AMICO CE SERIES SURGICAL LIGHTS, MIRA LED SERIES MINOR SURGICAL LIGHTS
K131893 · Ajw Technology Consultants, Inc. · FSY · Oct 30, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131893
Device Name
AMICO CE SERIES SURGICAL LIGHTS, MIRA LED SERIES MINOR SURGICAL LIGHTS
Applicant
Ajw Technology Consultants, Inc.
Product Code
FSY · General, Plastic Surgery
Decision Date
Oct 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The intended use of the Amico iCE LED series surgical lights are to provide high intensity, shadow free illumination to the patient surgical field during surgical procedures. The intended use of the Amico Mira LED series minor surgery lights is to provide high intensity, shadow free illumination to the patient surgical field during minor surgery procedures.
Device Story
Amico iCE and Mira LED series are surgical luminaires providing high-intensity, shadow-free illumination for surgical suites. iCE series (major surgery) and Mira series (minor surgery) utilize ceiling-mounted suspension systems with articulating extension and spring arms for positioning. Lamp heads feature plastic housings, polycarbonate lenses, and steel yokes. Mechanical (iCE) or electronic (Mira) focusing systems, controlled via a center handle, adjust the light field. Optional in-light camera modules with remote control allow recording/display of surgical sites on external monitors. Devices are operated by clinicians in surgical settings. Output is visual illumination of the surgical field; camera output provides visual feedback to assist surgical detail and accuracy. Benefits include improved visibility during procedures. Disposable sterile handles provide a barrier between the device and surgical staff.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), particular requirements for surgical luminaires (IEC 60601-2-41), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
LED-based surgical luminaires. Materials: plastic housing, polycarbonate lens, steel yoke. Mounting: ceiling, wall, or roll stand. Focusing: mechanical (iCE) or electronic (Mira). Connectivity: optional camera module with remote control and network interface. Service life: 50,000 hours. Power: 75-200W. Standards: IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2.
Indications for Use
Indicated for providing high-intensity, shadow-free illumination to the surgical field during surgical or minor surgical procedures for patients undergoing such interventions.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{1}------------------------------------------------
# K131893
# 510(k) SUMMARY (as required by 807.92)
**OCT**
**30 2013**
| Regulatory Correspondent: | AJW Technology Consultants, Inc.<br>445 Apollo Beach Blvd.<br>Apollo Beach, FL 33572 USA<br>Janet Douglas<br>Email: jdouglas@ajwtech.com |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of 510(k): | Amico Clinical Solutions Corporation<br>85 Fulton Way<br>Richmond Hill, ON L4B 2N4 Canada<br>Varun Chandan<br>Email: vchandan@amico.com<br>Phone: 905 764 0800 ex.5715<br>Fax: 905 764 0862 |
| Date of Summary: | September 17, 2013 |
| Trade/Proprietary Name: | Amico Clinical Solutions Corp. - iCE LED Series<br>and Amico Lights Corp. - Mira LED Series Minor<br>Surgical Lights |
| Common/Usual Name: | Surgical Lamp |
| Classification Name: | Class II |
| Product Code: | FSY |
# Intended Use:
#### Device Name: Amico iCE LED Light
The intended use of the Amico iCE LED series surgical lights are to provide high intensity, shadow free illumination to the patient surgical field during surgical procedures.
#### Device Name: Amico Mira LED Series Minor Surgical Light
The intended use of the Amico Mira LED series minor surgery lights is to provide high intensity, shadow free illumination to the patient surgical field during minor surgery procedures.
{2}------------------------------------------------
## Device Description:
## Amico iCE Series Device Description:
Amico iCE series consists of 3 models - iCE-30-6, iCE-25-6, & iCE-25-3. These are major surgical luminaires that provide illumination in a surgical suite allowing doctors to operate on a patient.. The device comes with several built-in fail safe's preventing any single fault from creating a hazardous situation.
