← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K110385

# ECHELON FLEX 45 / 60 POWERED ARTICULATING ENDOSCOPIC LINEAR CUTTER (K110385)

_Ethicon Endo-Surgery, Inc. · GDW · Mar 25, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K110385

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, Inc.
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** Mar 25, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The ECHELON families of endoscopic linear cutters are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.

## Device Story

Echelon Flex Powered Articulating Endoscopic Linear Cutter is a sterile, single-patient-use surgical instrument. It features an ergonomic handle with an integrated battery-powered motor and an insulated tubular shaft. The device delivers six staggered rows of staples while simultaneously cutting tissue. It includes an articulation mechanism allowing the distal end effector to pivot up to 45° in 15° increments to facilitate lateral access. The device is used in OR settings by surgeons. It incorporates a safety lock-out feature preventing firing without a cartridge or with a used cartridge. The powered firing system replaces manual force, potentially reducing surgeon fatigue and improving consistency during transection and resection procedures.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Insulated tubular shaft; ergonomic handle with integrated motor; battery-powered firing system; articulation mechanism (up to 45° in 15° increments); sterile, single-patient-use; delivers six staggered rows of staples.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Echelon Endoscopic Linear Cutters-Articulating ([K081146](/device/K081146.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K110385

MAR 2 5 2011

## 510(k) Summary

Ethicon Endo-Surgery, LLC Company 475 Calle C Guaynabo, PR 00969

Contact Asifa Vonhof, RAC Regulatory Affairs Associate Ethicon Endo-Surgery, Inc. Telephone: (513) 337-3118 (513) 337-2314 Fax: Email: avonhof@its.jnj.com

Date Prepared February 09, 2011

Device Name Trade Name: Echelon Flex Powered Articulating Endoscopic Linear Cutters Common or Usual Name: Cutter/Stapler Classification Name: Staple, Implantable ; Stapler, Surgical

Predicate Device Echelon Endoscopic Linear Cutters-Articulating (cleared under K081146)

The Echelon Flex Powered Articulating Endoscopic Linear Cutters Device Description are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in three shaft lengths: compact, regular and long. The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.

The instruments are packaged with a battery pack that must be installed prior to use.

The instruments are shipped without a cartridge and must be loaded prior to use. The instrument has a safety lock-out feature that is designed to prevent an instrument without a cartridge or a used cartridge from being fired.

Indications for Use The ECHELON families of endoscopic linear cutters are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.

{1}------------------------------------------------

Technological Characteristics The instrument utilizes an insulated tubular shaft, an ergonomic handle with an integrated motor, and battery power to simultaneously transect (cut) and staple tissue. The instrument also features an articulation system that can adjust the end effector in increments of 15° to a maximum of 45°.

Performance Data Bench testing was performed to demonstrate that the new devices will perform as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Ethicon Endo-Surgery, LLC % Ethicon Endo-Surgery, Inc. Asifa Vonhof 4545 Creek Road Cincinnati, Ohio 45242

MAR 2 5 % .. i

Re: K110385

Trade/Device Name: Echelon Flex Powered Articulating Endoscopic Linear Cutters Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW, GAG Dated: February 9, 2011 Received: February 10, 2011

Dear Asifa Vonhof:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Asifa Vonhof

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Aling B. Rohn

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): _KII0305

Device Name: Echelon Flex Powered Articulating Endoscopic Linear Cutters

Indications for Use:

The ECHELON families of endoscopic linear cutters are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daind Kucuefer MXM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K110385

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K110385](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K110385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
