K203186 · Medos International SARL · GAT · Mar 18, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203186
Device Name
DYNATAPE Suture
Applicant
Medos International SARL
Product Code
GAT · General, Plastic Surgery
Decision Date
Mar 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DYNATAPE suture is indicated for orthopedic procedural use in soft-tissue approximation, and or ligation, including use with allograft tissue.
Device Story
DYNATAPE Suture is a sterile, non-absorbable orthopedic suture featuring a double sheath and core design. It is composed of Ultra High Molecular Weight Polyethylene (UHMWPE) and Polyester, with a Silicone/NaCl core. The outer sheath is braided flat to increase tissue contact area. When placed in an aqueous environment, salt particles in the silicone core elute, creating a micro-porous structure that hydrates and expands radially. This radial expansion causes axial shortening of the suture, providing a self-tensioning property designed to resist laxity and minimize gap formation in soft-tissue repairs. The device is used by surgeons in orthopedic procedures. It functions as a mechanical fixation aid to maintain approximation force on tissue. Benefits include improved maintenance of tissue approximation and reduced risk of repair laxity.
Clinical Evidence
Bench testing only. Testing included USP knot tensile strength, in-vitro approximation force over time, and sterility testing. Biocompatibility, stability, and in-vivo testing were leveraged from the predicate device (K181182). Bacterial endotoxin testing was completed as part of K200949.
Technological Characteristics
Materials: Ultra High Molecular Weight Polyethylene (UHMWPE), Polyester, Silicone, NaCl. Design: Double sheath and core; flat braided outer sheath. Dimensions: Meets USP knot tensile strength for size 2; does not follow USP diameter requirements due to flat geometry. Principle: Self-tensioning via radial expansion of hydrated silicone core. Sterilization: Sterile.
Indications for Use
Indicated for orthopedic soft-tissue approximation and/or ligation, including use with allograft tissue.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2021
Medos International SARL % Tatyana Korsunsky Regulatory Affairs Technical Manager DePuy Mitek, a Johnson and Johnson company 325 Paramount Drive Raynham, Massachusetts 02767
Re: K203186
Trade/Device Name: DYNATAPE Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: October 25, 2020 Received: October 27, 2020
Dear Tatyana Korsunsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203186
Device Name DYNATAPETM Suture
Indications for Use (Describe)
DYNATAPE suture is indicated for orthopedic procedural use in soft-tissue approximation, and or ligation, including use with allograft tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
### DYNATAPE™ Suture
Date Prepared: February 09, 2021
| Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Tatyana Korsunsky<br>Regulatory Affairs Technical Manager<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>Telephone: 508-828-3122<br>e-mail: tkorsuns@its.jnj.com |
| Name of<br>Medical Device | Proprietary Name: DYNATAPE™ Suture<br><br>Classification Name: Polyethylene, suture, nonabsorbable, synthetic (21 CFR 878.5000)<br><br>Common Name: Suture |
| Substantial<br>Equivalence | The DYNATAPE™ Suture is substantially equivalent to:<br>K181182 DYNACORD™ Suture (MEDOS International SARL) Reference devices: K200949 HEALIX ADVANCE™ Anchor with DYNATAPE™ Suture (MEDOS International SARL) K183506 HEALIX ADVANCE™ Anchor with DYNACORD™ Suture (MEDOS International SARL) K021434, K041553 FiberWire® (Arthrex) |
| Device<br>Classification | DYNATAPETM Suture is classified as: |
| | Polyethylene, suture, nonabsorbable, synthetic, classified as Class II, product code<br>GAT, regulated under 21 CFR 878.5000. |
| Device<br>Description | DYNATAPE™ Suture is a sterile non-absorbable orthopedic suture. DYNATAPE™<br>Suture has a double sheath and core design, composed of Ultra High Molecular<br>Weight Polyethylene and Polyester, with or without colorants, as well as a<br>Silicone/NaCl core. DYNATAPE™ Suture's outer sheath is braided flat, with<br>rounded tips. |
| Technological<br>Characteristics | The DYNATAPE™ Suture is a surgical suture, that meets USP knot tensile strength<br>for size 2 non-absorbable sutures. As DYNATAPE™ Suture is flat, it does not follow<br>USP suture diameter requirements. DYNATAPE™ Suture's outer sheath is braided<br>flat to allow for a wider contact area with tissue. |
| | When DYNATAPE™ Suture is placed in an aqueous environment, the salt particles<br>within the silicone core elute out, leaving behind a micro-porous structure within the<br>silicone core. These small voids are consequently filled with surrounding fluid as the<br>core hydrates, resulting in a radial expansion of the suture. If laxity is present, this<br>radial expansion of the braid causes an axial shortening of the total suture length.<br>The self-tensioning property of DYNATAPE™ Suture is designed to resist laxity and<br>minimize gap formation in a repair, by maintaining approximation force<br>(compression) if laxity (looseness) is present. |
| Comparison to<br>the Predicate<br>Devices | The DYNATAPE™ Suture, similar to the predicate device DYNACORD®<br>(K181182), meets USP knot tensile requirements for size 2 suture. Both subject and<br>predicate device have the same indications, intended use and materials. Both<br>DYNATAPE™ and DYNACORE™ sutures are laxity resistant, as shown by<br>approximation force testing.<br>The sheath of the DYNATAPE™ Suture has been redesigned with a unique braiding<br>pattern of ticker fibers to create the flattened structure specific to DYNATAPE™ |
| | Suture, differentiating from the predicate device.<br>DYNATAPE™ Suture is offered as a suture component of the reference predicate<br>HEALIX ADVANCE™ Anchor with DYNATAPE™ Suture (K200949). |
| Indications for<br>Use | DYNATAPE suture is indicated for orthopedic procedural use in soft-tissue<br>approximation, and/ or ligation, including use with allograft tissue. |
| Non clinical<br>Testing | USP Knot Tensile Strength, <i>in-vitro</i> approximation force over time, and sterility<br>testing have been conducted. Biocompatibility, stability, <i>in-vivo</i> testing of the<br>predicate device were included by reference (K181182). Bacterial endotoxin testing |
{5}------------------------------------------------
{6}------------------------------------------------
| | has been completed as part of the HEALIX ADVANCE™ Anchor with DYNATAPE™ Suture (K200949), meeting the endotoxin limits. |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety and Performance | Results of non-clinical testing have demonstrated that the proposed devices are suitable for their intended use. Based on similarities in the indications for use, technological characteristics, and performance in comparison to the predicate device, the proposed DYNATAPE™ Suture has shown to be substantially equivalent to the predicate device under the Federal Food, Drug and Cosmetic Act. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.