K231183 · Shandong Haidike Medical Products Co., Ltd. · GAT · Aug 25, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231183
Device Name
Non absorbable Surgical Polyester Suture
Applicant
Shandong Haidike Medical Products Co., Ltd.
Product Code
GAT · General, Plastic Surgery
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Non-absorbable Surgical Polyester Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular, and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.
Device Story
Coated, braided, non-absorbable synthetic surgical suture composed of polyethylene terephthalate; dyed green with D&C Green 6; coated with bees wax. Available in USP sizes 6-0 through 2; lengths 45cm to 150cm; supplied with or without stainless steel needles. Used by clinicians for general soft tissue approximation and ligation. Device provides short-term wound support (7-10 days) and is removed within 10 days. Sterile, single-use device.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1:2018, including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity, and implantation. Performance testing verified compliance with USP <861> (diameter), USP <871> (needle attachment), and USP <881> (tensile strength).
Indicated for general soft tissue approximation and/or ligation in patients requiring short-term wound support (7-10 days). Contraindicated for ophthalmic, cardiovascular, and neurological procedures. Maximum duration of use is 10 days.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 25, 2023
Shandong Haidike Medical Products Co.,Ltd. Yan Wang Registration Manager Tianfu Road, Dongcheng District, Shan County Heze. Shandong 274300 China
Re: K231183
Trade/Device Name: Non absorbable Surgical Polyester Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: July 25, 2023 Received: July 25, 2023
Dear Yan Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N. Digitally signed by Lamichhane Tek N. Lamichhane -S.
Lamichhane Date: 2023.08.25 -5 10:53:57 -04'00' Tek N. Lamichhane, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231183
Device Name Non-absorbable Surgical Polyester Suture
### Indications for Use (Describe)
Non-absorbable Surgical Polyester Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular, and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary-K231183
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
#### 1. Administrative Information
Date of preparation: 07/24/2023
Shandong Haidike Medical Product Co., Ltd.
Tianfu Road, Dongcheng District, Shan County, 274300 Heze City, Shandong Province, China.
Establishment Registration Number: 3016426842
Contact Person: Yan Wang Position: Registration Manager Tel: +86-530-4660062 Fax: +86-530-4660055 Email: registration01(@suturescn.com
#### Identification of Subject Device 2.
Trade Name: Non-absorbable Surgical Polyester Suture Common Name: Nonabsorbable Poly (Ethylene Terephthalate) Suture
Regulatory Information Classification Name: Suture, Nonabsorbable, Synthetic, Polyester Classification: II Product Code: GAT Regulation Number: 21CFR 878.5000 Review Panel: General & Plastic Surgery
## 3. Identification of Predicate Device
510(k) Number: K172149 Regulation Number: 21CFR 878.5000 Classification: II Product Code: GAT Review Panel: General & Plastic Surgery Product name: MERICRON XL - Non-Absorbable Polyester Surgical Suture
## 4. Device Description
The subject device is a coated, braided, non-absorbable synthetic surgical suture composed of polyethylene terephthalate which is supplied sterile. The suture is coated with bees wax and dyed green. The color additive is D&C Green 6.
The subject device will be offered in diameters ranging from USP size 6-0 through 2 and available in length varying from 45cm to 150cm with or without needles attached.
{5}------------------------------------------------
## 5. Indications for Use:
Non-absorbable Surgical Polyester Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular, and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.
