K223625 · Pneumeric, Inc. · GAA · Mar 28, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223625
Device Name
CapnoSpot™ Pneumothorax Decompression Indicator
Applicant
Pneumeric, Inc.
Product Code
GAA · General, Plastic Surgery
Decision Date
Mar 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Indications for Use
Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.
Device Story
CapnoSpot is a colorimetric capnography device used during decompressive thoracostomy to guide needle placement. It consists of a polycarbonate tube containing CO2-detecting paper, a silicone one-way valve to prevent gas backflow, and luer connectors. During a pneumothorax decompression procedure, the device is connected to the decompression needle. As gas passes through the device, the CO2-detecting paper changes color if CO2 is present, providing real-time visual feedback to the clinician. This objective visual indicator replaces or supplements traditional auditory assessments, helping the clinician confirm successful entry into the pleural space. By confirming accurate needle placement, the device facilitates effective decompression of tension pneumothorax, potentially improving patient outcomes by reducing complications associated with improper needle positioning.
Clinical Evidence
Bench testing only. Design verification and shelf-life testing included visual inspection, dimensional analysis, backflow testing, cracking pressure, torque, and luer leakage. Testing confirmed the device meets specifications before and after accelerated aging.
Technological Characteristics
Cylindrical polycarbonate tube housing CO2-detecting paper; silicone one-way valve; male/female luer connectors. Non-sterile, single-use device. Operates via colorimetric chemical reaction to detect CO2 gas. No electronic components or software.
Indications for Use
Indicated for patients with suspected tension pneumothorax requiring decompression. Used to guide accurate placement of decompression needles/thoracostomy devices by detecting CO2 in gas passing through the device.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2023
Pneumeric, Inc. Johnathon Aho Chief Executive Officer 823 4th Street SW, c/o Johnathon Aho Rochester, Minnesota 55902
Re: K223625
Trade/Device Name: CapnoSpot™ Pneumothorax Decompression Indicator Regulation Number: 21 CFR 878.4800 Regulation Name: Manual Surgical Instrument For General Use Regulatory Class: Class I Product Code: GAA Dated: March 1, 2023 Received: March 2, 2023
Dear Johnathon Aho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.03.28<br>13:01:59-04'00' |
|---------------------|--------------------------------------------------------------------------------|
|---------------------|--------------------------------------------------------------------------------|
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223625
## Device Name
Pneumeric™ CapnoSpot™ Pneumothorax Decompression Indicator
## Indications for Use (Describe)
Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "PNEUMERIC" in bold, black letters, with a trademark symbol next to it. To the left of the word is a gray image of lungs. The lungs are stylized and have a smooth, rounded shape. The image is set against a white background.
# 6. 510(K) SUMMARY
# 6.1 ADMINISTRATIVE INFORMATION
Date of Summary Preparation:
# 6.1.1 CONTACT INFORMATION
| Submitter/Manufacturer | Johnathon M.E. Aho M.D., Ph.D.<br>Chief Executive Officer, Pneumeric, Inc.<br>823 4th St SW<br>Rochester, MN 55902<br>United States<br>507-206-1942<br>John.aho@pneumeric-medical.com<br>jaho@mtu.edu |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Submission Contact | Johnathon M.E. Aho M.D., Ph.D.<br>Chief Executive Officer, Pneumeric, Inc.<br>823 4th St SW<br>Rochester, MN 55902<br>United States<br>507-206-1942<br>John.aho@pneumeric-medical.com<br>jaho@mtu.edu |
| Secondary Submission Contact | Jonathan Sackner-Bernstein M.D.<br>Chief Regulatory Officer, Pneumeric, Inc<br>jsb@jsbmd.com<br>+1.212.888.2589<br><br>Ming Cheng Chew<br>Regulatory Consultant,<br>Libra Medical Inc.<br>8401 73rd Ave North, Suite 63 |
Minneapolis, MN 55428
Phone: 763-232-3701 Email: mcchew@libramed.com
Veronica McDougall
{4}------------------------------------------------
Regulatory Consultant, Nagl Medtech 3451 Commerce Pkwy Miramar F1 33025 Phone: 954-507-3830 Email: vmcdougall(@nmddo.com
#### 6.1.2 DEVICE INFORMATION
| Trade Name | CapnoSpot Pneumothorax Decompression<br>Indicator |
|---------------------------|---------------------------------------------------|
| Common Name | Decompression Indicator |
| Classification Name | Needle, Aspiration and Injection, Disposable |
| Classification Regulation | 878.4800 |
| Class | I |
| Panel | General & Plastic Surgery |
| Product Code | GAA |
| FDA Documents Related to | None |
| Modified Device | C210112 |
#### PREDICATE DEVICE 6.2
The Pneumeric CapnoSpot Pneumothorax Decompression Indicator ("CapnoSpot" or "device") is substantially equivalent to the Turkel Pneumothorax Kit (K923028).
