← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K211296

# Surgical face mask (K211296)

_Sunsmed Protective Products , Ltd. · FXX · Sep 11, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX/K211296

## Device Facts

- **Applicant:** Sunsmed Protective Products , Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Sep 11, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Surgical face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

## Device Story

Surgical face mask; single-use, three-layer, flat-pleated design; worn over nose and mouth; secured by elastic ear straps; includes HDPE nose piece for fit. Protects patient and healthcare personnel from microorganisms, body fluids, and particulates; reduces exposure to blood/body fluids. Used in clinical/infection control settings. Non-sterile, disposable.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing confirmed compliance with ASTM F2100-19 Level 2 standards: BFE (98.9-99.7%), differential pressure (4.1-5.0 mmH2O/cm²), PFE (98.03-98.86%), synthetic blood penetration resistance (passed at 120 mmHg), and Class 1 flammability. Biocompatibility testing (cytotoxicity, irritation, sensitization) passed.

## Technological Characteristics

Three-layer construction: nonwoven inner/outer layers, melt-blown middle layer. Materials: polypropylene, HDPE nose piece, polyester/spandex ear straps. Dimensions: 17.5cm x 9.5cm. ASTM F2100 Level 2 performance. Non-sterile, single-use, disposable.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K182515](/device/K182515.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

September 11, 2021

Sunsmed Protective Products Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K211296

Trade/Device Name: Surgical face mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 5, 2021 Received: August 10, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Clarence W. Murray III -S

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K211296

Device Name Surgical face mask

#### Indications for Use (Describe)

Surgical face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary — K211296

This 510(k) summary is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's information

Name: Sunsmed Protective Products Ltd. Address: No.18, Industrial park, Maozui Town, Xinatao City, 433000, Hubei Province, China Phone Number: +86-15271236397 Contact: Ms. Joyce Lee Date of Preparation: 16/03/2021

#### Designated Submission Correspondent

Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device information

Trade name: Surgical face mask Common name: Surgical face mask Classification name: Mask, Surgical Model(s): ear strap, 175×95mm

#### 3.0 Classification

Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Surgical apparel Panel:

#### 4.0 Predicate device information

Manufacturer: Wuhan Dymex Healthcare Co., Ltd Device: Surgical Face Mask 510(k) number: K182515

#### 5.0 Indication for Use Statement

Surgical face mask is intended to be worn to protect both the patient and healthcare

{4}------------------------------------------------

personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

#### 6.0 Device description

The Surgical face mask is single use, three-layer, flat-pleated style with ear straps and nose piece. The mask is manufactured with three layers, the inner and outer layers are made of nonwoven fabrics, and the middle layer is made of melt blown fabrics. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The elastic ear straps are not made with natural rubber latex. The nose piece on the layers of facemask is to allow the user to fit the facemask around their nose, which is made of HDPE (high density polyethylene). The Surgical face mask will be provided in blue. The masks are sold non-sterile and are intended to be single use, disposable devices.

| Table 1 - General Comparison |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                      |                          |         |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|---------|
| Item                         | Proposed device                                                                                                                                                                                                                                                                                                                                                                                     | Predicated device                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison               |         |
| Product Code                 | FXX                                                                                                                                                                                                                                                                                                                                                                                                 | FXX                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                     |         |
| Regulation No.               | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                     | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                      | Same                     |         |
| Class                        | II                                                                                                                                                                                                                                                                                                                                                                                                  | II                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                     |         |
| Product name                 | Surgical face mask                                                                                                                                                                                                                                                                                                                                                                                  | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                                   | -                        |         |
| 510(k) No.                   | K211296                                                                                                                                                                                                                                                                                                                                                                                             | K182515                                                                                                                                                                                                                                                                                                                                                                                                              | -                        |         |
| Models                       | ear strap, 175×95mm                                                                                                                                                                                                                                                                                                                                                                                 | ear strap                                                                                                                                                                                                                                                                                                                                                                                                            | -                        |         |
| Intended Use                 | The Surgical face mask is<br>intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. It is intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. This is a<br>single use, disposable<br>device(s), provided non<br>sterile. | The Surgical Face Masks are<br>intended to be worn to protect<br>both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This<br>is a single use, disposable<br>device(s),<br>provided non-sterile. | Same                     |         |
| OTC use                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                     |         |
| Composite                    | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                              | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                               | Similar                  |         |
| Material                     | Internal<br>layer                                                                                                                                                                                                                                                                                                                                                                                   | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                           | Spun-bond polypropylene  | Similar |
|                              | Middle<br>layer                                                                                                                                                                                                                                                                                                                                                                                     | Melt blown<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                          | Melt blown polypropylene | Similar |
|                              | External                                                                                                                                                                                                                                                                                                                                                                                            | Spun-bond                                                                                                                                                                                                                                                                                                                                                                                                            | Spun-bond polypropylene  | Similar |

