← Product Code [FMK](/submissions/SU/subpart-e%E2%80%94surgical-devices/FMK) · K223370

# NeatNick Heel Safety Lancet (K223370)

_Facet Technologies, LLC · FMK · Feb 13, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FMK/K223370

## Device Facts

- **Applicant:** Facet Technologies, LLC
- **Product Code:** [FMK](/submissions/SU/subpart-e%E2%80%94surgical-devices/FMK.md)
- **Decision Date:** Feb 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4850
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Pediatric

## Intended Use

The NeatNick Heel Safety Lancet is intended for use by medical professionals to obtain a sample of blood from neonates and infants for subsequent diagnostic testing. The device is for use only on a single patient. The NeatNick Heel Safety Lancet is a disposable, single use device. The device incorporates a spring-actuated sterile sweeping action blade that automatically retracts into the device housing after deployment and renders the device inoperable thereafter.

## Device Story

Single-use, sterile blood sampling device for neonates and infants; used by medical professionals. Device features spring-actuated, sweeping-action stainless steel blade; automatically retracts into housing post-deployment to prevent reuse and sharps injury. Available in Preemie and Full Term configurations. Sterile barrier provided by LDPE overmolding. Operated by manual trigger actuation; provides controlled incision depth and length for blood collection. Benefits include standardized, safe blood sampling with reduced risk of sharps injury and cross-contamination.

## Clinical Evidence

No clinical data. Bench testing only; mechanical design verification and validation performed to ensure compliance with 21 CFR 878.4850 and shelf-life requirements. Biocompatibility testing conducted per ISO 10993.

## Technological Characteristics

Single-use, spring-actuated safety lancet. Materials: Type 301 stainless steel blade, High Impact Polystyrene (HIPS) body, Tenac 7010 acetal (polyoxymethylene homopolymer) trigger. Sterilization: Gamma radiation (SAL 10^-6). Actuation force: 8N-36N. Cut dimensions: Full term (2.9-4.1mm length, 1.0-1.5mm depth); Preemie (1.8-3.0mm length, 0.65-1.25mm depth).

## Regulatory Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

## Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- NeatNick Heel Safety Lancet (510(k) Exempt)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 13, 2023

Facet Technologies LLC James Bonds Director Regulatory Affairs 3900 North Commerce Drive Atlanta, Georgia 30344

Re: K223370

Trade/Device Name: NeatNick Heel Safety Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: January 10, 2023 Received: January 11, 2023

### Dear James Bonds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Carr-S

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K223370

Device Name

NeatNick Heel Safety Lancet

Indications for Use (Describe)

The NeatNick Heel Safety Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants for subsequent diagnostic testing.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K223370

# NeatNick Heel Safety Lancets

# 510(k) Summary

## (Reference 21 CFR 807.92)

| Submitted by:           | Facet Technologies, LLC<br>3900 North Commerce Drive<br>Atlanta, GA 30344-8149<br>Phone Number: (770) 590-6462<br>Fax Number: (770) 590-6412 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                | James R. Bonds                                                                                                                               |
| Date of Preparation:    | October 17, 2022                                                                                                                             |
| Device Trade Name:      | NeatNick Heel Safety Lancets                                                                                                                 |
| Common Name:            | Blood Lancet                                                                                                                                 |
| Classification Name:    | Single use only blood lancet with an integral sharps injury<br>prevention feature                                                            |
| Regulation:             | 878.4850                                                                                                                                     |
| Product codes:          | FMK (Single use only blood lancet with an integral sharps injury<br>prevention feature)                                                      |
| Product Classification: | II                                                                                                                                           |
| Panel:                  | General & Plastic Surgery                                                                                                                    |
| Predicate Device:       | NeatNick Heel Safety Lancet (Product Code FMK, 510(k)<br>Exempt)                                                                             |

{4}------------------------------------------------

## Device Description

The NeatNick Heel Safety Lancet is a single use, blood sampling device with a sterile blade that are used by medical professionals to obtain a blood sample from the feet of neonates and infants for diagnostic purposes. The NeatNick Lancet incorporates an integral sharps injury prevention feature that also prevents reuse of the lancet.

