← Product Code [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW) · K030653

# SURGASSIST STRAIGHT LINEAR STAPLER AND STRAIGHT LINEAR 4 ROW NO KNIFE DIGITAL LOADING UNITS WITH RELOADS (K030653)

_Power Medical Interventions, Inc. · GDW · Mar 31, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K030653

## Device Facts

- **Applicant:** Power Medical Interventions, Inc.
- **Product Code:** [GDW](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW.md)
- **Decision Date:** Mar 31, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The SurgASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.

## Device Story

SurgASSIST™ Straight Linear Stapler and 4-Row No-Knife Digital Loading Units (DLUs) are single-patient-use surgical staplers. The device consists of a DLU attached to a flexible shaft (FlexShaft) connected to a Power Console (PC). The surgeon operates the device via a handheld electronic Remote Control Unit (RCU). The DLU features an anvil and staple cartridge that compresses tissue, forms staples, and (in stapler models) cuts tissue. The device is used in general and endoscopic surgery to secure tissue layers. Each DLU/Reload contains an electronic memory module that identifies the component type and prevents reuse of fired cartridges. The system provides mechanical power for stapling and cutting, allowing the surgeon to adjust the DLU attitude for optimal access. The output is a stapled (and optionally cut) tissue line, facilitating resection and anastomosis.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Materials: Titanium staples. Energy: Mechanical power via Power Console. Connectivity: Electronic memory module in DLU for identification and reuse prevention. Sterilization: Ethylene Oxide (ETO). Form factor: 15mm diameter DLU, axial to FlexShaft. Software: Embedded firmware for DLU identification and lockout.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- SurgASSIST™ System with Straight Linear Cutter Digital Loading Units® ([K020719](/device/K020719.md))
- Endopath ETS Flex45 No-Knife Articulating Linear Stapler ([K020779](/device/K020779.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030653 (1 of 7)

MAR 3 1 2003

## Special 510(k) Device Modification PREMARKET NOTIFICATION SAFETY AND EFFECTIVENESS SUMMARY

# SurgASSIST Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads

In Accordance with 21 CFR section 807.92 Power Medical Interventions, Inc., is submitting the following safety and effectiveness summary.

- 1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 267-775-8151 Ph 267-775-8123 Fax

Applicant:

Barbara J. Whitman

Date of Notification: February 28, 2003

- 2) Name of Device:

| Trade Name:  | SurgASSIST™<br>Straight Linear Stapler DLU with Reloads<br>and<br>Straight Linear 4 Row No Knife DLU<br>with Reloads |
|--------------|----------------------------------------------------------------------------------------------------------------------|
| Common Name: | Linear Stapler with Implantable Staples<br>and Reloads                                                               |

Classification Name: Staple, Implantable, GDW

- 3) Predicate Devices:
	- A. SurgASSIST™ System with Straight Linear Cutter Digital Loading Units®, 55mm and 30mm with Blue and Green Reloads, with Titanium Implantable Staple, Power Medical Interventions, Inc., New Hope, PA. REF SLC55B, SLC55G, SLCR55B, SLCR55G, SLC30B, SLC30G, SLCR30B, SLCR30G. (K020719).

# 000014

{1}------------------------------------------------

030653 (P.2 of 7)

- B. Endopath ETS Flex45 No-Knife Articulating Linear Stapler, Ethicon Endo-Surgery, Inc., Cincinnati, OH. REF NAB, NAG. (K020779).
- 4) Device Description:

The devices described here are reloadable Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® (DLUs) with Reloads for single patient use. All have a maximum diameter of .596" and can be used within a 15 mm cannula. The DLUs are supplied pre-sterilized and ready for use upon removal from their packaging. Reloadable stapling and cutting cartridges are referred to as "Reloads".

Each DLU contains a staple-forming anvil. The anvil acts with the staple cartridge to compress and position layers of tissue in readiness for stapling and cutting. At the same time, the anvil provides support and a means for correctly forming staples while they are closed sequentially along the tissue, followed by the cutting blade, when included. The 15 mm diameter DLUs are axial to the FlexShaft, forming a straight extension to the FlexShaft, to which they are connected.

Loaded Straight Linear Stapler and Straight Linear 4 Row No Knife DLUs are used to close otomies and other common and uncommon openings by applying staples through the tissue and forming the staples to a controlled closed condition to secure the layers of tissue together. Straight Linear Stapler and Straight Linear 4 Row No Knife Reloads contain staples and the means to sequentially force staples toward the anvil. The cutting blade, when included, is advanced immediately behind the staple pushers so that tissue is sequentially stapled and then cut, proceeding from one end of the cartridge to the other.

