K220980 · Tas Medical, Inc. · GAR · Jul 5, 2023 · General, Plastic Surgery
Device Facts
Record ID
K220980
Device Name
Tissue Approximation System (TAS)
Applicant
Tas Medical, Inc.
Product Code
GAR · General, Plastic Surgery
Decision Date
Jul 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.
Device Story
TAS is a non-absorbable polyamide (nylon) zip-tie device with a braided polyethylene terephthalate (PET) leader; designed for soft tissue approximation without surgical knots. Device includes hand-held insertion instruments. Used by trained healthcare providers in clinical settings (e.g., general surgery). Operator inserts device; one-way clasp secures tissue approximation; replaces traditional knot-tying. Output is mechanical fixation of tissue. Benefits include simplified closure procedure and secure tension maintenance comparable to traditional sutures. Evaluated via large animal GLP study and human factors validation.
Clinical Evidence
Large animal GLP-compliant study compared TAS to predicate in full-thickness midline abdominal wall incisions. Endpoints: incision healing, absence of serious tissue swelling, necrosis, dehiscence, hematoma, or extrusion. All safety/effectiveness endpoints met. Biocompatibility per ISO 10993; sterilization per ISO 11137 (SAL 10^-6, 25 kGy); 12-month shelf-life validated via accelerated aging. Human factors validation confirmed safe use by intended users.
Technological Characteristics
Materials: Polyamide (nylon) zip-tie, braided polyethylene terephthalate (PET) leader. Form factor: Zip-tie with one-way clasp. Sterilization: Gamma irradiation (25 kGy, ISO 11137). Biocompatibility: ISO 10993. Performance standards: USP <881> (Tensile Strength), USP <871> (Sutures-Needle Attachment).
Indications for Use
Indicated for soft tissue approximation in general surgery, including minimally invasive ventral hernia and abdominal wall closure <10cm. Contraindicated for ophthalmic, neurologic, or cardiovascular procedures.
Regulatory Classification
Identification
Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 5, 2023
TAS Medical, Inc. % Maureen O'connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K220980
Trade/Device Name: Tissue Approximation System (TAS) Regulation Number: 21 CFR 878.5020 Regulation Name: Nonabsorbable Polyamide Surgical Suture Regulatory Class: Class II Product Code: GAR Dated: March 8, 2023 Received: March 9, 2023
# Dear Maureen O'connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
David Krause, Ph.D. Deputy Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220980
Device Name Tissue Approximation System
#### Indications for Use (Describe)
The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
# TAS Medical, Inc. Tissue Approximation System (TAS)
#### 510(k) Owner
TAS Medical 1100 Industrial Road, Suite 16 San Carlos, CA 94070
#### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245
Date Prepared: June 28, 2023
Trade Name of Device Tissue Approximation System (TAS)
Common or Usual Name
Suture
#### Classification Name
Suture, Nonabsorbable, Synthetic, Polyamide 21 C.F.R. §878.5020 Class II Product Code: GAR
Predicate Device Sutumed Sutulon Suture cleared in K123034
#### Device Description
The Tissue Approximation System (TAS) consists of a polyamide (nylon) "zip-tie" attached to a braided polyethylene terephthalate (PET, or Dacron) leader to aid insertion. A one-way clasp allows for fixation of the tissue approximation without the need to tie surgical knots.
When packaged for commercial distribution, four zip-ties with integrated suture leaders are accompanied by hand-held instruments that can be optionally used for insertion. The kit will be provided sterile within a polyethylene/Tyvek pouch.
{5}------------------------------------------------
## Indications for Use
The Tissue Approximation System (TAS) is indicated for use in soft tissue approximation, including use in general surgery procedures such as minimally invasive ventral hernia and abdominal wall closure less than 10cm, but not for use in ophthalmic, neurologic, or cardiovascular procedures.
# Substantial Equivalence
The predicate device for TAS is the Sutumed Sutulon Suture cleared in K123034. Both devices are intended for use in soft tissue approximation and are prescription devices for use by trained healthcare providers. The TAS zip-tie shares important technological characteristics with the predicate device (both are non-absorbable nylon monofilament sutures), however, there are also some differences, such as the zip-tie having a rectangular cross-section, not a USP diameter, and the zip-tie using a clasp rather than a knot. The technological differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and the differences do not affect the safety and effectiveness of the device when used as labeled. Regarding fixation, the predicate device uses knots to maintain the tension set by the user. The subject device is maintained by the zip-tie's clasp. Both devices maintain the tension set by the user, thus they are substantially equivalent. Regarding cross-sectional shape, the primary function of the devices is to approximate soft tissue, which is not affected by cross-sectional shape. Both of these technological characteristic differences, as well as overall designs, were evaluated through performance testing, which demonstrated substantial equivalence.
## Performance Data
Performance data included a large animal. GLP compliant study comparing the closure of full thickness, midline abdominal wall incisions comparing TAS (Test Article) to the predicate device (Control Article) used as labeled (e.g., fixation methods, no surgical mesh). Study endpoints included comparable incision healing, absence of device-associated serious nontransient tissue swelling, necrosis, dehiscence, hematoma, or extrusion in the Test Article as compared to the Control Article. All predefined safety and effectiveness endpoints were met, demonstrating that the different technological characteristics do not affect the safety/effectiveness when used as labeled.
Biocompatibility testing was conducted for the TAS per ISO 10993 and demonstrated substantially equivalent safety. Sterilization processes were validated per ISO 11137, and sterility was validated to SAL 10° with a sterilization dose of 25 kGy by the validation method VDmax25. Individual device testing was conducted to demonstrate the ability of each device to perform its function. A shelf-life of 12 months was validated using accelerated aging for the devices, packaging, and the sterile barrier. All of the verification and validation testing was completed and passed for the Tissue Approximation System with 12-month shelf life.
Human factors validation testing was conducted which demonstrated that the device could be used by the intended users and under the expected use conditions without errors that could result in serious harm.
{6}------------------------------------------------
Testing based on the USP Monographs (including <881, Tensile Strength> and <871, Sutures -Needle Attachment>) showed that the TAS has a strength that meets or exceeds the requirement of USP Size 7, the length of the TAS zip-tie is no less than 95.0% of the length stated on the label and the secured TAS zip-tie is as secure or better than the predicate device tied in a standard square knot.
Therefore, the Tissue Approximation System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.