K143584 · Resorba Medical GmbH · GAM · Nov 2, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143584
Device Name
PGA RESORBA, PGA RESOQUICK, GLYCOLON
Applicant
Resorba Medical GmbH
Product Code
GAM · General, Plastic Surgery
Decision Date
Nov 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GLYCOLON® is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery. PGA RESORBA® is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not in cardiovascular surgery or neural tissue. PGA resoquick™ is indicated for use in general soft tissue approximation, including ophthalmic (e.g. conjunctiva) procedures, but not in cardiovascular or neurological procedures.
Device Story
Synthetic absorbable surgical sutures; provided sterile for single use. PGA RESORBA and PGA resoquick are braided multifilament polyglycolic acid (PGA) sutures coated with polycaprolactone and calcium stearate. GLYCOLON is a monofilament copolymer of glycolide and epsilon-caprolactone (PGA/CL), uncoated. Available in various lengths, diameters, and with/without attached needles. Used by surgeons for soft tissue approximation and ligation. Performance meets USP requirements for diameter, tensile strength, and needle attachment. Biocompatibility established per ISO 10993-1. Devices facilitate tissue healing through temporary support, eventually absorbing in the body.
Clinical Evidence
Bench testing only. Devices evaluated against USP monographs for synthetic absorbable sutures, including diameter <861>, tensile strength <881>, and needle attachment <871>. Resorption profiles evaluated via in-vitro and in-vivo data. Biocompatibility supported by ISO 10993-1 compliance and established history of material use.
Technological Characteristics
Materials: Polyglycolic acid (PGA) or glycolide/epsilon-caprolactone copolymer. Coating: Polycaprolactone/calcium stearate (PGA variants) or uncoated (GLYCOLON). Form: Braided multifilament or monofilament. Sterile, single-use. Meets USP requirements for synthetic absorbable sutures.
Indications for Use
Indicated for general soft tissue approximation and/or ligation. PGA RESORBA and PGA resoquick include ophthalmic procedures. All exclude cardiovascular and neurological tissues; GLYCOLON also excludes microsurgery and ophthalmic surgery.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
Resorba Medical GmbH % Mr. Karl Beck Quality Assurance/Regulatory Affairs Manager Am Flachmoor 16 90475 Nürnberg Germany
Re: K143584
Trade/Device Name: PGA RESORBA, PGA RESOQUICK, GLYCOLON Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: September 23, 2015 Received: September 28, 2015
Dear Mr. Beck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part
{1}------------------------------------------------
807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) unknown
### Device Name GLYCOLON Absorbable poly(glycolide/l-lactide) surgical suture
#### Indications for Use (Describe)
GLYCOLON® is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) unknown
#### Device Name
PGA resorba Absorbable poly(glycolide/l-lactide) surgical suture
#### Indications for Use (Describe)
PGA RESORBA® is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not in cardiovascular surgery or neural tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) unknown
#### Device Name
PGA resoquick Absorbable poly(glycolide/l-lactide) surgical suture
Indications for Use (Describe)
PGA resoquick™ is indicated for use in general soft tissue approximation, including ophthalmic (e.g. conjunctiva) procedures, but not in cardiovascular or neurological procedures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 5.0 TRADITIONAL 510(K) SUMMARY
Submitted by: Resorba Medical GmbH Am Flachmoor 16 90475 Nuremberg Germany Tel: +49 9128-9115-0 Fax: +49 9128-9115-10
Karl-Josef Beck/Rose Guang Contact Person:
Date of Summary: 30 October 2015
| Trade Name | Common Name | Classification Name | Regulation Classification | Product Code | Class of device | Predicate device |
|-------------------|--------------------------------------------------------|----------------------------------------------------------------|---------------------------|--------------|-----------------|----------------------------------------------------------------------------------------------------------------------|
| PGA<br>RESORBA® | Suture, absorbable,<br>synthetic,<br>polyglycolic acid | Absorbable<br>poly(glycolide/l-<br>lactide) surgical<br>suture | 21 CFR<br>§878.4493 | GAM | II | Surgisorb,<br>Samyang<br>K984374 |
| PGA<br>Resoquick™ | Suture, absorbable,<br>synthetic,<br>polyglycolic acid | Absorbable<br>poly(glycolide/l-<br>lactide) surgical<br>suture | 21 CFR<br>§878.4493 | GAM | II | Safil Quick,<br>Aesculap<br>K011372<br>*Surgisorb,<br>Samyang<br>K984374<br>*Optime<br>Peters<br>Surgical<br>K062366 |
| GLYCOLON® | Suture, absorbable,<br>synthetic,<br>polyglycolic acid | Absorbable<br>poly(glycolide/l-<br>lactide) surgical<br>suture | 21 CFR<br>§878.4493 | GAM | II | Monocryl<br>Ethicon<br>K960653 |
*Secondary predicates
{6}------------------------------------------------
#### The subject devices are synthetic absorbable surgical sutures. Device Description:
| | PGA Resorba | PGA resoquick | Glycolon |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|--------------------------------------------------|-------------------------------------------------------|
| Primary material | Polyglycolic acid<br>(PGA) | Polyglycolic acid<br>(PGA) | Copolymer glycolide<br>and E-caprolactone<br>(PGA/CL) |
| Material presentation | Multifilament,<br>braided | Multifilament,<br>braided | Monofilament |
| Dyed or undyed | Undyed and dyed<br>D&C violet No.2 | Undyed | Undyed and dyed<br>D&C violet No.2 |
| Coated or uncoated | Coated<br>Polycaprolactone +<br>calcium stearate | Coated<br>Polycaprolactone +<br>calcium stearate | Uncoated |
| Suture material is a<br>sterile, flexible, thread<br>offered in a variety of<br>lengths, reels, range of<br>diameters, with or<br>without needles<br>attached. | Yes | Yes | Yes |
Indication for Use: PGA RESORBA® is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not in cardiovascular surgery or neural tissue.
