CARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910 (fast absorbing) Surgical Suture

K170166 · Cpt Sutures Co., Ltd. · GAM · Sep 29, 2017 · General, Plastic Surgery

Device Facts

Record IDK170166
Device NameCARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910 (fast absorbing) Surgical Suture
ApplicantCpt Sutures Co., Ltd.
Product CodeGAM · General, Plastic Surgery
Decision DateSep 29, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4493
Device ClassClass 2
AttributesTherapeutic

Indications for Use

CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues. CARESORB® RAPID Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures.

Device Story

Multifilament, braided, synthetic absorbable surgical sutures composed of 90% glycolide and 10% L-lactide copolymer. Coated with poly(glycolide-co-L-lactide) and calcium stearate. Used by surgeons for soft tissue approximation and ligation. Sutures provide temporary wound support via tensile strength retention before undergoing hydrolysis and absorption. CARESORB RAPID is designed for faster absorption (42 days) for superficial skin/mucosa closure. Available in various USP sizes with or without stainless steel needles. Sterilized by ethylene oxide. Benefits include secure tissue approximation during healing and predictable absorption profiles, reducing need for suture removal.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), physical performance (USP <861>, <871>, <881>), and in-vivo/in-vitro resorption/breaking strength retention studies comparing subject devices to predicates.

Technological Characteristics

Multifilament braided suture; 90% glycolide/10% L-lactide copolymer; coated with poly(glycolide-co-L-lactide) and calcium stearate. USP sizes 6-0 to 1 (CARESORB) and 4-0 to 2-0 (CARESORB RAPID). Sterilized by ethylene oxide. Complies with USP <861> (diameter), <871> (needle attachment), and <881> (tensile strength).

Indications for Use

Indicated for general soft tissue approximation/ligation (CARESORB) or superficial soft tissue approximation of skin/mucosa requiring short-term support (CARESORB RAPID). Contraindicated for cardiovascular or neurological tissues; CARESORB RAPID also contraindicated for ligation and ophthalmic procedures.

