← Product Code [FXX](/productcode/FXX) · K210020

# Medical Mask (K210020)

_Zhende Medical Co., Ltd. · FXX · Apr 28, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210020

## Device Facts

- **Applicant:** Zhende Medical Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Apr 28, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids.

## Device Story

Single-use, three-layer, flat pleated face mask; consists of inner/outer spun-bond polypropylene layers and middle melt-blown polypropylene filter layer. Features nose clamp (metal core plastic) and either elastic ear loops (polyester/spandex) or tie-on straps (spun-bond polypropylene). Provided in both non-sterile and sterile (EO) forms. Used by healthcare personnel and patients in clinical settings for infection control to reduce exposure to blood, body fluids, and particulates. Device acts as physical barrier; performance verified via fluid resistance, filtration efficiency, and biocompatibility testing.

## Clinical Evidence

No clinical data. Bench testing only, including fluid resistance (ASTM F1862), particulate filtration efficiency (ASTM F2299 >98%), bacterial filtration efficiency (ASTM F2101 >98%), differential pressure (EN 14683), flammability (16 CFR 1610), and biocompatibility (ISO 10993-5, ISO 10993-10).

## Technological Characteristics

3-layer flat pleated mask; outer/inner layers: spun-bond polypropylene; middle layer: melt-blown polypropylene. Nose clamp: metal core plastic. Ear loops: polyester/spandex; ties: spun-bond polypropylene. Sterilization: EO (for sterile models). Standards: ASTM F2100-19, ASTM F1862-13, ASTM F2299-03, 16 CFR 1610, ISO 10993-5/10, ANSI/AAMI/ISO 11135:2014.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Single-use Surgical Mask ([K200923](/device/K200923.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 28, 2021

Zhende Medical Co., Ltd. % Joyce Yang Consultant Shenzhen Jovantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518100 China

Re: K210020

Trade/Device Name: Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 24, 2021 Received: March 29, 2021

Dear Joyce Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Ryan Ortega -S

Ryan Ortega PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210020

Device Name Medical Mask

Indications for Use (Describe)

The Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# 510(k) Number: K210020

# Date of Summary prepared: April 27, 2021

# 1. Submission Sponsor

| Applicant Name | Zhende Medical Co., Ltd.                               |
|----------------|--------------------------------------------------------|
| Address        | Gaobu Town, 312035, Shaoxing, Zhejiang, P.R.<br>China. |
| Contact person | Chen Ming                                              |
| Phone          | +86-575-88770363                                       |

#### Submission correspondent 2.

| Name           | Shenzhen Joyantech Consulting Co., Ltd                                            |
|----------------|-----------------------------------------------------------------------------------|
| Address        | 1713A, 17th Floor, Block A, Zhongguan Times<br>Square, Nanshan District, Shenzhen |
| Post Code      | 518000                                                                            |
| Phone No.      | +86-755-86069197                                                                  |
| Contact Person | Joyce Yang                                                                        |
| Email          | joyce@cefda.com                                                                   |

#### Device Identification 3.

| Type of 510(k) submission: | Traditional      |
|----------------------------|------------------|
| Trade Name:                | Medical Mask     |
| Classification name:       | Mask, Surgical   |
| Review Panel:              | Surgical Apparel |
| Product Code:              | FXX              |
| Device Class:              | II               |
| Regulation Number:         | 878.4040         |

# 4. Legally Marketed Predicate Device

| Trade Name        | Single-use Surgical Mask            |
|-------------------|-------------------------------------|
| Regulation number | 878.4040                            |
| Regulation class  | II                                  |
| Regulation name   | Surgical Apparel                    |
| 510(k) Number     | K200923                             |
| Product Code      | FXX                                 |
| Manufacturer      | BYD Precision Manufacturer Co.,Ltd. |

{4}------------------------------------------------

# 5. Device Description

The Medical Masks are single use, three-layer, flat masks with ear loops/straps and nose clamp. The Medical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle laver is made of melt blown polvpropylene filter.

The model of proposed device, Ear-loop, is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex.

The model of proposed device. Tie-on, is held in place over the user's mouth and nose by four ties welded to the face mask. The tie is made of spun-bond polypropylene.

The nose clamp contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of metal core plastic.

The proposed devices are single use, disposable device, provided in both non-sterile and sterile forms.

### 6. Intended Use/ Indications for Use

The Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids.

### 7. Technological characteristics comparison

The Medical Mask is compared with the predicate device Single-use Surgical Mask(K200923). The product characteristics are shown as follow:

