← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K080898

# AUTO SUTURE ENDO GIA STAPLERS WITH ENDO GIA SINGLE USE LOADING UNITS WITH STAPLE LINE REINFORCEMENT (K080898)

_Tyco Healthcare Group LP · GDW · May 5, 2008 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K080898

## Device Facts

- **Applicant:** Tyco Healthcare Group LP
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** May 5, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.

## Device Story

Surgical stapler system; places two triple-staggered rows of titanium staples while simultaneously dividing tissue; utilizes single-use loading units (SULU) with integrated synthetic absorbable film (glycolide, dioxanone, trimethylene carbonate) for staple line reinforcement; reinforcement material secured with BIOSYN™ synthetic absorbable suture; used by surgeons in OR settings; provides mechanical tissue closure and reinforcement to reduce potential leaks; reinforcement material retains strength for 14 days and absorbs within 90-110 days.

## Clinical Evidence

Bench testing only; performance evaluations conducted to verify safety and effectiveness.

## Technological Characteristics

Titanium staples; synthetic absorbable film (glycolide, dioxanone, trimethylene carbonate); BIOSYN™ synthetic absorbable suture; straight and articulating 45mm/60mm cartridges; 3.5mm or 4.8mm staple sizes; materials compliant with ISO 10993-1.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Autosuture™ ENDO GIA™ Stapler ([K061095](/device/K061095.md))
- Monofilament BIOSYN™ Suture ([K000037](/device/K000037.md))
- Gore SEAMGUARD® Bioabsorbable Staple Line Reinforcement ([K043056](/device/K043056.md))
- Synovis Peri-Strips® Staple Line Reinforcement ([K040119](/device/K040119.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 2

Autosuture™ ENDO GIA™ Surgical Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement

## 510(k) Summary of Safety and Effectiveness

| SUBMITTER:              | Surgical Devices, a global business unit<br>of Tyco Healthcare Group LP (d/b/a Covidien)<br>60 Middletown Avenue<br>North Haven, CT 06473<br>Tel. No.: (203) 492-6060                                                                                                                                                                                                                                                                                                                                                                                                                                    | MAY - 5 2008 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| CONTACT PERSON:         | Sharon Alexander<br>Senior Associate, Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| DATE PREPARED:          | March 28, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |              |
| TRADE/PROPRIETARY NAME: | Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single<br>Use Loading Units with Staple Line Reinforcement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| COMMON/USUAL NAME:      | Surgical Stapler with Implantable Staple                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| CLASSIFICATION NAME:    | Staple, Implantable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |              |
| PREDICATE DEVICE(S):    | Autosuture™ ENDO GIA™ Stapler (K061095)<br>Monofilament BIOSYN™ Suture (K000037)<br>Gore SEAMGUARD® Bioabsorbable Staple Line<br>Reinforcement (K043056)<br>Synovis Peri-Strips® Staple Line Reinforcement (K040119)                                                                                                                                                                                                                                                                                                                                                                                     |              |
| DEVICE DESCRIPTION:     | The Autosuture™ ENDO GIA™ Staplers with ENDO GIA™<br>Single Use Loading Units with Staple Line Reinforcement<br>place two, triple-staggered rows of titanium staples and<br>simultaneously divides the tissue between the two, triple-<br>staggered rows. The size of the staples is determined by the<br>selection of the 3.5 mm or 4.8 mm Single Use Loading Unit<br>(SULU). The ENDO GIA™ UNIVERSAL Single Use Loading<br>Units (SULUs) with Staple Line Reinforcement are only<br>available in the straight and articulating 45 mm and 60 mm<br>length cartridges with the 3.5 mm or 4.8 mm staples. |              |
|                         | The staple line reinforcement material is a synthetic<br>absorbable film prepared from synthetic polyester composed<br>of glycolide, dioxanone, and trimethylene carbonate. The<br>staple line reinforcement material is supplied on each SULU<br>undyed (natural) and is secured to the anvil and cartridge with<br>BIOSYN™ synthetic absorbable suture.                                                                                                                                                                                                                                                |              |
|                         | Studies indicate that the staple line reinforcement material can<br>be expected to retain measurable mechanical strength through<br>14 days and absorption should be complete between 90 and<br>110 days.                                                                                                                                                                                                                                                                                                                                                                                                |              |
| INTENDED USE:           | The Autosuture™ ENDO GIA™ Staplers with ENDO GIA™<br>Single Use Loading Units with Staple Line Reinforcement have<br>applications in abdominal, gynecologic, pediatric and thoracic<br>surgery for resection, transection, and creation<br>of<br>anastomoses. They may be used for transection and resection<br>of liver substance, hepatic vasculature and biliary structures.                                                                                                                                                                                                                          |              |

{1}------------------------------------------------

Page 2 of 2.

## Autosuture™ ENDO GIA™ Surgical Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement

- TECHNOLOGICAL CHARACTERISTICS: The Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement are substantially equivalent to the predicate devices with regard to the stapling and staple line reinforcement technologies. MATERIALS: All components of the Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement are comprised of materials which are in accordance with ISO Standard 10993-1. PERFORMANCE DATA: Performance evaluations were completed to verify that the Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement are safe
and effective and perform as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY -5 2008

Tyco Healthcare Group, LP % Ms. Sharon Alexander Senior Associate, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473

Re: K080898

Trade/Device Name: Autosuture™ ENDO GLATM Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW, GAG Dated: March 28, 2008 Received: March 31, 2008

Dear Ms. Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Sharon Alexander

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M. Mulleen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

## Indications For Use

510(k) Number (if known):

## Device Name: Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement

Indications For Use:

The Autosuture™ ENDO GIA™ Staplers with ENDO GIA™ Single Use Loading Units with Staple Line Reinforcement have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and billary structures.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General. Restorative. and Neurological Device

**510(k) Number** ll o 80 rar

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K080898](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K080898)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