These lamps are a Class 2 device (under FDA) providing max illumination of 160,000 lx, 160.0001x, and 130.000 Ix respectively. These surgical lights are coupled with ceiling mounted suspension systems supporting horizontally articulating extensions arms and horizontally and vertically articulating spring arms. Both arms can rotate either 340 degree (with stop) or infinitely without a stop. The spring arms vertically articulate 50° downwards and 45° upwards. This allows for ease of movement of the lamp head to any given position.
The lamp housing is made of high quality plastic wrapped around an aluminum structure. The front clear glass is made of high quality polycarbonate material. The yoke connecting the lamp head to the spring arms is made of steel.
The lamps also have a built-in mechanical focusing system that is controlled using the center Aluminum and/or plastic handle. Rotating this handle allows the user to expand and contract the lighted field as per the user's requirement. The lamps can also be sold with an optional in-light camera that doubles as a focusing handle. This provides the user with the option to record/display small areas of a surgical site on a large display allowing for greater detail and accuracy during surgeries. A separate remote control is used to control the various functions of the camera.
Available accessories for iCE series surgical lights
- Camera Module
- Remote control for Camera module
- Remote control with network interface for camera module
- Single monitor yoke for flat panel monitors
- Double monitor yoke for flat panel monitors
- Instrument trays
- Trays for CRT Monitors
- Low profile wall control unit
#### Amico Mira LED Series Device Description:
Amico Mira LED series consists of 3 models - Mira LED 90, Mira LED 65 & Mira LED 50. These are minor surgical luminaires that provide illumination in a surgical suite allowing doctors to perform minor surgical procedures on a patient.
These lamps are a Class 2 device (under FDA) providing max illumination of 90.000 1x. 65.000 Ix, and 50.000 Ix respectively. These surgical lights are coupled with ceiling
{3}------------------------------------------------
mounted suspension systems supporting horizontally articulating extensions arms and horizontally and vertically articulating spring arms. Both arms can rotate either 340 degree (with stop) or infinitely without a stop. The spring arms vertically articulate 50° downwards and 45° upwards. This allows for ease of movement of the lamp head to any given position.
The lamp housing is made of high quality plastic which connects to the front clear glass is made of high quality polycarbonate material. The yoke connecting the lamp head to the spring arms is made of steel.
The lamps also have a built-in electronic focusing system that is controlled using the center Aluminum/plastic handle. Rotating this handle allows the user to expand and contract the lighted field as per the user's requirement.
#### Predicate Device:
### K093009 Mach LED SC K093010 Mach LED MC
#### Substantial Equivalence:
The Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights are substantially equivalent in intended use and technological characteristics to the Mach LED SC and Mach LED MC surgical lights. Any difference that exists between the Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights and both predicate devices has no negative effect on safety or effectiveness.
| | Comparison of Technological Char | | | | | |
|----------------|----------------------------------|-----------------|-------------|-------------|--|--|
| Comparison | Applicant Device | | | Predicated | | |
| Elements | | | Device. | Device | | |
| 510(K) | | | K093010 | K093009 | | |
| Number | | | | | | |
| Device Name | Amico iCE LED | Amico Mira LED | Mach LED MC | Mach LED SC | | |
| Technical Data | | | | | | |
| Light | 130,000 - 160,000 | 50,000 - 90,000 | 160,000 | 130,000 | | |
| Intensity | | | | | | |
| (Central | | | | | | |
| Luminance) | | | | | | |
| Color | તેરે રે | તેરે | > 96 | તેરે રે | | |
| Rendering | | | | | | |
| Index (Ra) | | | | | | |
| Color | >90 | તેરે | તેરે રે | તેરે તે રે | | |
| rendering | | | | | | |
| index | | | | | | |
| R9(red) | | | | | | |
| Focusable | 7.6-12.0in | 7.0-12.0in | | | | |
| size of the | | | | | | |
| light field | | | | | | |
| | | | 17-28cm | 17-28cm | | |
{4}------------------------------------------------
