The medical surgical mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
Device Story
Three-layer, flat-pleated, single-use surgical mask; provides physical barrier against microorganisms, blood, body fluids, and particulates. Constructed from 25g/m2 PP non-woven cloth; features nylon/PU ear loops or nylon tie strings; includes iron nose clip with polypropylene covering. Used in operating rooms and clinical settings by healthcare personnel. Device provided sterile. Protects both patient and wearer by filtering air and blocking fluid splashes.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM F1862 (fluid resistance), ASTM F2299 (particulate efficiency), ASTM F2101 (bacterial filtration), and ISO 10993 (biocompatibility).
Technological Characteristics
Three-layer flat-pleated mask. Materials: 25g/m2 PP non-woven cloth, Nylon/PU ear loops, Nylon ties, Iron/PP nose clip. ASTM F2100 Level 2. Fluid resistance: 120mmHg. Sterilization: Ethylene Oxide (ISO 10993-7). Biocompatibility: ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for operating room personnel and general healthcare workers for single-use protection against microorganisms, blood, body fluids, and particulates.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) (K173062)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 16, 2020
Beijing Biosis Healing Biological Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co. Ltd P.O.Box 120-119 Shanghai, 200120 China
Re: K202029
Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 15, 2020 Received: July 22, 2020
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202029
Device Name Medical Surgical Mask
Indications for Use (Describe)
The medical surgical mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K202029
#### 1. Sponsor Identification
## Beijing Biosis Healing Biological Technology Co., Ltd.
No.6 plant west, Valley No.1 Bio-medicine Industry Park, Daxing District, 102600, Beijing, China.
Contact Person: Ting Jiang Position: RA Director Tel: +86-10-6125 2660-816 Fax: +86-10-6125 2030 Email: jiangting@biosishealing.com
Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
#### Mid-Link Consulting Co., Ltd.
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net
Date of Preparation: 12/16/2020
{4}------------------------------------------------
- 2. Identification of Proposed Device
Trade Name: Medical Surgical Mask Common Name: Surgical Mask Regulation Name: Surgical Apparel
- 3. Regulatory Information
Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040 Review Panel: General Hospital
- 4. Identification of Predicate Device
Manufacturer: V&Q Manufacturing Corporation 510(k) Number: K173062 Product Name: Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties))
- 5. Indication for use:
The medical surgical mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
- 6. Device Description:
The proposed device, Medical Surgical Mask is a three-layer, single-use, flat-pleated mask. The mask body is made of 25g/m2 PP non-woven cloth. The mask contains tie strings or ear loops to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. Ear loops are made of Nylon and PU, and tie strings are made of Nylon. The nose clip which is made of Iron strip covered by polypropylene covering. The device is provided in sterile.
{5}------------------------------------------------
- 7. Summary of Technological Characteristics
| Table 1 Comparison of Medical Surgical Mask | | | |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| ITEM | Proposed Device | Predicate Device | Remark |
| Product Code | K202029 | K173062 | |
| | FXX | FXX | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Indication for Use | The medical surgical mask<br>is intended for single use by<br>operating room personnel<br>and other general healthcare<br>workers to protect both<br>patients and healthcare<br>workers against transfer of<br>microorganisms, blood and<br>body fluids, and particulate<br>materials. | Non Woven Face Mask<br>(Models: VQN0185W (earloop)<br>and VQN0185B (ties)) is<br>intended for single use by<br>operating room personnel and<br>other general healthcare workers<br>to protect both patients and<br>healthcare workers against<br>transfer of microorganisms,<br>blood and body fluids, and<br>particulate materials. | Same |
| Mask style | Flat pleated | Flat pleated | Same |
| Mask color | Blue | Blue | Same |
| Design feature | Earloop or tie-on | Earloop or tie-on | Same |
| Dimension (mm) | 120mm×70mm<br>145mm×90mm<br>175mm×95mm | 175mm×95mm | Different |
| ASTM F2100 Level | Level 2 | Level 2 | Same |
| Performance | | | |
| Fluid resistance | Pass at 120mmHg | Pass at 120mmHg | Same |
| Particulate efficiency<br>level | Average 98.87% | Average 99.74% at 0.1µm | Similar |
| Bacterial filtration<br>level | Average 99.46% | Average 99.4% | Similar |
| Differential pressure | Average 3.72mmH2O/cm2 | Average 2.7mmH2O/cm2 | Similar |
| Flammability | Class 1 | Class 1 | Same |
| Label/Labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part 801 | Same |
| Patient Contacting Material | | | |
| ear strap | Nylon and PU | Urethane elastic fiber or<br>spun-bond polypropylene | Different |
| nose clip | Iron strip covered by polypropylene covering | White aluminum strip covered by PP covering | Different |
| mask body | 25g/m² PP non-woven cloth | Spun-bond polypropylene | Different |
| Biocompatibility | | | |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No Sensitization | No Sensitization | Same |
| Irritation | No Irritation | No Irritation | Same |
| Sterility | Sterile | Non-sterile | Different |
Table 1 Comparison of Medical Surgical Mask
{6}------------------------------------------------
{7}------------------------------------------------
- 8. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals
- A ASTM F88/F88M-15, Standard Test Method For Seal Strength Of Flexible Barrier Materials.(Sterility)
- A ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization
- ア ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- A ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres
- > ASTM F2101-2017 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- 9. Clinical Test Conclusion
No clinical study is included in this submission.
- 10. Conclusion
The summary includes the conclusions drawn from the nonclinical tests (discussed above) that demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the predicate device.