K240598 · Moon Surgical · FQO · Jun 3, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240598
Device Name
Maestro System (REF100)
Applicant
Moon Surgical
Product Code
FQO · General, Plastic Surgery
Decision Date
Jun 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4960
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.
Device Story
Maestro System is a 2-arm, electrically actuated surgical support device used in laparoscopic procedures. It utilizes DC motors to compensate for gravitational forces on instruments, allowing surgeons to maneuver tools freely; when hand force is removed, the system maintains the instrument's position and orientation. The system includes a surgical base, arms, control unit, and instrument coupling. New features include positioning guidance sensors, bedside joint control via a forearm collar, arm proximity indicators with audible/visual alerts, and tactile vibration feedback for hold status. The system is operated by surgeons in the OR. It benefits patients by providing stable instrument positioning, reducing surgeon fatigue. Output is physical support of the instrument; status is communicated via LEDs, touch screen, and tactile vibration.
Clinical Evidence
No clinical data from human trials was required. Evidence consists of bench-level design verification (payload capacity, force accuracy, collision detection, system latency, electrical safety, EMC) and design validation, including cadaver testing and human factors testing to confirm system usability and safety.
Technological Characteristics
Electrically actuated 2-arm system; DC motors for gravity compensation; sensors for positioning guidance and proximity detection; LED status indicators; tactile vibration feedback; touch screen interface. Sterilization for single-use coupling devices via Ethylene Oxide (EtO). Connectivity includes system console and bedside controls. Software-controlled motion and fault monitoring.
Indications for Use
Indicated for holding and positioning laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.
Regulatory Classification
Identification
Operating tables and accessories and operating chairs and accessories are AC-powered or air-powered devices, usually with movable components, intended for use during diagnostic examinations or surgical procedures to support and position a patient.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 3, 2024
Moon Surgical % Michael Daniel Consultant Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460
Re: K240598
Trade/Device Name: Maestro System (REF100) Regulation Number: 21 CFR 878.4960 Regulation Name: Operating tables and accessories and operating chairs and accessories Regulatory Class: Class I Product Code: QZB Dated: March 2, 2024 Received: March 4, 2024
Dear Michael Daniel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen =S Digitally signed by Long H. Chen-S
Digitally signed Date: 2024.06.03 14:50:58 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240589
Device Name
Maestro System (REF100)
Indications for Use (Describe)
The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.
Type of Use (Select one or both, as applicable)
(Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. Premarket Notification 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### 510(k) Number: K240589
#### Applicant Information:
| Date Prepared: | July 6, 2023 |
|-----------------|-------------------------------------------------------------|
| Name: | Moon Surgical |
| Address: | 9 rue d'Enghien<br>75010 Paris France |
| Contact Person: | Michael A Daniel, Consultant<br>madaniel@clinregconsult.com |
| Mobile Number: | (415) 407-0223 |
#### Device Information:
| Device Trade Name: | Maestro System |
|---------------------------|-----------------------------------|
| Common Name: | Maestro System |
| Classification Name(s): | Table, Operating-Room, AC-Powered |
| Product Code/ Regulation: | QZB 21 CFR 870.4960 |
| Classification: | Class I |
### Predicate Device:
Maestro System (K221410)
# Subject Device Description
The Moon Maestro System is a 2-arm system which utilizes software and hardware to provide support to surgeons for manipulating and maintaining instrument position. Motors compensate for gravitational force applied to laparoscopic instruments, while surgeon control is not affected. Conventional laparoscopic tools are exclusively controlled and maneuvered by the surgeon, who grasps the handle of the surgical laparoscopic instrument and moves it freely until the instrument is brought to the desired position. Once surgeon hand force is removed, the Maestro system reverts to maintenance of the specified tool position and instrument tip location. This 510(k) is being submitted to implement design changes to the previously cleared Maestro System. The following modifications have been implemented to the Maestro System:
- · System Positioning Guidance
- · System Hold Status Indication
- Instrument Coupling
- · System Setup
- Bedside Setup Joint Control
{4}------------------------------------------------
# Subject Device Intended Use / Indications for Use
The Maestro System is intended to hold and position laparoscopes and laparoscopic instruments during laparoscopic surgical procedures.
# Predicate Device and Subject Device Comparison
The table below compares the Maestro System to the predicate device.
