← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K192566

# Reach and CQ'ENCE Circular Staplers (K192566)

_Reach Surgical, Inc. · GDW · Oct 17, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K192566

## Device Facts

- **Applicant:** Reach Surgical, Inc.
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** Oct 17, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of end-to-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.

## Device Story

Sterile, single-use, hand-held, non-powered circular stapler; used in general surgery for alimentary tract anastomoses. Device places double-staggered circular row of titanium staples while simultaneously resecting tissue. Surgeon manually activates device via trigger; controls tissue compression by varying closed staple height. Features include anvil, cartridge, instrument shaft, trigger, safety release, wing nut, staple height indicator, and ancillary trocar. Audible indicator clicks when device is fully open for safe removal. Used in OR by surgeons; provides mechanical tissue joining and resection. Benefits include standardized anastomosis creation and adjustable compression for varying tissue requirements.

## Clinical Evidence

Bench testing only. Performance data demonstrated compliance with specifications for visual inspection, dimensional analysis, staple form, and safety mechanism performance, consistent with predicate K120179.

## Technological Characteristics

Materials: ASTM F 67 unalloyed titanium staples; polycarbonate and stainless steel end effector; stainless steel shaft; polycarbonate handle. Lubricant: PTFE and silicone grease. Energy: Manual. Dimensions: 21mm, 25mm, 28mm, 31mm, 34mm head diameters; 27cm or 35cm shaft lengths. Sterilization: EtO (SAL 10^-6). Connectivity: None (mechanical).

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- REACH Circular Staplers ([K120179](/device/K120179.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 17, 2019

Reach Surgical, Inc. % Rich Grant Official Correspondent SeCQure Surgical 4480 Lake Forest Dr. Cincinnati. Ohio 45242

Re: K192566

Trade/Device Name: Reach and CQ'ENCE Circular Staplers Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: September 6, 2019 Received: September 18, 2019

Dear Rich Grant:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192566

Device Name CQ'ENCE TM Circular Staplers

Indications for Use (Describe)

The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of end-to-end, side-toside and end-to-side anastomoses in both open and laparoscopic surgeries.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

#### I. SUBMITTER

Rach Surgical Inc. 120 Xinxing Road West Zone, TEDA Tianjin, Tianjin CHINA 300462

Contact Person: Richard Grant Phone: 513-608-4017 Email: rgrant@secquresurgical.com

Date Prepared: September 6, 2019

#### II. DEVICE

Name of Device: Reach™ and CQ ENCE™ Circular Stapler Common or Usual Name: Circular Stapler Classification Name: Staple, implantable (21 CFR 878.4750) Regulatory Class: II Product Code: GDW

III. PREDICATE DEVICE

REACHTM Circular Staplers, (K120179)

#### IV. DEVICE DESCRIPTION

The REACH™ and SeCQure CO'ENCE™ Circular Staplers (CO'ENCE™ Circular Staplers) are sterile, single-use, hand-held, non-powered stapling devices used during general surgical procedures for the creation of end-to-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.

The Circular Staplers place a double-staggered, circular row of titanium staples while simultaneously resecting tissue when the device is manually activated. The device is provided in five head diameter sizes: 34mm, 31mm, 28mm, 25mm, and 21mm (product codes CS34, CS31, CS28, CS25, and CS21 respectively). Each size device can be purchased with a standard 27cm length shaft. The CS21 and SC25 can also be purchased with a longer, 35cm shaft (product codes CS21L, CS25L) to facilitate surgeon access to desired tissues.

The CQ 'ENCE™ Circular Stapler allows the surgeon to control tissue compression by varying the closed staple height. The CQ ENCE Circular Stapler forms traditional B-shaped staples that are comprised of titanium. Safe removal of the device is facilitated by opening the instrument one full the Audible Indicator makes a clicking sound.

#### V. INDICATIONS FOR USE

{4}------------------------------------------------

The CQ ENCE Circular Staplers have application throughout the alimentary tract for the creation of endto-end, side-to-side and end-to-side anastomoses in both open and laparoscopic surgeries.

