K090097 · Lerner Medical Devices, Inc. · FTC · Feb 26, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090097
Device Name
LH-75T PHOTOTHERAPY SYSTEM
Applicant
Lerner Medical Devices, Inc.
Product Code
FTC · General, Plastic Surgery
Decision Date
Feb 26, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LH-75T Phototherapy System is intended for use, by or under the direction of a physician, in UVB phototherapy for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leucoderma. The LH-75T Phototherapy System is intended for use on all skin types (I -VI).
Device Story
LH-75T Phototherapy System is an ultraviolet light source and energy delivery system; emits UV-B (290-320 nm) light for dermatologic treatment. System comprises a console containing electronics and a handpiece containing the lamp and ballast, connected by electrical cable. Includes Spot Treatment Attachment (LP-3) and Fiber-optic Brush Attachment (LB-10) for targeted treatment of skin lesions while sparing healthy tissue. Powered by household AC outlet; dose selected via console controls and activated by panel switch or footswitch. Used in clinical settings by physicians or under their direction. Provides controlled UV-B exposure to manage skin conditions; benefits include targeted therapy for specific lesions. Includes power meter for calibration and output monitoring.
Clinical Evidence
Bench testing only. Performance data confirmed spectral output is within the 290-320 nm range, consistent with predicate devices. No clinical data provided.
Technological Characteristics
UV-B light source (290-320 nm); console-handpiece configuration with electrical cable; includes spot treatment and fiber-optic brush attachments. Features integrated power meter for calibration. Powered by standard AC outlet. Non-software controlled; manual dose selection via console panel.
Indications for Use
Indicated for patients of all skin types (I-VI) requiring UVB phototherapy for treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and leucoderma. Use by or under physician direction.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
Predicate Devices
TheraLight, Inc. UV1 20-2 UVA / UVB Phototherapy System (K022165, K024020)
Lerner Medical Devices, Inc. Levia Phototherapy System (K040062)
Daavlin Distributing Co. Dermapal with Digital Timer (K073587)
Submission Summary (Full Text)
{0}------------------------------------------------
K090097
FEB 2 6 2009
Page ① of ③
## 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
Date:
Submitter's name:
Submitter's Address:
Submitter's Telephone:
Submitter's Fax:
Contact Person:
Device Trade Name:
Device Common Name:
Device Classification Name:
Regulation Number:
Product Code:
Classification:
Establishment Reg. Number:
List of Predicate Devices:
January 14, 2009
Lerner Medical Devices, Inc.
1545 Sawtelle Ave. Suite 36 Los Angeles, CA 90025
(310) 914 0091
(310) 914 0095
Zafirios F. Gourgouliatos, Ph.D., Chief Science Officer Management Representative
LH-75T Phototherapy System
Targeted UVB Phototherapy System
Ultraviolet lamp for dermatologic / skin disorders
878.4630
FTC
Device Class II
3006793564
TheraLight, Inc. UV1 20-2 UVA / UVB Phototherapy System K022165, K024020,
Lerner Medical Devices, Inc. Levia Phototherapy System K040062
Daavlin Distributing Co. Dermapal with Digital Timer K073587
{1}------------------------------------------------
K090097
ge ② of ③
#### Device Description:
The LH-75T Phototherapy System is an ultraviolet Light Source and energy delivery system. The System emits UV-B (290-320 mm) light for use in phototherapy and allows delivery of controlled doses!
The LH-75T consists of a Console that contains the electronics and a Handpiece that contains the lamp and the ballast. The console and handpiece are connected with an electrical cable. A Spot Treatment Attachment (LP-3) jand a Fiber-optic Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Brush Attachment (LB-10) are connected to the handpiece for selective treatment of skin lesions without exposure of adjacent, healthy tissues. The desired dose is selected using controls on the panel of the console and activated with either a switch on the panel or foot-switch. The system is powered by a common household AC outlet. Protective eyewear is supplied with the system.
## Indications for Use:
The LH-75T Phototherapy System is intended for use, by or under the direction of a physician, in UVB phototherapy for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leucoderma.
The LH-75T Phototherapy System is intended for use on all skin types (I -VI).
## Substantial Equivalence:
The LH-75T Phototherapy System emits UVB light within the spectral band of 290 -320 nm, similar to that of all the predicate devices. This light has been shown to be safe and effective in the treatment of scalp psoriasis. The difference between the LH-75T Phototherapy System and the predicate devices is the configuration of the hardware >>>
The delivery components (Spot Treatment Attachment LP-3, Fiber-optic Brush Attachment LB-10) and control electronics are identical to those of the Levia Phototherapy System (Lerner Medical Devices, Inc., K040062).
The lamp is situated the Handpiece and is connected to the console with an electrical cable in a manner similar to the Dermapal with Digital Timer predicate device. (Daavlin Distributing Co., K073587).
The LH-75T Phototherapy System has a Power Meter embodied in the console to facilitate calibration, in a manner similar to the UV1 20-2 UVA / UVB Phototherapy System (TheraLight, Inc., K022165, and K024020). The Power Meter is used to indicate output of the LH-75T for operator convenience. The meter is similar to the meter of the Levia Phototherapy System.
Lemer Medical Devices. Inc. believes that these similarities provide added convenience to the user and do not raise new questions about safety or effectiveness.
{2}------------------------------------------------
090097 Page (3) of (3)
## Performance Data:
Performance data have been submitted as part of the 510(k) application to confirm that the spectral output of the LH-75T Phototherapy System is between 290 and 320 nm, which is similar to spectra emitted by predicate devices. Other performance data included in the application show that the technological specifications are similar to those of the predicate devices.
## Conclusion:
On the basis of the information provided in this summary, Lerner Medical Devices, Inc. believes the LH-75T Phototherapy System is substantially equivalent to legally commercialized predicate devices.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, representing the department's role in protecting the health of all Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Lerner Medical Devices, Inc. % Zafirios Gourgouliatos, Ph.D 1545 Sawtelle Boulevard, Suite 36 Los Angeles, California 90025
Re: K090097
FEB 2 6 2009
Trade/Device Name: LH-75T Phototherapy System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: II Product Code: FTC Dated: January 14, 2009 Received: January 15, 2009
Dear Dr. Gourgouliatos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Zafirios Gourgouliatos, Ph.D
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
y/m. D. fr
Mark N. Melkerson Director . Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
510(k) Number_K090097
Device Name: LH-75T Phototherapy System
The LH-75T Phototherapy System is intended for use, by or under the direction of a physician, in UVB phototherapy for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leucoderma.
The LH-75T Phototherapy System is intended for use on all skin types (I -VI).
Prescription Use X OR Over-the-Counter Use (per 21 CFR 801.109)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Milke Ryl for mtn
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K090097
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.