Regulatory Classification

Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a way that they also resemble an abstract bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 29, 2017 CPT Sutures Co., Ltd. % Ms. Natalya Valerio Official Correspondent for CPT Sutures Co., Ltd. MDI Consultants. Inc. 55 Northern Blvd.. Suite 200 Great Neck, New York 11021 Re: K170166 Trade/Device Name: CARESORB® - Polyglactin 910 Surgical Suture, CARESORB RAPID® - Polyglactin 910 (fast Absorbing) Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: August 31, 2017 Received: September 1, 2017 Dear Ms. Valerio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # David Krause -S - for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170166 Device Name CARESORB® - Polyglactin 910 Surgical Suture Indications for Use (Describe) CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170166 Device Name CARESORB® RAPID - Polyglactin 910 (fast absorbing) Surgical Suture ### Indications for Use (Describe) CARESORB® RAPID Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {4}------------------------------------------------ K170166 p. 1 of 7 # 510(k) SUMMARY The assigned 510(k) number is: K170166 #### 1. Submitter's Identification: CPT Sutures Co., Ltd. 8 Dao Tri Street Phu Thuan Ward District 7, Ho Chi Minh, Vietnam Date Summary Prepared: August 31, 2017 Mr. Khoa Do Contact: Vice Director CPT Sutures Co., Ltd. e-mail: khoa.do@cpt-medical.com phone: (+84) 28 38 222 228 #### 2. Name of the Device: | Trade Name: | CARESORB® - Polyglactin 910 Surgical Suture and<br>CARESORB® RAPID - Polyglactin 910 fast absorbing<br>Surgical Suture | |--------------------|------------------------------------------------------------------------------------------------------------------------| | Common Name: | Suture, Absorbable, Synthetic, Polyglycolic Acid | | Regulation Number: | 21CFR 878.4493 | | Regulation Name: | Absorbable Poly(glycolide/L-lactide) Surgical Suture | | Regulatory Class: | Class II | | Product Code: | GAM | #### 3. Information for the 510(k) Cleared Device (Predicate Device): Ethicon's Coated VICRYL™ (Polyglactin 910) Synthetic Absorbable Suture, 510(k) # K022269 Ethicon's Coated VICRYL™ Rapide (Polyglactin 910) Synthetic Absorbable Suture, 510(k) # K033746 #### 4. Device Description: The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are multifilament, braided, sterile synthetic absorbable surgical sutures composed of a copolymer of 90% glycolide and 10% L-lactide. The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are coated with copolymer of Poly(glycolide-co-L-lactide) (30/70) and calcium stearate. The CARESORB® - Polyglactin 910 suture is available dyed with FDA-approved color additive D&C Violet No. 2 - Cl 60725 or undyed in the natural beige color. {5}------------------------------------------------ The CARESORB® RAPID - Polyglactin 910 suture is available undyed in the natural beige color only. The CARESORB® sutures are available in USP sizes 6-0 through 1 and CARESORB® RAPID sutures in USP sizes 4-0 through 2-0, in different lengths, with or without a standard needle attached. The CARESORB® and CARESORB® RAPID sutures meet USP Monograph for Synthetic Absorbable Sutures, except for diameter. #### 5. Indications for Use: CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues. CARESORB® RAPID - Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures. #### 6. Comparison to the 510(k) Cleared Devices (Predicate Devices): | Comparison<br>Criteria | New Device<br>CARESORB® -<br>Polyglactin 910<br>Surgical Suture | New Device<br>CARESORB® RAPID -<br>Polyglactin 910<br>Surgical Suture | Predicate<br>Device(s) | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Predicate Device | Ethicon's Coated<br>VICRYL™ (Polyglactin<br>910) Synthetic<br>Absorbable Suture,<br>510(k) # K022269 | Ethicon's Coated<br>VICRYL™ Rapide<br>(Polyglactin 910)<br>Synthetic<br>Absorbable Suture,<br>510(k) # K033746 | Image: checkerboard pattern | | Classification | Synthetic<br>Absorbable<br>Surgical Suture | Synthetic<br>Absorbable<br>Surgical Suture | Same | | Indications<br>for Use | General soft tissue approximation and / or ligation, including use in ophthalmic procedures, but not in cardiovascular or neurological tissues | Superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required but not for use in ligation, ophthalmic, cardiovascular or neurological procedures | Same for each<br>respective<br>predicate | | Suture Material | Copolymer of 90%<br>glycolide and 10%<br>L-lactide | Copolymer of 90%<br>glycolide and 10%<br>L-lactide | Same | | Material<br>Presentation | Multifilament, braided | Multifilament, braided | Same for each<br>respective<br>predicate | | Coating | Coated with copolymer<br>of Poly(glycolide-co-L-<br>lactide) (30/70) and<br>calcium stearate | Coated with copolymer<br>of Poly(glycolide-co-L-<br>lactide) (30/70) and<br>calcium stearate | Same | | Color Additive | Available dyed with FDA-<br>approved color additive<br>D&C Violet No. 2 - Cl<br>60725 as well as undyed<br>in natural beige color | Available undyed only | Same for each<br>respective<br>predicate | | Sterilization | Sterilized by<br>Ethylene Oxide | Sterilized by<br>Ethylene Oxide | VICRYL™ is<br>sterilized by<br>Ethylene Oxide<br>VICRYL™ Rapide<br>is sterilized by<br>Gamma Irradiation | | Single Use | Yes | Yes | Same | | How Supplied | Sutures of various<br>lengths and gauge sizes<br>of USP 6-0 to 1, with or<br>without the standard<br>stainless steel needle<br>(series 302 with silicon<br>coating) attached.<br>Available as one-dozen<br>and three-dozen boxes. | Sutures of various<br>lengths and gauge sizes<br>of USP 4-0 to 2-0, with<br>or without the standard<br>stainless steel needle<br>(series 302 with silicon<br>coating) attached.<br>Available as one-dozen<br>and three-dozen boxes. | Sutures of various<br>gauge sizes and<br>lengths, with or<br>without standard<br>stainless steel<br>needle attached.<br>Available as one-<br>dozen and three-<br>dozen boxes. | | Absorption<br>Profile | Suture absorption begins<br>as a loss of tensile<br>strength followed by<br>eventual absorption by<br>hydrolysis. Complete<br>absorption occurs<br>between 56-70 days. | Suture absorption begins<br>as a loss of tensile<br>strength followed by<br>eventual absorption by<br>hydrolysis. Complete<br>absorption occurs in 42<br>days. | Same for each<br>respective<br>predicate | | Breaking<br>Strength<br>Retention Profile | Percent Breaking<br>Strength Retention (%<br>BSR): > 65% at 14 days;<br>> 40% at 21 days. | Percent Breaking<br>Strength Retention (%<br>BSR): 50% at 5 days;<br>0% at 14 days. | Same for each<br>respective<br>predicate | | Suture Diameter | Complies with<br>USP <861> except for<br>slight oversize in the<br>suture diameter | Complies with<br>USP <861> except for<br>slight oversize in the<br>suture diameter | Same | | Suture Length | ≥ 95% of the claimed<br>label length as required<br>by USP | ≥ 95% of the claimed<br>label length as required<br>by USP | Same | | Needle<br>Attachment | Complies with<br>USP <871> | Complies with<br>USP <871> | Same | | Tensile Strength | Complies with<br>USP <881> | Complies with<br>USP <881> | Same | | Biocompatibility | Complies with all ISO<br>10993-1 required tests<br>for absorbable surgical<br>sutures | Complies with all ISO<br>10993-1 required tests<br>for absorbable surgical<br>sutures | Same | | Packaging | Validated sterile barrier<br>packaging system | Validated sterile barrier<br>packaging system | Same | {6}------------------------------------------------ {7}------------------------------------------------ ### Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence: The below referenced testing demonstrated substantial equivalence of the CARESORB® - Polyglactin 910 Surgical Suture and CARESORB® RAPID -Polyglactin 910 (fast absorbing) Surgical Suture to the Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide predicate devices respectively. The testing was conducted on the final sterilized devices in accordance with the FDA Class II Special Controls Guidance Document: Surgical Sutures. # Performance Testing: - 1) ISO 10993-3 Biological Evaluation of Medical Devices Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity, 2014 - 2) ISO 10993-5 Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity, 2009 (R) 2014 - 3) ISO 10993-6 Biological Evaluation of Medical Devices Part 6: Tests for Local Effects after Implantation, 2007 (R) 2014 - 4) ISO 10993-7 Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals, 2008 (R) 2012 - 5) ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization, 2010 (R) 2014 {8}------------------------------------------------ - 6) ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity, 2006 (R) 2010 - 7) USP 39-NF33 Absorbable Surgical Suture, 2016 - 8) USP 39-NF33 <861> Sutures Diameter, 2016 - 9) USP 39-NF33 <871> Sutures Needle Attachment, 2016 - 10) USP 39-NF33 <881> Tensile Strength, 2016 - 11) ISO 11607-1 Section 5.2 Whole package microbial aerosol challenge test, 2014 - 12) ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials, 2015 - 13) ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages, 2013 - 14) ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test), 2011 - 15) USP39-NF 34 <71> Sterility Test, 2016 - 16) USP39-NF 34 <85> Bacterial Endotoxins Test, 2016 - 17) ISO 15223-1 Third Edition 2016-11-01, Medical