K160529 · Lsi Soultions · GAW · Sep 15, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160529
Device Name
Pledgeted COR-SUTURE QUICK LOAD
Applicant
Lsi Soultions
Product Code
GAW · General, Plastic Surgery
Decision Date
Sep 15, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pledgeted COR-SUTURE™ QUICK LOAD® is intended for use in the approximation of soft tissue and prosthetic material.
Device Story
Pledgeted COR-SUTURE™ QUICK LOAD® is a sterile, non-absorbable surgical suture system designed for use with LSI SOLUTIONS® suturing devices. The device consists of monofilament polypropylene or PTFE-coated braided polyester suture with a PTFE pledget pre-attached. Stainless steel ferrules are attached to each end of the suture to facilitate rapid, reliable loading into compatible suturing devices. The system includes a detachable clear tube to prevent tangling. Used in clinical settings by surgeons, the device simplifies suture placement during soft tissue or prosthetic material approximation. The pre-attached pledget reduces procedural steps, potentially improving efficiency and consistency in tissue repair. The device is MR safe and intended for single-patient use.
Clinical Evidence
No clinical studies were required. Substantial equivalence was demonstrated through bench testing, including USP monograph testing for diameter, tensile strength, and needle/ferrule attachment. Biocompatibility was established per ISO 10993-1:2009, including cytotoxicity, sensitization, irritation, toxicity, genotoxicity, implantation, and hemocompatibility testing. Data was leveraged from the suture material supplier and the reference device.
Technological Characteristics
Materials: monofilament polypropylene or PTFE-coated braided polyester; PTFE pledget; stainless steel ferrule. Energy: N/A (mechanical). Form factor: sterile tray with suture release feature and anti-tangle tube. Sterilization: Ethylene oxide. Connectivity: None. Software: None.
Indications for Use
Indicated for the approximation of soft tissue and prosthetic material in patients requiring surgical suture closure.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
LSI SOLUTIONS® Suture Placement Devices and Accessories (K100593)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2016
LSI Solutions Mr. Kevin Bentley Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd Victor, New York 14564
Re: K160529
Trade/Device Name: Pledgeted Cor-Suture Quick Load Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable Polypropylene Surgical Suture Regulatory Class: Class II Product Code: GAW, GAS Dated: August 9, 2016 Received: August 11, 2016
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160529
Device Name Pledgeted COR-SUTURE™ QUICK LOAD®
Indications for Use (Describe)
The Pledgeted COR-SUTURE™ QUICK LOAD is intended for use in the approximation of soft tissue and prosthetic material.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
| Submitted By: | LSI SOLUTIONS®, Inc.<br>7796 Victor-Mendon Road<br>Victor, NY 14564 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kevin Bentley<br>Executive Director of Regulatory Affairs and Quality<br>Phone: (585) 869-6669<br>Fax: (585) 742-3398<br>Email: kbentley@lsisolutions.com |
| Date Prepared: | February 24, 2016 |
| Trade Name: | Pledgeted COR-SUTURE™ QUICK LOAD® |
| Common Name: | Nonabsorbable Surgical Suture |
| Classification Name: | Nonabsorbable poly(ethylene terephthalate) surgical suture<br>(21 CFR 878.5000, Product Code GAS)<br>Nonabsorbable polypropylene surgical suture<br>(21 CFR 878.5010, Product Code GAW) |
| Device Classification: | Class II |
{5}------------------------------------------------
#### Predicate Device:
The predicate device, LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture, was selected to demonstrate substantial equivalence to the Pledgeted COR-SUTURE™ QUICK LOAD®. LSI SOLUTIONS® Suture Placement Devices and Accessories was cleared in 510(k) Premarket Notification K100593 on September 24, 2010.
The Bard® PTFE Felt Pledget, which was cleared under 510(k) Premarket Notification K770835 on October 5, 1977, is used as a reference device in this submission.