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Device Name | Maestro | Maestro | N/A |
| 510(k) Number | K221410 | Not yet assigned | N/A |
| Product Code /<br>Regulation | FQO / 878.4960<br>Table Operating-room, AC-<br>Powered. | FQO / 878.4960<br>Table Operating-room, AC-<br>Powered. | Identical |
| Brief<br>Description | The Moon Surgical Maestro<br>device is an electrically<br>actuated device with<br>movable components<br>intended for laparoscopic<br>surgical procedures to<br>support and position<br>laparoscopic instruments.<br>Its major components consist<br>of the Surgical Base,<br>Surgical Arms, Control Unit,<br>Instrument Coupling, and<br>Control Station. | The Moon Surgical Maestro<br>device is an electrically<br>actuated device with<br>movable components<br>intended for laparoscopic<br>surgical procedures to<br>support and position<br>laparoscopic instruments.<br>Its major components<br>consist of the Surgical<br>Base, Surgical Arms,<br>Control Unit, Instrument<br>Coupling, and Control<br>Station. | Identical |
| Intended Use /<br>Indications for<br>Use | The Maestro System is<br>intended to hold and position<br>laparoscopic Instruments<br>during laparoscopic surgical<br>procedures. | Same | Identical |
| Mechanism of<br>Action (System<br>Actuation /<br>Positioning) | The Moon Maestro System<br>utilizes DC motors to<br>compensate for gravitational<br>forces applied to<br>laparoscopic instruments. | Same | Identical |
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
| | Software and hardware are<br>designed to maintain<br>instrument position.<br>Surgeon control is achieved<br>by grasping the handle of<br>conventional laparoscopic<br>tools and moving the tool to<br>the desired position. Once<br>surgeon hand force is<br>removed, the Maestro<br>system reverts to<br>maintenance of the specified<br>location including<br>orientation. | | |
| Endoscope<br>Movement | Control algorithm applying<br>motor current for instrument<br>motion triggered by<br>physician direct movement. | Same | Identical |
| Platform /<br>Mechanism of<br>stable<br>attachment of<br>arms | Jointed arms are attached to<br>stages located on a portable<br>base that can be locked in<br>position. The bases are<br>locked in position during<br>system use. | Same | Identical |
| Cart Setup prior<br>to procedure<br>initiation | No positioning guidance | Sensors have been added to<br>improve system positioning<br>and setup prior to procedure<br>initiation. These sensors<br>monitor OR table position<br>and distance as the user<br>advances the system toward<br>the OR table. The user<br>always has the freedom to<br>position the system where<br>they want, but the guidance<br>provides the recommended<br>window of placement | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness. |
| System Setup | User-controlled powered x &<br>z translation of the arms, and<br>manual translation of the | User-selectable preset<br>position for each specific<br>clinical procedure to be | Substantially<br>equivalent with no<br>new or different |
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
| | shoulder roll via a toggle on<br>the system console. | performed, chosen off a<br>menu on the system<br>console. | questions of safety or<br>effectiveness. |
| | | Discrete manual control of<br>these degrees of freedom<br>always remains available to<br>the user on the system<br>console for fine adjustments<br>via user-controlled powered<br>x & z translation of the<br>arms, and user-controlled<br>powered translation of the<br>shoulder roll | |
| Bedside Setup<br>Control | Lack of complete control<br>over the setup joints from the<br>surgeon's bedside position.<br>Adjusting x & z setup joints<br>required interfacing with the<br>non-sterile console on the<br>system. | A control collar was added<br>to each Maestro arm's<br>forearm. Adjusting x, z and<br>shoulder roll setup joints no<br>longer requires interfacing<br>with the non-sterile<br>console.<br><br>In all cases, if an instrument<br>is attached during<br>positioning adjustments, the<br>system maintains the<br>instrument's tip position<br>during any setup joint<br>motion. | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness. |
| Arm Proximity<br>Indicator | No Arm Proximity Indicator | Sensors have been added to<br>improve arm positioning by<br>monitoring for objects<br>within an established range.<br>When an object enters that<br>range, the system emits an<br>audible chime, and provides<br>a notification on the<br>Console. | |
| Method of<br>instrument<br>attachment | Mechanical clamp around<br>laparoscopic instrument and<br>magnetic attachment to arm | Mechanical clamp around<br>laparoscopic instrument and | Substantially<br>equivalent with no<br>new or different |
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
| | | mechanical attachment to<br>arm | questions of safety or<br>effectiveness |