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Reach™ and CQ 'ENCE™ Circular Stapler has the same technological characteristics as the predicate device with the exception of an additional device size and minor technological characteristics as described below

| Product                            | Reach TM and SeCQure CQ'ENCETM<br>Circular Stapler (Subject)                                                                                       | Predicate ReachTM Circular<br>Stapler<br>(K120179)                                                                                  | Comparison                                                                                                                                                                                                                                                                                                                           |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Anatomical Site Used               | Various soft tissues accessible during<br>laparoscopic procedures                                                                                  | Various soft tissues accessible<br>during laparoscopic procedures                                                                   | Same                                                                                                                                                                                                                                                                                                                                 |
| Product Models Provided            | SeCQure CQ'ENCETM 1<br>CS34, CS31, CS28, CS25, CS25L,<br>CS21, CS21L<br>ReachTM 1<br>ACS34D, ACS31D, ACS28D,<br>ACS25D, ACS25DL, ACS21D<br>ACS21DL | RCS34D, RCS31G, RCS28G,<br>RCS25G, RCS23G,<br>RCS17G                                                                                | Equivalent. The subject<br>device is provided in a<br>21mm size which is<br>within the previously<br>cleared range and was<br>cleared in K071023. The<br>17 mm size is<br>discontinued. The models<br>are available in two brand<br>types, ReachTM or<br>SeCQure CQ'ENCETM,<br>which are the same with<br>the exception of branding. |
| Cutting Mechanism                  | Circular Knife                                                                                                                                     | Circular Knife                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                 |
| Method of Activation               | Single Trigger manually activated                                                                                                                  | Single Trigger manually activated                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                 |
| Disposable/Reusable                | Disposable                                                                                                                                         | Disposable                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                 |
| Safety Mechanism                   | Firing Safety with Out-of-Range<br>Lockout                                                                                                         | Firing Safety with Out-of-Range<br>Lockout                                                                                          | Same                                                                                                                                                                                                                                                                                                                                 |
| Main Device<br>Features/Components | Anvil<br>Cartridge                                                                                                                                 | Anvil<br>Cartridge                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                 |
|                                    | Instrument Shaft                                                                                                                                   | Instrument Shaft                                                                                                                    |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Trigger                                                                                                                                            | Trigger                                                                                                                             |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Safety Release                                                                                                                                     | Safety Release                                                                                                                      |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Wing Nut                                                                                                                                           | Wing Nut                                                                                                                            |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Staple Height Indicator                                                                                                                            | Staple Height Indicator                                                                                                             |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Ancillary Trocar                                                                                                                                   | Ancillary Trocar                                                                                                                    |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Shaft                                                                                                                                              | Shaft                                                                                                                               |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Staple                                                                                                                                             | Staple                                                                                                                              |                                                                                                                                                                                                                                                                                                                                      |
| Materials                          | Staples: ASTM F 67 Unalloyed<br>Titanium                                                                                                           | Staples: ASTM F 67 Unalloyed<br>Titanium                                                                                            | Equivalent. A<br>biocompatible lubricant<br>was added to the<br>manufacturing process in<br>the current configuration.                                                                                                                                                                                                               |
|                                    | End Effector: polycarbonate and<br>stainless steel                                                                                                 | End Effector: polycarbonate and<br>stainless steel                                                                                  |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Shaft: stainless steel                                                                                                                             | Shaft: stainless steel                                                                                                              |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Handle: polycarbonate                                                                                                                              | Handle: polycarbonate                                                                                                               |                                                                                                                                                                                                                                                                                                                                      |
|                                    | Lubricant: PTFE & Silicone Grease                                                                                                                  | Lubricant: Silicone Grease                                                                                                          |                                                                                                                                                                                                                                                                                                                                      |
| Staple Shape                       | Standard "B" shaped staple                                                                                                                         | Standard "B" shaped staple                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                 |
| Staple Height (open)               | 3.5 and 4.8 mm                                                                                                                                     | 3.5 and 4.8 mm                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                 |
| Staple Height (closed)             | 1.0 to 1.8 for 3.5 mm Staples<br>1.2 to 2.2 mm for 4.8 mm staples                                                                                  | 1.5 and 2.2 mm                                                                                                                      | Same staple height. The<br>Staple Height labeling has<br>been amended to align<br>with industry standard to<br>describe the range of<br>staple heights to be<br>selected by the user. The<br>staple height remains<br>adjustable in the control<br>of the user.                                                                      |
| End Effector Diameters             | 34mm, 31mm, 28mm, 25mm, and<br>21mm                                                                                                                | 34mm, 31mm, 28mm, 25mm,<br>23mm and 17 mm                                                                                           | Equivalent. The subject<br>device is provided in a<br>21mm size which is<br>within the previously<br>cleared range. The 17 mm<br>size is being discontinued.                                                                                                                                                                         |
| Shaft Length                       | Standard (27cm), Long (35 cm)                                                                                                                      | Standard (27cm), Long (35 cm)                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                 |
| Markings on Handle                 | Printed instructions for opening                                                                                                                   | None                                                                                                                                | Equivalent. Printed instructions were added for ease of use.                                                                                                                                                                                                                                                                         |
| Anvil Tilt                         | Anvil Tilt Feature not offered                                                                                                                     | Anvil Tilt Feature offered                                                                                                          | Equivalent. The Anvil Tilt feature is no longer offered as non-tilting is a standard for circular staplers.                                                                                                                                                                                                                          |
| Opening Method                     | Turn the Wing Nut counterclockwise one half to a full turn until the Audible Indicator clicks                                                      | Turn the Wing Nut counterclockwise one half to a full turn                                                                          | Equivalent. Opening is the same. An audible click has been added to indicate to the user when the device is fully open.                                                                                                                                                                                                              |
| Biocompatibility                   | Stapler is biocompatible for blood/bone/tissue contact for limited duration<br>Staples are biocompatible for permanent implantation                | Stapler is biocompatible for blood/bone/tissue contact for limited duration<br>Staples are biocompatible for permanent implantation | Same                                                                                                                                                                                                                                                                                                                                 |
| Endotoxin Limit                    | 0.5 EU/mL (20 EU/Device)                                                                                                                           | 0.5 EU/mL (20 EU/Device)                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                 |
| Sterilization Method               | EtO, SAL 10-6                                                                                                                                      | EtO. SAL 10-6                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                 |
| Packaging                          | Tyvek covered tray in cardboard box                                                                                                                | Tyvek covered tray in cardboard box                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                 |
| Shelf Life                         | 5 years                                                                                                                                            | 5 years                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                 |

