← Product Code [FZP](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP) · K150840

# Ligaclip 10mm M/L Endoscopic Rotating Multiple Clip Applier (K150840)

_Ethicon Endo-Surgery, LLC · FZP · Sep 10, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K150840

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, LLC
- **Product Code:** [FZP](/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP.md)
- **Decision Date:** Sep 10, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Ligaclip® Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

## Device Story

Sterile, single-patient use endoscopic instrument; used for ligation of tubular structures or vessels. Device features pistol handle, 360° rotating shaft (10mm diameter), and 20 medium-large titanium clips. Operated by surgeon via trocar; one-handed technique enabled by handle-mounted rotation knob. Mechanism advances clips individually after each firing. Includes visual indicator (yellow bar) when 3 or fewer clips remain; trigger resistance increases when empty. Provides secure vessel occlusion; benefits patient by facilitating efficient, reliable hemostasis during endoscopic procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance assessed via ex-vivo testing including trocar compatibility, clip retention strength, ratchet mechanism strength, ergonomic torque, tactile lockout function, clip indicator performance, and clip security. Hydrodynamic testing on excised porcine vessels demonstrated clip leak resistance. Biocompatibility evaluated per ISO 10993-1.

## Technological Characteristics

Pistol handle, rotation knob, 10mm diameter shaft (28.9 cm length). Low-glare shaft material. Contains 20 medium-large titanium clips. Mechanical sensing for clip count (yellow indicator bar) and empty-device lockout (increased trigger force). Manual actuation. Sterile, single-use.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier (submitted as Ligaclip® 20/20 Multiple Clip Applier, [K864102](/device/K864102.md))

## Reference Devices

- Ligamax™ 5 ([K110699](/device/K110699.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 10, 2015

Ethicon Endo-Surgery, LLC % Ms. Emily Kruetzkamp Ethicon Endo-Surgery, Incorporated 4545 Creek Road Cincinnati. Ohio 45245

Re: K150840

Trade/Device Name: LIGACLIP 10mm M/L Endoscopic Rotating Multiple Clip Applier Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: August 7, 2015 Received: August 11, 2015

Dear Ms. Kruetzkamp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{1}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K150840

Device Name

LIGACLIP 10mm M/L Endoscopic Rotating Multiple Clip Applier

Indications for Use (Describe)

The Ligaclip® Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

#### Company

Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969

#### Contact

Emily Kruetzkamp, Manger, Regulatory Affairs Ethicon Endo-Surgery, Inc. Telephone: 513.337.1546 Email: ekruetzk@its.jnj.com

| Date Prepared:             | March 26, 2015                                                                                                           |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Trade Name:                | Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier                                                             |
| Common Name:               | Clip Applier                                                                                                             |
| Classification Name:       | Clip, Implantable                                                                                                        |
| Device Class:              | Class II                                                                                                                 |
| Classification Regulation: | 878.4300                                                                                                                 |
| Panel:                     | 79, General and Plastic Surgery                                                                                          |
| Classification Code:       | FZP                                                                                                                      |
| Predicate Device           | Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier submitted as Ligaclip® 20/20 Multiple Clip Applier, K864102 |
| Reference Device           | Ligamax™ 5, K110699                                                                                                      |

#### Device Description

The Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier is a sterile, single-patient use device, designed to provide a means of ligation on tubular structures or vessels through an appropriately-sized trocar. The device contains 20 medium-large titanium clips that individually advance after each firing. The device shaft rotates 360° in either direction. The rotating knob is located on the handle to allow for a one-handed technique.

#### Indications for Use

The Ligaclip Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

#### Technological Characteristics

The instrument configuration consists of a pistol handle, a rotation knob, and a shaft with an outer diameter of approximately 10 mm and length of 28.9 cm. The shaft is made of a low glare material that minimizes reflective distortion. At the distal end of the shaft are the jaws, which form ligating clips. The force to squeeze the trigger increases when no clips remain in the device. The shaft contains a yellow clip counter indicator bar, which appears yellow when only 3 clips or fewer remain in the device.

{4}------------------------------------------------

## Performance Data

Ex-vivo tests were performed to verify that the performance of the Subject Device, Ligaclip 10mm M/L Endoscopic Rotating Multiple Clip Applier instrument, meets the definition of substantial equivalence to the Predicate Device, Ligaclip 10mm M/L Endoscopic Rotating Multiple Clip Applier submitted as Ligaclip® 20/20 Multiple Clip Applier instrument. Device performance was assessed against design requirements. Bench testing includes trocar compatibility, retention strength of unformed clip in jaws, ratchet mechanism strength, ergonomic torque to fire device, tactile indication torque for "no-clip lockout" function, low clip indicator bar performance, clip occlusion of the formed clip, and clip security of the formed clip. Hydrodynamic testing on excised porcine vessels was included to demonstrate clip leak resistance.

## Biocompatibility Testing

The biocompatibility evaluation for the Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier device was conducted in accordance with the FDA Blue Book Memorandum #G95- 1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. Pyrogen Testing was conducted.

The conclusions of the testing criteria demonstrate that the Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier device is substantially equivalent to the legally marketed Predicate Device, K864102.

This submission does not include data from Clinical Studies.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K150840](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FZP/K150840)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