Straight Linear Stapler DLUs, 55 mm with Blue and Green Reloads have 2 staggered rows of staples on one side of the cutter, coded Blue (1.2 mm) for normal tissue and Green (2.0 mm) for thick tissue. DLUs are available with Blue or Green Reloads installed. Individual Blue and Green Reloads are available.

Straight Linear 4 Row No Knife DLUs, 55 mm with Gray Reloads have two, double-staggered rows of staples, coded Gray (1.2 mm) for normal tissue (same as 55 mm Blue Straight Linear Cutter without the blade). DLUs are available with Gray cartridges installed. "No Knife" distinctively labeled. Individual Gray Reloads are available.

The DLUs are attached to the end of the flexible shaft assembly that contains a pair of flexible rotary drive shafts within an overall flexible tube called the FlexShaft, hereafter referred to as FS. The other end of the FS is connected to a Power Console (PC), which applies mechanical power to the drive shafts.

{2}------------------------------------------------

DLUs have all functions powered by the PC. The FS has a short steerable section at the distal end (near the attached DLU) so that the angle of attack (attitude) of the DLU can be adjusted by the surgeon to optimize patient accessibility.

The surgeon operates a DLU via a hand held electronic Remote Control Unit (RCU) that is plugged into the front panel of the PC.

The DLUs have quick attach and release means of coupling to the FS. No tools are required. DLUs are pushed onto the FS end, snapping and locking into place. To remove a DLU from the FS, a sleeve on the DLU at the junction with the FS, is rotated by hand.

DLU designs allow for attachment of Reloads, but will inhibit attachment of incompatible Reloads. Each Reload has an integral electronic memory module. This identifies the type and size of the Reload being used and prevents a used Reload from being refired.

- 5) Indications For Use
The SuraASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.

- 6) Comparison to Predicate Devices
The following table compares the subject Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads to the previously cleared predicate Straight Linear Cutter Digital Loading Units® with Reloads (K020719).

