STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
K221744 · ETHICON, Inc. · GAM · Nov 15, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221744
Device Name
STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
Applicant
ETHICON, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Nov 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.
Device Story
Single-use, antibacterial, monofilament, synthetic absorbable surgical suture; composed of poliglecaprone 25 copolymer (glycolide and epsilon-caprolactone) with IRGACARE® MP (triclosan) and D&C Violet No.2 colorant. Features unidirectional barbs and an anchoring fixation loop; eliminates need for surgical knots. Used by physicians in clinical settings for soft tissue approximation. Barbs allow tissue holding; loop provides anchor. Benefits include secure tissue approximation without knot-related complications. Sterilized via ethylene oxide.
Clinical Evidence
Bench testing only. Evaluated suture loop tensile strength, straight tensile strength, diameter, needle attachment strength, and in-vivo breaking strength retention (BSR). Testing performed per FDA Guidance Document: "Class II Special Controls Guidance Document: Surgical Sutures" (2003). Stability evaluation supported equivalence to primary predicate.
Indicated for soft tissue approximation in patients requiring absorbable sutures.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device (K182873)
Reference Devices
STRATAFIX™ Spiral MONOCRYL™ Plus Bidirectional Knotless Tissue Control Device (K192580)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
11/15/2022
Ethicon, Inc. Jenny Wang Regulatory Affairs Specialist 1000 Us-202 Raritan, New Jersey 08869-1425
#### Re: K221744
Trade/Device Name: STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/l-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: June 15, 2022 Received: June 16, 2022
Dear Jenny Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Deborah A. Fellhauer -S
Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K221744
Device Name
STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device
Indications for Use (Describe)
STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Ethicon Inc. a Johnson & Johnson company<br>1000 US-202 Raritan, New Jersey 08869 |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jenny Wang<br>Ethicon, Inc. a Johnson & Johnson company<br>Phone: 908-440-7850<br>e-mail: jwang347@its.jnj.com |
| Date Prepared: | November 14, 2022 |
| Device Trade Name: | STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional<br>Knotless Tissue Control Device |
| Device Common Name: | Surgical Suture |
| Class: | II |
| Classification Name: | Suture, Absorbable, Synthetic, Polyglycolic Acid<br>(21 CFR 878.4493) |
| Product Code: | GAM |
| Predicate Devices | 510(k) Number | Purpose |
|-------------------------------------------------------------------------------------------------------------------------|---------------|-----------------------------------------------------------------------------------------------|
| <b>Primary Predicate Device:</b><br>STRATAFIX™ Spiral MONOCRYL™ Plus<br>Knotless Tissue Control Device | K182873 | Predicate for all<br>technological<br>characteristics, Indication<br>for Use, and Performance |
| <b>Reference Device for Dye:</b><br>STRATAFIX™ Spiral MONOCRYL™ Plus<br>Bidirectional Knotless Tissue Control<br>Device | K192580 | Reference device for Dye |
{4}------------------------------------------------
#### Device Description:
The STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is an antibacterial monofilament, synthetic absorbable single-use surgical suture device composed of a copolymer of glycolide and (epsilon) s-caprolactone. The device contains IRGACARE®: MP (triclosan), a broad-spectrum antibacterial agent, at no more than 2360 ug/m. The colorant employed is D&C Violet No.2, in <0.050% w/w concentration. Poliglecaprone 25 copolymer has been found to be nonpyrogenic and elicits only a slight tissue reaction during absorption. The subject device is intended for professional use only.
Same as the currently marketed undyed version, the subject device consists of a unidirectional barbed suture material, armed with a surgical needle on one end and an anchoring fixation loop at the opposite end. The barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
While the formation of barbs in the subject STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Device reduces the tensile strength relative to non-barbed suture material of the same size, tying of knots in non-barbed suture materials also reduces their effective strength. For this reason, the strength of the STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Device can be compared to USP knot strength of non-barbed sutures. The actual diameter of the non-barbed section fiber is one size greater than the designated size with a maximum overage of 0.1 mm.
The subject device is sterilized by Ethylene Oxide, and is available in lengths of 6, 9, and 12 inches, with a USP diameter size 2-0 and 3-0 (metric sizes 3.0 and 2.0), and single armed with various needle sizes.
#### Indications for Use:
STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.
#### Performance Data:
Non-clinical laboratory performance testing on suture loop tensile strength and suture straight tensile strength was performed demonstrating that the subject device conforms to the current USP Monograph for absorbable surgical sutures, except for diameter. The testing was performed in accordance with FDA's Guidance Document: "Class II Special Controls Guidance Document: Surgical Sutures" issued on June 3, 2003. In addition, Suture diameter, Needle Attachment Strength, and In-vivo Breaking Strength Retention (BSR) were evaluated to show that the device performs as intended and as claimed. Stability evaluation was performed to support equivalence to the primary predicate device.
The added colorant in the subject device is the same as that used in the secondary predicate device for dye and does not raise new questions on safety or effectiveness. The subject device meets all testing criteria to demonstrate substantial equivalence to the primary predicate device.
{5}------------------------------------------------
#### Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use:
100979694| Rev. 1
The intended use, technological characteristics, and performance of the subject device are consistent with those of the primary predicate device. The subject and primary predicate (K182873) devices are identical except for the addition of the colorant and the subject device is supplied in a subset of sizes. The dye and its concentration used in the subject device is the same as that used in the reference device for dye (K192580), which is comprised of the identical base material but which does not include a welded loop.
The following table summarizes the similarities and differences between the subject device and the predicate and reference devices.
| Characteristic | Subject Device<br>STRATAFIXTM Spiral<br>MONOCRYLTM Plus<br>Unidirectional Device<br>(Dyed) | Primary Predicate<br>Device<br>STRATAFIXTM Spiral<br>MONOCRYLTM Plus<br>Knotless Tissue Control<br>Device (K182873) | Reference Device for<br>Dye<br>STRATAFIXTM Spiral<br>MONOCRYLTM Plus<br>Bidirectional Knotless<br>Tissue Control Device<br>(K192580) |
|-------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Barb<br>Orientation | Unidirectional Barbed<br>suture with a welded loop | Same | Bidirectional Barb<br>suture without a welded<br>loop |
| Absorbable<br>Suture Material | Copolymer<br>poliglecaprone 25<br>composed of glycolide and<br>ε-caprolactone | Same | Same |
| Colorant | D&C Violet No.2 (21CFR<br>§74.3602) in <0.050% w/w<br>concentration | Undyed | Same |
#### Substantial Equivalence:
The subject device has the same intended use and indications for use as the primary predicate device. There is no polymer formulation, construction, specification, body/fluid contact. manufacturing, or sterilization process change to the currently marketed device.
The added colorant in the subject device is the same as that used in the reference device for dye and does not raise new questions on safety or effectiveness.
#### Conclusion:
Based on the intended use, technological characteristics, and safety and performance testing, and biocompatibility evaluation, the STRATAFIX™ Spiral MONOCRYL™Plus Unidirectional Knotless Tissue Control Device has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.