NOBLESTITCH EL NOBLESTITCH MR NOBLESTITCH TA

K113494 · Nobles Medical Technologies Ii, Inc. · GAW · Apr 19, 2012 · General, Plastic Surgery

Device Facts

Record IDK113494
Device NameNOBLESTITCH EL NOBLESTITCH MR NOBLESTITCH TA
ApplicantNobles Medical Technologies Ii, Inc.
Product CodeGAW · General, Plastic Surgery
Decision DateApr 19, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NobleStitch™ EL, MR, and TA is indicated for use in performing vascular stitching in general surgery, including endoscopic procedures. The NobleStitch™ EL, MR, and TA is not intended for blind vascular closure.

Device Story

NobleStitch™ EL, MR, and TA are hand-held, manually operated vascular suturing devices. Used by physicians to place sutures at remote sites directly, through cannulas/introducers, or via laparoscopic access. Components include suture delivery device, monofilament polypropylene suture, KnotPusher™ for knot advancement, and KwiKnot™ accessory. Device facilitates vascular stitching; improves handle ergonomics compared to predicate. Used in general surgery and endoscopic settings. Provides mechanical assistance for suturing; does not involve automated processing or clinical decision support.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Hand-held, manually operated suturing device. Materials include monofilament polypropylene suture. Includes KnotPusher™ and KwiKnot™ accessories. No energy source, software, or connectivity. Sterilization method not specified.

Indications for Use

Indicated for vascular stitching in general surgery and endoscopic procedures. Not intended for blind vascular closure. Patient population not specified.

Regulatory Classification

Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113494 /2 #### 510(k) SUMMARY ## Nobles Medical Technologies II, Inc.'s NobleStitch™ EL, MR, and TA ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Nobles Medical Technologies II, Inc. 17080 Newhope St. Fountain Valley, California 92708 Phone: (714) 427-0398 Facsimile: (714) 427-0399 Contact Person: Anthony Nobles Date Prepared: November 11, 2011 Alternate Contact: Maria Hategan ## Name of Device and Name/Address of Sponsor NobleStitch™ EL, MR, and TA Nobles Medical Technologies II, Inc. 17080 Newhope St. Fountain Valley, California 92708 #### Common or Usual Name NobleStitch™ Vascular Suturing Device #### Classification Name Suture, Nonabsorbable, Synthetic, Polypropylene #### Predicate Devices SuperStitch® Vascular Suturing Device #### Intended Use/Indications for Use The NobleStitch™ EL, MR, and TA sizes are indicated for use in performing vascular stitching in general surgery, including endoscopic procedures. The NobleStitch™ EL, MR, and TA are not intended for blind vascular closure. {1}------------------------------------------------ ## Technological Characteristics The NobleStitch™ EL, MR. and TA versions are hand-held and manually operated suturing devices designed to allow a physician to place a suture to a remote site either directly, through a cannula/introducer, or through a laparascopic access device. The device contains the following components and accessories: a suture delivery device, monofilament polypropylene suture, a KnotPusher™ accessory for advancing the knot to the wound site, and/or a KwiKnot™ accessory. KI13494 2/2 ## Substantial Equivalence The NobleStitch™ EL, MR, and TA has the same intended use and indications for use, principles of operation, and fundamental technological characteristics as the cleared SuperStitch®, except that the NobleStitch™ EL, MR, and TA provide the option of different diameters or lengths with improved handle ergonomics for use by the physician. The minor modifications to the NobleStitch™ do not raise any new questions of safety or effectiveness. Thus, the NobleStitch™ EL, MR, and TA are substantially equivalent. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 APR 1 9 2012 Nobles Medical Technologies, Inc. % Mr. Anthony Nobles 17080 Newhope Street Fountain Valley, California 92708 Re: K113494 Trade/Device Name: NobleStitch™ EL, Vascular Suturing Device NobleStitch™ MR. Vascular Suturing Device NobleStitch™ TA, Vascular Suturing Device Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: II Product Code: GAW Dated: March 14, 2012 Received: March 22, 2012 Dear Mr. Nobles: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ ### Page 2 - Mr. Anthony Nobles or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable; the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Mark A. Milliman Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number: (if known) Device Name: NobleStitch™ EL, Vascular Suturing Device NobleStitch™ MR, Vascular Suturing Device NobleStitch™ TA, Vascular Suturing Device # Indications for Use: Indications for Use: The NobleStitch™ EL, MR, and TA is indicated for use in performing vascular stitching in general surgery, including endoscopic procedures. The NobleStitch™ EL, MR, and TA is not intended for blind vascular closure. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 C.F.R. 801.109) OR Over-The-Counter Use_ # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keane for MKM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113494
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...