The Visual-ICE Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures. The Visual-ICE Cryoablation System has the following specific indications: • Urology: Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) • Oncology: Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention • Dermatology: Ablation or freezing of skin cancers and other cutaneous disorders Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin • Gynecology: Ablation of malignant neoplasia or benign dysplasia of the female genitalia • General surgery: Palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas • ENT: Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth • Thoracic surgery: Ablation of arrhythmic cardiac tissue cancerous lesions • Proctology: Ablation of benign or malignant growths of the anus or rectum, and hemorrhoids
Device Story
Mobile console system; performs minimally invasive cryoablation. Uses Joule-Thomson effect via high-pressure argon gas circulating through closed-tip needles to freeze tissue; active thawing via helium gas or i-Thaw heating element. Operated by clinicians in clinical settings. System provides touch-screen interface for procedure monitoring and control. Software revision 1.3.1 adds usability enhancements: movable timers, configurable gas alerts, flexible freeze-thaw cycle programming, and improved diagnostic messaging. Output allows clinicians to precisely target and destroy diseased tissue; benefits include effective tumor/lesion ablation and palliative care.
Clinical Evidence
Bench testing only. No clinical data provided. Software verification and validation testing confirmed that all modifications met performance requirements and did not introduce new risks.
Technological Characteristics
Mobile console; Joule-Thomson cryoablation principle; high-pressure argon (freezing) and helium (thawing) gas; i-Thaw heating element; touch-screen interface; software-controlled; no hardware changes from predicate.
Indications for Use
Indicated for patients requiring cryosurgical tissue destruction in general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. Specific conditions include prostate cancer, BPH, malignant/benign tumors, skin cancers, warts, lesions, gynecological dysplasia, hemorrhoids, and arrhythmic cardiac tissue.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
K123865 — VISUAL-ICE CRYOABLATION SYSTEM · Galil Medical, Inc. · Feb 6, 2013
K113860 — VISUAL-ICE CRYOABLATION SYSTEM · Galil Medical, Inc. · Mar 12, 2012
K230551 — Visual-ICE Cryoablation System · Boston Scientific Corporation · Aug 8, 2023
K201588 — CRYOCARE TOUCH System and Accessories, Endocare 1.7mm Round Ice PerCryo Cryoprobe, Short, Round Ice · Varian Medical Systems, Inc. · Aug 11, 2020
K234002 — ICEfx Cryoablation System (FPRCH8000-02) · Boston Scientific · Mar 13, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half. Inside the circle is an abstract design featuring three stylized human profiles facing to the right, with flowing lines that resemble both hair and wings. The overall design is simple and conveys a sense of health, well-being, and government service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2015
Galil Medical Incorporated Ms. Amy McKinney Vice President, Global Regulatory Affairs & Quality 6518 Tamarind Sky Lane Fulshear, Texas 77441
Re: K143564
Trade/Device Name: Visual-ICE® Cryoablation System, Software Revision 1.3.1 Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: February10, 2015 Received: February 11, 2015
Dear Ms. McKinney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143564
Device Name
Visual-ICE Cryoablation System, Software Revision 1.3.1
#### Indications for Use (Describe)
The Visual-ICE Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures. The Visual-ICE Cryoablation System has the following specific indications:
· Urology: Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH)
- · Oncology: Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention
- · Dermatology: Ablation or freezing of skin cancers and other cutaneous disorders, destruction of warts or lesions,
angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors,
dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin
- · Gynecology: Ablation of malignant neoplasia or benign dysplasia of the female genitalia
- · General surgery: Palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas
- ENT: Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth
- · Thoracic surgery: Ablation of arrhythmic cardiac tissue cancerous lesions
- · Proctology: Ablation of benign or malignant growths of the anus or rectum, and hemorrhoids
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" in blue, with a green dot inside the "G". To the right of the "G" is the text "GALILMEDICAL" in blue. Below the logo is the text "Leading Through Innovation" in a smaller font size and also in blue.
