BD Microtainer® Contact-Activated Lancets; BD Microtainer® Quikheel™ Lancets

K223243 · Becton, Dickinson and Company · FMK · Dec 16, 2022 · General, Plastic Surgery

Device Facts

Record IDK223243
Device NameBD Microtainer® Contact-Activated Lancets; BD Microtainer® Quikheel™ Lancets
ApplicantBecton, Dickinson and Company
Product CodeFMK · General, Plastic Surgery
Decision DateDec 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4850
Device ClassClass 2
AttributesPediatric

Intended Use

BD Microtainer® Contact-Activated Lancets The BD Microtaine® Contact-Activated Lancet is a sterile, single-use, permanently retracting, lancing device used to perform fingerstick punctures in in adults, adolescents, children, and infants (greater than 6 months of age and 10 kg [22 lbs}). The device is intended to be used by a healthcare professional in a clinical setting to obtain capillary blood specimens for testing utilizing small amounts of blood. BD Microtainer® Quikheel™ Lancets The BD Microtainer® Quikheel™ Lancet is a sterile, single-use medical device with a push-button activated, permanently retracting blade. The device is intended to be used by healthcare professionals to perform heel sticks in neomates, pre-term (preemie) and full-term infants (non-walking) to obtain capillary blood specimens for testing utilizing small amounts of blood.

Device Story

Sterile, single-use, permanently retracting lancing devices; used by healthcare professionals in clinical settings to obtain capillary blood specimens. Contact-Activated Lancets: pressure-activated needle or blade; auto-disables after use to prevent reuse and accidental sticks. Quikheel™ Lancets: push-button activated blade; designed for low-trauma sampling in infants. Both devices utilize stainless steel needles/blades and silicone lubricant. Output is a capillary blood sample for diagnostic testing. Benefits include standardized, safe, and low-trauma blood collection while minimizing sharps injury risk.

Clinical Evidence

No clinical data submitted; per 21 CFR 878.4850(a), clinical testing is not required for blood lancets.

Technological Characteristics

Materials: Polystyrene, Polypropylene, Polyethylene, ABS, steel tinned wire, stainless steel, silicone lubricant. Principle: Mechanical spring-actuated needle/blade retraction. Energy: Mechanical (spring). Form factor: Handheld, single-use. Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 10993 (biocompatibility), ISO 23908 (sharps protection), ISO 11137 (sterilization).

Indications for Use

Indicated for capillary blood specimen collection via fingerstick in adults, adolescents, children, and infants (>6 months, >10 kg) using BD Microtainer® Contact-Activated Lancets; or via heel stick in neonates, pre-term, and full-term infants (non-walking) using BD Microtainer® Quikheel™ Lancets.