Devices Symbols To Be Used With Medical Device Labels, Labelling, And Information To Be Supplied - Part 1: General Requirements. (General I (QS/RM)) The biocompatibility test results demonstrated no evidence of cytotoxicity, irritation, delayed dermal sensitization or acute systemic toxicity for CARESORB® Surgical Suture. In the short-term, mid-term and long-term Intramuscular Implantation Tests with Histopathology, the CARESORB® Surgical Suture was considered a non-irritant as compared to the predicate VICRYL™ used as a control. In the mid-term Intramuscular Implantation Test with Histopathology, the CARESORB® Rapid Surgical Suture was considered a non-irritant as compared to the predicate VICRYL™ Rapide used as a control. The CARESORB® Suture was also found to be non-pyrogenic, non-mutagenic and non-genotoxic. {9}------------------------------------------------ # Resorption Profile: The resorption profiles of the final sterilized sutures were demonstrated in vivo. The CARESORB® sutures were compared to the predicate VICRYL™ sutures in the in-vivo implantation study. The study results demonstrated that CARESORB® sutures have similar tensile strength retention profile in comparison to the marketed predicate VICRYL™ sutures. The in-vivo breaking strength retention profile of CARESORB® RAPID was based on the data provided by the suture material supplier from the in-vivo implantation study. The resorption profile of CARESORB® RAPID was determined to be similar to that of the marketed predicate VICRYL™ Rapide. The resorption profile of the final sterilized CARESORB® RAPID suture was also demonstrated in vitro. The CARESORB® RAPID. sterilized by EtO. were compared to the predicate VICRYL™ Rapide sutures, sterilized by gamma irradiation, and to the raw material supplier's sutures, sterilized by gamma irradiation. All sutures were immersed in the phosphate buffer solution (pH 7.4) for 5. 14. 21 and 28 days and then tested for breaking strength retention at each time point. The study results demonstrated that all three sutures had essentially identical resorption profiles under the same study conditions and that tensile strength and tensile strength retention of CARESORB® RAPID was comparable to those of VICRYL™ Rapide and the raw material supplier's sutures. Additionally, the rate of absorption of CARESORB® sutures was compared to that of predicate VICRYL™ sutures in the in-vivo implantation study. The study results demonstrated that CARESORB® sutures and VICRYL™ sutures have very similar absorption in tissue. The rate of absorption of CARESORB® RAPID was based on the data provided by the suture material supplier from the in-vivo implantation study. The CARESORB RAPID rate of absorption in tissue was determined to be similar to that of the marketed predicate VICRYL™ Rapide. ## Shelf Life: CARESORB® and CARESORB® RAPID sutures have been tested in real time stability studies as well as in the accelerated aging studies. The real time stability studies for both sutures were performed at 18°C to 26°C temperature and relative humidity of no more than 65%. The accelerated aging stability studies were performed at 68°C to 72°C temperature and relative humidity of no more than 65% for CARESORB® and at 53°C to 57°C temperature and relative humidity of no more than 65% for CARESORB® RAPID. The aged sutures were tested for parameters specified in the USP Monograph for Absorbable Surgical Sutures and other applicable requirements: {10}------------------------------------------------ - Diameter, USP 39 <861> - Needle Attachment, USP 39 <871> ● - Tensile Strength, USP 39 <881> ● - In vitro Breaking Strength Retention (BSR) - Visual Inspection ● - . Moisture content - Sterility Testing, USP 39-NF34 <71> ● The study results demonstrated conformance to the specified acceptance criteria and supported the 5 year expiration date assignment for CARESORB® and CARESORB® RAPID sutures. #### 8. Discussion of Clinical Tests Performed: Not Applicable #### 9. Conclusions: The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 Surgical Sutures are composed of the same material, have the same intended use and main technological characteristics as their respective predicate devices Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide Surgical Sutures. Based on the results of completed performance testing inclusive of physical testing, biocompatibility testing and absorption profile, it can be concluded that CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are substantially equivalent in terms of safety and effectiveness to Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide respectively.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...