SOLARC / SOLRX E-SERIES ULTRAVIOLET PHOTOTHERAPY LAMP UNIT FAMILY

K103204 · Solarc Systems, Inc. · FTC · Apr 27, 2011 · General, Plastic Surgery

Device Facts

Record IDK103204
Device NameSOLARC / SOLRX E-SERIES ULTRAVIOLET PHOTOTHERAPY LAMP UNIT FAMILY
ApplicantSolarc Systems, Inc.
Product CodeFTC · General, Plastic Surgery
Decision DateApr 27, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Solarc/SolRx E-Series Ultraviolet Phototherapy Lamp Family is a wall-mounted or self-supporting ultraviolet phototherapy lamp unit used for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use as a spot or full-body treatment device. It is intended for use on all skin types. (I -VI) This device is intended for Prescriptive Use only, per Part 21 CFR 801 Subpart D.

Device Story

Solarc/SolRx E-Series is a family of expandable ultraviolet (UV) phototherapy lamp units; utilizes low-pressure mercury vapor fluorescent bulbs to emit UV radiation (UVA, UVA1, UVB-Broadband, or UVB-Narrowband). Devices are modular; can be used singly or ganged together. Features include digital countdown timer, keyed switchlock/password entry, and protective goggles. Used in home, physician office, or clinic settings; operated by patients or healthcare providers. Input is manual timer setting; output is controlled UV exposure to skin. Benefits include localized or full-body phototherapeutic treatment for dermatologic conditions. Safety features include wire guards or acrylic windows to prevent bulb breakage.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, comparison of technical specifications, and established safety profiles of predicate devices.

Technological Characteristics

Sheet metal construction; low-pressure mercury vapor fluorescent bulbs; digital countdown timer; keyed switchlock/password security; wire guards or acrylic windows; UV protective goggles. Modular design allows ganging of units. Operates on standard electrical power. Complies with medical device standards for high voltage, leakage, and electromagnetic compatibility.

Indications for Use

Indicated for treatment of psoriasis, vitiligo, and atopic dermatitis (eczema) in patients of all skin types (I-VI). Intended for spot or full-body treatment. Prescription use only.