### DEVICE DESCRIPTION 6.3
The Pneumeric Capnospot Pneumothorax Decompression Indicator is a capnography device that provides objective, real-time feedback during decompressive thoracostomy. Specifically, the device detects the presence of carbon dioxide in the gas from the decompressed pneumothorax, permitting more accurate placement of the needle or thoracostomy device compared to the current standard of care based upon auditory assessments.
The Pneumeric Capnospot is a therapeutic device intended to guide the placement of pneumothorax decompression needles and thoracostomy devices.
The product consists of a cylindrical polycarbonate tube containing a carbon dioxide detecting paper. The device has a male luer connector on the distal end of the device to connect to decompression needles and thoracostomy devices, a diaphragm to prevent backflow of gases, a color changing indicator and a female luer on the proximal end of the device that can be used to connect accessories.
The product is shipped non-sterile and labeled for single use only.
### 6.4 INDICATIONS FOR USE/INTENDED USE
Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a
{5}------------------------------------------------
colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.
### PERFORMANCE DATA 6.5
Testing was performed before and after accelerated aging. The relevant design verification and shelf life tests included:
- Visual inspection ●
- Dimensionals ●
- Backflow
- Cracking Pressure ●
- Torque ●
- Luer leakage
- Animal Data ●
The design verification and shelf life testing showed that the device meets specifications before and after aging. This indicates that the device is as safe and effective as the predicate device.
### SUBSTANTIAL EQUIVALENCE 6.6
# Table 6-1 Substantial Equivalence
| Table 6-1 Substantial Equivalence | | |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device (K923028) | CapnoSpot (Subject Device) |
| Intended Use | Evacuate fluid or gas from<br>either a body cavity or the<br>pleural space while providing<br>visualization of needle<br>placement for safety | Provide a visual indication for<br>the accurate placement of a<br>pneumothorax needle |
| Indications for<br>Use | The Turkel Safety Fluid System<br>has application in:<br>-Thoracentesis Procedures –<br>Diagnostic and Therapeutic<br>-Paracentesis Procedures –<br>Diagnostic and Therapeutic | The Pneumothorax<br>Decompression Indicator is a<br>colorimetric capnography<br>device indicated to detect the<br>CO2 in the gas passing through<br>the device used to treat<br>Pneumothorax with suspected<br>tension physiology. |
| Regulation Class<br>&<br>Product Code | Class I, GCB<br>878.4200 | Class I, GAA<br>878.4800 |
| Included<br>Components | -Blunt cannula housed within a<br>sharp hollow needle<br>-In-line stopcock<br>-Color Change indicator<br>-Ball Valve<br>-One-way Valve | - Color Change indicator<br>- One-way valve |
| Presence of One-<br>Way Valve to<br>prevent backflow | Yes | Yes |
| Indicator Tube<br>and Luer Material | Unknown | Polycarbonate |
| One Way Valve<br>Material | Unknown | Silicone |
{6}------------------------------------------------
| Characteristic | Predicate Device (K923028) | CapnoSpot (Subject Device) |
|------------------|----------------------------|----------------------------|
| Indicator Method | Mechanical Spring | Colorimetric Paper |
| Sterilization | SAL of 10-6 via EtO | Non-sterile |
| Biocompatibility | ISO 10993-1 | N/A |
| Single-use only? | Yes | Same |
The differences between this device and its predicate do not raise new questions of safety or efficacy.
### 6.7 CONCLUSION
The Pneumeric CapnoSpot Pneumothorax Decompression Indicator is substantially equivalent to the predicate device (K923028) in design, function, and intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.