K230966 · A2 Medical Systems, LLC · FZP · Sep 19, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230966
Device Name
angiOCCLUDE Ligating Clips
Applicant
A2 Medical Systems, LLC
Product Code
FZP · General, Plastic Surgery
Decision Date
Sep 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
angiOCCLUDE ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of vessel or tissue structure to be ligated based upon their experience, judgment, and needs.
Device Story
Permanent, non-absorbable, sterile titanium surgical clips; used for vessel or tissue ligation. Device consists of clips loaded into color-coded cartridges; used with compatible appliers. Surgeon loads clip from cartridge into applier; applier closes clip around vessel/tissue. Symmetrical legs pivot from centered apex during closure. Used in surgical settings by surgeons. Provides secure mechanical occlusion of vessels/tissues. Benefits include reliable hemostasis or tissue closure.
Clinical Evidence
Bench testing only. Performance data includes dimensional/geometry comparisons using precise overlays, clip closure testing, and material safety analysis. Results demonstrate equivalent function, secure clip retention in appliers, and identical release characteristics compared to the predicate device.
Technological Characteristics
Material: Implantable grade titanium. Form factor: Six sizes (micro, small-wide, medium, medium-large, large). Mechanism: Symmetrical leg closure via centered apex pivot. Sterilization: Ethylene Oxide (EO). Packaging: Peel-open blister packs (6 or 24 clips).
Indications for Use
Indicated for ligation of vessels or tissue structures during surgical procedures. Contraindicated for use as a contraceptive tubal occlusion device and for ligation of the renal artery during minimally invasive donor nephrectomies.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Teleflex Medical Horizon Ligating Clips (K901303, K982313)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 19, 2023
A2 Medical Systems, LLC Robert Hall Chief Operating Officer 6387 Technology Ave. Kalamazoo, Michigan 49009
Re: K230966
Trade/Device Name: angiOCCLUDE Ligating Clips Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP, DSS Dated: August 15, 2023 Received: August 16, 2023
Dear Robert Hall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N.
Lamichhane -S
Digitally signed by
Tek N. Lamichhane -S
Date: 2023.09.19
11:09:21 -04'00'
Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230966
Device Name angiOCCLUDE Ligating Clip
Indications for Use (Describe)
angiOCCLUDE ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of vessel or tissue structure to be ligated based upon their experience, judgment, and needs.
Type of Use (Select one or both, as applicable)
| <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# 510(k) Number: K230966
## Name, Address, Phone and Fax Number of Applicant
Owner: Andy Ambro Phone #: 336-706-6027 Address: A2 Medical Systems, LLC
6387 Technology Ave. Suite F Kalamazoo, MI 49009
### Contact Person
Robert Hall Chief Operating Officer Address: A2 Medical Systems, LLC 6387 Technology Ave. Suite F Kalamazoo, MI 49009
Date Prepared: September 18, 2023
### A. Device Name
| Trade Name: | angiOCCLUDE Ligating Clip |
|----------------------|--------------------------------|
| Common Name: | Hemostatic Clip, Ligating Clip |
| Classification Name: | Clip, Implantable |
| B. Device Classification: | Class II |
|---------------------------|--------------------------------------|
| Classification Panel: | Cardiovascular & General Surgery |
| Product Codes: | FZP per 878.4300<br>DSS per 870.3250 |
# C. Predicate Devices
Teleflex Medical Horizon Ligating Clips (510ks; K901303, K982313), Product Codes: DSS (870.3250) and FZP (878.4300)
# D. Clip Device Description
AngiOCCLUDE ligating clips are permanent implant, non-absorbable, sterile, surgical clips made from an implantable grade of Titanium and are available in six different sizes (micro, small-wide, medium, medium-large and large).
#### نن Clip Indications for Use
angiOCCLUDE ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size of vessel or tissue structure to be ligated based upon their experience, judgment, and needs.
{4}------------------------------------------------
- F. Contraindications: angiOCCLUDE ligating clips are never intended for use as a contraceptive tubal occlusion device. This product is also contraindicated when ligating the renal artery during minimally invasive donor nephrectomies.
# G. Characteristic Comparison to Predicate Device Equivalence
The A2 Medical Systems titanium ligating clips are substantially equivalent to the predicate device Horizon in design and use.
- a. Substantially Equivalent areas are:
- i. Both will close over vessels
- ii. Both have the same clip material; implantable grade of titanium
- iii. Both have the same clinical application for indications and contraindications
- iv. Both have the same mechanism of action – cartridge and clip design allows for an applier to easily load a clip and clips are then removed for use. When the clip is closing the symmetrical legs pivots in from a centered apex as clips are closed.
- > Of the five sizes provided MC, SML, M, M/L, and LG; they have the same geometry.
- vi. Both have cartridges that are color coded to match applier ring colors
- vii. Both use traditional EO sterilization processing
- viii. Both use peel open blister packs with either 6 or 24 clips each.
- ix. Both use 4-slides to form the clips
- b. Differences:
- i. A2 Medical Systems titanium clips are made in the USA and Horizon is made in Mexico.
## H. Summary of Performance Data
AngiOCCLUDE ligating clips are made of an implantable grade of Titanium with similar dimensions and metallurgical characteristics as the Teleflex Medical Horizon ligating clips.
- a. Substantial Equivalence analyses preformed:
- i. A2 Medical Systems angiOCCLUDE clip to predicate clip dimensional comparison was completed at an open and closed state within an applier that resulted in having the same function as Horizon when clips are applied and closed.
- ii. Both material comparisons prove that the formed and sterilized material used for the angiOCCLUDE clip is as safe and the same as the material used in Teleflex Horizon.
- iii. Quantitative - angiOCCLUDE clip closure testing proved that the clips are secure in the applier, and release from applier's jaws equivalent to Horizon.
- iv. Qualitative - dimensional/geometry comparisons using precise overlays prove that angiOCCLUDE and Teleflex Horizon clips are the same size.
- v. Mechanism of action is the same between angiOCCLUDE and Horizon; cartridge and clip design allows for an applier to easily load a clip, and clips are then removed for use. When the clip is closing the symmetrical legs pivots in from a centered apex as clips are closed.
In summary, the A2 Medical Systems titanium Ligation clips are the same in design, material, and function to the predicate device.
#### l. Statement of Substantial Equivalence
The angiOCCLUDE® Ligating Clips are substantially equivalent to their predicate device; the Teleflex Medical Horizon® Ligating Clips (K901303, K982313), based upon similarities in intended use, design, principles of operation and performance specifications.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.