# Device Description:
Each LSI SOLUTIONS® Pledgeted COR-SUTURE™ QUICK LOAD® sterile surgical suture is held in a customized tray, with suture release feature designed to enable the rapid, easy and reliable loading of suture into compatible LSI SOLUTIONS® suturing devices. Pledgeted COR-SUTURE™ QUICK LOAD® suture is available as a non-absorbable monofilament polypropylene and PTFE coated, braided, polyester surgical suture with attached PTFE pledget. A short length of modified surgical stainless steel tubing, called a "ferrule", is attached to each end of the suture. The Pledgeted COR-SUTURE™ QUICK LOAD® suture also includes a detachable clear suture tube to keep the suture from tangling. Pledgeted COR-SUTURE™ QUICK LOAD® surgical suture is packaged for single patient use. The polypropylene suture is dyed blue with the FDA approved colorant [phthalocyaninato(2-)] copper. The polyester suture is offered undyed, or dyed green with the FDA approved colorant D&C Green No. 6. For both polypropylene and polyester suture, there is no known significant change in tensile strength retention to occur in vivo. The Pledgeted COR-SUTURE™ QUICK LOAD® surgical suture is MR safe.
#### Intended Use:
The Pledgeted COR-SUTURE™ QUICK LOAD® is intended for use in the approximation of soft tissue and prosthetic material.
{6}------------------------------------------------
The indications for use statement is identical to that of the predicate device, LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture.
# Technological Characteristics (comparison to Predicate Device):
The Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate device. Both the modified and predicate devices are non-absorbable surgical suture available as either monofilament polypropylene or PTFE coated, braided polyester. The Pledgeted COR-SUTURE™ QUICK LOAD® and the predicate device both have a short length of modified surgical stainless steel tubing, called a ferrule, attached to each end of the suture to facilitate use with LSI SOLUTIONS® suture placement devices. The only technological difference between the Pledgeted COR-SUTURE™ QUICK LOAD® and the predicate device is that the modified device will be packaged as surgical suture with a pledget preattached directly on the suture for ease of use. The pledget is composed of polytetrafluoroethylene (PTFE) and has identical material characteristics to the reference device, Bard® PTFE Felt Pledget. The modified device is packaged and sterilized in an equivalent manner and has equivalent labeling claims to the predicate device, including indications, contraindications, warnings, cautions and precautions. The modified device is considered substantially equivalent and has the same technological characteristics to its predicate device through comparison in design, intended use, material composition, function, and range of sizes.
# Performance Testing Summary:
As recommended by the FDA Guidance for Industry, Class II Special Control Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the Pledgeted COR-SUTURE™ QUICK LOAD® was subject to the requirements of the United States Pharmacopeia (USP) monograph for Non-Absorbable Sutures. Performance testing includes:
- . Sutures - Diameter <861>
- Sutures Tensile Strength <881> .
{7}------------------------------------------------
- Ferrule Attachment which complies with the requirements for standard needle . attachments as defined in USP section Sutures – Needle Attachment <871>.
Biocompatibility evaluation for the Pledgeted COR-SUTURE™ QUICK LOAD® was conducted in accordance with ISO 10993-1:2009: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and FDA 510(k) Blue Book Memorandum - #G95-1 Table 1 Initial Evaluation Tests for Consideration, to demonstrate that the Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent to the predicate device. The Pledgeted COR-SUTURE™ QUICK LOAD® is a permanent contact duration device that will be implanted in circulating blood, which is the identical type and duration of patient contact as the predicate device. Biocompatibility data has been leveraged from the manufacturer that supplies the surgical suture material to LSI SOLUTIONS®. The testing included the following:
- Cytotoxicity ●
- Sensitization .
- Irritation .
- Sub-chronic Toxicity and Sub-acute Toxicity .
- Genotoxicity ●
- Implantation .
- Hemocompatibility .
All biocompatibility testing passed.
The Bard® PTFE Felt Pledget, which has identical material characteristics to the pledget used in the Pledgeted COR-SUTURE™ QUICK LOAD®, was tested and found biocompatible based on the materials, manufacturing processes, body contact, sterilization properties, acceptable biocompatibility test results and long clinical history of this type of use.
The Pledgeted COR-SUTURE™ QUICK LOAD® will be sterilized using a validated ethylene oxide cycle.
{8}------------------------------------------------
#### Clinical Testing:
The technological characteristics, material, manufacturing and sterilization processes are the same as the predicate device, therefore, no clinical studies were required to demonstrate the safety or effectiveness of the modified device.
## Substantial Equivalence:
With the addition of the pledget there are no new risks introduced and no negative impact to the safety and effectiveness of the surgical suture. The indications for use remain exactly the same. Both the Non-Absorbable QUICK LOAD® Suture and the Pledgeted COR-SUTURE™ QUICK LOAD® are indicated for use in the approximation of soft tissue and prosthetic material. The Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent to the currently marketed LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture, cleared in 510(k) Premarket Notification K100593 on September 29, 2010.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.