>rectum, and<br>hemorrhoids)<br>The system may be used<br>with an ultrasound device<br>to provide real-time<br>visualization of the<br>cryosurgical procedure |                                                                                                                                                                               |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cooling<br>system<br>location and<br>refrigerant                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Internal small [1 or 3]<br>dewar(s) of liquid nitrogen.<br>[1 or 3] dewar(s) used per<br>procedure                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Internal small dewar of<br>liquid nitrogen.<br>1 dewar used per<br>procedure                                                                                                                                                                                                                                                    | Internal cylinders of<br>argon.                                                                                                                                               |
| Thawing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Active thawing: nitrogen gas<br>is heated by an electrical<br>heating element.<br>Reaching up to 35 degrees<br>Celsius.                                                                                                                                                                                                                                                                                                                                                                                                                                     | Active thawing: nitrogen<br>gas is heated by an<br>electrical heating<br>element.<br>Reaching around 35<br>degrees Celsius.                                                                                                                                                                                                     | Active thawing: electric<br>element or helium gas<br>reaching 35 degrees<br>Celsius.                                                                                          |
| Operating<br>mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Freeze, thaw, active thaw.<br>Manual/pre-programmed<br>freezing cycles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Freeze, thaw, active<br>thaw.<br>Manual/pre-programmed<br>freezing cycles                                                                                                                                                                                                                                                       | Freeze, low freeze, active<br>thaw, off, stick                                                                                                                                |
| Parameters<br>controlled by<br>the user                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Automatic or manual<br>treatment mode; freeze or<br>thaw time and active thaw<br>mode.<br>Operated from the touch<br>screen control panel, and in<br>addition for the IceSense™3<br>model only: the cryohandle<br>buttons or the foot pedal                                                                                                                                                                                                                                                                                                                 | Automatic or manual<br>treatment mode; freeze<br>or thaw time and active<br>thaw mode.<br>Operated from the touch<br>screen control panel, the<br>cryohandle buttons or the<br>foot pedal.                                                                                                                                      | Probe temperature, freeze<br>thaw stick and active thaw<br>mode.<br>Operated from the probe<br>handle buttons, remote<br>control unit.                                        |
| Software<br>controls                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Valves, pump, heaters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Valves, pump, heaters                                                                                                                                                                                                                                                                                                           | Valves, pump, heater                                                                                                                                                          |
| Parameters<br>monitored by<br>the software                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | The controller constantly<br>monitors the status of:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | The controller constantly<br>monitors the status of:                                                                                                                                                                                                                                                                            | The controller constantly<br>monitors the status of:                                                                                                                          |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Microswitches, pressure and<br>flow gauges, temperature<br>sensors, buttons.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Microswitches, pressure<br>and flow gauges,<br>temperature sensors,<br>buttons.                                                                                                                                                                                                                                                 | Gauges, temperature<br>sensors, buttons.                                                                                                                                      |
| System<br>display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Online display: time, mode,<br>status, messages, errors.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | On line display: time,<br>mode, status, messages,<br>errors.                                                                                                                                                                                                                                                                    | On line display: time,<br>pressure, temperature,<br>status, errors, messages.                                                                                                 |
| Procedure<br>track record                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                           |
| Number of<br>probes<br>controlled by<br>the device                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 1 probe for ProSense™;<br>3 probes for MultiSense™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1 probe                                                                                                                                                                                                                                                                                                                         | 1-5 probes                                                                                                                                                                    |
| Probes<br>characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 2.4 mm, 3.4 mm pencil or<br>trocar tips, for small or<br>medium, elliptical or spheric<br>iceball.<br>Shaft lengths for 2.4 mm<br>probes include 124 mm and<br>134 mm; Shaft lengths for<br>3.4 mm probes include 127<br>mm, 140mm, and 185 mm.<br>Made from 304 and 316<br>stainless steel.                                                                                                                                                                                                                                                                | 2.0 mm, 2.7 mm, 3.0<br>mm, 3.4 mm sharp or<br>blunt tips, for small or<br>medium iceball.<br>Made from 304 and 316<br>stainless steel.                                                                                                                                                                                          | 1.5 mm, 2.0 mm, 3.2 mm,<br>3.4 mm, 5.0 mm, 6.0 mm<br>for small medium or large<br>iceball, sharp or blunt and<br>surface probes.<br>Made from 304 and 316<br>stainless steel. |
| Probe<br>connection to<br>the system                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Probe is screwed to the<br>handle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Probe is screwed to the<br>handle                                                                                                                                                                                                                                                                                               | Quick connection of the<br>pipe to the system or<br>screwing the probe to the<br>handle                                                                                       |
| Safety<br>measures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Alarms, pretest, microswitch.<br>Safety valve, pressure relief<br>valves, Emergency stop<br>button.<br>Automatic abortion of the<br>procedure in case of<br>technical malfunction.<br>One dewar content is for up<br>to 15 minutes of continuous<br>operation of freeze.                                                                                                                                                                                                                                                                                    | Alarms, pretest,<br>microswitch.<br>Safety valve, pressure<br>relief valves, Emergency<br>stop button.<br>Automatic abortion of the<br>procedure in case of<br>technical malfunction.<br>One dewar content is for<br>up to 16 minutes of<br>continuous operation of<br>freeze.                                                  | Emergency stop button<br>Safety valve, pressure<br>relief valves, Emergency<br>stop button, alarms.                                                                           |
| Screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Touch Screen mounted on the system. Tilt and swivel as needed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Touch Screen mounted on the system. Tilt and swivel as needed                                                                                                                                                                                                                                                                   | Screen mounted on the system. Tilt and swivel as needed                                                                                                                       |
| Dimensions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Height: 120cm/47.24"<br>Depth: 50cm /19.68" // 70 cm / 27.56"<br>Width: 50 cm /19.68"<br>Weight: 70 Kg / 154 lb (with 1 dewar) // 150 Kg / 330 lb (with 3 dewars)                                                                                                                                                                                                                                                                                                                                                                                           | Height: 120cm/47.24"<br>Depth: 50cm /19.68"<br>Width: 50 cm /19.68"<br>Weight: 70 Kg / 154 lb                                                                                                                                                                                                                                   | Height: 58"<br>Depth: 34"<br>Width: 23"<br>Weight (without gas cylinders): 265 lb.                                                                                            |
| Cryoprobe pretest                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                           |
| Use Environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Hospital - operating room<br>Office procedure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Hospital - operating room<br>Office procedure                                                                                                                                                                                                                                                                                   | Hospital - operating room<br>Office procedure                                                                                                                                 |