Regulatory Classification

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. December 16, 2022 Becton, Dickinson and Company Alexandra Kirby Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417 # Re: K223243 Trade/Device Name: BD Microtainer® Contact-Activated Lancets; BD Microtainer® Quikheel™ Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: October 19, 2022 Received: October 20, 2022 Dear Alexandra Kirby: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Carr -S for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223243 #### Device Name BD Microtainer® Contact-Activated Lancets and BD Microtainer® Quikheel™ Lancets #### Indications for Use (Describe) BD Microtainer® Contact-Activated Lancets The BD Microtaine® Contact-Activated Lancet is a sterile, single-use, permanently retracting, lancing device used to perform fingerstick punctures in in adults, adolescents, children, and infants (greater than 6 months of age and 10 kg [22 lbs}). The device is intended to be used by a healthcare professional in a clinical setting to obtain capillary blood specimens for testing utilizing small amounts of blood. #### BD Microtainer® Quikheel™ Lancets The BD Microtainer® Quikheel™ Lancet is a sterile, single-use medical device with a push-button activated, permanently retracting blade. The device is intended to be used by healthcare professionals to perform heel sticks in neomates, pre-term (preemie) and full-term infants (non-walking) to obtain capillary blood specimens for testing utilizing small amounts of blood. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Research Use Only (RUO) / For Further Manufacturing</label> | | <label><input type="checkbox"/> For <i>In-Vitro</i> Diagnostic Use (IVD) / Intended Use</label> | |X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K223243 #### 510(K) SUMMARY 5 BD Microtainer® Contact-Activated Lancets and BD Microtainer® Ouikheel™ Lancets ## Summary Preparation Date: December 16, 2022 ## Submitted by: Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417-1885 Phone: (201) 847-6800 #### Contact: Alexandra Kirby Staff Regulatory Affairs Specialist Email: Alexandra.Kirby@bd.com Phone: (862) 774-2318 Proprietary Names: BD Microtainer® Contact-Activated Lancets, BD Microtainer® Quikheel™ Lancets Common or Usual Names: Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature ## Regulatory Information Classification Name: Blood Lancets Classification Regulation: 21 CFR 878.4850(a) Regulatory Class: Class II Product Code: FMK Classification Panel: General & Plastic Surgery Predicate Device: BD Microtainer® Contact-Activated Lancets, BD Microtainer® Quikheel™ Lancets (previously Class I exempt, reclassified to Class II devices requiring a 510(k) under Docket No. FDA-2016-N-0040 on November 22, 2021) ## Device Establishment: Becton, Dickinson and Company ## Registration Number: 2243072 ## Performance Standards: - 1. EN ISO 13485:2016 Medical devices Quality management systems Requirements for regulatory purposes - 2. EN ISO 14971:2019 Medical Devices Application of risk management to medical devices {4}------------------------------------------------ - 3. EN ISO 23908:2013 Sharps injury protection Requirements and test methods Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling - 4. ISO 11607-1:2019 Packaging for terminally sterilized medical devices Part 1: Requirements for materials, sterile barrier systems and packaging systems* - 5. ISO 11607-2:2019 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes* - 6. EN ISO 15223-1:2021 Medical devices Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements - 7. EN ISO 20417:2021 Medical devices Information to be supplied by the manufacturer - 8. EN 556-1:2001/AC:2006 Sterilization Of Medical Devices Requirements For Medical Devices To Be Designated "Sterile" - Part 1: Requirements For Terminally Sterilized Medical Devices - 9. EN ISO 11137-1:2015/A2:2019 Sterilization of health care products Radiation Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices - 10. EN ISO 11137-2:2015 Sterilization of health care products Radiation Part 2: Establishing the sterilization dose - 11. EN ISO 11137-3:2017 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation and routine control - 12. EN ISO 11737-1:2018/A1:2021 Sterilization of health care products Microbiological methods - Part 1: Determination of a population of microorganisms on products - 13. EN ISO 11737-2:2020 Sterilization of medical devices Microbiological methods Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process - 14. ISO/TS 13004:2013 Sterilization of health care products Radiation Substantiation of selected sterilization dose: Method VDmax D - 15. ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process - 16. EN ISO 10993-2:2006 Biological evaluation of medical devices Part 2: Animal welfare requirements - 17. ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood {5}------------------------------------------------ BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management Becton, Dickinson and Company - 18. ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity - 19. ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization - 20. ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity - 21. ISO 10993-12:2012 Biological evaluation of medical devices Part 12: Sample preparation and reference materials - 22. ISO 10993-17:2002 Biological evaluation of medical devices Part 17: Establishment of allowable limits for leachable substances - 23. ISO 10993-18:2020 Biological evaluation of medical devices Part 18: Chemical characterization of materials - 24. ISO 10993-23:2021, Biological Evaluation of Medical Devices Part 23: Tests for irritation. - 25. ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - 26. ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Svstems - 27. ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) - 28. ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials - 29. EN ISO 22442-1:2020 Medical devices utilizing animal tissues and their derivatives Part 1: Application of risk management - 30. EN ISO 22442-2:2020 Medical devices utilizing animal tissues and their derivatives Part 2: Controls on sourcing, collection and handling - 31. EN ISO 22442-3:2007 Medical devices utilizing animal tissues and their derivatives Part 3: Validation of the elimination and/or inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents - 32. CLSI GP42-A7: Collection of Capillary Blood Specimens. 7th Edition For the BD Microtainer® Contact-Activated Lancets, only partial compliance is claimed for standards marked with an asterisk (*). ## Indications for Use {6}------------------------------------------------ ## BD Microtainer® Contact-Activated Lancets The BD Microtainer® Contact-Activated Lancet is a sterile, single-use, permanently retracting. lancing device used to perform fingerstick punctures in adults, adolescents, children, and infants (greater than 6 months of age and 10 kg [22 lbs]). The device is intended to be used by a healthcare professional in a clinical setting to obtain capillary blood specimens for testing utilizing small amounts of blood. ## BD Microtainer® Quikheel™ Lancets The BD Microtainer® Quikheel™ Lancet is a sterile, single-use medical device with a push-button activated, permanently retracting blade. The device is intended to be used by healthcare professionals to perform heel sticks in neonates, pre-term (preemie) and full-term infants (nonwalking) to obtain capillary blood specimens for testing utilizing small amounts of blood. ## Device Description ## BD Microtainer® Contact-Activated Lancets The BD Microtainer® Contact-Activated Lancets are sterile, single-use lancets with a permanently retracting stainless steel needle or blade used for the purpose of obtaining capillary blood specimens for testing utilizing small amounts of blood. The BD Microtainer® Contact-Activated Lancet needle/blade is never visible and remains sterile until the protective tab cap is removed. The BD Microtainer® Contact-Activated Lancet is activated by applying pressure against the puncture site upon contact; after skin puncture, the needle/blade returns into the device. Once activated, the device is auto-disabled and cannot be re-used (when used as intended) to minimize the chance of an accidental fingerstick and contamination. The BD Microtainer® Contact-Activated Lancets are made up of the following key functional components: - 1. Rear Cap - 2. Drive Spring - 3. Lever Element - 4. Needle Carrier with Protective Tab - 5. Return Spring - 6. Shield - 7. Main Housing - 8. Tab Cap - 9. Silicone Lubricant ## BD Microtainer® Quikheel™ Lancets The BD Microtainer® Quikheel™ Lancets are sterile, single-use lancets with a permanently retracting stainless steel blade used for the purpose of obtaining capillary blood specimens for testing utilizing small amounts of blood. The BD Microtainer® Quikheel™ Lancets consist of a front and back housing, a trigger, and a spring onto which a stainless-steel blade is mounted. The stainless-steel blade is non-redeployable to minimize the chance of accidental blade stick injury {7}------------------------------------------------ BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management Becton, Dickinson and Company and contamination. The device is designed to produce a very rapid triggering of the blade along a predictable path, for low-trauma sampling. The BD Microtainer® Quikheel™ Lancets are made up of the following key functional components: - 1. Front Housing - 2. Back Housing - 3. Trigger - 4. Spring - 5. Blade - 6. Silicone Lubricant ## Substantial Equivalence1 The subject and predicate device are substantially equivalent as described in Table 1 and Table 2. | Table 1 | BD Microtainer® Contact-Activated Lancet Substantial Equivalence<br>Comparison | | | |-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristic | Subject Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets | Predicate Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets² | Comparison | | Device<br>Classification | Class II requiring a 510(k) | Class I 510(k) exempt | The subject BD Microtainer®<br>Contact-Activated Lancets<br>devices were reclassified from<br>Class I exempt devices not<br>requiring a 510(k) to Class II<br>devices requiring a 510(k)<br>under Docket No. FDA-2016-<br>N-0040, effective November<br>22, 2021. | | Characteristic | Subject Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets | Predicate Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets2 | Comparison | | Intended Use /<br>Indications for Use | The BD Microtainer®<br>Contact-Activated Lancet<br>is a sterile, single-use,<br>permanently retracting,<br>lancing device used to<br>perform fingerstick<br>punctures in in adults,<br>adolescents, children, and<br>infants (greater than 6<br>months of age and 10 kg<br>(22 lbs)). The device is<br>intended to be used by a<br>healthcare professional in<br>a clinical setting to obtain<br>capillary blood specimens<br>for testing utilizing small<br>amounts of blood. | The BD Microtainer®<br>Contact-Activated Lancet<br>is a sterile, single-use,<br>permanently retracting,<br>lancing device used to<br>perform fingerstick<br>punctures in in adults,<br>adolescents, children, and<br>infants. The device is<br>intended to be used by a<br>healthcare professional in a<br>clinical setting to collect<br>capillary blood specimens<br>for diagnostic testing. | The current intended use<br>statement has been converted<br>to an indications for use<br>statement for this initial<br>510(k) submission. The<br>subject devices' intended<br>use/indications for use also<br>incorporates a minimum age<br>and weight indication<br>consistent with<br>recommendations outlined for<br>fingerpricks in CLSI GP42-<br>A7: Collection of Capillary<br>Blood Specimens, 7th Edition<br>(FDA