Table 1: Technological Characteristics

<sup>4</sup> Note: The Reach and SeCQure CQ ENCE™ products are the same products with different branding i.e. the CS34 is the 34mm size SeCQure branded CQ ENCE™ Stapler and the ACS34D is the 34mm size Reach™ branded CQ ENCE™ Stapler. The devices are the same other than the branding on the labels and pad printed branding on the handles.

{5}------------------------------------------------

{6}------------------------------------------------

### PERFORMANCE DATA VII.

Reach™ Circular Stapler performance bench testing was provided in K120179. Testing for the Reach™ and CQ 'ENCE™ Circular Stapler is provided in Section 018 - Performance Testing - Bench of this submission to the same specifications as the Reach™ Circular Stapler previously provided in K120179. The data demonstrates that the Reach™ and CQ ENCE™ Circular Stapler meets the same specifications for visual inspection, dimensional analysis, staple form and safety mechanism performance. The materials used in the device, packaging, and sterilization method remain unchanged.

### VIII. CONCLUSIONS

The CQ 'ENCE™ Circular Stapler has the same intended use and principles of operation as its predicate device. The minor differences in technological characteristics do not change the device use or application and have been tested to the same standards as the currently marketed devices. There are no new issues of safety or effectiveness. The CQ ENCE™ Circular Stapler is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K192566](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K192566)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
