← Product Code [FZW](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZW) · K972109

# SPACEMAKER SURGICALBALLOON DISSECTOR/EXPANDER (K972109)

_General Surgical Innovations · FZW · Nov 14, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZW/K972109

## Device Facts

- **Applicant:** General Surgical Innovations
- **Product Code:** [FZW](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZW.md)
- **Decision Date:** Nov 14, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4800
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the Spacemaker® Surgical Balloon Dissector/Expander is to create a surgical space anywhere in the body by dissecting layers of connective tissue along natural planes of separation and provide intraoperative and subsequent tissue expansion. The Spacemaker® Surgical Balloon Dissector/Expander is a temporary subcutaneous or submuscular implant device. The indications for use include the following: breast reconstruction, limb reconstruction, correction of congenital deformities and cosmetic defects, and scar revision.

## Device Story

Handheld, manually manipulated surgical balloon device; used for tissue dissection along natural planes and tissue expansion. Input: saline or air for inflation (1-2500cc). Operation: physician inserts balloon subcutaneously or submuscularly; inflates to dissect tissue or expand; injection port/connector allows controlled inflation. Used in general, laparoscopic, endoscopic, plastic, and reconstructive surgeries. Output: physical creation of surgical space or expanded tissue volume. Benefits: facilitates surgical access and tissue expansion for reconstructive procedures. Single-use, disposable, provided sterile.

## Clinical Evidence

Bench testing only. Specific ASTM tests performed to evaluate design differences compared to predicates; results indicated no new safety or efficacy issues.

## Technological Characteristics

Materials: polymers, silicone, stainless steel. Principle: manual balloon dissection/expansion. Energy: manual inflation (saline/air). Form factor: handheld balloon with injection port/connector. Sterilization: provided sterile. Connectivity: none.

## Regulatory Identification

A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.

## Predicate Devices

- Spacemaker® and Spacemaker® II Surgical Balloon Dissector with and without Cannula ([K951878](/device/K951878.md))
- Inamed Ruiz-Cohen Intraoperative Expander ([K904265](/device/K904265.md))
- McGhan Magna-Site™ Tissue Expander ([K862203](/device/K862203.md))
- McGhan Tissue Expander ([K843704](/device/K843704.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

10/13/2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

General Surgical Innovations Cathleen Mantor Regulatory Affairs/ Quality Assurance Manager 10460 Bubb Rd. Cupertino, California 95014

Re: K972109

Trade/Device Name: Spacemaker® Surgical Balloon Dissector/expander Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: Class I Product Code: FZW

Dear Cathleen Mantor:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 14, 1997. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code FZW.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Deborah Fellhauer, OHT4: Office of Surgical and Infection Control Devices, 301-796-9570, Deborah.Fellhauer@fda.hhs.gov.

Sincerely,

Deborah A. Fellhauer - S Deborah A. Fellhauer -S 2022.10.13 13:13:29 -04'00'

Deborah A. Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an emblem. The emblem is a stylized image of a human figure in profile, with three overlapping figures suggesting a sense of community or support.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Cathleen Mantor ·Regulatory Affairs/Quality Assurance Manager General Surgical Innovations 10460 Bubb Road Cupertino, California 95014

NOV 1 4 1997

K972109 Re:

Trade Name: Spacemaker® Surgical Balloon Dissector/Expander Regulatory Class: Unclassified Product Code: LCJ Dated: September 26, 1997 Received: September 29, 1997

Dear Ms. Mantor:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the " current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions. or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Ms. Cathleen Mantor

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely vours.

Sincerely yours,

- Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

8.1 Statement of Indications for Use

510(k) Number (if known):

K972109

## Device Name

Spacemaker® Surgical Balloon Dissector/Expander

# Indications for Use

The intended use of the Spacemaker® Surgical Balloon Dissector/Expander is to create a surgical space anywhere in the body by dissecting layers of connective tissue along natural planes of separation and provide intraoperative and subsequent tissue expansion. The Spacemaker® Surgical Balloon Dissector/Expander is a temporary subcutaneous or submuscular implant device. The indications for use include the following: breast reconstruction, limb reconstruction, correction of congenital deformities and cosmetic defects, and scar revision.

