← Product Code [GAM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM) · K130083

# DEMECAPRONE (POLIGLECAPRONE 25) SYNTHETIC MONOFILAMENT (PGCL) ABSORBABLE SUTURE (K130083)

_Demetech Corp. · GAM · Jul 17, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K130083

## Device Facts

- **Applicant:** Demetech Corp.
- **Product Code:** [GAM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM.md)
- **Decision Date:** Jul 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4493
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

DemeCAPRONE (Poliglecaprone 25) Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

## Device Story

DemeCAPRONE is a synthetic, absorbable, sterile, flexible monofilament surgical suture composed of poly(glycolic-co-caprolactone) copolymer (PGCL). It is supplied in sizes 6-0 through 1, either undyed or dyed with D&C Violet #2. The device is used by surgeons for soft tissue approximation and ligation. It functions as a mechanical support for tissue healing, eventually undergoing absorption in the body. The device is provided sterile for clinical use. It benefits patients by providing temporary wound closure and tissue support during the healing process.

## Clinical Evidence

No clinical trials were conducted. Substantial equivalence is supported by non-clinical bench testing, including USP performance requirements (tensile strength, needle attachment, suture length, sterility) and biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity, subchronic toxicity, and muscle implantation) per ISO 10993.

## Technological Characteristics

Material: Poly(glycolic-co-caprolactone) copolymer (PGCL). Form: Synthetic monofilament. Sizes: 6-0 to 1 (metric 0.7-4). Standards: Meets USP requirements for synthetic absorbable sutures. Biocompatibility: ISO 10993 compliant. Sterilization: Sterile. Connectivity: None.

## Regulatory Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Riverpoint's Mono Q PGCL Absorbable Suture ([K100461](/device/K100461.md))
- Sutures India PVT.LTD Monoglyde Poliglecaprone 25 Absorbable Suture ([K081002](/device/K081002.md))
- Ethicon's Monocryl Synthetic Absorbable Poliglecaprone 25 suture ([K960653](/device/K960653.md) & [K964072](/device/K964072.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DemeTECH.

### JUL 1 7 2013

# Section 1 - 510(k) Summary

This summary is submitted in accordance with the Safe Medical Device Act (SMDA) of 1990 and Title 21 CFR § 807.92. This summary demonstrates the equivalence of Demetech Sutures to those of the legally marked devices listed.

| A. Applicant:                                                                                                                                            | Demetech Corporation,<br>14175 NW 60th Ave.<br>Miami Lakes FL. 33014                  |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| B. Contact Person:                                                                                                                                       | A. J. Dimercurio<br>e-mail: tony@demetech.us<br>Phone # 305-824-1048 Ext 115          |
| C. Date Prepared:                                                                                                                                        | July 15th, 2013                                                                       |
| Trade Name:                                                                                                                                              | DemeCaprone (Poliglecaprone<br>25) Synthetic Monofilament<br>(PGCL) Absorbable Suture |
| Common Name:                                                                                                                                             | DemeTECH Synthetic PGCL<br>Monofilament Poliglecaprone 25<br>Absorbable Suture        |
| Classification Name:                                                                                                                                     | Absorbable poly (glycolide/l-<br>lactide) Surgical Suture braided<br>or monofilament  |
| D. Device Classification                                                                                                                                 |                                                                                       |
| FDA Class:<br>Product Classification:                                                                                                                    | II<br>878.4493, Absorbable<br>Poly(glycolide/I-lactide) Surgical<br>Suture            |
| Product Code:                                                                                                                                            | GAM                                                                                   |
| E. Predicate Devices: DemeCAPRONE (Poliglecaprone 25) Synthetic<br>Absorbable Surgical Suture is substantially equivalent to these predicate<br>devices: |                                                                                       |
| • Riverpoint's Mono Q PGCL Absorbable Suture reference 510K number<br>K100461, Riverpoint Medical. Portland OR.                                          |                                                                                       |

Sutures India PVT.LTD Monoglyde Poliglecaprone 25 Absorbable Suture . reference 510K number K081002, Sutures India Private Limited Bangalore India.

. . . . .

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{1}------------------------------------------------

K130083 page 2/5

- Ethicon's Monocryl Synthetic Absorbable Poliglecaprone 25 suture, . reference 510K number, K960653 & K964072, Ethicon Inc. Somerville NJ.
- F. Device Description:

DemeCAPRONE (Poliglecaprone 25) is a synthetic monofilament absorbable surgical suture composed of Poly (glycolic-co-caprolactone) copolymer (PGCL) and is supplied un-dyed and dyed with D&C Violet #2 below 0.1wt%. DemeCAPRONE (Poliglecaprone 25) synthetic absorbable suture is available in sizes 6-0 through 1 (metric sizes 0.7 - 4). DemeCAPRONE (Poliglecaprone 25) Surgical Suture meets the requirements established by the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures except for diameter.

- G. Intended Use:
DemeCAPRONE (Poliglecaprone 25) Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

- H. Non-Clinical Tests Performed:
Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and invivo and in-vitro resorption to further demonstrate substantial equivalence to the predicate devices. Physical properties and functionality testing assured that the device conformed with suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterility to methods outlined in USP 35.

