← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K103263

# ENDO GIA SINGLEUSE DUET TRS RELOAD WITH TRI0STAPLE TECHNOLOGY (K103263)

_Covidien, Formerly US Surgical A Divison of Tyco H · GDW · Nov 22, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K103263

## Device Facts

- **Applicant:** Covidien, Formerly US Surgical A Divison of Tyco H
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** Nov 22, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Endo GIA™ Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.

## Device Story

Endo GIA™ Duet TRS™ Reload is a surgical stapler accessory used with Endo GIA™ Staplers. Device places two sets of triple-staggered titanium staples while simultaneously dividing tissue and integrated absorbable reinforcement material. Reinforcement material consists of synthetic polyester (glycolide, dioxanone, trimethylene carbonate) secured with BIOSYN™ suture. Stepped cartridge configuration provides varying staple heights for different tissue thicknesses. Used by surgeons in OR settings for resection, transection, and anastomosis in abdominal, gynecologic, pediatric, and thoracic procedures. Provides mechanical tissue closure and hemostasis; reinforcement material supports staple line integrity during healing; absorbs over 90-110 days.

## Clinical Evidence

No clinical data. Evidence consists of bench and in vivo performance evaluations. Bench testing assessed firing force, staple line pull-apart strength, and staple formation. In vivo testing evaluated staple formation, hemostasis, burst strength, and lung air leak. Materials comply with ISO 10993-1.

## Technological Characteristics

Titanium staples; synthetic absorbable film (glycolide, dioxanone, trimethylene carbonate); BIOSYN™ synthetic absorbable suture. Stepped cartridge configuration with triple-staggered rows. Materials conform to ISO 10993-1. Mechanical actuation via Endo GIA™ Stapler handle.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Endo GIA™ Stapler ([K083519](/device/K083519.md))
- Endo GIA™ Stapler ([K093410](/device/K093410.md))
- Endo GIA™ Stapler ([K101444](/device/K101444.md))
- Endo GIA™ Stapler ([K080898](/device/K080898.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4103262

page 1(

### 510(k) Summary of Safety and Effectiveness

| SUBMITTER: | Surgical Devices, a global business unit<br>of Tyco Healthcare Group LP (d/b/a Covidien)<br>60 Middletown Avenue<br>North Haven, CT 06473 |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------|
|            | Tel. No.: (203) 492-5352                                                                                                                  |

NOV 2 2 2010

- CONTACT PERSON: Frank Gianelli Senior Associate, Regulatory Affairs
DATE PREPARED: November 2, 2010

TRADE/PROPRIETARY NAME: Endo GIA™ Stapler

COMMON/USUAL NAME: Surgical Stapler with Implantable Staples

CLASSIFICATION NAME: Staples, Implantable

PREDICATE DEVICE(S). Endo GIA™ Stapler (K083519, K093410, K101444, K080898)

DEVICE DESCRIPTION: This 510(k) shall describe a new Reload cartridge with Tri-Staple™ technology for use with the currently marketed Endo GIA™ Stapler. The new Duet TRS™ Reload with Tri-Staple™ Technology bears the identical design characteristics as the Endo GIA™ Reloads with Tri-Staple™ Technology described in predicate submissions K083519 and K090419 with the exception that the Reloads have pre-attached staple line reinforcement material which is a synthetic absorbable film prepared from synthetic polyester composed of glycolide, dioxanone, and trimethylene carbonate. The staple line reinforcement material is supplied on each Reload undyed (natural) and is secured to the

> Studies indicate that the staple line reinforcement material can be expected to retain measurable mechanical strength through 14 days and absorption should be complete between 90 and 110 days.

anvil and cartridge with BIOSYN™ synthetic absorbable suture.

The Endo GIA™ Duet TRS™ Reload with Tri-Staple™ Technology when used with the Endo GIA™ Stapler places two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows.

INTENDED USE:

The Endo GIA™ Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.

Note: The Endo GIA™ Duet TRS™ Reload with Tri-Staple™ Technology is intended for use with the Endo GIA™ Stapler and does not carry a separate indication from the stapling device.

{1}------------------------------------------------

#### TECHNOLOGICAL CHARACTERISTICS:

Endo GIA™ Duet TRS™ Reloads with Tri-Staple™ Technology are substantially equivalent to the predicate devices with regard to the stapling and staple line reinforcement technologies.

Duet TRS™ Reloads with Tri-Staple™ Technology place two sets of triple-staggered rows of titanium staples along with two layers of absorbable tissue reinforcement material (one layer on cartridge side and one layer on anvil side) and simultaneously divide the tissue and tissue reinforcement material between the two sets of triple-staggered rows. These Reloads have a stepped cartridge configuration whereby for each set of triple-staggered rows, the staples in the outer row are a taller height than the staples in the middle row which in turn are a taller height than the staples in the inner row closest to the cut line. The size of the staples is determined by the selection of the appropriate single use Reload.

#### MATERIALS:

#### PERFORMANCE DATA:

All components of the Endo GIA™ Duet TRS™ Reload with Tri-Staple™ Technology are comprised of materials that are in accordance with ISO Standard 10993-1.

Bench and in vivo performance evaluations were conducted to demonstrate that the Endo GIA™ Duet TRS™ Reload with Tri-Staple™ Technology when used with the Endo GIA™ Stapler is safe and effective and performs as intended.

The, bench tests performed evaluated firing force, staple line pull apart strength and staple formation. The in vivo tests performed evaluated staple formation, hemostasis, burst strength and lung air leak.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Covidien % Mr. Frank Gianelli Senior Associate, Regulatory Arfairs 60 Middletown Avenue North Haven, Connecticut 06473

NOV 2 2 2010

Re: K103263

Trade/Device Name: Endo GIA™ Single Use Duet TRS™ Reload with Tri-Staple" Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: November 2, 2010 Received: November 4, 2010

Dear Mr. Gianelli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Frank Gianelli

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Pan 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR) Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Molloy

Mark N. Melkers Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure .

{4}------------------------------------------------

# Indications for Use

NOV 2 2 2010

KIO 3263 510(k) Number (if known):

Endo GIA™ Single Use Duet TRS™ Reload with Tri-Staple™ Device Name: Technology

## Indications For Use:

The Endo GIA™ Staplers with Endo GIA™ Single Use Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Krone for MKM
(Division Sign Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K103263

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K103263](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K103263)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