| Comparison of Technological Characteristics | | | | |
|---------------------------------------------|------------------|------------------------------|------------------------------------|---------------------------------|
| Comparison<br>Elements | Applicant Device | Applicant Device | Predicated<br>Device<br>K093010 | Predicated<br>Device<br>K093009 |
| 510(k)<br>Number | | | | |
| Device Name | Amico iCE LED | Amico Mira LED | Mach LED MC | Mach LED SC |
| Color<br>temperature<br>(Kelvin) | 4500K | 4500K | 3750, 4000,<br>4250,<br>4500, 4750 | 4500K |
| Diameter of<br>the lamp<br>head | 25-30in | 18in | 57cm | 57cm |
| Total power<br>consumption | 150-200 | 75 | 120 W | 45 W |
| Brightness<br>Control | 5-100% | 5-100% | 5-100% | 5-100% |
| # of other<br>LED's | 208-320 | 144 | 112 | 28 |
| LED service<br>life | 50,000 | 50,000 | > 40.000 h | > 40.000 h |
| Mounting<br>Options | Ceiling, Wall, | Ceiling, Wall,<br>Roll Stand | Ceiling, Wall,<br>Roll Stand | Ceiling, Wall,<br>Roll Stand |
#### Sterilization and Shelf-Life
The Disposable Light Handle (K020304) is used as an accessory to the Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights. The Disposable Light Handle provides a sterile interface between the Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights, and the Surgeon or Scrub Nurse.
#### Non-Clinical Testing:
Electrical Safety testing was carried out on the Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights. Below is a summary of the performed testing:
Attachment A: IEC 60601-1 Medical electrical equipment
Part 1: General requirements for basic safety and essential performance
Report Reference No .: 100874632MTL-002
· Result: Pass
Attachment B: IEC 60601-2-41 Medical electrical equipment
Part 2-41: Particular requirements for the safety of surgical luminaires and luminaires for diagnosis
Report Reference No .: 100874632MTL-003
Result: Pass
Attachment C: Constructional Data Report
{5}------------------------------------------------
Report Reference No .: 100874632MTL-001 Result: Pass
Attachment D: IEC 60601-1-2 (Ed. 3): 2007 - Medical electrical equipment Part 1-2:General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
Report Reference No.: 100874632ATL-003
Result: Pass
Attachment E: IEC 60601-1-2 (Ed. 3): 2007 - Medical electrical equipment Part 1-2:General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
Report Reference No .: 100874632ATL-004 Result: Pass
#### Electrical Safety Testing Conclusions:
Based on the performed testing it has been shown that Amico Clinical Solutions Corp. iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights has been proven to be safe and effective and poses no foreseen hazards. No Open Items remain from the planned Amico Clinical Solutions Corp. - iCE LED Series and Amico Lights Corp. - Mira LED Series Minor Surgical Lights Electrical Safety Testing.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Conter - W 066-G609 Silver Spring, MD 20993-0002
October 30, 2013
Amico Clinical Solutions Corporation % Ms. Janet Douglas AJW Technology Consultants, Incorporated 445 Apollo Beach Boulevard Apollo Beach, Florida 33572
Re: K131893
Trade/Device Name: Amico iCE LED Light Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: September 17, 2013 Received: September 27, 2013
Dear Ms. Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Ms. Janet Douglas
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indications for Use
510(k) Number (if known): K131893
Device Name: Amico iCE LED Light
The intended use of the Amico iCE LED series surgical lights are to provide high intensity, shadow free illumination to the patient surgical field during surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen -A for MXM Page I of I (Division Sign-Off) Division of Surgical Devices 510 (k) Number: K131893
{9}------------------------------------------------
Indications for Use
510(k) Number (if known): K131893
Device Name: Amico Mira LED Series Minor Surgical Light
The intended use of the Amico Mira LED series minor surgery lights is to provide high intensity, shadow free illumination to the patient surgical field during minor surgery procedures
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lona H. Chen - M for MXM A Page I of I ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(Division Sign-Off) Division of Surgical Devices 510(k) Number: K131893
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.