← Product Code [GAT](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT) · K181772

# NovoStitch Pro Meniscal Repair System, CTX-A004 (size 0) (K181772)

_Ceterix Orthopaedics, Inc. · GAT · Aug 1, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K181772

## Device Facts

- **Applicant:** Ceterix Orthopaedics, Inc.
- **Product Code:** [GAT](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT.md)
- **Decision Date:** Aug 1, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.5000
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The NovoStitch Pro Mensical Repair System is intended for approximation of soft tissue in meniscal repair procedures.

## Device Story

Handheld, disposable arthroscopic surgical instrument; used by orthopedic surgeons for meniscal repair. System consists of delivery handle and suture cartridges. Surgeon manually operates device to pass size 0 braided, nonabsorbable polyethylene suture through soft tissue. Device retains tissue via upper and lower jaws; delivers up to six sutures per procedure. Output is placed suture for tissue approximation; facilitates healing by securing meniscal tears. Benefits include reliable, minimally invasive suture delivery during arthroscopic procedures.

## Clinical Evidence

Bench testing only. Simulated use environment testing validated device strength, reliability, insertion, suture deployment, and removal. Biocompatibility testing conducted per ISO 10993-1; suture met requirements for permanent implants, and handle components met requirements for limited duration contact.

## Technological Characteristics

Handheld, manually operated arthroscopic suture placement system. Materials: ABS/PC blend handle, stainless steel distal end, Nitinol needle, polyethylene nonabsorbable suture (size 0). Single-patient use, disposable. Sterilization: Ethylene oxide. Connectivity: None.

## Regulatory Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- NovoStitch Pro Meniscal Repair System, CTX-A003 (size 2-0) ([K180531](/device/K180531.md))

## Reference Devices

- Force Fiber® Green Ultra High Molecular Weight Polyethylene Non-Absorbable Surgical Suture ([K100506](/device/K100506.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 1, 2018

Ceterix Orthopaedics, Inc. % Mr. Tarhan Kayihan Director of Quality and Regulatory 6500 Kaiser Drive. Suite 120 Fremont, California 94555

# Re: K181772

Trade/Device Name: NovoStitch Pro Meniscal Repair System, CTX-A004 (size 0) Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: July 2, 2018 Received: July 3, 2018

Dear Mr. Kayihan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

#### K181772

Device Name

NovoStitch Pro Mensical Repair System, CTX-A004 (size 0)

Indications for Use (Describe)

The NovoStitch Pro Mensical Repair System is intended for approximation of soft tissue in meniscal repair procedures.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K181772 P. 1 of 3" on the left side. To the right of the text is an orange graphic and the word "CETERIX" in red, with the word "ORTHOPAEDICS" in orange below it. The image appears to be a logo or header for a document.

# 510(K) SUMMARY

### Manufacturer and Submitter

| Company Name:    | Ceterix Orthopaedics, Inc.                                                                                                       |
|------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | 6500 Kaiser Drive<br>Suite 120<br>Fremont, CA 94555<br>TEL: (650) 316-8660<br>FAX: (650) 618-2779<br>Email: tkayihan@ceterix.com |
| Contact Person:  | Tarhan Kayihan                                                                                                                   |

### Device Name and Classification

| Trade/Proprietary Name: | NovoStitch® Pro Meniscal Repair System, CTX-A004 (size 0)                                     |
|-------------------------|-----------------------------------------------------------------------------------------------|
| Common Name:            | Suture, Nonabsorbable, Synthetic, Polyethylene                                                |
| Classification Name:    | Suture, Nonabsorbable, Synthetic, Polyethylene                                                |
| Regulation Number:      | 21 CFR 878.5000                                                                               |
| Regulatory Class:       | II                                                                                            |
| Product Code:           | GAT                                                                                           |
| Panel:                  | General & Plastic Surgery                                                                     |
| Predicate Device        |                                                                                               |
| Device Name:            | NovoStitch Pro Meniscal Repair System, CTX-A003 (size 2-0)                                    |
| 510(k) Number:          | K180531                                                                                       |
| Applicant:              | Ceterix Orthopaedics                                                                          |
| Reference Device        |                                                                                               |
| Device Name:            | Force Fiber® Green Ultra High Molecular Weight<br>Polyethylene Non-Absorbable Surgical Suture |
| 510(k) Number:          | K100506                                                                                       |
| Applicant:              | Teleflex Medical                                                                              |

