Disposable Linear Stapler, Disposable Linear Cutter

K172538 · Changzhou Kangdi Medical Stapler Co., Ltd. · GDW · Nov 30, 2017 · General, Plastic Surgery

Device Facts

Record IDK172538
Device NameDisposable Linear Stapler, Disposable Linear Cutter
ApplicantChangzhou Kangdi Medical Stapler Co., Ltd.
Product CodeGDW · General, Plastic Surgery
Decision DateNov 30, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Disposable Linear Stapler: The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues. Disposable Linear Cutter: The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.

Device Story

Disposable linear staplers and cutters; manual surgical instruments used by surgeons in OR settings. Devices deliver two staggered rows of titanium staples to approximate internal tissues; used for resection, transection, and anastomosis. Staple height adjustable for varying tissue thickness. Modifications include updated firing knob and handle designs. Devices function via manual mechanical actuation; no electronic or software components. Benefit: provides secure tissue approximation and closure during surgical procedures.

Clinical Evidence

Bench testing only. Testing demonstrated consistent energy requirements for staple delivery and extraction compared to the predicate. Biocompatibility testing performed per ISO 10993-1, -3, -5, -6, -10, and -11. Sterility and packaging verification completed.

Technological Characteristics

Manual surgical stapling instruments. Materials: Titanium staples. Design: Staggered row staple delivery; adjustable staple height. Connectivity: None. Sterilization: Not specified, but process unchanged from predicate. Software: None.

Indications for Use

Indicated for patients undergoing alimentary tract, thoracic, abdominal, or gynecological surgical procedures requiring resection, transection, or anastomosis of internal tissues. Applicable to pediatric patients.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 30, 2017 Changzhou Kangdi Medical Stapler Co. Ltd. % Mr. Charlie Mack Principal Engineer IRC 2091 Oak Drive Lake Havasu City, Arizona 86406 Re: K172538 Trade/Device Name: Disposable Linear Stapler, Disposable Linear Cutter Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: August 12, 2017 Received: August 22, 2017 Dear Mr. Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172538 Device Name Disposable Linear Stapler Disposable Linear Cutter Indications for Use (Describe) Disposable Linear Stapler: The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues. Disposable Linear Cutter: The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: August 13, 2017 ## Company making the submission: Name - Changzhou Kangdi Medical Stapler Co., Ltd Address - No. 16 Kunlun Road, XinBei Zone, Changzhou City, Jiangsu Province, China 213022 Telephone - 86-519-85162780 Fax - 86-519-85139853 Contact-Ms. EveZhou Email - charliemack@irc-us.com ### Correspondent: Name: Charles Mack Address: 2091 Oak Drive, Lake Havasu City, Arizona 86406 Telephone: 931-625-4938 Email: charliemack@irc-us.com #### Device: Trade/proprietary names: | Disposable Linear Stapler | KFD-30L, KFD-30H, KFD-45L, KFD-45H, KFD-60L, KFD-60H, KFD-90L, KFD-90H | |---------------------------|--------------------------------------------------------------------------------| | | KYFA-30L, KYFA-30H, KYFA-45L, KYFA-45H, KYFA-60L, KYFA-60H, KYFA-90L, KYFA-90H | | | KYFB-30L, KYFB-30H, KYFB-45L, KYFB-45H, KYFB-60L, KYFB-60H, KYFB-90L, KYFB-90H | | Disposable Linear Cutter | KYQ-60L, KYQ-60H, KYQ-80L, KYQ-80H, KYQ-100L, KYQ-100H | | | KYQB-60L, KYQB-60H, KYQB-80L, KYQB-80H, KYQB-100L, KYQB-100H | Common Name: Implantable Staple Classification Name: Staple, implantable {5}------------------------------------------------ #### Predicate Device: This submission is a modification to the previously cleared Changzhou Kangdi Medical Stapler Co., Ltd Disposable Linear Cutter and Disposable Linear Stapler under 510(k) number K121474 with a decision date on 07/16/2012. ## Classification 21 CFR 878.4750, Implantable staple, Class II, GDW ### Description: #### General #### Disposable Linear Stapler The Disposable Linear Stapler delivers two staggered rows of titanium staples to approximate internal tissues. The instrument is available in a variety of sizes to accommodate different tissues. Staple height is adjustable to compensate for various tissue thickness. The model number defines the staple line. For example, the 30 mm instruments create a 30-mm staple line, the 45 mm instruments create a 45-mm staple line, the 60 mm instruments create a 60-mm staple line and the 90 mm instruments create a 90-mm staple line. #### Disposable Linear Cutter The disposable linear cutter and the cartridge for disposable linear cutter have applications in abdominal gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. #### Indications For Use #### Disposable Linear Stapler: The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues. #### Disposable Linear Cutter: The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. {6}------------------------------------------------ ### Technical Characteristics: The Disposable Linear cutter and Disposable Linear cutter have not changed in basic design. There is no change to the staples in material or structure. The only difference in the current submitted device is the firing knob and the handles. Testing of the staples during insertion and extraction showed that there are no appreciable differences in the energy needed to insert or extract the staples. #### Test Summary: In order to establish substantial equivalence to the identified predicate device. the following tests were performed: Bench testing was performed to demonstrate the energy to deliver staples and also to remove staples was consistent with the predicate device. Biocompatibility testing was performed to ensure the stapler and cutters were constructed of material which passes the ISO 10993 benchmarks. The devices successfully passed tests for ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10 and ISO 10993-11. The modifications to the devices have not affected the sterility process, materials or packaging. For verifications, packaging and sterility verfications were completed. ### Conclusions: In accordance with the Federal Food. Drug and Cosmetic Act. 21 CFR Part 807 and based on the information provided in this premarket notification Changzhou Kangdi Medical Stapler Co.. Ltd. concludes that Disposable Linear Stapler. Model KYFB and Disposable Linear Cutter, Model KYQII are substantially equivalent to predicate devices as described herein. END
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...