| Instrument<br>Coupling | Self-orienting sterile<br>couplers attached to the<br>instrument's shaft. Two<br>coupler sizes were used, 10<br>mm and 5 mm, to<br>accommodate the most<br>typical surgical instrument<br>shaft calibers | A modified Coupler design<br>maintains the same<br>functionality and sizes (5<br>and 10 mm).<br>Alignment and constraints<br>are achieved through new<br>mechanical methods on the<br>modified coupler. | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness |
| Mechanisms<br>involved in<br>physical<br>positioning | Electro-mechanical jointed<br>arms driven by internal<br>motors that follow the<br>physician movements. | Same | Identical |
| Method of<br>control | Direct manipulation of<br>conventional laparoscopic<br>instrument handles.<br>Endoscope and Instrument<br>positions are unlocked via<br>surgeon applied force to<br>laparoscope or instrument<br>handles. This unlocking<br>force is on the order of<br>magnitude of what would be<br>expected in holding or<br>repositioning the endoscope<br>or instruments themselves.<br>The endoscope or<br>instruments are repositioned<br>as desired by the surgeon.<br>Once the surgeon releases<br>the instruments or scope the<br>system maintains the new<br>position with motor driven<br>forces sufficient to<br>compensate for gravity. | Same | Identical |
| System Hold<br>Status Indicator | Two LED indicators on the<br>system forearms are used to<br>indicate transitions between | The three LED locations<br>have been expanded and<br>pushed further proximal on | Substantially<br>equivalent with no<br>new or different |
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
| | co-manipulation and hold<br>states as they change<br>between purple and blue,<br>respectively | the device. LED indicators<br>are present at the elbow and<br>shoulder of the support arm.<br>In addition to the visual<br>indicators, the system now<br>has light tactile feedback<br>upon transition from co-<br>manipulation to hold for<br>information purposes to the<br>user. The system provides<br>short vibrations along the<br>axis of the instrument,<br>which the surgeon feels<br>while holding the<br>instrument. | questions of safety or<br>effectiveness. |
| Alarm and<br>Warning<br>Condition<br>Indicators | The Maestro has two<br>primary LEDs for<br>communication to the<br>surgeon. These LEDs change<br>colors and pulse to<br>communicate system status.<br>In the case of a critical<br>system fault, the LEDs will<br>change to red. | Same | Identical |
| | The Maestro is also<br>equipped with a touch screen<br>graphical interface. This<br>screen provides messages to<br>the operating room staff<br>regarding system state. | | |
| Single Fault<br>Tolerance | Automatic System<br>Performance Monitoring in<br>real-time with multiple fault<br>tiers + visual supervision by<br>surgeon via LED indicating<br>system status. | Same | Identical |
| | Examples include:<br>- Redundant encoders on<br>motorized axes | | |
| ASPECT | MOON SURGICAL<br>(Predicate Device) | MOON SURGICAL<br>(Subject Device) | COMPARISON |
| | - Velocity, acceleration,<br>current and torque limits<br>- Brakes engage if power is<br>removed | | |
| Back-up fault<br>response | Brakes engage on motorized<br>axis in the event of a fault<br>state to prohibit any arm<br>motion | Same | Identical |
| Maximum<br>Applied Load | 4.4 lbs tested. | Same | Identical |
| Sterilization<br>Method | Steam - for coupling devices<br>used to attach laparoscopic<br>instruments | Ethylene Oxide (EtO) – for<br>single use coupling devices<br>used to attach laparoscopic<br>instruments | Substantially<br>equivalent with no<br>new or different<br>questions of safety or<br>effectiveness |
| Sterility barrier | Drape | Same | Identical |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Testing Performed
Design verification testing included the following:
- . Payload capacity
- Single fault condition ●
- Force accuracy ●
- Drape integrity ●
- Hold position accuracy.
- Positioning guidance and collision detection ●
- System positioning accuracy ●
- Bedside joint control ●
- System end-to-end workflow ●
- Design inspection ●
- System setup ●
- System latency ●
- System cleaning ●
- LED status ●
- Emergency stop ●
- Coupler performance
- IFU inspection ●
- Software verification .
{10}------------------------------------------------
- Electrical safety
- EMC
Design validation testing included the following:
- Cadaver testing ●
- Human factors testing ●
Testing described in this 510(k) consisted of verification of all system input requirements and product specifications. All clinical input requirements were validated.
#### Summary
Based upon the Intended Use, Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, the Maestro System has been shown to be substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.