Regulatory Classification

Identification

An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K107.204 ## APR 2 7 2011 Image /page/0/Picture/2 description: The image shows the logo for Solar Arc Systems Inc. The logo features the word "SOLARC" in bold, sans-serif font, with a stylized swoosh extending from the bottom of the "S" and curving upwards over the word. Below "SOLARC", the words "SYSTEMS INC." are printed in a smaller, sans-serif font. The logo is black and white. ## 510(k) Summary Date: Oct26, 2010 Trade Name: Solarc / SolRx E-Series Ultraviolet Phototherapy Lamp Unit Family A new family of medical devices that can be used singly, or ganged together in any number and combination of models. The devices use various numbers and lengths of tubular ultraviolet fluorescent bulbs. The model numbering system for single devices is as follows: Ehn0t-waveband-suffix , where: E = A constant indicating the E-Series and standing for "Expandable". h = The nominal height of the device in inches divided by ten; where h=2 for 24" (2-foot) high devices, h=4 for 48" (4-foot) high devices, h=7 for 72" (6-foot) high devices, and h=8 for 80" (2-metre) high devices. n = The number of bulbs in the device; with the only exceptions being "9" for 10-bulb devices (such as E790), and "92" for 12-bulb devices (such as E792). 0 = A constant (zero); except for 12-bulb devices where this is a "2", such as the "E792". t = The device type, either "M" for a "Master" device, or "A" for an "Add-on" device. waveband = either "UVBNB" for UVB-Narrowband, "UVB" for UVB-Broadband, "UVA" for UVA, and "UVA1" for UVA-1. suffix = any or all of: "CR" for Clinic Rated devices, "AW" for devices equipped with a clear Acrylic Window in lieu of wire guards, "SD" for reflective Side Doors and "RB" for internal Reflector Bulbs. Common Name: Ultraviolet Phototherapy Lamp Unit Classification: 21 CFR 878.4630, Product Code FTC Intended Use: Phototherapeutic treatment of psoriasis, vitiligo, and atopic dermatitis (eczema) Solarc Systems Inc., 1515 Snow Valley Road, Minesing, ON, Canada LOL 1Y0 Applicant: Contact: Bruce Elliott, P.Eng., President Solarc Systems Inc. Contact. Bruce Elliott, P.Eng., President Solarc Systems Inc. Phone: 705 730 2048 Fax: 705 730 0684 Email: belliott@solarcsys.ca Email: belliott@solarcsystems.com Phone: 705-739-3048 = Fax: 705-739-9684 Solarc Systems Inc. declares that, to the best of its knowledge, the proposed device family has the same intended use and similar technical characteristics as the following predicate devices: Solarc/SolRx 1000 Series [510(k)# K935572] Solarc/SolRx 500 Series [510(k)# K031800] Solarc/SolRx 100 Series [510(k)#: K061589] Daavlin "7-Series" and "2-Series" [510(k)# K854498 "Spectra 724"] National Biological Corp. (NBC) "Foldalite-32" [510(k)#: K827890] National Biological Corp. (NBC) "Panosol 3D" [510(k)#: uncertain] National Biological Corp. "Panosol II… with optional wings" [510(k)#: uncertain] lt can be demonstrated that the proposed device is as safe and effective as the legally marketed devices and does not raise different questions regarding safety and effectiveness than the predicate devices. This is based on the following areas of comparison between the proposed device and the referenced predicate devices: ## Intended Use As with all the predicate devices, the proposed device is intended for medical ultraviolet phototherapy delivered in the patient's home, at a physician's office or in a hospital's clinic; for the treatment of psoriasis, vitiligo, and eczema (atopic dermatitis). {2}------------------------------------------------ K 103204 Image /page/1/Picture/1 description: The image shows the logo for SOLARc SYSTEMS INC. The logo is black and white and features the company name in bold, sans-serif font. Above the company name is a curved line. The words "SYSTEMS INC." are in a smaller font size and are located below the main company name. ### Design Concept ੰ The proposed device is constructed primarily of sheet metal, and has provision for connecting additional lighting banks, as do two of the predicate devices. ### Treatment Area & Anatomical Sites The proposed device can be made in 2-foot, 4-foot and 2-metre lengths; for partial or full body treatment excluding the eyes, as are the predicate devices. The proposed device has provision for ultraviolet reflective side panel doors, as does at least one of the predicate devices. The treatment area of the proposed device can be reduced by using the "Face Shield" accessory, which is similar to the aperture plate system of the Solarc/SolRx 100 Series predicate device. #### Type of Ultraviolet Bulbs The proposed device uses low pressure mercury vapor fluorescent ultraviolet bulbs that are comparable to that of the predicate devices, as are the spectral distribution, irradiance levels, ballasts and related circuitry. #### Quantity of Ultraviolet Bulbs The proposed device uses 2 to 12 builss per device, or multiples thereof when devices are ganged together, within electrical current limits. The predicate devices use 1 to 48 bulbs. ### Electrical Rating The voltage and frequency ratings are comparable. High voltage breakdown, current leakage, and electro-magnetic compatibility specifications meet current industry and medical device standards. #### Timer The proposed device uses a digital countdown timer that is technically comparable to that of the predicate devices. #### User's Manual The User's Manual is written for the layman user and provides comprehensive exposure guidelines. It is at least as effective as the predicate devices. #### Safety Features The proposed device has a keyed switchlock or timer password entry to prevent unauthorized usage, wire guards or an acrylic window to reduce the chance of bulb breakage, and FDA compliant ultraviolet protective goggles; all comparable to one or more of the predicate devices. #### Requiatory Requirements The proposed device is designed and will be manufactured according to the FDA Good Manufacturing Practices and ISO13485. Solarc Systems Inc. is ISO-13485:2003 certified. #### Conclusion On the basis of the information provided, Solarc Systems Inc. believes that the proposed device is substantially equivalent to legally commercialized predicate devices. End of 510(k) Summary {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the circle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Solarc Systems Inc. % Mr. Bruce Elliott 1515 Snow Valley Road Minesing, Ontario Canada LOL 1 Y 3 APR 2 7 2011 Re: K103204 Trade/Device Name: Solarc/SolRx E-Series Ultraviolet Phototherapy Lamp Unit Family and Accessories Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: Class II Product Code: FTC Dated: April 11, 2011 Received: April 12, 2011 Dear Mr. Elliott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interest commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandijng and adulteration. Please note: CDRH does not evaluate information related to contractly and warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Bruce Elliott Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as st forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050, If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Prati 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollorn the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Ais B. nh Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K103204 510(k) Number (if known): Device Name: Solarc / SolRx E-Series Ultraviolet Phototherapy Lamp Unit Family ## Indications for Use: The Solarc/SolRx E-Series Ultraviolet Phototherapy Lamp Family is a wall-mounted or self-supporting ultraviolet phototherapy lamp unit used for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use as a spot or full-body treatment device. It is intended for use on all skin types. (I -VI) This device is intended for Prescriptive Use only, per Part 21 CFR 801 Subpart D. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R. Dahlen for mxm (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K03204 Page 1 of /
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...