{3}------------------------------------------------

Pow. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 20

Straight Linear Stapler DLU Product Features Comparison Cha

| Features & Description                                                                                                                              | Name                                                                                                                                                                                                                | Predicate                                                                                                                                                                                                           | Predicate                                                                                                                                                                                                                                                                                                                                                                         | Features & Description                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate                                                                                                                                                      | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                      |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| SurgASSIST™<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs) | SurgASSIST™<br>Straight Linear Stapler DLUs, 55 mm<br>with Blue and Green Reloads and<br>Straight Linear 4 Row No Knife DLUs                                                                                        | SurgASSIST™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads                                                                                                                       | Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler                                                                                                                                                                                                                                                                                         | SurgASSISTTM<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | SurgASSISTTM<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads                                                                 | Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                      |
|                                                                                                                                                     |                                                                                                                                                                                                                     | SurgASSIST™<br>Straight Linear Cutter DLUs, 55 mm and<br>30 mm with Blue and Green Reloads                                                                                                                          | Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler                                                                                                                                                                                                                                                                                                                       | Do not use the SurgASSISTTM Straight<br>Linear Stapler DLU, 55mm Blue, on any<br>tissue that requires excessive force to<br>compress to 1.5 mm or on any tissue<br>that compresses easily below 1.5 mm.<br><br>Do not use the SurgASSISTTM Straight<br>Linear Stapler DLU, 55 mm Green, on<br>any tissue that requires excessive force<br>to compress to 2.3 mm or on any tissue<br>that compresses easily below 2.3 mm.<br><br>Do not use the SurgASSISTTM Straight<br>Linear 4 Row No Knife DLU, (Gray), on<br>any tissue that requires excessive force<br>to compress to 1.5 mm or on any tissue<br>that compresses easily below 1.5 mm. | Do not use the SurgASSISTTM System<br>Straight Linear Cutter DLU, 55 mm,<br>30 mm where combined tissue<br>thickness exceeds 2.3 mm or is less<br>than 1.0 mm. | Do not use the instrument with blue<br>reload on any tissue that requires<br>excessive force to compress to 1.5 mm<br>or on any tissue that compresses easily<br>to below 1.5 mm.<br><br>Do not use on ischemic or necrotic<br>tissue.<br><br>Do not use the instruments on the aorta.<br>Do not use any linear cutter on major<br>vessels without making provision for<br>proximal and distal control.<br>Do not use the instrument on solidorgans, such as liver or spleen, where<br>attempted compression would be<br>destructive.<br>These instruments are not intended for<br>use when surgical stapling is<br>contraindicated.<br>Do not use the instruments with green<br>reload on any tissue that requires<br>excessive force to compress to 2.0 mm<br>or on any tissue that compresses easily<br>to below 2.0 mm. |                                      |
| Manufacturer of Record                                                                                                                              | Power Medical Interventions, Inc.                                                                                                                                                                                   | Power Medical Interventions, Inc.                                                                                                                                                                                   | Ethicon Endo-Surgery, Inc                                                                                                                                                                                                                                                                                                                                                         | FDA Class (System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |
| Contract Manufacturer                                                                                                                               | The MedTech Group<br>South Plainfield, NJ                                                                                                                                                                           | The MedTech Group<br>South Plainfield, NJ                                                                                                                                                                           | Ethicon Endo-Surgery, Inc                                                                                                                                                                                                                                                                                                                                                         | Number of Staples                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | SLS55B, SLSR55B - 28 staples<br>SLS55G, SLSR55G - 28 staples<br>SLS55B4, SLSR55B4 - 56 staples                                                                 | SLC55B, SLCR55B - 56 staples<br>SLC55G, SLCR55G - 56 staples<br>SLC30B, SLCR30B - 32 staples<br>SLC30G, SLCR30G - 32 staples                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | NAB - 44 staples<br>NAG - 44 staples |
| 510(k) Clearance Numbers                                                                                                                            | Subject of this Notification                                                                                                                                                                                        | K020719                                                                                                                                                                                                             | K020779                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |
| Product Codes                                                                                                                                       | SLS55B, SLSR55B, SLS55G,<br>SLSR55G, SLS55B4, SLSR55B4                                                                                                                                                              | SLC55B, SLCR55B, SLC55G,<br>SLCR55G, SLC30B, SLCR30B,<br>SLC30G, SLCR30G                                                                                                                                            | NAB45 and NAG45                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |
| Intended use                                                                                                                                        | Have applications for general and<br>endoscopic surgery in gastrointestinal,<br>gynecological, general abdominal and<br>thoracic surgical procedures for<br>resection, transection, and creation of<br>anastomoses. | Have applications for general and<br>endoscopic surgery in gastrointestinal,<br>gynecological, general abdominal and<br>thoracic surgical procedures for<br>resection, transection, and creation of<br>anastomoses. | Intended for transection, resection,<br>and/or creation of anastomoses. The<br>instruments have application in multiple<br>open or minimally invasive general,<br>gynecologic, urologic (including radical<br>prostatectomy), thoracic, and pediatric<br>surgical procedures. They can be used<br>with staple line or tissue buttressing<br>materials such as bovine pericardium. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |

K 030653 (P.4 of 7)

000017

{4}------------------------------------------------

Pow. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 2

# Straight Linear Stapler DLU Product Features Comparison Cham

K030653 (P.Sofm)

{5}------------------------------------------------

Pow.. Medical Interventions, Inc.
SurgASSIST™ Straight Linear Stapler DLUs with Reloads
Special 510(k) Device Modification PreMarket Notification, February 28, 20

