SAPPHIRE LINE FLAT DRAIN/KIT WITH SPI-ARGENT II W/OR W/OUT TROCAR

K961295 · Sil-Med Corp. · GBX · Jun 26, 1996 · General, Plastic Surgery

Device Facts

Record IDK961295
Device NameSAPPHIRE LINE FLAT DRAIN/KIT WITH SPI-ARGENT II W/OR W/OUT TROCAR
ApplicantSil-Med Corp.
Product CodeGBX · General, Plastic Surgery
Decision DateJun 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4200
Device ClassClass 1
AttributesTherapeutic

Indications for Use

For post-operative drainage of serosanguineous fluids and blood away from operative site; for use with suction reservoir as a closed wound drain system

Device Story

Silicone drainage catheter; flat perforated, ribbed section for fluid collection; round tubing connects to suction reservoir; optional introduction trocar; silver ion surface treatment (Spi-Argent II) to facilitate insertion/removal; blue radiopaque stripe for visualization. Used in clinical/post-operative settings by healthcare providers to manage wound exudate; closed system prevents contamination; aids healing by removing fluid accumulation.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Materials: Silicone; silver ion surface treatment. Features: Flat perforated/ribbed drain section, blue radiopaque stripe, optional trocar. Connectivity: None. Energy: None (passive suction).

Indications for Use

Indicated for post-operative patients requiring drainage of serosanguineous fluids and blood from surgical sites using a closed wound drain system.

Regulatory Classification

Identification

An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions. Examples include irrigation and drainage catheters, pediatric catheters, peritoneal catheters (including dialysis), and other general surgical catheters. An introduction/drainage catheter accessory is intended to aid in the manipulation of or insertion of the device into the body. Examples of accessories include adaptors, connectors, and catheter needles.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K961295 JUN 26 1996 # 510 (k) Summary | Company: | Sil-Med Corporation 700 Warner Boulevard Taunton, MA 02780 | | --- | --- | | Telephone: | (508) 823-7701 | | Fax: | (508) 823-1438 | | Contact Person: | Karen K. Sylvia QA/QC Manager | **Trade Name:** Flat Suction Drain/Kit with Spi-Argent™II (with and without Trocar) **Common Name:** Silicone Drainage Catheter **Classification Name:** 79 GBW Introduction/drainage catheter and accessories. (878.4200) **Description:** Two joined silicone tubing sections; flat drain section is partially perforated for fluid collection, is ribbed to prevent collapse as suction is applied, and has blue radiopaque stripe; round tubing is connected to suction reservoir inlet; offered with an introduction Trocar attached as an option; silver ion surface treated to ease insertion and removal. **Predicate Device:** Mediastinal Silicone Drain: K942709<br/>Sil-Slip Flat Suction Drain/Kit (with and without Trocar): K926408 **Intended Use:** For post-operative drainage of serosanguineous fluids and blood away from operative site; for use with suction reservoir as a closed wound drain system Sil-Med Corporation’s Flat Suction Drain/Kit with Spi-Argent™II, (with and without Trocar) is the same drain as the Sil-Slip Flat Suction Drain/Kit (with and without Trocar), K926408. The surface treatment is the same as the surface treatment on Sil-Med Corporation’s Mediastinal Silicone Drain, K942709. No further claims will be made for the Flat Suction Drain/Kit with Spi-Argent™II than made on K942709. Sil-Med Corporation<br/>510(k) Submission: Flat Suction Drain/Kit with Spi-Argent™II (with and without Trocar) 03 April 1996
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...