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | X | OR | Over-the-Counter Use     |
|----------------------|---|----|--------------------------|
| (per 21 CFR 801.109) |   |    | (Optional format 2Jan96) |

(Division Sign-Off)

Division of General Restorative Devices

| 510(k) Number | K972109 |
|---------------|---------|
|---------------|---------|

- .

{4}------------------------------------------------

K972109

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in

accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).

#### 1.Contact Person

Cathleen Mantor RA/QA Manager General Surgical Innovations, Inc. 10460 Bubb Road Cupertino, CA 95014

(408)863-2531 (408)863-1000 - Fax

# 2.Date Summary Prepared

6/4/97

#### 3.Trade Name

Spacemaker® Surgical Balloon Dissector/Expander

### 4.Common Name

Surgical Balloon Dissector/Expander

# 5.Predicate Devices

| K951878 | Spacemaker® and Spacemaker® II Surgical Balloon Dissector with and<br>without Cannula |
|---------|---------------------------------------------------------------------------------------|
| K904265 | Inamed Ruiz-Cohen Intraoperative Expander                                             |
| K862203 | McGhan Magna-Site™ Tissue Expander                                                    |
| K843704 | McGhan Tissue Expander                                                                |

#### 6.Device Description

The Spacemaker® Surgical Balloon Dissector/Expander device is a handheld,

manually manipulated surgical balloon device which dissects tissue planes,

intraoperatively and subsequently expands the overlying tissue. This device is a

temporary subcutaneous or submuscular implant device. The Spacemaker® Surgical

Balloon Dissector/Expander is a single use, disposable device provided sterile to the

{5}------------------------------------------------

physician. The major components of the device are a balloon dissector and an injection port/connector.

# 7.Intended Use

The intended use of the Spacemaker® Surgical Balloon Dissector/Expander is to create a surgical space anywhere in the body by dissecting layers of connective tissue along natural planes of separation and provide intraoperative and subsequent tissue expansion. The Spacemaker® Surgical Balloon Dissector/Expander is a temporary subcutaneous or submuscular implant device. The indications for use include the following: breast reconstruction, limb reconstruction, correction of congenital deformities and cosmetic defects, and scar revision.

# 8.Comparison to Predicate Devices

The claim of substantial equivalence is based on the fact that the GSI Spacemaker® Surgical Balloon Dissector/Expander has the same principle of operation and intended uses as the predicate device. These differences do not raise additional questions regarding safety or efficacy compared to the predicate devices. Therefore, it can be concluded that the Spacemaker® Surgical Balloon Dissector/Expander is substantially equivalent to the predicate devices.

The following table lists the similarities and differences between the Spacemaker® Surgical Balloon Dissector/Expander and predicate devices.