# raight Linear Stapler DLU Product Features Comparison Cham

| Features & Description | SurgASSIST™<br>Straight Linear Stapler, 55 mm<br>With Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUs) | Predicate<br>SurgASSIST™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads                                           | Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Rows of Staples        | SLS55B, SLSR55B - 2 rows<br>SLS55G, SLSR55G - 2 rows<br>SLS55B4, SLSR55B4 - 4 rows                                                                  | SLC55B, SLCR55B - 4 rows<br>SLC55G, SLCR55G - 4 rows<br>SLC30B, SLCR30B - 4 rows<br>SLC30G, SLCR30G - 4 rows                                         | NAB - 4 rows<br>NAG - 4 rows                                                                           |
| Knife/Scalpel          | SLS55B, SLSR55B - scalpel included<br>SLS55G, SLSR55G - scalpel included<br>SLS55B4, SLSR55B4 - NO KNIFE                                            | SLC55B, SLCR55B - scalpel included<br>SLC55G, SLCR55G - scalpel included<br>SLC30B, SLCR30B - scalpel included<br>SLC30G, SLCR30G - scalpel included | NAB - NO KNIFE<br>NAG - NO KNIFE                                                                       |
| Staple Diameter        | SLS55B, SLSR55B - .21 mm<br>SLS55G, SLSR55G - .23 mm<br>SLS55B4 - .21 mm                                                                            | SLC55B, SLCR55B - .21 mm<br>SLC55G, SLCR55G - .23 mm<br>SLC30B, SLCR30B - .23 mm<br>SLC30G, SLCR30G - .23 mm                                         | NAB - .23 mm<br>NAG - .23 mm                                                                           |
| Staple Height          | SLS55B, SLSR55B - 1.2 mm<br>SLS55G, SLSR55G - 2.0 mm<br>SLS55B4 - 1.2 mm                                                                            | SLC55B, SLCR55B - 1.2 mm<br>SLC55G, SLCR55G - 2.0 mm<br>SLC30B, SLCR30B - 1.2 mm<br>SLC30G, SLCR30G - 2.0 mm                                         | NAB - 1.5 mm<br>NAG - 2.0 mm                                                                           |
| DLU Internal Power     | None                                                                                                                                                | None                                                                                                                                                 | None                                                                                                   |
| Digital Information    | Memory module containing digital data<br>for identification, etc.                                                                                   | Memory module containing digital data<br>for identification, etc.                                                                                    | None                                                                                                   |
| How Supplied           | Sterile - Single Patient Use                                                                                                                        | Sterile - Single Patient Use                                                                                                                         | Sterile - Single Patient Use                                                                           |

# 000019

03d6S3(f.6of7)

{6}------------------------------------------------

Pow

# traight Linear Stapler DLU Product Features Comparison Cha
(continued from continued from previous page)

| 1<br>1 |  |                                                                                                                                                                                |
|--------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|        |  |                                                                                                                                                                                |
| t<br>1 |  |                                                                                                                                                                                |
| 1      |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
| 1      |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
| 1      |  | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|        |  |                                                                                                                                                                                |
| )      |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
| 1      |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
| .      |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |
|        |  |                                                                                                                                                                                |

| Features & Description                                                                                                                               | Method of Sterilization  | Digital Loading Unit®          | Reloads                     |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------|-----------------------------|
| SurgASSIST™™<br>Straight Linear Stapler, 55 mm<br>with Blue and Green Reloads and<br>Straight Linear 4 Row No Knife<br>Digital Loading Units® (DLUS) | Ethylene Oxide Gas (ETO) | Blister Tray<br>with Tyvek Lid | Tyvek Pouch                 |
| Predicate<br>SurgASSIST™™<br>Straight Linear Cutter DLUs,<br>55 mm and 30 mm with Blue and<br>Green Reloads                                          | Ethylene Oxide Gas (ETO) | Blister Tray with<br>Tyvek Lid | Tyvek Pouch                 |
| Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ETS Flex45 No-Knife<br>Articulating Linear Stapler                                               | Irradiation              | Blister Tray with Tyvek Lid    | Blister Tray with Tyvek Lid |

Image /page/6/Picture/5 description: The image contains handwritten text that appears to be an alphanumeric code or identifier. The text reads "KU30653(F.70P7)". The characters are written in a cursive style, and the text is slightly slanted.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAR 3 1 2003

Ms. Barbara J. Whitman Regulatory Affairs Manager Power Medical Interventions, Inc. 110 Union Square Drive New Hope, Pennsylvania 18938

Re: K030653

Trade/Device Name: SurgASSIST™ Straight Linear Stapler Digital Loading Unit® with Reloads and Straight Linear 4 Row No Knife Digital Loading Unit® with Reloads

Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: February 28, 2003 Received: March 3, 2003

Dear Ms. Whitman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic' Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2 - Ms. Barbara J. Whitman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

(Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Power Medical Interventions, Inc. New Hope, PA 18938

510(k) No. K K934653

Device Name:

SurgASSIST™ Straight Linear Stapler Digital Loading Unit® with Reloads And Straight Linear 4 Row No Knife Digital Loading Unit® with Reloads

### INDICATIONS FOR USE:

The SurgASSIST™ Straight Linear Stapler and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Prescription Use Per 21CFR §801.109

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

Miriam C. Phrost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

OR

510(k) Number_________________________________________________________________________________________________________________________________________________________________

000012

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K030653](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDW/K030653)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
