K143644 · Covidien · FZP · Sep 14, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143644
Device Name
Premium Surgiclip¿; Endo Clip¿; AcuClip¿
Applicant
Covidien
Product Code
FZP · General, Plastic Surgery
Decision Date
Sep 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Premium Surgiclip™ (All models) The Premium Surgiclip™ clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings Endo Clip™ M, ML, L The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings Endo Clip™ 5mm The Endo Clip™ clip applier has applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings Endo Clip™ II The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings AcuClip™ The instrument is primarily indicated for patients undergoing laparoscopic surqical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.
Device Story
Manual surgical clip appliers; used in open or endoscopic/laparoscopic procedures. Devices consist of shaft, handle, and integrated cartridge containing titanium clips. Physician places jaw around vessel/tubular structure; squeezing handle closes clip around tissue; releasing handle automatically loads next clip. AcuClip features right-angle hook for tissue access. Used by surgeons to achieve hemostasis or occlude ducts; benefits include secure vessel ligation and anatomical marking. No electronic or software components.
Clinical Evidence
No clinical data provided. Bench testing and in-vivo performance testing were not conducted as the submission pertains to a labeling change.
Technological Characteristics
Manual mechanical clip appliers. Components constructed from materials compliant with ISO 10993-1. Features titanium ligating clips. Available in various shaft lengths and clip sizes. Designed for use through trocar sleeves or open surgical access. No energy source, connectivity, or software.
Indications for Use
Indicated for patients undergoing surgical or endoscopic procedures requiring occlusion of vessels, ducts, or other tubular structures, and for radiographic marking or anatomical structure identification during surgery.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 14, 2015
Covidien Mr. Michael Koczocik Product Specialist, Regulatory Affairs 60 Middletown Ave North Haven, Connecticut 06473
Re: K143644
Trade/Device Name: Premium Surgiclip™; Endo Clip™; AcuClip™ Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: August 4, 2015 Received: August 7, 2015
Dear Mr. Koczocik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square made up of smaller squares of varying shades of blue on the left. To the right of the square is the word "COVIDIEN" in a dark blue sans-serif font.
Form Approved: OMB No. 0910-0120
to occlude vessels and other
| | Food and Drug Administration | Expiration Date: January 31, 2017 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| | Indications for Use | See PRA Statement below. |
| 510(k) Number (if known) | K143644 | |
| Device Name | Premium Surgiclip | |
| Indications for Use (Describe) | The Premium Surgiclip <sup>TM</sup> clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings | |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Servaces (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square made up of smaller squares on the left and the word "COVIDIEN" in blue on the right. The word "COVIDIEN" is in a sans-serif font.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### Indications for Use
510(k) Number (if known) K143644
Device Name Endo Clip™M
#### Indications for Use (Describe)
Endo Clip™M ML, L - The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings.
Endo Clip™ 5mm - The Endo Clip™ clip applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings.
Endo Clip™ II - The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "COVIDIEN" in blue font. To the left of the word is a blue square with a white rectangle in the middle. The square is divided into four smaller squares, with the top left and bottom right squares being a darker shade of blue than the top right and bottom left squares. The logo is simple and modern, and the word "COVIDIEN" is easy to read.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
### Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### Indications for Use
510(k) Number (if known)
#### K143644
Device Name AcuClip™
#### Indications for Use (Describe)
The instrument is primarily indicated for patients undergoing laparoscopic surgical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}------------------------------------------------
# 510(k) Summary
This 510(k) summary of data used to demonstrate substantial equivalence is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR § 807.98
| NAME: | COVIDIEN |
|----------------------------|------------------------------------------------------------------------------------------------------|
| ADDRESS: | 60 Middletown Avenue<br>North Haven, Connecticut 06473 USA |
| CONTACT PERSON: | Michael Koczocik<br>Product Specialist, Regulatory Affairs |
| PHONE NUMBER: | (203) 492-6312 |
| FAX NUMBER: | (203) 492-5029 |
| DATE PREPARRED: | March 13, 2015 |
| TRADE/PROPRIETARY NAME: | Premium Surgiclip™<br>Endo Clip™<br>AcuClip™ |
| COMMON/USUAL NAME: | Surgical clip appliers with implantable clips |
| CLASSIFICATION NAME: | Implantable Clip per 21 CFR § 878.4300 |
| PRODUCT CODE: | FZP |
| CLASSIFICATION PANEL NAME: | General and Plastic Surgery |
| FDA PANEL NUMBER: | 79 |
| DEVICE CLASS: | Pursuant to 21 CFR § 878.4300 implantable clips<br>are Class II devices |
| PREDICATE DEVICE(S): | Premium Surgiclip™ III (K853650), (K142869)<br>Endo Clip™ (K061288), (K883018)<br>AcuClip™ (K920599) |
{6}------------------------------------------------
## INTENDED USE:
Premium Surgiclip™ (All models) The Premium Surgiclip™ clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings
Endo Clip™ M, ML, L The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings
Endo Clip™ 5mm The Endo Clip™ clip applier has applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings
Endo Clip™ II The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings
AcuClip™ The instrument is primarily indicated for patients undergoing laparoscopic surqical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.