# Section 5. 510(k) Summary
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Galil Medical Inc.<br>4364 Round Lake Road W<br>Arden Hills, MN 55112<br>USA |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Phone:<br>Fax:<br>Email: | Amy E. McKinney<br>Vice President, Global Regulatory Affairs & Quality<br>Galil Medical Inc.<br>651-287-5096<br>651-287-5097<br>amy.mckinney@galilmedical.com |
| Alternate Contact Person:<br>Phone:<br>Email: | Lynne Davies<br>Sr. Manager, Global Regulatory Affairs<br>Galil Medical Inc.<br>651-287-5098<br>Lynne.davies@galilmedical.com |
| Device Name: | Visual-ICE® Cryoablation System, Software Revision 1.3.1<br>IceRod 1.5 CX Cryoablation Needle |
| Device Classification<br>Name: | Cryosurgical unit and accessories (GEH)<br>21 CFR 878.4350 |
| Predicate Device: | Visual-ICE® Cryoablation System (K113860)<br>Visual-ICE® Cryoablation System, Software Revision 1.2.2<br>(K123865)<br>IceRod CX Cryoablation Needle (K121251) |
| Date of Preparation: | March 3, 2015 |
#### Device Description:
The Visual-ICE Cryoablation System is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively, by the use of Galil Medical i-Thaw® technology in which a heating element inside the cryoablation needle can be
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Galil Medical. The logo consists of a stylized letter "G" with a green dot inside, followed by the text "GALILMEDICAL" in a bold, sans-serif font. Below the company name is the tagline "Leading Through Innovation" in a smaller, lighter font.
energized to cause thawing.
This Special 510(k) is being submitted to modify the software with a variety of changes to enhance usability of the system. The functions of the system that users use to deliver the cryoablation treatment remain unchanged. The table below provides a summary comparison of the changes included in this Special 510(k).
| User Interface Changes | Added information to be displayed on Big Timers and made timers movable Modified wording on all displayed messages for clarity Minor appearance modifications Minor bug fixes |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Service and<br>Maintenance Changes | Minor updates used only by service personnel |
| Functional Changes | Modified wording for clarity. Removed lockout for procedure at maintenance due and end of life; will display warning upon startup. Allow for user configurable time parameter for displayed low gas message. Removed restriction for programming up to three freeze thaw cycles. Added ability to load and save pre-programmed freeze-thaw cycles. Allow cautery and freeze in any order. Added a 30 second flush after a freeze. Added ability to display internal gas temperature of i-Thaw capable needles. Added gas line flush at start of every procedure Minor bug fixes |
## Intended Use:
The Visual-ICE Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures. The Visual-ICE Cryoablation System has the following specific indications:
| • Urology | Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) |
|------------|-----------------------------------------------------------------------------------------------|
| • Oncology | Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Galil Medical. The logo features a stylized letter "G" in blue, with a green circle inside the "G". To the right of the "G" is the text "GALIL MEDICAL" in blue, with "GALIL" being larger than "MEDICAL". Below the logo is the text "Leading Through Innovation" in a smaller font.
| ● | Dermatology | Ablation or freezing of skin cancers and other cutaneous disorders |
|---|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin |
| ● | Gynecology | Ablation of malignant neoplasia or benign dysplasia of the female genitalia |
| ● | General surgery | Palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas |
| ● | ENT | Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth |
| ● | Thoracic surgery | Ablation of arrhythmic cardiac tissue cancerous lesions |
| ● | Proctology | Ablation of benign or malignant growths of the anus or rectum, and hemorrhoids |
# Summary of Performance Data and Substantial Equivalence:
The Visual-ICE System and software were evaluated in accordance with Galil Medical's risk management plan. No new unacceptable risks were identified based on the changes incorporated into Software Revision 1.3.1. Complete software verification and validation testing was completed on the new software revision. Software Revision 1.3.1 passed all verification and validation testing. No changes were made to the Visual-ICE System hardware.
## Conclusion:
The information and data provided in this Special 510(k) establish that the Visual-ICE Cryoablation System Software Revision 1.3.1 is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.