Coldplay CryoBalloon® Ablation System

K153541 · C2 Therapeutics, Inc. · GEH · Jan 7, 2016 · General, Plastic Surgery

Device Facts

Record IDK153541
Device NameColdplay CryoBalloon® Ablation System
ApplicantC2 Therapeutics, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateJan 7, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Coldplay CryoBalloon® Ablation System is intended for use as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.

Device Story

The Coldplay CryoBalloon® Ablation System is an endoscopic cryosurgical device used to ablate unwanted tissue. The system consists of a sterile, single-use catheter with a distal balloon, a reusable handle, and a single-use nitrous oxide cartridge. The catheter is inserted through the working channel of a standard endoscope. The physician visualizes the treatment site through the pre-inflated balloon, positions the device using the handle, and triggers the release of nitrous oxide. The gas inflates the balloon and cools it to extreme temperatures, freezing and destroying the target tissue. The nitrous oxide is exhausted through the handle. The device is intended for use in clinical settings by trained physicians. It provides a minimally invasive method for tissue ablation, potentially reducing the need for more invasive surgical procedures.

Clinical Evidence

Bench testing only. Design verification testing was performed to evaluate physical, reliability, and safety specifications to support substantial equivalence.

Technological Characteristics

Cryosurgical unit using nitrous oxide-cooled balloon. Components: sterile single-use catheter, reusable handle, single-use cartridge. Compatible with endoscopes (min 3.7mm ID, max 100cm length). Features 8mm diameter distal balloon. Battery-operated handle controller with software. User-controlled trigger for cryogen release.

Indications for Use

Indicated for destruction of unwanted tissue in general surgery via endoscopic application.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, overlaid on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. January 7, 2016 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 C2 Therapeutics, Inc. Theresa Brandner-Allen VP Of Regulatory Affairs And Quality Assurance 303 Convention Way, Suite 1 Redwood City, California 94063 Re: K153541 Trade/Device Name: Coldplay CryoBalloon® Ablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: December 9, 2015 Received: December 10, 2015 Dear Theresa Brandner-Allen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. Joshua Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153541 Device Name Coldplay CryoBalloon® Ablation System ### Indications for Use (Describe) The Coldplay CryoBalloon® Ablation System is intended for use as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 5. 510(k) Summary This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. #### I. SUBMITTER C2 Therapeutics, Inc. 303 Convention Way, Suite 1 Redwood City, CA 94063 Phone: 650-521-5921 Fax: 650-556-1145 | Contact Person: | Theresa Brandner-Allen | |-----------------|------------------------------------------------| | | VP of Regulatory Affairs and Quality Assurance | | Date Prepared: | December 09, 2015 | #### II. DEVICE Name of Device: Coldplay CryoBalloon® Ablation System Common Name: Cryosurgical Unit, Cryogenic Surgical Device Classification Name: Cryosurgical Unit, Cryogenic Surgical Device 21 CFR§878.4350(a)(2) Regulatory Class: Class II Product Code: GEH #### III. PREDICATE DEVICE Coldplay CrvoBalloon® Full Ablation System, K151054 This predicate has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION The modified device is the Catheter used as part of a cryosurgical unit with a nitrous oxide cooled balloon that is compatible with commercially available endoscopes with a minimum working channel inner diameter (ID) of 3.7 mm and maximum length of 100 cm. The modified Catheter is part of a system comprised of three components including a Catheter (sterile, single use), Handle (non-sterile, reusable), and Cartridge (non-sterile, single use). The modified device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter is comes in contact with tissue and is {4}------------------------------------------------ inflated with nitrous oxide. Tissue is visualized through the pre-inflated balloon, and the treatment site is selected is by adjusting the endoscope and Handle. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Handle. A detailed comparison of the modified device to the predicate device is presented in detail in Section 11 and Section 12. #### V. INDICATIONS FOR USE The Coldplay CryoBalloon® Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE Cryoablation is the fundamental technological principle for both the modified Coldplay CryoBalloon® Ablation System and the predicate Coldplay CryoBalloon® Full Ablation System. Both the modified device and predicate device are based on the same endoscopic instrumentation for removing unwanted tissue using extreme cold. The modified Coldplay CryoBalloon® Ablation System has similar technological characteristics to the legally marketed predicate. The modified device and predicate device are based on the following same technological elements: - Inserted through an endoscope to access the treatment site - Apply a cryogen to ablate (freeze) the unwanted tissue ● - Use of a compliant balloon to position the treatment diffuser and to contain ● and exhaust the cryogen - . User-controlled (trigger) activates release of the cryogen - Battery-operated Handle controller and software The only modifications that were made include: - The Catheter was modified to have a smaller (8mm diameter) distal balloon. . #### VII. PERFORMANCE DATA Performance data were provided in support of the substantial equivalence determination. Design verification testing was performed on the Coldplay CryoBalloon® Ablation System to evaluate physical, reliability, and safety specifications. # VII. CONCLUSION The modified Coldplay CryoBalloon® Ablation System has the same clinical attributes, technological characteristics, and expected performance as the legally marketed Coldplay CryoBalloon® Full Ablation System (K151054) predicate. The {5}------------------------------------------------ performance data results demonstrate that the modified Coldplay CryoBalloon® Ablation System should perform as intended in the specified use conditions and should perform comparably to the legally marketed predicate that is currently marketed for the same intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...