← Product Code [GAR](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAR) · K220980

# Tissue Approximation System (TAS) (K220980)

_Tas Medical, Inc. · GAR · Jul 5, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAR/K220980

## Device Facts

- **Applicant:** Tas Medical, Inc.
- **Product Code:** [GAR](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAR.md)
- **Decision Date:** Jul 5, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5020
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.

## Device Story

TAS is a non-absorbable polyamide (nylon) zip-tie device with a braided polyethylene terephthalate (PET) leader; designed for soft tissue approximation without surgical knots. Device includes hand-held insertion instruments. Used by trained healthcare providers in clinical settings (e.g., general surgery). Operator inserts device; one-way clasp secures tissue approximation; replaces traditional knot-tying. Output is mechanical fixation of tissue. Benefits include simplified closure procedure and secure tension maintenance comparable to traditional sutures. Evaluated via large animal GLP study and human factors validation.

## Clinical Evidence

Large animal GLP-compliant study compared TAS to predicate in full-thickness midline abdominal wall incisions. Endpoints: incision healing, absence of serious tissue swelling, necrosis, dehiscence, hematoma, or extrusion. All safety/effectiveness endpoints met. Biocompatibility per ISO 10993; sterilization per ISO 11137 (SAL 10^-6, 25 kGy); 12-month shelf-life validated via accelerated aging. Human factors validation confirmed safe use by intended users.

## Technological Characteristics

Materials: Polyamide (nylon) zip-tie, braided polyethylene terephthalate (PET) leader. Form factor: Zip-tie with one-way clasp. Sterilization: Gamma irradiation (25 kGy, ISO 11137). Biocompatibility: ISO 10993. Performance standards: USP <881> (Tensile Strength), USP <871> (Sutures-Needle Attachment).

## Regulatory Identification

Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Sutumed Sutulon Suture ([K123034](/device/K123034.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 5, 2023

TAS Medical, Inc. % Maureen O'connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180

Re: K220980

Trade/Device Name: Tissue Approximation System (TAS) Regulation Number: 21 CFR 878.5020 Regulation Name: Nonabsorbable Polyamide Surgical Suture Regulatory Class: Class II Product Code: GAR Dated: March 8, 2023 Received: March 9, 2023

# Dear Maureen O'connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

David Krause, Ph.D. Deputy Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220980

Device Name Tissue Approximation System

#### Indications for Use (Describe)

The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.

| Type of Use (Select one or both, as applicable)                                                                                                                     |                                                                                                               |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <b>                         ☑                     </b>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# TAS Medical, Inc. Tissue Approximation System (TAS)

#### 510(k) Owner

TAS Medical 1100 Industrial Road, Suite 16 San Carlos, CA 94070

#### Submission Correspondent

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245

Date Prepared: June 28, 2023

Trade Name of Device Tissue Approximation System (TAS)

Common or Usual Name

Suture

#### Classification Name

Suture, Nonabsorbable, Synthetic, Polyamide 21 C.F.R. §878.5020 Class II Product Code: GAR

Predicate Device Sutumed Sutulon Suture cleared in K123034

#### Device Description

The Tissue Approximation System (TAS) consists of a polyamide (nylon) "zip-tie" attached to a braided polyethylene terephthalate (PET, or Dacron) leader to aid insertion. A one-way clasp allows for fixation of the tissue approximation without the need to tie surgical knots.

When packaged for commercial distribution, four zip-ties with integrated suture leaders are accompanied by hand-held instruments that can be optionally used for insertion. The kit will be provided sterile within a polyethylene/Tyvek pouch.

{4}------------------------------------------------

## Indications for Use

The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.

# Substantial Equivalence

The predicate device for TAS is the Sutumed Sutulon Suture cleared in K123034. Both devices are intended for use in soft tissue approximation and are prescription devices for use by trained healthcare providers. The TAS zip-tie shares important technological characteristics with the predicate device (both are non-absorbable nylon monofilament sutures), however, there are also some differences, such as the zip-tie having a rectangular cross-section, not a USP diameter, and the zip-tie using a clasp rather than a knot. The technological differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and the differences do not affect the safety and effectiveness of the device when used as labeled. Regarding fixation, the predicate device uses knots to maintain the tension set by the user. The subject device is maintained by the zip-tie's clasp. Both devices maintain the tension set by the user, thus they are substantially equivalent. Regarding cross-sectional shape, the primary function of the devices is to approximate soft tissue, which is not affected by cross-sectional shape. Both of these technological characteristic differences, as well as overall designs, were evaluated through performance testing, which demonstrated substantial equivalence.

## Performance Data

Performance data included a large animal. GLP compliant study comparing the closure of full thickness, midline abdominal wall incisions comparing TAS (Test Article) to the predicate device (Control Article) used as labeled (e.g., fixation methods, no surgical mesh). Study endpoints included comparable incision healing, absence of device-associated serious nontransient tissue swelling, necrosis, dehiscence, hematoma, or extrusion in the Test Article as compared to the Control Article. All predefined safety and effectiveness endpoints were met, demonstrating that the different technological characteristics do not affect the safety/effectiveness when used as labeled.

Biocompatibility testing was conducted for the TAS per ISO 10993 and demonstrated substantially equivalent safety. Sterilization processes were validated per ISO 11137, and sterility was validated to SAL 10° with a sterilization dose of 25 kGy by the validation method VDmax25. Individual device testing was conducted to demonstrate the ability of each device to perform its function. A shelf-life of 12 months was validated using accelerated aging for the devices, packaging, and the sterile barrier. All of the verification and validation testing was completed and passed for the Tissue Approximation System with 12-month shelf life.

Human factors validation testing was conducted which demonstrated that the device could be used by the intended users and under the expected use conditions without errors that could result in serious harm.

{5}------------------------------------------------

Testing based on the USP Monographs (including <881, Tensile Strength> and <871, Sutures -Needle Attachment>) showed that the TAS has a strength that meets or exceeds the requirement of USP Size 7, the length of the TAS zip-tie is no less than 95.0% of the length stated on the label and the secured TAS zip-tie is as secure or better than the predicate device tied in a standard square knot.

Therefore, the Tissue Approximation System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAR/K220980](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAR/K220980)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
