← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K171561

# Dextera MicroCutter 30 Blue Reload, Dextera MicroCutter 30 Curved Tip Blue Reload, Dextera MicroCutter 30 White Reload, Dextera MicroCutter 30 Curved Tip White Reload (K171561)

_Dextera Surgical, Inc. · GDW · Aug 28, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K171561

## Device Facts

- **Applicant:** Dextera Surgical, Inc.
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** Aug 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The MicroCutter 5/80 Stapler and MicroCutter 30 White, Blue, and Blue Reloads are intended for transection and resection in multiple open or minimally invasive urologic, thoracic, and pediativ surgical procedures, as well as application for transection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix. The MicroCutter 5/80 Stapler and MicroCutter 30 White and Curved Tip White Reloads may additionally be used in conjunction with open solid organ parenchymal dissection techniques on veins 4-5 mm in diameter and arteries 3-7 mm in diameter.

## Device Story

The Dextera MicroCutter 30 Reloads are cartridge-based components used with the MicroCutter 5/80 Stapler. The system delivers two double-staggered rows of 316L stainless steel staples while simultaneously transecting tissue between the rows. Used in open or minimally invasive urologic, thoracic, pediatric, and intestinal surgeries. Operated by surgeons. The device provides mechanical tissue closure and transection, facilitating surgical procedures like anastomoses and organ resection. Benefits include reliable tissue approximation and vessel ligation during complex procedures.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical performance data, including functional tissue burst pressure testing on porcine veins and arteries, and previously submitted chronic animal studies (unilateral nephrectomies and lobectomies) comparing subject staples to predicate staples. Results showed no demonstrable differences in performance.

## Technological Characteristics

Staples constructed of 316L stainless steel. Reloads provide two double-staggered rows of staples with a 28.0mm cut length and 30mm staple line. Staple sizes are specific to tissue compression thicknesses of 1.0mm (White) or 1.5mm (Blue). Mechanical actuation via the MicroCutter 5/80 Stapler. Single-patient use.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Ethicon ENDOPATH® ETS-Flex45 Endoscopic Articulating Linear Cutters and (Reloads) ([K070887](/device/K070887.md))

## Reference Devices

- MicroCutter 30 Blue Reload, MicroCutter 30 Curved Tip Blue Reload, MicroCutter 30 White Reload and MicroCutter 30 Curved Tip White Reload used with the MicroCutter 5/80 Stapler ([K151081](/device/K151081.md) and [K161137](/device/K161137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2017

Dextera Surgical, Inc. Ms. Pamela Segale Consulting VP, Regulatory Affairs 900 Saginaw Drive Redwood City, CA 94063

Re: K171561

Trade/Device Name: Dextera MicroCutter 30 Blue Reload, Dextera MicroCutter 30 Curved Tip Blue Reload, Dextera MicroCutter 30 White Reload, Dextera MicroCutter 30 Curved Tip White Reload Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: May 27, 2017 Received: May 30, 2017

Dear Ms. Segale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Binita S. Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171561

### Device Name

MicroCutter 30 Blue Reload, MicroCutter 30 Curved Tip Blue Reload, MicroCutter 30 Curved Tip White Reload, Dextera MicroCutter 5/80 Stapler

### Indications for Use (Describe)

The MicroCutter 5/80 Stapler and MicroCutter 30 White, Blue, and Blue Reloads are intended for transection and resection in multiple open or minimally invasive urologic, thoracic, and pediativ surgical procedures, as well as application for transection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix.

The MicroCutter 5/80 Stapler and MicroCutter 30 White and Curved Tip White Reloads may additionally be used in conjunction with open solid organ parenchymal dissection techniques on veins 4-5 mm in diameter and arteries 3-7 mm in diameter.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

#### Date Prepared A.

August 25, 2017

#### B. 510(k) Owner

Dextera Surgical Inc. 900 Saginaw Drive Redwood City, California 94063 Phone: 650-364-9975 Fax: 650-364-3134

#### C. Contact Person

Pamela Segale Consulting VP, Regulatory Affairs (650) 331-7131 (office) (650) 799-7162 (cell) psegale@dexterasurgical.com