> PGA resoquick™ is indicated for use in general soft tissue approximation, including ophthalmic (e.g. conjunctiva) procedures, but not in cardiovascular or neurological procedures.
GLYCOLON® is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
- Each absorbable suture has the same intended use and similar design, Substantial Equivalence: materials, labeling, performance characteristics to their predicate device.
y
{7}------------------------------------------------
| Absorbable sutures | Page 5-3 |
|-------------------------------------------|----------|
| Traditional 510(k) Premarket Notification | |
Technological Each absorbable suture is substantially equivalent to the predicate device listed when compared to the technological characteristics and are characteristics supplied sterile for single use. All meet USP requirements except for minor variations in diameter for Glycolon® monofilament absorbable suture.
### Comparison of technological characteristics to predicate device:
| | Intended use | Material | Design | Performance<br>Diameter/<br>Needle<br>attachment/<br>Tensile<br>strength |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| PGA<br>RESORBA | Indicated for use in<br>general soft tissue<br>approximation and/or<br>ligation, including use in<br>ophthalmic surgery, but<br>not in cardiovascular<br>surgery or neural tissue | Material:<br>Polyglycolic acid<br>(PGA)<br>Coating:<br>Polycaprolactone +<br>calcium stearate<br><br>Dye:<br>D&C violet No.2 | Absorbable,<br>multifilament,<br>dyed, coated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P |
| Predicate | Same | Same | Same | Same |
| PGA<br>RESOQUICK | Indicated for use in<br>general soft tissue<br>approximation, including<br>ophthalmic (e.g.<br>conjunctiva) procedures,<br>but not in cardiovascular<br>or neurological<br>procedures. | Material:<br>Polyglycolic acid<br>(PGA)<br>Coating:<br>Polycaprolactone +<br>calcium stearate<br><br>Dye:<br>undyed | Absorbable,<br>multifilament,<br>undyed, coated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P |
| Predicate | Same | Similar, different<br>coating | Same | Same |
| Secondary<br>predicates | Same | Same | Same | Same |
| GLYCOLON | Indicated for use in<br>general soft tissue<br>approximation and/or<br>ligation, but not for use in<br>cardiovascular or<br>neurological tissues, | Material:<br>glycolide and ε-<br>caprolactone<br>(PGA/CL)<br><br>Coating:<br>Uncoated | Absorbable,<br>monofilament,<br>dyed, uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P<br>except for<br>diameter |
| | ophthalmic surgery. | Dye:<br>D&C violet No.2 | | |
| Predicate | Same | Same | Same | Same |
to
{8}------------------------------------------------
Performance Testing As per the FDA's Class II Special Control Guidance Document for Surgical Summary: Sutures, the devices were subjected to the requirements of the United States Pharmacopeia (U.S.P) monograph for Synthetic Absorbable Sutures. Testing included:
- . Diameter <861>
- 0 Tensile strength <881>
- . Needle attachment <871>
Resorption profiles have been evaluated for each material. Resorption data in-vitro and in-vivo were provided for PGA Resorba, Glycolon and in-vivo only for PGA resoquick.
Results of USP performance and resorption testing demonstrate that each absorbable suture meets the current USP performance requirements for absorbable sutures (with the exception of GLYCOLON® that has minor variations in diameter from USP) and they are substantially equivalent to each corresponding predicate device listed.
Biocompatibility evaluation in accordance with ISO 10993-1 has been supported for each material by leveraging data from each supplier and each material having an established history of use for the same intended use.
Based on the information provided within this 510(k) submission, Conclusion Resorba Medical GmbH concludes that the proposed suture products are substantially equivalent to each corresponding predicate device listed.
y