{7}------------------------------------------------
Device Descriptions
## Reorder Codes: Premium Surgiclip™
134031 - Premium Surgiclip™ M-11.5 134044 - Premium Surgiclip™ M-9.75 134046 - Premium Surgiclip™ S-9.0 134048 - Premium Surgiclip™ L-13.0 134051 - Premium Surgiclip™ II M-9.75 134053 - Premium Surgiclip™ II M-11.5
## Description of Premium Surgiclip™ Devices
The Premium Surgiclip™ clip applier consists of an applier shaft with attached handles and integrated cartridge containing 15 or 20 titanium clips. The clip applier jaw is placed around a vessel or other tubular structure. As the handles of the applier are brought together, the clip is closed around the vessel or structure. As the handles are released, a new clip is automatically loaded into the clip applier jaw.
Models available are:
Premium Surqiclip™ M – 11.5 and Premium Surgiclip™ II M-11.5 - medium sized clips with an 11.5 inch shaft
Premium Surgiclip™ M - 9.75 and Premium Surgiclip™ II M-9.75 - medium sized clips with an 9.75 inch shaft
Premium Surgiclip™ S-9.0 – small sized clips with an 9.0 inch shaft
Premium Surgiclip™ L-13.0 - large sized clips with an 13.0 inch shaft
## Reorder Codes: Endo Clip™
176615 – Endo Clip™ ML 176619 – Endo Clip™ M 176625 — Endo Clip™ L
The Endo Clip™ clip applier contains 20 titanium clips in the ML and M sizes, and 15 titanium clips in the L size. They are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.
{8}------------------------------------------------
176620 – Endo Clip™ 5mm
The Endo Clip™ 5mm clip applier contains 12 titanium clips. The applier is designed for introduction and use through an appropriately sized Covidien™ trocar sleeve, or larger with the use of a converter. The overall length of the shaft is approximately 28 cm (11").
176657 – Endo Clip™ II
The Endo Clip™ II clip applier contains 20 titanium clips. The appliers are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.
## Reorder Codes: AcuClip™
OMS-A8 - AcuClip™
The AcuClip™ right angle clip applier is loaded for delivery of twenty (20) 8 mm titanium ligating clips. It is designed for introduction and use through all appropriate sized Covidien™ trocar sleeves or larger trocar sleeves with the use of a converter. The distal portion of the AcuClip™ right angle clip applier forms a hook that is placed around the vessel or other tissue structure to be occluded. The clips are advanced into the jaws by squeezing the handle approximately half-way. The clips are then closed by fully depressing the handle.
SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTICS OF THE PROPOSED AND PREDICATE DEVICE(S)
MATERIALS:
PERFORMANCE DATA:
The clip appliers are designed to occlude vessels and other tissue and tubular structures. The technological characteristics of the proposed devices remain the same as the 510(k) cleared predicate devices. (Premium Surgiclip™ III K853650, K142869),(EndoClip™ K883018 & EndoClip™ III K061288), & (Acuclip K920599)
All components of the Premium Surgiclip™ Endo Clip™, and AcuClip™ are comprised of materials which are in accordance with ISO 10993-1
In-vitro and In-vivo performance testing were not conducted as a requirement of the labeling change to add a contraindication.
{9}------------------------------------------------
# CONCLUSION:
The addition of the contraindication does not alter the performance of the devices.The Premium Surgiclip™, Endo Clip™, and AcuClip™ are
determined to be substantially equivalent to predicate devices (Premium Surgiclip™ III) K142869),(EndoClip K061288), & (Acuclip K920599)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.