← Product Code [GAT](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT) · K171680

# Infinity-Lock 3, Infinity-Lock 5 (K171680)

_Xiros, Ltd. · GAT · Jun 30, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K171680

## Device Facts

- **Applicant:** Xiros, Ltd.
- **Product Code:** [GAT](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT.md)
- **Decision Date:** Jun 30, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.5000
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Infinity-Locks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.

## Device Story

Infinity-Lock 3 and 5 are non-absorbable, sterile, woven poly(ethylene terephthalate) surgical tapes; designed for soft tissue approximation, specifically Achilles tendon repair. Devices feature an integral loop for cow-hitch knotting, tapered ends for easier tissue passage, and green-striped tails for identification. Used by surgeons in clinical settings; tapes are sutured through tendon ends (e.g., Bunnell technique) to draw separated ends together, facilitating healing. Devices are single-use implants. Clinical benefit derived from improved usability features compared to predicate, aiding surgical workflow without altering fundamental mechanical performance or safety profile.

## Clinical Evidence

Bench testing and cadaveric validation only. Mechanical testing (ultimate tensile strength) performed on knotted and unknotted samples per internal method QT010 and USP 39 NF33:2016. Clinician validation using cadaveric materials confirmed suitability for intended use. No clinical trial data presented.

## Technological Characteristics

Material: Poly(ethylene terephthalate) (polyester) fibers. Construction: Woven textile tape. Form factor: Two sizes (3 and 5) with integral loop, tapered ends, and green identification stripe. Sterilization: Sterile. Performance: Complies with USP 39 NF33:2016 requirements for non-absorbable surgical sutures.

## Regulatory Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Xiros, Ltd, Poly-Tapes ([K002172](/device/K002172.md))

## Reference Devices

- Teleflex Medical Inc., polyethylene terephthalate suture (Cottony Il green suture polyester) ([K021019](/device/K021019.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 30, 2017

Xiros, Ltd Mr. Stephen Seed Springfield House Whitehouse Lane, Leeds LS19 7UE England

Re: K171680

Trade/Device Name: Infinity-Lock 3, Infinity-Lock 5 Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(ethylene terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: June 2, 2017 Received: June 6, 2017

Dear Mr. Seed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K171680

Device Name Infinity-Lock 3 Infinity-Lock 5

#### Indications for Use (Describe)

The Infinity-Looks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of the Achilles tendon.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is colored in gray. The background is white. There is a small line extending from the right side of the rounded rectangle.

#### SECTION 7. Special 510(k) SUMMARY

In accordance with 21 CFR 807.92, the following information constitutes the Special 510(k) summary for the Infinity-Locks 3 and 5.

## 1.SUBMITTER INFORMATION

| Submitter`s Name: | Xiros Limited                                                     |
|-------------------|-------------------------------------------------------------------|
| Address:          | Springfield House<br>Whitehouse Lane<br>Leeds LS19 7UE<br>England |
| Telephone Number: | +44 (0)1132387200                                                 |
| Fax Number:       | +44 (0)113 2387201                                                |
| Contact Person 1: | Stephen Seed (Compliance Director)                                |
| Contact Person 2: | Manuela Rodrigues (Principal Regulatory Affairs Associate)        |
| Date:             | 02 May 2017                                                       |

### II. DEVICE IDENTIFICATION

| Name of Device(s):    | Infinity-Lock 3<br>Infinity-Lock 5              |
|-----------------------|-------------------------------------------------|
| Trade Name(s):        | Infinity-Lock 3<br>Infinity-Lock 5              |
| Common or usual name: | Non-absorbable Polyester Surgical Tape          |
| Classification Name:  | Suture, Non-absorbable, Synthetic, Polyethylene |
| Regulatory Class:     | Class II                                        |
| Product Code:         | GAT                                             |
| Panel:                | General & Plastic Surgery                       |

### III. PREDICATE/UNMODIFIED DEVICE

Predicate: Xiros, Ltd, Poly-Tapes K002172.

This predicate has not been subject to a design related recall.

Reference device: Teleflex Medical Inc., polyethylene terephthalate suture (Cottony Il green suture polyester) K021019.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is a light gray color. The rounded rectangle has a thick, light gray border. The background is white.

## IV. DEVICE DESCRIPTION

The Infinity-Locks are non-absorbable, sterile, poly(ethylene terephthalate) sutures. They are prepared from fibres of high molecular weight, long chain, linear polyesters having recurrent aromatic rings as an integral component. Infinity-Locks differ from USP Sutures in being in the form of woven tapes and in exceeding all USP sizes.

The product comprises two sizes of textile implant based on the Neoligaments poly-tape design. The tapes have an integral loop at one end which are used to tie a cowhitch through one end of a ruptured tendon. The tapes have two tails which are then sutured through the separated end (using methods such as the Bunnell technique), thus drawing the two ends together in order to facilitate healing of the tear. One of the tails has a green stripe to aid identification and both taper into a cord to assist with passing through soft tissues.

#### V. INTENDED USE

The Infinity-Locks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.

The device subject to modification (Poly-Tapes) has exactly the same intended use as the modified device. The Poly-Tapes are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.

### VI. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The substantial equivalence of the Infinity-Locks is supported by the fact that they have the same intended use and they have common technological characteristics to the Poly-Tapes. The same material is used in the manufacture of the proposed modified device (both are made of polyester). Both Poly-Tapes and Infinity-Locks have the same construction; they are woven from polyester, and have the same performance characteristics in accordance with USP requirements.

The modified device retains all the features of the original poly-tapes but includes some new ones to aid usability. These are the: -

- . Tapered ends that are intended to aid threading which was previously achieved by adding sutures in theatre.
- . Tails that are marked with green weft to aid identification, previously both were plain.
- Addition of an integral loop to make it easier for surgeons to tie a cow hitch knot . when required. The previous device was un-looped and the modified device can be simply un-looped when required.

These minor differences in the design of the Infinity-Lock variants do not bring any new questions of safety and effectiveness in accordance with performance data hence do not constitute a change in the fundamental technological characteristics.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is gray. The background is white. The rounded rectangle has a thick border.

### VII. PERFORMANCE DATA

Mechanical testing was performed on the modified device and original poly-tape in both knotted and unknotted conditions to determine the ultimate tensile strengths using an internal method, QT010, and following USP 39 NF33:2016 for non-absorbable surgical sutures, where applicable. The modified devices were found to be equivalent or superior in all cases.

Validation was also undertaken by clinicians using cadaveric materials to demonstrate the suitability of the modified device for the intended use which was found to be satisfactory.

The performance data benefit/risk analysis concluded that the differences encountered do not affect the safety and efficacy of the new device variants in relation to the predicate.

Summary: Based on the pre-clinical testing performance, the Infinity-Lock 3 and Infinity-Lock 5 are found to have the same safety and effectiveness profile to the predicate device (Poly-Tapes).

#### VIII.CONCLUSION

Xiros conclude the Infinity-Lock 3 and Infinity-Lock 5 are substantially equivalent to the Poly-Tapes.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K171680](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAT/K171680)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