| Comparison<br>item                                      | Subject Device (K210020)                                                                                                                                                                        | Predicate Device (K200923)                                                                                                                                                                                                  | Comments  |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product<br>Code                                         | FXX                                                                                                                                                                                             | FXX                                                                                                                                                                                                                         | Same      |
| Regulation<br>Number                                    | 21 CFR § 878.4040                                                                                                                                                                               | 21 CFR § 878.4040                                                                                                                                                                                                           | Same      |
| Classification                                          | Class II                                                                                                                                                                                        | Class II                                                                                                                                                                                                                    | Same      |
| OTC use                                                 | Yes                                                                                                                                                                                             | Yes                                                                                                                                                                                                                         | Same      |
|                                                         | The Medical Masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face | The Single-use Surgical Face<br>Masks are intended to be worn<br>to protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate                                 |           |
| Indications for<br>Use                                  | masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids.                                                                   | material. These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single use,<br>disposable device, provided non-<br>sterile. | Same      |
| Level                                                   | Level 3                                                                                                                                                                                         | Level 3                                                                                                                                                                                                                     | Similar   |
| Design feature                                          | Ear-loop, Tie-on Flat Pleated, 3<br>layers                                                                                                                                                      | Ear-loop Flat Pleated, 3 layers                                                                                                                                                                                             | Similar   |
| Usage                                                   | Single use                                                                                                                                                                                      | Single use                                                                                                                                                                                                                  | Same      |
| Color                                                   | Blue                                                                                                                                                                                            | Blue                                                                                                                                                                                                                        | Same      |
| Size                                                    | Width: 18.0cm+/-1.0cm,<br>14.0cm +/-1.0cm<br>Length: 9.5cm+/-1.0cm                                                                                                                              | Width: 17.5cm+/-0.4cm<br>Length: 9.5cm+/-0.4cm                                                                                                                                                                              | Different |
| Sterile                                                 | Non-sterile, Sterile                                                                                                                                                                            | Non-sterile                                                                                                                                                                                                                 | Different |
| Sterilization<br>method                                 | EO                                                                                                                                                                                              | /                                                                                                                                                                                                                           | Different |
| Material                                                | Outer layer: Spun-bond<br>polypropylene                                                                                                                                                         | Outer layer: Spun-bond<br>polypropylene                                                                                                                                                                                     | Same      |
|                                                         | Middle layer: Melt blown<br>polypropylene                                                                                                                                                       | Middle layer: Melt blown<br>polypropylene                                                                                                                                                                                   | Same      |
|                                                         | Inner layer: Spun-bond<br>polypropylene                                                                                                                                                         | Inner layer: Spun-bond<br>polypropylene                                                                                                                                                                                     | Same      |
|                                                         | Nose clamp: Metal Core Plastic<br>Ear loops: Polyester and<br>spandex/Elastic laminate Straps:<br>Spun-bond polypropylene                                                                       | Nose piece: Metal Core Plastic<br>Ear loops: Polyester                                                                                                                                                                      | Different |
| Fluid Resistance<br>Performance<br>(ASTM F1862)         | Pass at 160 mmHg                                                                                                                                                                                | Pass at 160 mmHg                                                                                                                                                                                                            | Same      |
| Particulate<br>Filtration<br>Efficiency<br>(ASTM F2299) | > 98%                                                                                                                                                                                           | > 98%                                                                                                                                                                                                                       | Same      |
| Bacterial<br>Filtration<br>Efficiency<br>(ASTM F2101)   | > 98%                                                                                                                                                                                           | > 98%                                                                                                                                                                                                                       | Same      |
| Differential<br>Pressure (EN<br>14683:2019)             | 5.0mmH2O/cm2<br>( < 6.0mm H2O/cm2)                                                                                                                                                              | 5.62 mmH2O/cm2<br>( < 6.0mm H2O/cm2)                                                                                                                                                                                        | Similar   |
| Flammability<br>(16 CFR 1610)                           | Class 1                                                                                                                                                                                         | Class 1                                                                                                                                                                                                                     | Same      |
| Cytotoxicity<br>(ISO 10993-5)                           | Under the conditions of the study,<br>non cytotoxicity effect                                                                                                                                   | Under the conditions of the study,<br>non cytotoxicity effect                                                                                                                                                               | Same      |
| Irritation (ISO<br>10993-10)                            | Under the conditions of the study,<br>non irritation                                                                                                                                            | Under the conditions of the study,<br>non irritation                                                                                                                                                                        | Same      |
| Sensitization<br>(ISO 10993-10)                         | Under the conditions of the study,<br>non sensitization                                                                                                                                         | Under the conditions of the study,<br>non sensitization                                                                                                                                                                     | Same      |

Table 1: General Comparison

{5}------------------------------------------------

{6}------------------------------------------------

# 8. Summary of Non-clinical Testing

Non-clinical tests were conducted and conformed to the following standards and the requirements state in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 05, 2004.

Standards:

- ASTM F2100-19 Standard Specification For Performance of Materials used in Medical Face Masks.
- ASTM F1862-13 Standard Test Method For Resistance of Medical Face Masks to Penetration by Synthetic Blood.
- ASTM F2299-03 Stand Test Method For Determining The Initial Efficiency Of ● Materials Used In Medical Face Masks To Penetration By Particulates Using Latex Spheres.
- 16 CFR 1610 Standard For The Flammability Of Clothing Textiles.
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- ANSI/AAMI/ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for Development, validation, and routine control of a sterilization process for Medical devices
- ISO 10993-5: 2009 Biological Evaluation Of Medical Devices – Part 5: Tests For In Vitro Cytotoxicity.
- ISO 10993-10: 2010 Biological Evaluation Of Medical Devices – Part 10: Tests For Irritation And Skin Sensitization.

#### Brief discussion of clinical tests 9.

No clinical tests were performed.

### 10. Conclusions

The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(K) submission K210020, the Medical Mask is as safe, effective, and performs as well as or better than the legally marketed predicate device cleared under K200923.

---

**Source:** [https://fda.innolitics.com/device/K210020](https://fda.innolitics.com/device/K210020)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
