K121194 · Covidien · FZP · May 3, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121194
Device Name
ENDO CLIP III 5MM CLIP APPLIER
Applicant
Covidien
Product Code
FZP · General, Plastic Surgery
Decision Date
May 3, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ENDO CLIP™ III 5mm Clip Applier is primarily indicated for patients undergoing laparoscopic surgical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures, and for radiographic marking.
Device Story
ENDO CLIP™ III 5mm Clip Applier is a manual surgical instrument used in laparoscopic procedures. Device contains 16 titanium clips; features trigger handle, shaft rotation knob, clip counter window, and 33 cm shaft with distal jaws. Operated by surgeons in clinical/OR settings. Surgeon squeezes handle to load titanium clip into jaws and close them onto target vessel or structure. Provides mechanical occlusion of vessels/ducts and radiographic marking. Device is introduced through trocar sleeve or converter. Benefits include reliable vessel ligation during minimally invasive surgery.
Clinical Evidence
No clinical data. Evidence consists of comprehensive bench testing (in-vitro) and animal studies (in-vivo). In-vitro tests verified mechanical performance (firing force, jaw deflection, clip formation, burst pressure). In-vivo studies evaluated hemostasis, jaw trauma, and blunt dissection capabilities. Results confirmed performance equivalent to predicate.
Technological Characteristics
Manual surgical clip applier; 33 cm shaft; 16 titanium clips. Materials comply with ISO 10993-1. Mechanical operation via trigger handle and jaw mechanism. No energy source or software.
Indications for Use
Indicated for patients undergoing laparoscopic surgical procedures requiring dissection and occlusion of blood vessels, ducts, and other tubular structures, or for radiographic marking.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
K12494
page 1/a
# Covidien 510(k) Premarket Notification
## 510(k) Summary of Safety and Effectiveness
MAY - 3 2012
| SUBMITTER: | Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473<br>Tel. No.: (203) 492-5496<br>Fax No.: (203) 492-5029 |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Angela Van Arsdale<br>Regulatory Affairs Product Specialist |
| DATE PREPARED: | April 02, 2012 |
| TRADE/PROPRIETARY NAME: | ENDO CLIP™ III 5mm Clip Applier |
| COMMON/USUAL NAME: | Implantable Clip |
| CLASSIFICATION NAME: | Implantable Clip |
| PREDICATE DEVICE(S): | ENDO CLIP™ III 5mm Clip Applier (K071406) |
| DEVICE DESCRIPTION: | The ENDO CLIP™ III 5mm Clip Applier contains 16 titanium<br>clips. The applier is designed for introduction and use through<br>an appropriately sized trocar sleeve, or larger with the use of a<br>converter. The ENDO CLIP™ III 5mm Clip Applier consists of<br>a trigger handle, shaft rotation knob, clip counter window and a<br>33 cm shaft with jaws at its distal end. Squeezing the handle<br>places a titanium clip in the jaws and closes the jaws to close<br>the clip on the vessel or structure. |
| INTENDED USE: | The ENDO CLIP™ III 5mm Clip Applier is primarily indicated<br>for patients undergoing laparoscopic surgical procedures<br>involving dissection and occlusion of blood vessels, ducts and<br>other tubular structures, and for radiographic marking. |
| TECHNOLOGICAL<br>CHARACTERISTICS: | The ENDO CLIP™ III 5mm Clip Applier is<br>identical to the predicate device in terms of intended use and it<br>operates in a similar manner as the predicate device. |
| MATERIALS: | All patient contact components of the ENDO CLIP™ III 5mm<br>Clip Applier are comprised of materials which are in<br>accordance with ISO Standard 10993-1. |
·
{1}------------------------------------------------
K121194
### Covidien 510(k) Premarket Notification
P
#### PERFORMANCE DATA:
In-vitro and in-vivo tests were performed to verify that the ENDO CLIP™ III 5mm Clip Applier was substantially equivalent to the predicate device (K071406) in occluding vessels and other tubular structures and to validate that the ENDO CLIP™ III 5mm Clip Applier performed as intended.
The following describes the testing performed:
- · In Vitro Insertion / Removal force
- · In Vitro Counter Reliability Test
- · In Vitro Firing Force / Clip Formation
- · In Vitro Interlocking Break Force
- · In Vitro Clip Formation Test
- · In Vitro Perpendicular Clip Pulls
- · In Vitro Paralle! (Axial) Clip Pulls
- · In Vitro Jaw Deflection Resistance
- In Vitro Top Load on Jaw .
- In Vitro Bottom Load on Jaw with wedge plate . engaged prior to clip leading
- In Vitro Bottom Load on Jaw with wedge plate � disengaged prior to clip loading
- In Vitro Side Load on Jaw (Vessel approximated . first prior to trigger squeeze)
- . In Vitro Side Load on Jaw (Vessel approximated after wedge plate feed)
- In Vitro Media! Load on Jaw .
- · In Vitro Jaw Twist Resistance
- · In Vitro Clip next to clip proximity test
- · Ex Vivo Burst Evaluation
- · Ex Vivo Perpendicular Clip Pull
- · Ex Vivo Parailel Clip Pull
- · Ex Vivo Clip Twist Normal Firing (jaw tips vertical to vessel and no underlying tissue)
- · Ex Vivo Clip Twist with Shaft Rotation Normal firing with shaft rotation (jaw tips twisted 90 to vessel and no underlying tissue)
- · Ex Vivo Clip Twist Firing embedded tissue (jaw tips vertical to vessel with underlying tissue)
- · In Vivo Hemostasis Evaluation
- · In Vivo Jaw Trauma
- · In Vivo Blunt Dissection Evaluation
- · In Vivo Clip Trauma
- · In Vivo Jaw Deflection (Twist) Resistance
Based on the testing data, substantial equivalence has been established.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Covidien % Ms. Angela Van Arsdale Regulatory Affairs Product Specialist 60 Middletown Avenue North Haven, Connecticut 06473
MAY - 3 2012
Re: K121194
Trade/Device Name: ENDO CLIP™ III 5mm Clip Applier Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: FZP, GDO Dated: April 17, 2012 Received: April 19, 2012
Dear Ms. Arsdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must corply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
. Page 2 - Ms. Angela Van Arsdale
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yo
Oth. Der Lurr
Dr.
Mark N. Melkerson . Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: ENDO CLIP™ III 5mm Clip Applier
Indications For Use:
The ENDO CLIP™ III 5mm Clip Applier is primarily indicated for patients undergoing laparoscopic surgical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures, and for radiographic marking.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|-----------------------------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Surgical, Orthopedic, and Restorative Devices | |
| 510(k) Number | K121194 |
|---------------|---------|
|---------------|---------|
Page 1 of ____