PENROSE DRAIN MODELS 6001 - 6007

K960099 · Sil-Med Corp. · GCY · Mar 11, 1996 · General, Plastic Surgery

Device Facts

Record IDK960099
Device NamePENROSE DRAIN MODELS 6001 - 6007
ApplicantSil-Med Corp.
Product CodeGCY · General, Plastic Surgery
Decision DateMar 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4680
Device ClassClass 1
AttributesTherapeutic

Intended Use

Post-operative passive drainage of serosanguineous fluids away from soft tissue operative site.

Device Story

Passive wound drain; silicone construction; flat design; internal diameter teeth to prevent collapse; radiopaque stripe for visualization. Used post-operatively in clinical settings to facilitate drainage of serosanguineous fluids from soft tissue sites. Operated by clinicians; device functions via gravity-assisted drainage. Sold sterile for single-patient use.

Clinical Evidence

Bench testing only. Biocompatibility testing performed: Rabbit Pyrogen (non-pyrogenic), Hemolysis (non-hemolytic), Salmonella typhimurium reverse mutation assay (non-mutagenic), Kligman Maximization (Grade I), USP Class VI (satisfied), Elution test (non-cytotoxic), and 90-day USP muscle implantation study in rabbits (met requirements).

Technological Characteristics

Material: Silicone. Features: Flat geometry, internal diameter teeth, radiopaque stripe. Dimensions: 18" length. Sterilization: Sterile. Energy source: None (passive).

Indications for Use

Indicated for post-operative patients requiring passive drainage of serosanguineous fluids from soft tissue operative sites.

Regulatory Classification

Identification

A nonpowered, single patient, portable suction apparatus is a device that consists of a manually operated plastic, disposable evacuation system intended to provide a vacuum for suction drainage of surgical wounds.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K960099 1 of 2 ![img-0.jpeg](img-0.jpeg) # SIL-MED ## Sil-Med Corporation 700 WARNER BLVD. TAUNTON, MA 02780 (508) 823-7701 FAX: (508) 823-1438 MAR 11 1996 ## 510 (k) Summary Company: Sil-Med Corporation 700 Warner Boulevard Taunton, MA 02780 Telephone: (508) 823-7701 Fax: (508) 823-1438 Contact Person: Karen K. Sylvia QA/QC Manager Trade Name: Penrose Drain Common Name: Flat Silicone Drain Classification Name: Apparative, Single Patient Use, Portable, Non-Powered (878.4680) Description: Passive Wound Drain Predicate Device: Davol Gravity Drain Intended Use: Post-operative passive drainage of serosanguineous fluids away from soft tissue operative site. ### Biocompatibility Testing: | Test | Results | | --- | --- | | Rabbit Pyrogen Test | non-pyrogenic | | Hemolysis-Rabbit Blood | non-hemolytic at 2.02% Hemolysis | | Salmonella typhimurium REVERSE Mutation ASSAY | non-mutagenic | | Kligman Maximization Study (Cottonseed Oil Extract) | Grade I - weak allergenic potential (not considered significant) | | Kligman Maximization Study (Sodium Chloride Extract) | Grade I - weak allergenic potential (not considered significant) | | USP CLASS VI | Satisfied the requirements of USP XXII Biological Test for Plastics, Class VI-121°C | | ELUTION TEST | non-cytotoxic | | USP MUSCLE IMPLANTATION STUDY IN RABBITS WITH HISTOPATHOLOGY (90 Days) | met USP requirements | {1} K960099 2012 Sil-Med Corporation 510 (k) Summary Penrose Drain page 2 Sil-Med Corporation's Penrose Drain is dimensionally equivalent to the Davol Gravity Drain, except Davol offers both 12" and 18"; Sil-Med Corporation offers 18" only. Both have inner diameter teeth to prevent the drain from collapsing. Both devices have a Radiopaque stripe. Davol Gravity Drains are sold sterile and non sterile. Sil-Med Penrose Drains are sold sterile only. The intended use of both drains is equivalent. Date: 1/10/96
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...