Recognition Number 7-<br>301) and WHO Guidelines On<br>Drawing Blood: Best Practices<br>In Phlebotomy. The proposed<br>indications for use only add<br>specificity and clarity about<br>what minimum age/weight<br>criteria are appropriate for the<br>fingerstick; the changes do not<br>result in a new intended use. | | Target Population | Adults, adolescents,<br>children, and infants<br>(greater than 6 months of<br>age and 10 kg (22 lbs)) | Adults, adolescents,<br>children, and infants | This proposed modification<br>incorporates a minimum age<br>and weight indication<br>consistent with<br>recommendations outlined for<br>fingerpricks in CLSI GP42-<br>A7: Collection of Capillary<br>Blood Specimens, 7th Edition<br>(FDA Recognition Number 7-<br>301) and WHO Guidelines On<br>Drawing Blood: Best Practices<br>In Phlebotomy. The proposed<br>indications for use only add<br>specificity and clarity about<br>what minimum age/weight<br>criteria are appropriate for the<br>fingerstick; the changes do not<br>result in a new target<br>population. | | Device Design | | | | | Models | 366592<br>366593<br>366594<br>366599 | 366592<br>366593<br>366594<br>366599 | Same | | Colors | Purple, Pink, Blue | Purple, Pink, Blue | Same | | Types of Lancing<br>Device | Needle, Blade | Needle, Blade | Same | | Characteristic | Subject Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets | Predicate Device<br>BD Microtainer®<br>Contact-Activated<br>Lancets2 | Comparison | | Needle/Blade<br>Depths | 1.5 mm, 1.8 mm, 2.0 mm | 1.5 mm, 1.8 mm, 2.0 mm | Same | | Needle/Blade<br>Gauges/Widths | 30G (0.31 mm), 21G<br>(0.81 mm), 1.5 mm | 30G (0.31 mm), 21G (0.81<br>mm), 1.5 mm | Same | | Blood Flows | Low, Medium, High | Low, Medium, High | Same | | <b>Device Materials</b> | | | | | Needle Carrier with<br>Protective Tab | Polystyrene | Polystyrene | Same | | Main Housing/Tab<br>Cap/Rear Cap | Polypropylene | Polypropylene | Same | | Shield | Polyethylene | Polyethylene | Same | | Lever Element | Acrylonitrile butadiene<br>styrene (ABS) | Acrylonitrile butadiene<br>styrene (ABS) | Same | | Drive Spring | Steel tinned wire | Steel tinned wire | Same | | Return Spring | Steel tinned wire | Steel tinned wire | Same | | Needle/Blade | Stainless steel | Stainless steel | Same | | Needle/Blade<br>Lubricant | Silicone fluid | Silicone fluid | Same | | Materials | Compliant with ISO<br>10993 series | Compliant with ISO 10993<br>series | Same | | <b>Packaging and Sterility</b> | | | | | Sterile Needle<br>Carrier and<br>Needle/Blade | Yes | Yes | Same | | Sterility Assurance<br>Level (SAL) 10-6 | Yes | Yes | Same | | Sterilization Method | Gamma irradiation | Gamma irradiation | Same | | Shelf Life | 5 years | 5 years | Same | | Contains Packaging<br>Support Insert | No | Yes | To accommodate the new,<br>standalone IFU created for the<br>subject BD Microtainer®<br>Contact-Activated Lancets,<br>the shelf carton height and the<br>shelf box material density will<br>be increased. The increased<br>shelf carton height and density<br>will also result in the removal<br>of a support insert previously<br>used to hold the box together.<br>Ship testing was conducted to<br>support this packaging change. | <sup>1</sup> The term "substantial equivalence" as used in this 510(k) Notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts. 2 Per Docket No. FDA-2016-N-0040, " . . any 510(k)-exempt blood lances legally offered for sale on or before November 22, 2021, can serve as predicates for substantial equivalence purposes.' {8}------------------------------------------------ BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management Becton, Dickinson and Company {9}------------------------------------------------ BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management Becton, Dickinson and Company {10}------------------------------------------------ BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management Becton, Dickinson and Company | | Subject Device | Predicate Device | | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristic | BD Microtainer®<br>Quikheel™ Lancets | BD Microtainer®<br>Quikheel™ Lancets3 | Comparison | | Device<br>Classification | Class II requiring a 510(k) | Class I 510(k) exempt | The subject BD<br>Microtainer®<br>Quikheel™ Lancets<br>devices were<br>reclassified from Class<br>I exempt devices not<br>requiring a 510(k) to<br>Class II devices<br>requiring a 510(k)<br>under Docket No.<br>FDA-2016-N-0040,<br>effective November<br>22, 2021. | | Intended Use /<br>Indications for Use | The BD Microtainer®<br>Quikheel™ Lancet is a sterile,<br>single-use medical device with<br>a push-button activated,<br>permanently retracting blade.<br>The device is intended to be<br>used by healthcare<br>professionals to perform heel<br>sticks in neonates, pre-term<br>(preemie) and full-term infants<br>(non-walking) to obtain<br>capillary blood specimens for<br>testing utilizing small amounts<br>of blood. | The BD Microtainer®<br>Quikheel™ Lancet is a sterile,<br>single-use medical device with<br>a push-button activated,<br>permanently retracting blade.<br>The device is intended to be<br>used by healthcare<br>professionals to perform heel<br>sticks in neonates, pre-term<br>(preemie) and full-term infants<br>(non-walking) to obtain<br>capillary blood specimens for<br>testing utilizing small amounts<br>of blood. | The current intended<br>use statement has been<br>converted to an<br>indications for use<br>statement for this<br>initial 510(k)<br>submission. | | Target Population | Neonates, Premature And Full<br>Term Infants | Neonates, Premature And Full<br>Term Infants | Same | | Device Design | | | | | Models | 368100<br>368101 | 368100<br>368101 | Same | | Colors | Pink, Teal | Pink, Teal | Same | | Incision Depths | 0.85 mm, 1.00 mm | 0.85 mm, 1.00 mm | Same | | Incision Lengths | 1.75 mm, 2.50 mm | 1.75 mm, 2.50 mm…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...