SURGICAL MASKS

K111559 · Shanghai Neo-Medical Import & Export Co., Ltd. · FXX · Sep 16, 2011 · General, Plastic Surgery

Device Facts

Record IDK111559
Device NameSURGICAL MASKS
ApplicantShanghai Neo-Medical Import & Export Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateSep 16, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Surgical Masks are indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism, body fluid.

Device Story

Non-sterile, single-use 3-ply surgical masks; models NM-MA-008, NM-MA-010, NM-MA-011, and NM-MA-018. Configurations include ear loops, tie-ons, eye shields, and foam. Used by healthcare workers and patients in clinical settings to provide a physical barrier against microorganisms and body fluids. Protects wearer from fluid splashes and reduces transmission of particulates.

Clinical Evidence

Bench testing only. Performance verified via ASTM F2100-07 (materials), ASTM F1862-07 (fluid resistance), EN 14683:2005 (BFE), ASTM F2299-03 (particulate filtration), 16 CFR 1610 (flammability), and Mil M36954C (differential pressure). Biocompatibility testing conducted per ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2002/2006 (irritation and sensitization).

Technological Characteristics

3-ply construction; non-sterile; single-use. Configurations: ear loop, tie-on, eye shield, foam. Standards: ASTM F2100-07, ASTM F1862-07, ASTM F2299-03, EN 14683:2005, 16 CFR 1610, Mil M36954C. Biocompatibility: ISO 10993-5, ISO 10993-10.

Indications for Use

Indicated for healthcare workers and patients as a protective nose and mouth covering during medical and surgical procedures where there is a risk of exposure to microorganisms or body fluids.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SEP 1 6 2011 #### AI I for K111559 Surgical Masks #### Exhibit #1 510(k) Summary ### Exhibit # 1 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________ - 1. Date of Submission: May 20, 2011 - 2. Sponsor Shanghai Neo-Medical Import & Export Co.,Ltd Room 1503, No.2368, ZhongShanxi Road, Shanghai, 200030, China Establishment Registration Number: 3005327291 Contact Person: Pei Zhang Position: General Manager Tel: +86-21-54652962 Fax: +86-21-64673414 Email: Peter999@neo-medical.net ## 3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net - 4. Proposed Device Identification Proposed Device Name: Surgical Masks Proposed Model: NM-MA-008, NM-MA-010, NM-MA-011 and NM-MA-018 Classification: II Product Code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: General & Plastic Surgery . Intended Use Statement: {1}------------------------------------------------ The Surgical Masks are indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism, body fluid. - 5. Predicate Device Identification 510(k) Number: K061864 Product Name: Non Sterile Surgical Mask Manufacturer: Dukal Corporation - 6. Device Description The Surgical Masks are indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism, body fluid. The proposed devices are available in four models with same mask (3-ply) and different configurations, such as ear loop, tie on, eye shield and foam. Detail configurations of them are presented in Table V-1 Surgical Masks Description. | Product<br>Model: | Model Description | | | | |-------------------|-------------------|----------|--------|--------| | | Mask | Ear loop | Tie on | Shield | | NM-MA-010 | X | X | | | | NM-MA-011 | X | | X | | | NM-MA-008 | X | | X | X | | NM-MA-018 | X | X | | X | Table III-1 Surgical Masks Description All of the proposed devices are non-sterile, and only for single use. - 7. Non-Clinical Test Conclusion Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies: with the following standards: - ASTM F 2100-07 Standard Specification for Performance of Materials Used in Medical Face a. Masks. - ف Fluid Resistance per ASTM F 1862-07 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity) 2 {2}------------------------------------------------ - Bacterial Filtration Efficiency (BFE) Test per EN 14683: 2005 Surgical masks, Requirements and ن test methods. - Particulate Filtration Efficiency (Latex Particle Challenge) Test per ASTM F 2299-03 Standard Test di Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres - Flammability , per 16 CFR 1610, Standard for the Flammability of Clothing Textiles; Corrections e. - Differential Pressure (Delta-P) Test per Mil M36954C Military Specification- Mask, Surgical, f. Disposable - Biocompatibility :Cytotoxicity, Sensitization, Irritation g. - Including: ISO 10993-5:2009 "Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity. - ISO 10993-10:2002 Standard, "Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity" - ISO 10993-10:2002 Standard and Amendment 1, 2006. "Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity AMENDMENT I" - 8. Substantially Equivalent Conclusion The proposed devices are determined to be Substantially Equivalent (SE) to the predicate device, Non Sterile Surgical Mask (K061864) in respect of safety and effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 1 6 2011 Shanghai Neo Medical Import & Export Company, Limited C/O Ms. Diana Hong General Manager Mid-Link Consulting Company, Limited P.O. Box 237-023 Shanghai China 200237 Re: K111559 Trade/Device Name: Surgical Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: August 9, 2011 Received: August 26, 2011 Dear Ms. Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Hong Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, hn fo Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Exhibit #3 IFU Statement 510(k) Number: K111559 Device Name: Surgical Masks Models: NM-MA-008, NM-MA-010, NM-MA-011, NM-MA-018 Indications for Use: The Surgical Masks are indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism, body fluid. □PRESCRIPTION USE (Part 21 CFR 801 Subpart D) ZOVER-THE-COUNTER USE (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Elliott F. Currie W.K. (ISON 2008) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K111559
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...