### Device Description

The NovoStitch Pro Meniscal Repair System, size 0 is a new configuration of the predicate device. While the predicate device passes size 2-0 suture, this new configuration passes size 0 braided, nonabsorbable, surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with suture are attached.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of an abstract orange shape on the left, followed by the word "CETERIX" in red, and the word "ORTHOPAEDICS" in a smaller, lighter orange font below it. The logo is clean and modern, with a focus on the company name.

# Intended Use/Indication for Use

The NovoStitch Pro Meniscal Repair System, size 0 is intended for approximation of soft tissue in meniscal repair procedures.

# Technological Characteristics

The NovoStitch Pro Meniscal Repair System, size 0 is a handheld, disposable surgical instrument coupled with an implantable suture for the approximation of soft tissue in meniscal repair procedures. The system is comprised of a delivery handle and a suture cartridge, both of which are provided sterile.

| Characteristic      | NovoStitch Pro Meniscal Repair System, CTX-A004 (size 0)                                  | NovoStitch Pro Meniscal Repair System, CTX-A003 (size2-0)                          |
|---------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| 510(k) #            | K181772                                                                                   | K180531                                                                            |
| Method of Use       | Handheld, manually operated, single procedure arthroscopic suture placement system        | Handheld, manually operated, single procedure arthroscopic suture placement system |
| Delivery Instrument | Identical to predicate device, NovoStitch Pro Meniscal Repair System, CTX-A003 (size 2-0) | As described in K180531                                                            |
| Handle Material     | ABS/PC blend                                                                              | ABS/PC blend                                                                       |
| Distal end material | Stainless steel                                                                           | Stainless steel                                                                    |
| Usability           | Single-patient use, can deliver up to six sutures, disposable                             | Single-patient use, can deliver up to six sutures, disposable                      |
| Tissue interaction  | Tissue retained by upper and lower jaw                                                    | Tissue retained by upper and lower jaw                                             |
| Needle material     | Nitinol                                                                                   | Nitinol                                                                            |
| Suture material     | Polyethylene, nonabsorbable surgical suture, size 0, white with green co-braid            | Polyethylene, nonabsorbable surgical suture, size 2-0, white with blue co-braid    |
| Suture supplier     | Teleflex Medical                                                                          | Teleflex Medical                                                                   |
| Sterilization       | Ethylene oxide                                                                            | Ethylene oxide                                                                     |
| Knot tying method   | Manually tied knot                                                                        | Manually tied knot                                                                 |

#### Table 3-1. Comparison of technological characteristics to predicate device

### Performance Data

Product performance testing for the NovoStitch Pro Meniscal Repair System, size 0 was performed in a simulated use environment. Device strength, device reliability, device insertion, suture deployment, device removal, and the ability to reload the device with new suture cartridges to perform additional device insertions, deployments and removals were validated. Bench testing was performed to verify that the device and the implantable suture meet all pre-established acceptance criteria. The verification testing has demonstrated that the device can reliably deliver up to six stitches. This includes the preloaded handle assembly and up to five additional cartridges. The verification testing

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of an abstract graphic on the left, resembling an orange leaf with curved lines. To the right of the graphic is the word "CETERIX" in red, with the word "ORTHOPAEDICS" in a smaller, tan font underneath.

included device joint strength and device performance evaluations. Sterilization, packaging and shelf life were either validated or adopted from previous testing where appropriate. In addition, biocompatibility testing for each element of the NovoStitch Pro Meniscal Repair System, size 0 was conducted to demonstrate compliance with ISO 10993-1. The suture implant met all biocompatibility requirements for permanent implants in contact with tissue and/or bone. The patient contact portions of the delivery handle met all requirements for limited duration contact with tissue and/or bone.

### Conclusion

The NovoStitch Pro Meniscal Repair System, size 0 is substantially equivalent to the predicate device with respect to safety, effectiveness and reliability.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K181772](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K181772)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