#### D. Device Information

Trade name: MicroCutter 30 Blue Reload, MicroCutter 30 Curved Tip Blue Reload, MicroCutter 30 White Reload, and MicroCutter 30 Curved Tip White Reload

Common name: Staple Classification Name: Implantable staple Regulation Number: 21 CFR §878.4750 Product Code: GDW

#### E. Marketed Predicate Devices

Ethicon ENDOPATH® ETS-Flex45 Endoscopic Articulating Linear Cutters and (Reloads) (K070887)

## Reference Devices

MicroCutter 30 Blue Reload, MicroCutter 30 Curved Tip Blue Reload, MicroCutter 30 White Reload and MicroCutter 30 Curved Tip White Reload used with the MicroCutter 5/80 Stapler (K151081 and K161137)

#### F. Device Description

Each MicroCutter 30 Reload contains two double rows of 316L stainless steel staples. The size of the MicroCutter 30 Blue Reload and MicroCutter 30 Curved Tip Blue Reload staple is compatible with tissue that can be easily compressed to 1.5mm. The size of the MicroCutter30 White Reload and MicroCutter 30 Curved Tip White Reload staple is compatible with tissue that can be easily compressed to 1.0mm. If tissue cannot comfortably compress to the Maximum Tissue Thickness listed for each reload, or easily compresses to less than the Maximum Tissue Thickness listed for each reload, the tissue

{4}------------------------------------------------

is contraindicated as it may be too thick or too thin for the selected staple size. The cut length of the staple line is 28.0mm (1.10in) and the staple line length is 30mm long. The MicroCutter 30 Reloads are deployed using the Dextera MicroCutter 5/80TM Stapler. The MicroCutter 5/80 Stapler is a single patient use stapler that delivers two, double staggered rows of 316L stainless steel staples while simultaneously transecting tissue between staple rows.

#### G. Indications for Use

The MicroCutter 5/80 Stapler and MicroCutter 30 White, Blue, and Curved Tip White and Blue Reloads are intended for transection and resection in multiple open or minimally invasive urologic, thoracic, and pediatric surgical procedures, as well as application for transection, resection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix.

The MicroCutter 5/80 Stapler and MicroCutter 30 White and Curved Tip White Reloads may additionally be used in conjunction with open solid organ parenchymal dissection techniques on veins 4-5 mm in diameter and arteries 3-7 mm in diameter.

#### H. Comparison to Predicate Device

A comparison of the subject devices to the predicate devices demonstrates that the devices have the same intended use, which is to transect, resect and create anastomoses in patients during surgery in a variety of surgical procedures including urologic, thoracic, pediatric, and intestinal surgery. Additionally, they have the same mechanism of action and similar technologies given both the subject and the predicate devices achieve their intended use by means of cartridge based metal staples that are delivered using stapling devices.

Chronic animal studies, previously submitted by Dextera Surgical (K151081), were conducted in which unilateral nephrectomies and unilateral lobectomies were performed comparing the Dextera Surgical subject staple and the predicate Ethicon staple. The endpoints of the studies were met with no demonstrable differences noted between the subject devices and the predicate devices. Non-clinical performance data was also previously submitted (K151081) in which the Dextera Surgical subject staple and the Ethicon predicate staple were compared during Tissue Burst Pressure testing in porcine arteries. The test data demonstrated the subject devices to be substantially equivalent to the predicate devices.

#### I. Performance Data

### Non-Clinical Performance Data

Functional performance testing was conducted using the subject staple and the predicate staple on porcine veins during Tissue Burst Pressure testing. The testing simulated transection of vessels as may be encountered during surgery on solid organs. The results demonstrated the subject devices to be substantially equivalent to the predicate devices.

#### J. Conclusion

The subject devices and the predicate devices have intended use and similar indications for use. Additionally, the subject devices and the predicate devices have the same mechanism of action and utilize similar technologies. Functional performance testing (Tissue Burst Pressure) was conducted with the results demonstrating the subject 

{5}------------------------------------------------

devices to be substantially equivalent to the predicate devices. Thus, based on the data and the comparison of the devices, the subject devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K171561](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K171561)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