{6}------------------------------------------------

| DEVICE<br>POINT OF<br>COMPARISON                                                                                             |                                                                                                                        | INAMED RUIZ-<br>COHEN<br>INTRAOPERATIVE<br>TISSUE EXPANDER<br>(K904265)                                                             | MCGHAN MAGNA-<br>SITE™ TISSUE<br>EXPANDER (K862203)                     | MCGHAN TISSUE<br>EXPANDER (K843704)                                     | SPACEMAKER®<br>SURGICAL BALLOON<br>DISSECTOR/ TISSUE<br>EXPANDER                                                                                |
|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                                                                                 | SPACEMAKER® &<br>SPACEMAKER®II<br>DISSECTOR WITH &<br>W/OUT CANNULA<br>(K951878)                                       | Intraoperative tissue<br>expansion                                                                                                  | Tissue expansion                                                        | Tissue expansion                                                        | Dissect layers of<br>connective tissue along<br>natural planes of<br>separation & provide<br>intraoperative &<br>subsequent tissue<br>expansion |
| Indications                                                                                                                  | Dissect layers of<br>connective tissue<br>along natural planes of<br>separation                                        | Where rapid creation<br>of a tissue flap is<br>needed in a single<br>surgical session for<br>scar revision or other<br>skin defects | Post mastectomy<br>reconstruction,<br>augmentation and scar<br>revision | Post mastectomy<br>reconstruction,<br>augmentation and scar<br>revision | Breast reconstruction,<br>limb reconstruction,<br>correction of congenital<br>deformities and<br>cosmetic defects, and<br>scar revision         |
| Examples of Under-<br>lying Procedures<br>Associated with<br>Surgical Dissection<br>(Expansion) Specified<br>in the Labeling | Any area of the body<br>where it is necessary<br>to separate the layers<br>of connective tissue<br>for surgical access | Unknown                                                                                                                             | Unknown                                                                 | Unknown                                                                 | General, laparoscopic,<br>endoscopic, plastic and<br>reconstructive,<br>including aesthetic,<br>surgery                                         |
| How Supplied                                                                                                                 | General, laparoscopic,<br>endoscopic, plastic<br>and reconstructive,<br>including aesthetic,<br>surgery                | Unknown                                                                                                                             | Single use                                                              | Single use                                                              | Single patient use,<br>disposable                                                                                                               |
| Principle of Operation                                                                                                       | Single patient use,<br>disposable                                                                                      | Handheld, manually<br>manipulated tissue<br>expander                                                                                | Handheld, manually<br>manipulated tissue<br>expander                    | Handheld, manually<br>manipulated tissue<br>expander                    | Handheld, manually<br>manipulated surgical<br>dissector/expander                                                                                |
| Balloon Volume<br>Range                                                                                                      | Handheld, manually<br>manipulated surgical<br>dissector<br>10-2500cc saline or air<br>inflation                        | 0.8-40cc saline (from<br>catalog and<br>Instructions for Use)                                                                       | 50-800cc                                                                | Range of sizes (and<br>shapes)                                          | 1-2500cc saline or air<br>inflation                                                                                                             |
| Balloon Shapes                                                                                                               | Include kidney, clover,<br>horseshoe, semi-round<br>and U-shaped profiles                                              | Cylindrical and<br>spherical (from<br>catalog)                                                                                      | Round and rectangular                                                   | Range of (sizes and)<br>shapes                                          | Range of (sizes and)<br>shapes                                                                                                                  |
| Balloon Function                                                                                                             | Balloon dissector                                                                                                      | Balloon expander                                                                                                                    | Balloon expander                                                        | Balloon expander                                                        | Balloon<br>dissector/expander                                                                                                                   |
| Balloon Cover                                                                                                                | Yes                                                                                                                    | No                                                                                                                                  | No                                                                      | No                                                                      | Optional                                                                                                                                        |
| Filling                                                                                                                      | Not applicable                                                                                                         | Not applicable                                                                                                                      | Injection port and connector                                            | Injection port and connector                                            | Injection port and connector                                                                                                                    |
| Communication for<br>Temporary Implant                                                                                       | Not applicable                                                                                                         | Not applicable                                                                                                                      | Palpation and magnetic                                                  | Palpation                                                               | Palpation                                                                                                                                       |
| Port Detection Method                                                                                                        | Not applicable                                                                                                         | Intraoperative, single surgical session                                                                                             | Temporary implantation; typically less than 90 days                     | Temporary implantation; less than 60 days                               | Temporary implantation; typically less than 90 days                                                                                             |
| Materials                                                                                                                    | Polymers, silicone and stainless steel                                                                                 | Silicone, polymers                                                                                                                  | Silicone, stainless steel                                               | Silicone, stainless steel                                               | Polymers, silicone and stainless steel                                                                                                          |
| Packaging<br>Configuration                                                                                                   | With or without cannula                                                                                                | With blunt needle or stopcock                                                                                                       | With or without injection port and connector                            | With or without injection port and connector                            | Without cannula, with or without injection port and connector                                                                                   |

4June97

{7}------------------------------------------------

.

:

{8}------------------------------------------------

#### Testing in Support of Substantial Equivalence: 9.

Specific ASTM tests relating to the design differences between the Spacemaker® Surgical Balloon Dissector/Expander and the predicate devices were performed to evaluate if the differences would raise new questions of safety or efficacy. The results of this testing indicated no new issues were raised.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZW/K972109](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZW/K972109)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
