K223500 · Arcuro Medical , Ltd. · GAT · Aug 10, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223500
Device Name
SuperBall Meniscal Repair System
Applicant
Arcuro Medical , Ltd.
Product Code
GAT · General, Plastic Surgery
Decision Date
Aug 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures such as meniscal repair procedures.
Device Story
SuperBall™ 18ºUp, 24ºUp, and 12ºReverse are all-inside, all-suture meniscal repair systems; used for percutaneous or endoscopic soft tissue repair. System consists of two non-absorbable polyester/UHMWPE suture implants and a securing element preloaded in a curved needle delivery system. Operated by surgeons in clinical settings. Surgeon uses delivery system to place implants into meniscal tissue; sequential deployment mechanism secures the repair. Device provides mechanical fixation for meniscal tears; allows suture cutting to length. Benefits include minimally invasive repair of meniscal tissue, potentially reducing recovery time compared to open procedures.
Clinical Evidence
Bench testing only. Evaluations included functional sequential operation, deployment force characterization, USP-881 tensile strength (knot-pull), USP-871 needle attachment force, cyclic load pullout testing, ISO 10555-1:2013 corrosion resistance, and bond strength testing. Results confirmed performance equivalent to predicate and compliance with predefined specifications.
Technological Characteristics
All-inside suture retention system; materials: polyester and UHMWPE. Delivery system includes stainless steel components (corrosion tested per ISO 10555-1:2013). Features integrated depth limiter (10-18mm). Sterile (EtO), single-use. Packaging: PET-G blister tray with Tyvek® pouch. Manufacturing utilizes injection molding.
Indications for Use
Indicated for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures, specifically meniscal repair procedures in the lateral and medial meniscus (outer 2/3 region).
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
August 10, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Arcuro Medical Ltd. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103
Re: K223500
Trade/Device Name: SuperBall™ 18° Up, "24° Up" and "12° Reverse" Meniscal Repair Systems Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: July 11, 2023 Received: July 11, 2023
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir -S
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 510(k) Number (if known) K223500
Device Name
SuperBall™ "18º-Up", "24º-Up" and "12º-Reverse" Meniscal Repair System
#### Indications for Use (Describe)
The SuperBall Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures such as meniscal repair procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223500 510(k) SUMMARY
# Arcuro Medical Ltd.'s SuperBall™ 18º Up, "24º Up" and "12º Reverse" Meniscal Repair Systems
#### Submitter:
Arcuro Medical Ltd. 17 Tchelet Street Misgav 2017400 Israel Contact: Lee Ranon, CEO (972) 72 260-7000 lee.ranon@arcuromedical.com
#### Regulatory Correspondent:
Janice Hogan Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, PA 19102 (T) (267) 675-4611 (F) (267) 675-4601 Janice.hogan@hoganlovells.com
Date Prepared: 11-July-2023
Name of Device: SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" Meniscal Repair Systems
Common or Usual Name: Suture Retention Device
Classification Name per 21CFR878.5000: Nonabsorbable poly(ethylene terephthalate) surgical suture.
Regulatory Class: II
Product Code: GAT
Predicate Device: Arcuro Medical Ltd., SuperBall™ Meniscal Repair System (K180191)
{4}------------------------------------------------
## Device Description
The SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" Meniscal Repair System(s) is a suture retention device comprised of two non-absorbable, soft suture implants along with a SuperBall securing element preloaded within a curved needle delivery system. The SuperBall System is provided sterile for single use only.
The SuperBall implants and sutures are composed of polyester and ultra-high molecular weight polyethylene. The system allows for repair procedures in the lateral and medial meniscus, located within the outer 2/3 region of the meniscal zone (i.e., medial and posterior).
## Intended Use / Indications for Use
The SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures such as meniscal repair procedures.
## Technological Characteristics
The SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" Meniscal Repair Systems have near identical technological characteristics to the predicate device. While there are some differences in the technological characteristics of the devices related specifically to ergonomic design and manufacturing processes, these differences do not raise different questions of safety or effectiveness. For example, there are minor differences in materials between the subject devices and the predicate device utilizes polymeric materials used in 3D printing and manual grinding technologies; the subject devices are manufactured utilizing materials that are ideal for injection molding technologies. In addition, the some of the materials comprising the subject devices' implant have been sourced from different suppliers when compared to the predicate device. Therefore, while there are differences in materials, these differences do not raise new types of safety or effectiveness questions. In both cases, the key questions are whether the implant provides sufficient strength for its intended use and whether it is biocompatible for implantation. Performance testing and biocompatibility testing confirm that the SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" materials are appropriate for their intended use, and the differences from the predicate do not adversely impact performance.
Similarly, although there are minor differences in design and dimensions between the devices, specifically with the delivery system needle angulation, these also do not raise any new or different questions of safety or effectiveness. All devices are "all inside" designs. All system utilizes two implants that are different in length and width. The subject devices' implants are identical in size and are within the range of the predicate implants' size. All devices also allow the sutures to be cut to length. In all cases, the question of whether the dimensions are appropriate to perform the desired repair remains the same, and performance testing demonstrates that the dimensions are appropriate for the indications for use.