## 6. Summary of Technological Characteristics
| ITEM | Subject Device<br>(K231183) | Predicate Device<br>K172149 | Remark |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Product Code | GAT | |
| Regulation Number | 21CFR 878.5000 | 21CFR 878.5000 | Same |
| Class | II | II | Same |
| Sterile | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Same |
| Indication for Use | Non absorbable Surgical<br>Polyester Suture is indicated for<br>use in general soft tissue<br>approximation and/or ligation, but<br>not for use in ophthalmic,<br>cardiovascular, and neurological<br>procedures. The device is limited<br>to use where short term wound<br>support (7-10 days) is required<br>and can be left in place for a<br>maximum of 10 days. | MERICRON XL suture is<br>intended for use in general soft<br>tissue approximation and/or<br>ligation including<br>cardiovascular surgery,<br>neurosurgery, and ophthalmic<br>procedures. | Analysis 1 |
| | Configuration | Polyester Suture with or without<br>needle | 1、MERICRON XLTM Suture<br>with or without needle.<br>2、MERICRON XLTM Suture is<br>available with or without<br>PTFE<br>(Polytetrafluoroethylene)<br>Pledget |
| Suture | | | |
| Material | polyethylene terephthalate | poly (ethylene terephthalate) | Same |
| Structure | braided | braided | Same |
| Coating | Bees wax | Bees wax | Same |
| Dyed. Un-dyed | Dyed | Undyed/Dyed | Same |
| Colorant | D&C Green 6 | D & C Green No.6 | same |
| Length | 45cm to 150cm | Unknown | |
| Diameter | 6-0 through 2 | 6-0 through 5 | Analysis 3 |
| Needle | | | |
| Material | Stainless Steel | Stainless Steel | Same |
| Performance Test | | | |
| Diameter of suture | Comply with USP <861> | | |
| Needle Attachment | Comply with USP <871> | All characteristics meet USP<br>Requirement | |
| Tensile Strength | Comply with USP <881> | | Same |
| Length | Not less than 95.0% of the length<br>stated on the label | Unknown | Analysis 3 |
| Biocompatibility | | | |
| Cytotoxicity (ISO<br>10993-5) | Pass | The Surgical Suture has been<br>Evaluated:<br>a.In Vitro Cytotoxicity Study<br>b. Skin Sensitization Study<br>c.Intracutaneous Reactivity Test<br>d.Acute Systemic Toxicity<br>Study<br>e.Sub Chronic Toxicity Study<br>f. Intramuscular Implantation<br>Test<br>g. Bacterial Reverse Mutation<br>Test<br>h. Mammalian Erythrocyte<br>Micronucleus Test<br>i. In Vitro Hemolysis Test<br>j. Pyrogen Test | |
| Sensitization (ISO<br>10993- 10) | Pass | | |
| Intracutaneous<br>Reactivity (ISO<br>10993- 10) | Pass | | |
| Acute systemic<br>toxicity (ISO<br>10993- 11) | Pass | | |
| Pyrogen (USP 43<br>NF38 <151>) | Pass | | |
| Subacute<br>Systemic Toxicity<br>(ISO 10993- 11) | Pass | | Analysis 4 |
| Bacterial Reverse<br>Mutation (ISO<br>10993-3) | Pass | | |
| Chromosome<br>Aberration (ISO<br>10993-3) | Pass | | |
| Gene Mutation (ISO<br>10993-3) | Pass | | |
| Implantation (ISO<br>10993-6) | Pass | | |
| Hemolysis ( ASTM<br>F756-17) | Pass | | |
Table 1: Comparison of Technological Characteristics
{6}------------------------------------------------
Analysis 1-Indications for Use
The indications for use for the subject device is not exactly the same as the predicate device. The subject device is indicated for general tissue approximation but not for use in ophthalmic, cardiovascular, and neurological procedures. In addition, the longest duration of use for the subject device is up to 10 days,
{7}------------------------------------------------
while the predicate device is a permanent contact device per the contact duration. However, the biocompatibility tests have been conducted on the subject device, and the test results showed that the material and colorant of the subject device will not have any adverse effects when used for up to 10 days.
#### Analysis 2-Configuration
The configuration of the subject device is different from the predicate device. However, the configuration of Polyester Suture with or without a needle is the same as the predicate device.
#### Analysis 3-Length & Diameter
The length and diameter of the subject device is different from the predicate device. However, the length and diameter of the proposed device is within the range of that of the predicate device. In addition, the performance test about the length and diameter has been conducted on the subject device and the test result shows that the length and diameter of the subject device met the acceptance criteria. Therefore, the difference will not affect the safety and effectiveness of the subject device.
#### Analysis 4-Biocompatibility
The subject device is a prolonged contact device, while the predicate device is a permanent contact device. Based on the identified contact duration and indication for use of subject device, the biocompatibility tests were performed in accordance with ISO 10993- 1: 2018 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, and FDA guidance document entitled Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020.
Although the Biocompatibility contact duration is not the biocompatibility test has been conducted on the subject device and the test result showed that the material of the subject device will not have adverse effects.
## 7. Performance Data
#### Non-clinical Testing a.
The biocompatibility evaluation for the subject device was conducted in accordance with the ISO 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process".
The tests were conducted following these standards:
- > ISO 10993-3:2014 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- > ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity
- > ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects
- A after implantation
- > ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
- > ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin Sensitization
- > ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic
{8}------------------------------------------------
Toxicity
- > USP<151> Pyrogen Test (USP Rabbit Test)
- > ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the subject device. The test results demonstrated that the subject device complies with the following standards:
- > ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ハ ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- > USP<85>Bacterial Endotoxins Test
- > USP<861>Sutures - Diameter
- > USP<871>Sutures - Needle Attachment
- > USP<881>Tensile Strength
## c. Animal Study
No animal study is included in this submission.
## d. Clinical Test Conclusion
No clinical study is included in this submission.
## 7. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device, Non-absorbable Surgical Polyester Suture (K231183), is as safe, as effective, and performs as well as the legally marketed predicate device, K172149.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.