## 7.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

|  | layer                | polypropylene                |                                |         |
|--|----------------------|------------------------------|--------------------------------|---------|
|  | Nose<br>piece        | High Density<br>Polyethylene | Malleable polyethylene<br>wire | * Gap 1 |
|  | ear strap            | Polyester, spandex           | spandex                        | * Gap 2 |
|  | Color                | Blue                         | Yellow                         | * Gap 3 |
|  | Dimension (Length)   | $17.5cm\pm0.5cm$             | $17.5cm\pm0.2cm$               | * Gap 4 |
|  | Dimension<br>(Width) | $9.5cm\pm0.5cm$              | $9.5cm\pm0.2cm$                | * Gap 5 |

| Dimension<br>(Width) |             |             |      |
|----------------------|-------------|-------------|------|
| Sterility            | Non-Sterile | Non-Sterile | Same |
| Single Use           | Yes         | Yes         | Same |
| Sterile              | No          | No          | Same |
| ASTM F2100<br>Level  | Level 2     | Level 2     | Same |

* Gap analysis:

Gap 1-3: the two devices have some difference in materials and product color. Gap 4-5: the two devices share same dimensions otherwise the tolerance is different, the little deviation in tolerance..

#### 8.0_Clinical Test Conclusion

No clinical study implemented for the Surgical face mask.

#### 9.0Non-Clinical Test Conclusion

The proposed device was tested and conformed to the related recognized standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical face masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.

| Table 2 - Performance Testing |  |
|-------------------------------|--|
|-------------------------------|--|

| Items                                        | Purpose                                                                                             | Acceptance Criteria<br>(Level 2, ASTM<br>F2100-19) | Result                    |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------|---------------------------|
| Bacterial filtration<br>efficiency (BFE) (%) | The purpose of<br>the test is to<br>evaluate the<br>Bacterial filtration<br>efficiency (BFE)<br>(%) | ≥98                                                | 98.9~99.7%<br>Pass        |
| Different pressure<br>(mmH2O/cm²)            | The purpose of<br>the test is to<br>evaluate the<br>Different pressure<br>(mmH2O/cm²)               | <6.0 mmH2O/cm²                                     | 4.1~5.0 mmH2O/cm²<br>Pass |

{6}------------------------------------------------

| Sub-micron particulate<br>filtration efficiency at 0.1<br>micron, % (PFE)                       | The purpose of<br>the test is to<br>evaluate the Sub-<br>micron particulate<br>filtration efficiency<br>at 0.1<br>micron, % (PFE)                         | ≥98                                         | 98.03~98.86%<br>Pass                                                                                                                                                                       |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Resistance to penetration<br>by synthetic blood,<br>Minimum pressure in<br>mmHg for pass result | The purpose of<br>the test is to<br>evaluate the<br>Resistance to<br>penetration by<br>synthetic blood,<br>Minimum<br>pressure in mmHg<br>for pass result | 29 of 32 test articles<br>passed at 120mmHg | Test 1: 30 of 32<br>test articles<br>passed at<br>120mmHg;<br>Test 2: 31 of 32<br>test articles<br>passed at<br>120mmHg;<br>Test 3: 31 of 32<br>test articles<br>passed at<br>120mmHg Pass |
| Flammability                                                                                    | The purpose of<br>the test is to<br>evaluate the<br>Flammability                                                                                          | Class 1                                     | Class 1, Non Flammable<br>Pass                                                                                                                                                             |

#### Table 3 - Biocompatibility Testing

| Item          | Proposed Device                                                 | Result |
|---------------|-----------------------------------------------------------------|--------|
| Cytotoxicity  | Under the conditions of the study, the device is noncytotoxic.  | Pass   |
| Irritation    | Under the conditions of the study, the device is nonirritating. | Pass   |
| Sensitization | Under the conditions of the study, the device is nonsensitizing | Pass   |

#### 10.0 Conclusion

The conclusions drawn from the nonclinical test demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX/K211296](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX/K211296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