{9}------------------------------------------------

{10}------------------------------------------------

### Performance Data

Performance data were provided in support of the substantial equivalence determination. System was tested for conformance with relevant safety and performance standards, such as electrical testing, electromagnetic compatibility and software validation and verification, formulated in the device specifications. In all instances, the systems functioned as intended and the performance observed met acceptance criteria for all tests.

### Sterilization

The IceCure Family cryoablation system accessories (probes, introducers and temperature sensor) are all sterilized utilizing an Ethylene Oxide sterilization cycle validated in accordance with ISO 11135 - Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization.

### Biocompatibility

The biocompatibility evaluation of patient contacting materials of the IceCure Family cryoablation system accessories was conducted in accordance with FDA's "Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"". Biocompatibility testing was not required as there were no changes in patient contacting materials.

### EMC and Electrical Safety

Electromagnetic compatibility (EMC) and electrical safety (ES) of the Ice Cure Family cryoablation system were tested and determined to be compliant with the following standards:

{11}------------------------------------------------

- . ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD).
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

### Bench testing

Bench testing focused on demonstrating the performance of the MultiSense system using multiple probes. The performance of the MultiSense devices was found to be similar to that of the predicate devices IceSense™3 System (K102360) and Gall Medical Seednet family (K052530). Specifically, testing documented that in the range of up to 15 minutes freeze, the probes can create an iceball of at least 40 mm diameter for a single probe, and 94 mm for a three-probe system. Performance testing included testing of the shaft temperature during freeze, thaw and warm phases. Results demonstrated that temperatures along the shaft remained within the device's specifications during all phases. The time change of iceball size and probe tip temperature were similar for the candidate and predicate devices. Iceball isotherm measurements were conducted and the results met the device's specifications.

### Performance Testing - Clinical

Clinical testing was not required to demonstrate the safety and effectiveness of the device.

### Substantial Equivalence

The IceCure Family devices are as safe and effective as the cleared IceSense™3 System (K102360) and the cleared Galil Medical Seednet family (K052530).

The IceCure Family devices have the same intended uses and indications, technological characteristics, and principles of operation as the predicate devices. The minor technological differences between the IceCure devices and their predicate devices do not raise different issues of safety or effectiveness. Performance data demonstrate that the IceCure devices are as safe and effective as the cleared IceSense™ System (K102360) and the cleared Galil Medical Seednet family (K052530).

Thus, the IceCure Family of cryotherapy systems is substantially equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K183213](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K183213)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
