K972918 · Apex Medical Products, LLC · GAH · Oct 27, 1997 · General, Plastic Surgery
Device Facts
Record ID
K972918
Device Name
RAPID FIRE HAIR IMPLANTER CAROUSEL
Applicant
Apex Medical Products, LLC
Product Code
GAH · General, Plastic Surgery
Decision Date
Oct 27, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Rapid Fire Hair Implanter Carousel™ is to be used for hair transplantation on the scalp. Its use is indicated where patients require one or more hair grafting sessions during which multiple grafts have to be implanted.
Device Story
Disposable mechanical hair transplant facilitator; hand-held, ball-point pen size. Features circular carousel for hair graft storage, sharp end for scalp incision, and button-activated internal moving parts. Used by clinicians during hair transplantation procedures. Device stores grafts and facilitates implantation, reducing procedure time and bleeding compared to manual methods. Stainless steel blade contacts patient tissue; body composed of Lustran 348 resin. Provides uniform graft depth in dermis.
Clinical Evidence
Bench testing included in vitro porcine skin studies and a human subject study (n=1). In vitro study compared manual vs. Carousel implantation; histologic evaluation showed no significant difference in tissue integrity, with Carousel providing more uniform graft depth. Human study (400 grafts) compared Carousel to manual Nokor needle/forceps method; results showed reduced bleeding, shorter procedure time, and no difference in post-operative wound healing or hair growth rates.
Technological Characteristics
Manual, non-powered surgical instrument. Materials: stainless steel (blade), Lustran 348 resin (body). Form factor: hand-held, ball-point pen size. Mechanical operation via button-activated internal components for graft storage and implantation. No energy source required.
Indications for Use
Indicated for patients requiring one or more hair grafting sessions involving multiple graft implantations on the scalp.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
Predicate Devices
Calivitron Hair Transplant System
Padgett Hair Transplant Punch
Submission Summary (Full Text)
{0}------------------------------------------------
## K972918
OCT 2 7 1997
Apex Medical Products 9911 W. Pico Boulevard, Suite 301 Los Angeles, CA 90035 (800) 683-8673
# XI. SUMMARY OF SAFETY AND EFFECTIVENESS
#### SPONSOR IDENTIFICATION
#### ESTABLISHMENT REGISTRATION NUMBER:
MANUFACTURER IDENTIFICATION:
#### OFFICIAL CONTACT PERSON:
Apex Medical Products, L.L.C. 9911 West Pico Blvd., Suite301 Los Angeles, CA 90035
Pending
Surgical Technology Inc. 292 East Lafayette Road St. Paul, MN 55107
Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac . MD 20854
Tel. : (301) 279 -2899 Fax : (301) 294-0126
DATE OF PREPARATION OF THIS SUMMARY:
TRADE NAME:
COMMON NAME:
CLASSIFICATION NAME AND REFERENCE:
PROPOSED REGULATORY CLASS:
DEVICE PRODUCT CODE:
PANEL CODE:
August 6, 1997
RAPID FIRE HAIR IMPLANTER CAROUSEL™
Hair Transplant Facilitator
Manual Surgical Instrument (21 CFR, Section 878.4800) Class I
GAH
80 SU
000113
{1}------------------------------------------------
### DESCRIPTION OF DEVICE:
INTENDED USE:
#### INDICATIONS FOR USE:
#### PREDICATE DEVICE:
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
#### The RAPID FIRE HAIR IMPLANTER CAROUSEL™
(Carousel) is a disposable mechanical device designed to facilitate a hair transplant. It consists of a circular Carousel designed to house hair grafts, a sharp end designed to create an opening in the skin of the scalp and a button end that activates the moving parts within the Carousel.
The RAPID FIRE HAIR IMPLANTER CAROUSEL™ is in ended to be used for hair transplantation on the scalp.
#### The RAPID FIRE HAIR
IMPLANTER CAROUSEL™ is indicated where patients require one or more hair grafting sessions during which multiple grafts have to be implanted.
The RAPID FIRE HAIR IMPLANTER CAROUSEL™ is substantially equivalent to the Calivitron Hair Transplant System (manufactured by Medicamat, S.A.)
and the Padgett Hair Transplant Punch (manufactured by Padgett Instruments){
Both the PADGETT and the CAROUSEL are non-powered simple units that use a stainless steel device for graft cutting. The CAROUSEL differs from the PADGETT in having a carousel feature that facilitates storage of the graft and speeds up the surgical procedure. These differences between these devices do not impact the safety and effectiveness in a significant way except for the improvements relating to the carousel noted above.
The CALVITRON also uses a four bladed scalpel device but utilizes a complex electrically powered system to
{2}------------------------------------------------
#### SUMMARY OF STUDIES:
provide for aspiration, compression and micromotors. The CAROUSEL is much simpler in design than the CALVITRON. In contrast the 48 kg CALVITRON, which lists 11 components to its system, including a control panel, vacuum and compression connectors, sterile arm unit, etc., the CAROUSEL is a hand-held unit roughly the size of a ball point pen.
The CAROUSEL consists of a stainless steel blade and the body composed of Lustran 348 resin. The predicate devices also have stainless steel blades which act as the patient contacting surfaces.
> Performance test results on the CAROUSEL including in vitro tests on porcine skin and a test on a human subject, validated its safety and effectiveness. In the in vitro study, visual inspection of the biopsy specimens of groups implanted manually and with the Carousel under loop magnification showed no difference in the gross morphology of the implants. In both the test and control groups the implants were positioned identically in the dermis and in both groups the implants appeared intact and undamaged. Histologic evaluation revealed no significant difference between the implanted tissue of both groups with the exception that the hair follicles implanted with the Carousel were at a somewhat more uniform depth within the porcine tissue than those of the control group.
400 grafts were implanted on the right side of a patient's scalp approximately 1 cm behind the right frontal hairline. The left side was implanted manually, using a Nokor needle and manual forceps. There was much less bleeding by the Carousel
{3}------------------------------------------------
method, resulting in a much shorter time nectical for grafi placement. The post operative wounds showed no difference between those made with the Carousel and those made with the conventional appeared. No difference in the rate of growth or the number of hairs was seen villes each of the two groups were : ーで使
.
000116
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 7 1997
Apex Medical Products, L.L.C. c/o Norman F. Estrin, Ph.D., RAC Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854
K972918 Re:
> Trade Name: Rapid Fire Hair Implanter Carousel™ Regulatory Class: I Product Code: GAH Dated: August 7, 1997 Received: August 7, 1997
Dear Dr. Estrin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Norman F. Estrin, Ph.D., RAC
devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely vours.
Acolle
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
**510(k) Number (if known):** K9722918
#### Rapid Fire Hair Implanter Carousel™ Device Name:
Indications for Use:
The Rapid Fire Hair Implanter Carousel™ is to be used for hair transplantation on the scalp. Its use is indicated where patients require one or more hair grafting sessions during which multiple grafts have to be implanted.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR Over-the-Counter U
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General Restorative Devices
972918
Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.