ECHELON CIRCULAR Powered Stapler

K163523 · Ethicon Endo-Surgery, LLC · GDW · Apr 18, 2017 · General, Plastic Surgery

Device Facts

Record IDK163523
Device NameECHELON CIRCULAR Powered Stapler
ApplicantEthicon Endo-Surgery, LLC
Product CodeGDW · General, Plastic Surgery
Decision DateApr 18, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ECHELON CIRCULAR™ Powered Staplers have applications throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.

Device Story

ECHELON CIRCULAR Powered Stapler is a sterile, single-patient use surgical instrument; used for creating anastomoses in alimentary tract. Device features battery-powered firing system to simultaneously staple and cut tissue; delivers two rows of staples. Surgeon operates device; places detachable anvil at target site; installs battery pack prior to use. Rotatable adjustment knob allows tissue compression and selection of target staple height. Powered mechanism facilitates staple formation and lumen cutting. No energy passes to patient. Output is a stapled, cut anastomosis. Benefits include consistent staple formation and tissue cutting during surgical procedures.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench and animal testing. Bench tests: staple line integrity, staple form quality, leak onset pressure, force to fire, formed staple height, battery output, and fluid ingress. Animal studies: acute hemostasis, device removal, insertion tissue effects, healing of stapled anastomosis, and staple line perfusion.

Technological Characteristics

Battery-powered stapling and cutting device. Features 24 cm curved shaft; 23mm, 25mm, 29mm, and 31mm end-effector sizes. Detachable anvil. Rotatable adjustment knob for tissue compression. No software; no energy transfer to patient.

Indications for Use

Indicated for patients requiring alimentary tract anastomoses (end-to-end, end-to-side, side-to-side).

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 18, 2017 Ethicon Endo-Surgery, LLC Ms. Rubina Dosani Ethicon Endo-Surgery, LLC 4545 Creek Road Cincinnati, Ohio 45242 Re: K163523 Trade/Device Name: ECHELON CIRCULAR Powered Stapler Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: March 20, 2017 Received: March 22, 2017 Dear Ms. Dosani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163523 Device Name ECHELON CIRCULAR™ Powered Staplers Indications for Use (Describe) The ECHELON CIRCULAR™ Powered Staplers have applications throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Submitter Information: | Ethicon, LLC | |---------------------------|-------------------------------------| | | 475 Calle Street | | | Guaynabo, PR 00969 | | Application Correspondent | | | | Rubina Dosani | | | Sr. Regulatory Affairs Program Lead | | | Ethicon Endo-Surgery, LLC. | | Telephone: | (513) 337-3566 | | Fax: | (513) 337-2314 | | Email: | rdosani@its.jnj.com | Date Prepared December 14, 2016 | Device Trade Name: | ECHELON CIRCULAR™ Powered Staplers | |--------------------------------|------------------------------------| | Device Common Name: | Circular Stapler | | Classification Regulation: | 21 CFR 878.4750 | | Device Class: | II | | Panel: | General & Plastic Surgery | | Classification (Product) Code: | GDW | Legally Marketed Predicate Device: Ethicon Endo-Surgery® Curved Intraluminal Staplers (cleared under K983536) ## Device Description The ECHELON CIRCULAR™ Powered Staplers are sterile, single-patient use devices that simultaneously staple and cut tissue with a battery powered firing system to create an anastomosis. The devices deliver 2 rows of staples on the outside of the cut line. The ECHELON CIRCULAR™ Powered Staplers are available in a 24 cm curved shaft length in 4 end-effector sizes: 23mm, 25mm, 29mm and 31mm. Each device has a detachable anvil that allows a surgeon to place the anvil in the desired location. The devices are packaged with a battery pack that must be installed prior to use. ## Indications for Use The ECHELON CIRCULAR™ Powered Staplers have applications throughout the alimentary tract for end-to-end, end-to side, and side-to-side anastomoses. ## Technological Characteristics The subject device utilizes battery power to allow powered staple formation and lumen cutting to create the anastomosis. The subject device has a similar design as the predicate with the addition of tighter staple height range and staple pockets extensions to the end-effector of the device as {4}------------------------------------------------ well as a slightly different shape of the formed staples. Similar to the predicate, a rotatable adjustment knob enables the compression of tissue and selection of a target staple height based on the tissue compression within the green zone. No energy passes through to the patient in the use of the subject or predicate devices. Neither the subject device or predicate device uses software. # Performance Data Performance data demonstrate that the subject device is substantially equivalent to the predicate device and the differences between the devices were found not to affect safety or performance. The following bench and animal tests were performed to demonstrate substantial equivalence to the predicate: - staple line integrity and staple form quality equivalency ● - leak onset pressure equivalency - force to fire - o formed staple height - battery output - o fluid ingress Animal testing performed included acute hemostasis evaluation, device removal acceptability, device insertion tissue effects, healing of stapled anastomosis and staple line perfusion. The conclusions of the testing demonstrate that the subject device performs substantially equivalent to the predicate device and does not raise new questions of safety and effectiveness. This submission does not include data from Clinical Studies.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...