← Product Code [FZP](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP) · K172742

# AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (PRO2) (K172742)

_AtriCure, Inc. · FZP · Oct 11, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K172742

## Device Facts

- **Applicant:** AtriCure, Inc.
- **Product Code:** [FZP](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP.md)
- **Decision Date:** Oct 11, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The AtriClip LAA Exclusion System is indicated for the occlusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.

## Device Story

AtriClip LAA Exclusion System consists of sterile, single-use, self-closing implantable clip preloaded on disposable clip applier; used by cardiac surgeons during open cardiac procedures. Device provides uniform pressure across LAA to ensure secure occlusion; clip remains as permanent implant. Applier features handle, shaft, suture anchors, and articulation controls for deployment. Modifications to PRO2 applier address spring return force consistency and manufacturability. Surgeon uses direct visualization (with or without endoscopic assistance) to position and deploy clip. Device benefits patient by providing consistent, reproducible LAA occlusion.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance verified through mechanical and reliability testing.

## Technological Characteristics

Implantable clip and manual applier system. Clip available in 35, 40, 45, and 50 mm lengths. Materials include medical-grade metals and plastics. Mechanical actuation via handle-controlled deployment loop. Sterile, single-use device.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- AtriClip LAA Exclusion System ([K163261](/device/K163261.md))

## Reference Devices

- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ([K093679](/device/K093679.md))
- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ([K131107](/device/K131107.md))
- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ([K142120](/device/K142120.md))
- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ([K150996](/device/K150996.md))
- AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ([K160454](/device/K160454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2017

AtriCure, Inc. Melissa Smallwood Regulatory Affairs Specialist 7555 Innovation Wav Mason, Ohio 45040

Re: K172742

Trade/Device Name: AtriClip® LAA Exclusion System with preloaded Gillinov-Cosgrove® Clip (PRO2) Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP Dated: September 8, 2017 Received: September 12, 2017

Dear Ms. Smallwood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172742

Device Name

AtriClip®LAA Exlusion System with Gillinov-Cosgrove® Clip

Indications for Use (Describe)

The AtriClip LAA Exclusion System is indicated for the occlusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.

Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)        </span> </span> </span> </span> </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)        </span> </span> </span> </span> </span> </span>                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the AtriCure logo. The logo is in blue and orange. The word "Atri" is in blue, and the word "Cure" is in orange. There is a blue dot in the middle of the "C" in "Cure". There is a registered trademark symbol next to the "e" in "Cure".

# 510(k) Summary

### l. Submitter

| Manufacturer:        | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 1 (513) 755-4100 |
|----------------------|-----------------------------------------------------------------------------------|
| Contact Person:      | Melissa Smallwood<br>Regulatory Affairs Specialist                                |
| Alternate Contact:   | Jonathan McElwee<br>Manager, Regulatory Affairs                                   |
| Date Prepared:       | 09/08/2017                                                                        |
| II.<br>Device        |                                                                                   |
| Name of Device:      | AtriClip® LAA Exclusion System with preloaded Gillinov-Cosgrove® Clip (PRO2)      |
| Common Name:         | Implantable Clip and Clip Applier                                                 |
| Classification Name: | Implantable Clip and Clip Applier (21 CFR 878.4300)                               |
| Regulatory Class:    | Class II                                                                          |
| Product Code:        | FZP                                                                               |

### lll. Predicate Device

The device proposed for modification in this submission is the AtriClip LAA Exclusion System cleared via K163261 on May 19, 2017.

The following reference devices were also used in this submission:

- . K093679 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip
- K131107 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ●
- K142120 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip ●
- K150996 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip
- . K160454 AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip

### IV. Device Description

The AtriClip PRO2 LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier along with a selection guide. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure occlusion of the left atrial

{4}------------------------------------------------

appendage (LAA). The clip is then deployed and is left as a permanent implant. The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm.

The Clip Applier is a disposable device with a handle, shaft, suture anchors, articulation controls, and deployment loop which contains the Clip. This Special 510(k) contains modifications to the AtriClip LAA Exclusion System (PRO2) which are intended to 1.) mitigate inconsistencies of spring return force post deployment of the clip, and 2.) aid in manufacturability of the device.

### V. Indications For Use

The AtriClip LAA Exclusion System is indicated for the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.

Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.

### VI. Comparison Of Technological Characteristics With The Predicate Device

Minor dimensional modifications were made to internal components of the Clip Applier and a minor internal component was replaced with an equivalent component in the handle of the Clip Applier.

- The devices have the same intended use, and;
- No changes were made in operating principle, or specifications of performance, and;
- . Both the previously cleared PRO2 Appliers and PRO2 with proposed modifications use the Gillinov-Cosgrove Clip, and;
- . Both the previously cleared proposed modifications of the PRO2 are made of similar patient contacting materials (medical grade metals and plastics) with long and safe histories of use.
- The results of the verification and validation testing:
	- o Demonstrated equivalency in performance
	- Did not raise any new issues of safety o

The modifications to the proposed AtriClip LAA Exclusion System are intended to 1.) mitigate inconsistencies of spring return force post deployment of the clip, and 2.) aid in manufacturability of the device.

### VII. Performance Data

## Non-clinical Bench Testing

- Mechanical Testing ●
- Reliability Testing

### VIII. Conclusions

The modified AtriClip LAA Exclusion System (PRO2) is equivalent to the previously cleared AtriClip LAA Exclusion System (PRO2) as there is no change to intended use, operating principals, or function of the device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K172742](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K172742)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
