RedDrop ONE (One)

K234081 · Reddrop DX · FMK · Mar 21, 2024 · General, Plastic Surgery

Device Facts

Record IDK234081
Device NameRedDrop ONE (One)
ApplicantReddrop DX
Product CodeFMK · General, Plastic Surgery
Decision DateMar 21, 2024
DecisionSESU
Submission TypeTraditional
Regulation21 CFR 878.4850
Device ClassClass 2

Indications for Use

The RedDrop ONE is a single-use blood lancing device intended for producing microliter capillary whole blood samples. It does not collect or transport such samples.

Device Story

RedDrop ONE is a single-use, spring-actuated blood lancing device for clinical use. Adhered to the upper arm, the device is activated by sliding an actuator, which releases an energized spring to deploy a lancet array (max depth 2 mm) into the skin to access capillaries. The same action releases spring-loaded pistons to create a gentle vacuum, facilitating capillary blood emergence. The lancet array automatically retracts to a locked, safe position to prevent sharps injury and re-activation. The device does not collect or transport blood. It is intended for use by healthcare professionals in clinical settings to produce blood samples for subsequent collection into compatible tubes.

Clinical Evidence

Clinical usability study conducted with human subjects. The device demonstrated a total success rate of over 95.0% in producing capillary blood samples from the upper arm when used according to instructions. Head-to-head clinical and usability studies were performed against the predicate device to confirm substantial equivalence. Results indicated the device is safe and effective for its intended use, with all residual risks deemed acceptable.

Technological Characteristics

Single-use, spring-actuated mechanical lancing device. Features include a lancet array, spring-loaded vacuum pistons, and a sharps injury prevention retraction/lock-out mechanism. Materials and design validated via ASTM F2096 (bubble leak), ASTM F88 (seal strength), ASTM F1980 (accelerated aging), and ANSI/AAMI ST72 (bacterial endotoxins). No electronic components or software.

Indications for Use

Indicated for producing microliter capillary whole blood samples via upper arm lancing. Not for at-home use or lay-user collection. Must be used with compatible collection tubes.

Regulatory Classification

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 20, 2024 RedDrop Dx Kris Buchanan CEO 2401 Research Blvd. #206 Fort Collins. Colorado 80526 Re: K234081 Trade/Device Name: RedDrop ONE (One) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: December 22, 2023 Received: December 22, 2023 Dear Kris Buchanan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. The OHT4: Office of Surgical and Infection Control Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions/Warnings/Contraindications section of the device's labeling in addition to being placed prominently immediately after any images or references to a collection tube: - 1. This device is only for use with compatible collection tubes that are cleared for use with this device: - 2. This device is not intended for use as a blood collection kit; and {2}------------------------------------------------ - 3. This device is not intended for at-home collection or collection by lay-users. Furthermore, the indication for use "The RedDrop ONE is a single-use blood lancing device intended for producing microliter capillary whole blood samples. It does not collect or transport such samples." must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {3}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Binita S. Ashar -S Binita Ashar, M.D., M.B.A., F.A.C.S. Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K234081 Device Name RedDrop ONE (One) Indications for Use (Describe) The RedDrop ONE is a single-use blood lancing device intended for producing microliter capillary whole blood samples. It does not collect or transport such samples. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ 510(k) Summary – RedDrop ONE This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92 | Submitter: | RedDrop Dx, Inc.<br>2401 Research Blvd. #206<br>Fort Collins, CO 80526<br>Phone: 970-443-8118<br>Contact: Kris Buchanan, CEO | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Summary: | December 20, 2023 | | Device Trade Name: | RedDrop ONE Lancet | | Common or Usual<br>Name: | Blood Lancet<br>Single Use Only Blood Lancet With An Integral Sharps Injury<br>Prevention Feature | | Regulation Number: | 21 CFR 878.4850 | | Device Class: | ll | | Product Code: | FMK | | Panel: | General Surgery OHT4 | | Predicate Device: | TAP Lancet (K223201) | | Device Description: | The RedDrop ONE Lancet is a single-use blood lancing<br>device with an integral sharps injury prevention feature<br>intended for producing microliter capillary whole blood<br>samples. The device is adhered to the skin in the selected<br>position on the upper arm. The device is actuated by sliding<br>the actuator causing an energized spring to release and<br>deploy the lancet array. The lancet array is deployed to a<br>maximum depth of 2 mm to access capillaries. The lancet<br>array is then automatically retracted to a safe position that<br>prevents sharps injury and re-activation. Sliding the actuator<br>also releases two spring loaded pistons that create a gentle<br>vacuum to the skin to facilitate the emergence of capillary<br>blood from the sample site. | | Intended Use/<br>Indications for Use: | The RedDrop ONE Lancet is a single-use blood lancing<br>device intended for producing microliter capillary whole blood<br>samples. It does not collect or transport such samples. | | Indications for Use<br>Comparison: | The indications for use of the candidate device are the same<br>as the indications for use of the predicate device. | | Technological<br>Comparison: | Penetration of capillary blood vessels and application of<br>vacuum to draw out whole blood samples is the underlying<br>principle of both the subject and predicate devices. At a high<br>level, the subject and predicate device are both based on the<br>same technological elements:<br>A spring-loaded array of sharp needles to penetrate<br>skin to reach capillary vessels. A retraction mechanism to withdraw the needle array<br>from the skin to a location where it can no longer be<br>accessed. A lock-out feature to permanently prevent the needles<br>from being re-deployed. A mechanical means of manual activation. A means of sealing the device to the skin. A means of using spring force to create vacuum at the<br>incision site. The following technological differences exist between the<br>subject and predicate devices: The subject device stores some spring energy in the<br>device that the predicate device requires the user to<br>supply. The predicate device requires an actuation force that is<br>directed toward the body. A lighter actuation force of<br>the subject device is applied parallel to the body. | | Non-Clinical<br>Performance Testing: | Non-Clinical performance testing was conducted to confirm<br>the device met its design specifications, intended use and<br>substantial equivalence. Additional comparison testing to the<br>predicate device was conducted to further support substantial<br>equivalence. Testing also included: | | | Package Integrity Testing including bubble leak per ASTM<br>F2096 and Seal Strength per ASTM F88. | | | Accelerated aging per ASTM F1980. | | | Bacterial Endotoxins testing per ANSI/AAMI ST72. | | Clinical Testing,<br>Summary and<br>Conclusions | The usability of the RedDrop ONE device with the instructions<br>for use was evaluated in an actual-use study and is discussed<br>in the clinical testing section of this submission. The device, in<br>conjunction with the instructions for use, was found to be safe | | and effective for the intended use, and all residual risks were<br>deemed acceptable for this type of device. An assessment of<br>clinical performance data for the RedDrop ONE device<br>successfully demonstrated its ability to produce blood<br>samples from the upper arm of human subjects according to<br>the device labeling. Subjects produced their blood samples<br>following the instructions for use. The devices had a total<br>success rate of over 95.0% and demonstrated that they<br>performed as intended. | | | In addition, clinical and useability studies were conducted<br>head-to-head with the predicate device to further confirm<br>substantial equivalence. | | {6}------------------------------------------------ {7}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...