{5}------------------------------------------------
A table comparing the key features of the subject and predicate devices is provided below.
| | SuperBall™ (K180191) | SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | The SuperBall Meniscal Repair System is an all-inside, all-suture meniscal repair device. Each device includes two non-absorbable, soft suture implants preloaded within a curved needle delivery system along with the SuperBall securing element. The SuperBall System is provided sterile for single use only. | The SuperBall Meniscal Repair System is an all-inside, all-suture meniscal repair device. Each device includes two non-absorbable, soft suture implants preloaded within a curved needle delivery system along with the SuperBall securing element. The SuperBall System is provided sterile for single use only. |
| Intended Use/Indications for Use | The SuperBall Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures such as meniscal repair procedures. | The SuperBall Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures such as meniscal repair procedures. |
| Implant Size | L5.60mm x 1.30mm | L5.60mm x 1.30mm |
| Securing Element | One-way self-locking suture implant (SuperBall) securing element tied with 2 suture bundle implants | One-way self-locking suture implant (SuperBall) securing element tied with 2 suture bundle implants |
| Suture Size | Multi strand composition (incorporating suture diameters ranging between USP 2-0 -1) | Multi strand composition (incorporating suture diameters ranging between USP 2-0 -1) |
| Delivery | Sequential delivery using a designated curved needle delivery system and related accessories | Sequential delivery using a designated curved needle delivery system and related accessories |
| Depth Limiter | Integrated with delivery system | Integrated with delivery system |
| Depth Adjustment limiter | 10mm-18mm from limiter to the tip of the needle | 10mm-18mm from limiter to the tip of the needle |
| How Provided | Sterile by EtO, Single Use Only | Sterile by EtO, Single Use Only |
{6}------------------------------------------------
## Performance Data
Comprehensive bench testing has been performed to confirm that the SuperBall™ 188Up, "24ºUp" and "12ºReverse" have appropriate strength for the stated intended use and perform in an equivalent manner to the predicate. These tests include the following:
- . Functional evaluation - Conducted for the verification of the sequential operation of the subject device(s) per the supplied IFU and the proposed mode of operation. These evaluations were conducted upon the subject devices as they were operated throughout routine, sequential operation that results in the deployment of the SuperBall implant in a meniscal simulating material.
- Evaluation of deployment force – Force(s) characterization tests determined the forces applied to the system throughout said sequential operation. These evaluations were conducted upon the subject devices as they were operated throughout routine, sequential operation that results in the deployment of the SuperBall implant in a meniscal simulating material. Each operating step has been evaluated using a designated force measuring gauge suitable for the applied force.
- . Suture knot-pull - Suture knot-pull was conducted in accordance with USP-881 – Tensile Strength for the suture strands in the implant. The results demonstrated compliance with the required limits on average knot-pull tensile strength.
- . Implant and suture detachment force - Implant and suture detachment force was conducted in accordance with USP-871 – Suture Needle Attachment for the suture strands comprising the SuperBall implant. The results demonstrated compliance with the required limits on Needle Attachment.
- . Implant pullout force following Cyclic load – Implant pullout following cyclic load evaluations were conducted to demonstrate that the subject device can withstand cyclical tensile forces and maintain a minimal displacement, as compared to the predicate device under identical loading conditions.
- . Corrosion resistance - A corrosion resistance test was conducted in accordance with ISO 10555-1:2013, Intravascular catheters - Sterile and single-use catheters, Annex A. The results demonstrated that the devices' stainless steel components are corrosion resistant.
- . Bond strength - Bonded components detachment tests were conducted to verify that the bonded elements of the subject device(s) delivery system(s) can withstand forces greater than those that may be experienced during clinical use and as derived from the functional evaluations conducted, i.e., the forces required for the sequential operation of the system.
All tests, including those performed on aged product supporting the shelf life period for the subject device, confirm that the device performs as intended and has the appropriate physical characteristics and strength for its intended use.
The SuperBall™ "18-Up", "24º-Up" and "12º-Reverse" devices shall be provided sterile by EtO for single use; packaged in a rigid PET-G Blister tray and lid sealed within a Tyvek® pouch.
Validation processes conducted indicated qualified the subject devices' stability throughout transportation and shelf life (storage), verifying that the system complies with predefined specifications and acceptance criteria following a shelf-life period of 2 years (24 months).
{7}------------------------------------------------
## Conclusions
The SuperBall™ "18ºUp", "24ºUp" and "12ºReverse" are as safe and effective as the SuperBall™ device(s). Said share the same intended uses and identical indications, technological characteristics, and principles of operation as its predicate device. The minor differences between the SuperBall™ 189Up, "249Up" and "12ºReverse" and their predicate device do not alter the intended surgical use of the device and do not raise different questions of safety or effectiveness. Thus, the SuperBall™ 18ºUp, "24ºUp" and "12ºReverse" are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.