The NeatNick Lancet has been in commercial distribution since 2006.

NeatNick Lancets are available in two configurations (Preemie and Full Term) are intended to be used by medical professionals to obtain a blood sample for diagnostic testing. The lancets consist of a stainless steel blade overmolded with low density polyethylene (LDPE) which protects the blade until use and functions as a sterile barrier. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-6.

Biocompatibility testing has been conducted per ISO 10993 at a GLP testing facility. The lancet is classified as Class II.

The lancet is currently in commercial distribution in the United States and Canada. Since commercial distribution of the current lancets began, there have been no design changes.

### Indication for Use

Indication for Use: The NeatNick Heel Safety Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants for subsequent diagnostic testing.

### Technological Characteristics

The primary technological characteristics and intended use of the NeatNick Heel Safety Lancetss are the same as the identified predicate devices.

{5}------------------------------------------------

As indicated in Table 1, the NeatNick Heel Safety Lancet is substantially equivalent to characteristics of the identified predicate device, the NeatNick Heel Safety Lancet previously marketed as a 510(k) exempt device.

| Characteristic          | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Subject Device |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Indication for Use      | The NeatNick Heel Safety Lancet is a<br>single use device with a sterile blade that<br>is used by medical professionals to obtain<br>a blood sample from the heel of neonates<br>and infants for subsequent diagnostic<br>testing.                                                                                                                                                                                                                                                                                    | Same           |
| Intended Use            | The NeatNick Heel Safety Lancet is<br>intended for use by medical professionals<br>to obtain a sample of blood from neonates<br>and infants for subsequent diagnostic<br>testing. The device is for use only on a<br>single patient. The NeatNick Heel Safety<br>Lancet is a disposable, single use device.<br>The device incorporates a spring-actuated<br>sterile sweeping action blade that<br>automatically retracts into the device<br>housing after deployment and renders the<br>device inoperable thereafter. | Same           |
| Manufacturer            | Facet Technologies, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same           |
| 510(k) Number           | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same           |
| Product Code            | FMK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same           |
| Trigger/spring material | Tenac 7010 acetal (polyoxymethylene<br>homopolymer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same           |
| Blade material          | Type 301 stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same           |
| Trigger actuation force | 8N-36N                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same           |
| Length of cut           | Full term: 2.9mm-4.1mm<br>Preemie: 1.8mm-3.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same           |
| Depth of cut            | Full term: 1.0mm-1.5mm<br>Preemie: 0.65mm-1.25mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same           |
| Body Material           | High impact polystyrene (HIPS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same           |

|  | Table 1:      Comparison of Subject Device and Predicate Device |
|--|-----------------------------------------------------------------|
|--|-----------------------------------------------------------------|

{6}------------------------------------------------

| Characteristic   | Predicate Device                                 | Subject Device |
|------------------|--------------------------------------------------|----------------|
| Biocompatibility | Conforms to ISO10993-1                           | Same           |
| Blade Sterility  | Sterilized by Gamma Radiation<br>$SAL = 10^{-6}$ | Same           |

## Non-clinical Testing Summary and Conclusions

Non-clinical bench testing was performed to ensure predetermined criteria were met and the special controls (21 CFR 878.4850) were satisfied. This includes mechanical design verification and validation testing in order to ensure the risks were appropriately managed in addition to verifying that the mechanical functions of the device are suitable for use over the shelf life of the device.

Clinical testing is not applicable as the risk analysis confirmed that all identified risks were addressed and mitigated appropriately. Residual risks after mitigation were acceptable. There were no special performance of safety concerns identified.

In summary, the results of nonclinical testing demonstrate that the candidate device is substantially equivalent to the predicate device.

### Conclusion

The intended use, technology, non-clinical testing, and functionality of the NeatNick Heel Safety Lancets demonstrate a substantially equivalent safety and effectiveness profile to the predicate device and should perform as well as the predicate in the specified use conditions.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FMK/K223370](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/FMK/K223370)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
