← Product Code [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH) · K020605

# SANARUS VISICA TREATMENT SYSTEM (K020605)

_Sanarus Medical, Inc. · GEH · Mar 8, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K020605

## Device Facts

- **Applicant:** Sanarus Medical, Inc.
- **Product Code:** [GEH](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH.md)
- **Decision Date:** Mar 8, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Sanarus Visica™ Treatment System is intended for use in general surgery, gynecology and oncology. The system is designed to destroy tissue by the application of extreme cold temperatures. In addition, the system is intended for use in the following indications: General Surgery - Ablation of breast fibroadenoma. Gynecology - Ablation of malignant neoplasia or benign dysplasia of the female genitalia. Oncology - Ablation of cancerous or malignant tissue. - Ablation of benign tumors. - Palliative intervention.

## Device Story

Visica™ Treatment System is a cryosurgical unit using inert argon gas to destroy tissue via extreme cold. System comprises a software-controlled console and 1-8 disposable cryoprobes. Cryoprobes utilize Joule-Thompson principle for cooling at distal tip; integrated thermocouples monitor tip temperature. System also supports up to 8 independent T-type needle thermocouples for monitoring surrounding tissue temperatures. Operated by clinicians in clinical settings; console displays system status and allows control via keys or remote keypad. Provides localized tissue destruction for surgical, gynecological, and oncological applications. Enables precise thermal control to facilitate tumor ablation and palliative care.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Cryosurgical unit; Joule-Thompson cooling principle; inert argon gas; 1-8 disposable cryoprobes; T-type needle thermocouples for tissue monitoring; 486 IBM-compatible microprocessor control; 110/230 VAC power; software-controlled operation.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Endocare Cryocare® Surgical System ([K003811](/device/K003811.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K020605

### 8 2002 MAR

# Section 2 - Summary of Safety and Effectiveness

### Contact Information (1)

Vincent Cutarelli Vice President, Regulatory Affairs Telephone: (949) 768-1184 ext. 105 FAX: (925) 460-6084

### Company Information (2)

Sanarus Medical, Inc. 5880 W. Las Positas Blvd., Suite 52 Pleasanton, CA 94588 Telephone: (925) 460-6080 FAX: (925) 460-6084

### Device Name (3)

Sanarus Visica™ Treatment System

#### (4) Device Description

The Visica™ Treatment System consists of a control unit that controls one to eight single-use, disposable cryoprobes. The control unit is software-controlled and operates off standard 110/230 VAC wall power. A 486 IBM-compatible microprocessor serves as the host computer and a screen displays the status of the system. System control is accomplished either directly through keys on the console itself (e.g., 1-probe system) or through a remote control keypad (e.g., 4 and 8-probe system). The cryoprobes operate on the Joule-Thompson principle and the refrigerative capacity is limited only to the distal tip of the probe. The cryoprobes incorporate a thermocouple to measure temperatures at the probe tip. The thermocouple is mounted inside each cryoprobe tip and its signal is used to monitor and control some operations of the system. The control unit can also control one to eight independent temperature probes to monitor temperatures in surrounding tissues. The temperature probes are standard T-type needle thermocouples.

The system utilizes inert argon gas as a cooling agent. The system is available in 1, 4 and 8-Cryoprobe configurations. The performance characteristics and internal design of each model are equivalent. The primary differences are the number of valves to control the cryoprobes (e.g., 1-8), number of thermocouple inputs (e.g., 1-8) and the size of the outer case.

{1}------------------------------------------------

KO20605

### (ട) Indications for Use

The Sanarus Visica™ Treatment System is intended for use in general surgery, gynecology and oncology. The system is designed to destroy tissue by the application of extreme cold temperatures. In addition, the system is intended for use in the following indications:

General Surgery

- Ablation of breast fibroadenoma .
### Gynecology

- Ablation of malignant neoplasia or benign dysplasia of the female genitalia ●
### Oncology

- Ablation of cancerous or malignant tissue ●
- Ablation of benign tumors .
- Palliative intervention ●

### Name of Predicate or Legally Marketed Device (6)

Endocare Cryocare® Surgical System

#### Substantial Equivalence (7)

The Sanarus Visica™ Treatment System is substantially equivalent to the Endocare Cryocare® Surgical System that was determined to be substantially equivalent on October 11, 2001 (reference K003811).

#### (8) Technological Characteristics

The Visica™ Treatment System has the same technological characteristics and performance specifications as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure embracing another, with the figures forming a bird-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 8 2002

Mr. Vincent Cutarelli Vice President, Regulatory Affairs Sanarus Medical, Inc. 5880 West Las Positas, Suite 52 Pleasanton, California 94588

Re: K020605

Trade Name: Visica™ Treatment System Regulation Number: 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: February 22, 2002 Received: February 25, 2002

Dear Mr. Cutarelli:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rottened your have determined the device is substantially equivalent (for the ications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have May 26, 1970, the chargence with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, Condons provisions or the ice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may or subject to basil additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 loase be actived a determination that your device complies with other requirements of the Act that I Dr Has intact and regulations administered by other Federal agencies. You must or any I vict all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K Fat 6077, idoomig (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Vincent Cutarelli

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Cor Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

U20605

## Indications For Use

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

Device Name: Visica™ Treatment System

Indications for Use: The Visica™ Treatment System is intended for use in general moncations for OSC. The Piersa - The system is designed to destroy tissue by the surgery, gyllecology and oncology. "The bytellers and one of the system is intended for use in the following indications:

General Surgery

- Ablation of breast fibroadenoma .
Gynecology

- Ablation of malignant neoplasia or benign dysplasia of the female genitalia ●
### Oncology

- Ablation of cancerous or malignant tissue .
- Ablation of benign tumors ●
- Palliative intervention .

Concurrence of CDRH, Office of Device Evaluation (ODE):

iriam C. Provost

General, Restorative and Neurological Devic

510(k) Number K620605

Prescription Use: __X (Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K020605](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEH/K020605)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