Poliglecaprone 25 was selected based on known biocompatibility (per ISO 10993) and established history of use in the medical device industry for implantable devices, and are identical or substantially equivalent to the material used in the predicate devices listed above. Biocompatibility testing performed on Poliglecaprone 25 sutures within the submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Genotoxicity - Bacterial Reverse Mutation and Chromosomal Aberration, Bone Marrow Micronucleus, Subchronic Toxicity (4-week, following subcutaneous implantation), Muscle Implantation (12-week).

| COMPARISON TABLE DEMETECH POLIGLECAPRONE 25 TO PREDICATE                                                                                                                                                 |                                         |                                          |                                         |                               |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------------------------------------------|-----------------------------------------|-------------------------------|
| DEVICES                                                                                                                                                                                                  |                                         |                                          |                                         |                               |
| Comparison Items                                                                                                                                                                                         | Demetech<br>Poliglecaprone<br>25 Suture | Riverpoint's<br>Mono Q<br>PGCL<br>Suture | Sutures<br>India<br>Monoglyde<br>Suture | Ethicon<br>Monocryl<br>Suture |
| DemeCAPRONE<br>(Poliglecaprone 25) suture is a<br>synthetic absorbable surgical<br>suture. It is a sterile flexible<br>monofilament thread, composed<br>of Poly (glycolic-co-<br>caprolactone) copolymer | Same                                    | Same                                     | Same                                    | Same                          |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| Same                                                                                                                                                                                                     | Same                                    | Same                                     | Same                                    |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         | Same                                     |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
|                                                                                                                                                                                                          | Same                                    | Same                                     |                                         | Same                          |
|                                                                                                                                                                                                          |                                         |                                          |                                         |                               |
| DemeCAPRONE meets or<br>exceeds the performance<br>requirements defined in the<br>United States Pharmacopeia for<br>"Tensile Strength" < 881 >                                                           | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE meets or<br>exceeds the performance<br>requirements defined in the<br>United States Pharmacopeia and<br>the current edition USP for<br>"Needle Attachment" < 871 >                           | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE meets or<br>exceeds the performance<br>requirements defined in the<br>United States Pharmacopeia for<br>"Suture Length Requirement"<br>(95% of stated label length)                          | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE meets the<br>performance requirements<br>defined in the United States<br>Pharmacopeia current edition<br>U.S.P. for sterility                                                                | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE is packaged in<br>a same or equivalent manner<br>with sterile single or double<br>packaging having labeling<br>conforming to 21 CFR and<br>Current edition of USP.                           | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE<br>(Poliglecaprone 25) suture is<br>biologically compatible when<br>tested as per ISO-10993                                                                                                  | Same                                    | Same                                     | Same                                    | Same                          |
| DemeCAPRONE<br>(Poliglecaprone 25) suture is<br>tested and proved to be non<br>toxic, when tested as per ISO-<br>10993 for toxicity                                                                      | Same                                    | Same                                     | Same                                    | Same                          |

{2}------------------------------------------------

K130063 page 3/5

{3}------------------------------------------------

K130083 page 4/5

#### Clinical Tests Performed: l.

.

No clinical trials were conducted

{4}------------------------------------------------

K130083 page 5/5

### J. Conclusion:

DemeCAPRONE (Poliglecaprone 25) is composed of the same material, as are the predicated devices and the same design being a sterile, flexible, monofilament threads meeting all the requirements of the United States Pharmacopeia. DemeCAPRONE (Poliglecaprone 25) Suture is manufactured in the same manner as the predicate devices, being composed of composition of absorbable flexible, monofilament thread prepared from Poly (glycolic-co-caprolactone) copolymer (PGCL) and produced in operations considered standard in the fiber industry to form the finished suture fiber. The manufacturer supplies to Demetech the same suture materials as it does to other suture manufacturers including some of those listed above.

The biocompatibility data and the results of performance testing presented demonstrate the substantial equivalence of DemeCAPRONE (Poliglecaprone 25) Synthetic Absorbable Suture to that of the predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

July 17, 2013

DemeTECH Corporation % Anthony J. Dimercurio Vice President RA/OA 14175 NW 60th Avenue Miami Lakes, Florida 33014

Re: K130083

Trade/Device Name: DemeCaprone (Poligecaprone 25) Synthentic (PGCL) Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: May 29, 2013 Received: May 31, 2013

Dear Mr. Dimercurio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 - Anthony J. Dimercurio

CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

For

Peter D. Rumm -S Mark N. Melkerson

Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DemeTECH. The logo consists of a stylized globe-like symbol on the left, followed by the company name "DemeTECH" in a bold, italicized font. The "CH" portion of the name is slightly smaller and raised, with a small circle as a subscript.

## Section 6 - Indications for Use Statement

Indication for use

510K Number: _________________________________________________________________________________________________________________________________________________________________

Device Name: Demetech Absorbable Poliglecaprone 25 Surgical Suture.

Indication for Use:

Demetech Absorbable Poliglecaprone 25 Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

| Prescription Use             | "X" |
|------------------------------|-----|
| Use                          |     |
| (Part 21 CFR 801; Subpart D) |     |
| And/Or                       |     |
| Over the-Counter             |     |
| (21 CFR 801; Subpart C)      |     |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF

NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

### David Krause -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130083

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K130083](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GAM/K130083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
