G-CATH TISSUE ANCHOR DELIVERY CATHETER

K102916 · Usgi Medical · GAT · Oct 20, 2010 · General, Plastic Surgery

Device Facts

Record IDK102916
Device NameG-CATH TISSUE ANCHOR DELIVERY CATHETER
ApplicantUsgi Medical
Product CodeGAT · General, Plastic Surgery
Decision DateOct 20, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.5000
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The g-Cath Tissue Anchor Deliver Catheter is intended for approximation of soft tissue in minimally invasive gastroenterology procedures, e.g. fistula closure, perforation/leak closure and repair of dilated gastric tissue.

Device Story

The g-Cath Tissue Anchor Delivery Catheter is a sterile, single-patient-use device designed for minimally invasive gastroenterology procedures. It contains a pair of tissue anchors made of nitinol, polyester, and titanium within its catheter lumen. During a procedure, the clinician deploys the anchor pair through the catheter to compress and approximate soft tissue. The device is used to facilitate fistula closure, perforation/leak closure, and the repair of dilated gastric tissue. It functions as a mechanical delivery system for tissue approximation.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Materials: nitinol, polyester, titanium. Form factor: catheter-based delivery system. Single-patient use. Sterile.

Indications for Use

Indicated for soft tissue approximation in minimally invasive gastroenterology procedures, including fistula closure, perforation/leak closure, and repair of dilated gastric tissue.

Regulatory Classification

Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102916 page 1/2 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92 #### 1. SUBMITTER INFORMATION OCT 2 0 2010 - a. Company Name: USGI Medical b. Company Address: - c. Telephone: Fax: d. Contact Person: 1140 Calle Cordillera San Clemente, CA 92673 (949) 369-3890 (949) 369-3891 October 21, 2010 Mary Lou Mooney Vice President of Clinical, Regulatory & Quality e. Date Summary Prepared: ### DEVICE IDENTIFICATION 2. a. Trade/Proprietary Name: g-Cath Tissue Anchor Delivery Catheter b. Common Name: c. Classification Name: Non-absorbable surgical suture Non-absorbable (PET surgical suture, 878.5000 #### 3. IDENTIFICATION OF PREDICATE DEVICES g-Cath Tissue Anchor Delivery Catheter USGI Medical (K061276, K100251) #### DESCRIPTION OF THE DEVICE 4. The g-Cath Tissue Anchor Delivery Catheter is a sterile, single patient use device that contains a nitinol/polyester/titanium tissue anchor pair within the catheter lumen. The anchor pair is deployed through the catheter lumen to compress and approximate tissue. {1}------------------------------------------------ K102916 USGI Medical Special 510(k) Device Modification USGI g-Cath Tissue Anchor Delivery Catheter #### ડ. STATEMENT OF INTENDED USE The g-Cath Tissue Anchor Deliver Catheter is intended for approximation of soft tissue in minimally invasive gastroenterology procedures, e.g. fistula closure, perforation/leak closure and repair of dilated gastric tissue. ## COMPARISON WITH PREDICATE DEVICES 6. The g-Cath Tissue Anchor Delivery Catheter is comparable to the predicate devices in terms of intended use, technology, and materials. Bench testing was conducted to ensure that the modified device performs as intended when used according to its instructions for use. ### 7. SUMMARY OF PERFORMANCE DATA Design control activities for the device modifications were completed in accordance with 21CFR 820.30 and USGI Medical documented design control procedures. Human factors considerations were also incorporated into the device modifications. Risk analysis was performed in accordance with USGI Medical procedures and ISO 14971 to identify any risks associated with the modifications. Bench testing confirmed that the device met its performance specifications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 USGI Medical % Ms. Mary Lou Mooney VP, Clinical, Regulatory & Quality 1140 Calle Cordillera San Clemente, California 92673 OCT 2 0 2010 Re: K102916 Trade/Device Name: g-Cath Tissue Anchor Delivery Catheter Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: II Product Code: GAT, GDW, HET Dated: September 30, 2010 Received: October 1, 2010 Dear Ms. Mooney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms. Mary Lou Mooney CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115800 htm fast the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollofre number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Malleus Mark N Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use OCT 2 0 2010 KIO2916 510(k) Number (if known): . Device Name: g-Cath Tissue Anchor Delivery Catheter # Indications For Use 510(k) Number (if known): The g-Cath Tissue Anchor Delivery Catheter is intended for approximation of soft tissue in minimally invasive gastroenterology procedures, e.g., fistula closure, perforation/leak closure and repair of dilated gastric tissue. V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Kramer her MIVIM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of __ 510(k) Number K102916
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...