K161002 · Cook Incorporated · BTA · Sep 28, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161002
Device Name
Aprima Smartesis Centesis Pump
Applicant
Cook Incorporated
Product Code
BTA · General, Plastic Surgery
Decision Date
Sep 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aprima Smartesis™ Centesis Pump is a motorized pump to be used in conjunction with the appropriate disposable Aprima Smartesis Centesis Collection Tray for paracentesis drainage procedures.
Device Story
Portable, reusable peristaltic suction pump; used for paracentesis drainage. Input: mechanical displacement of fluid via centesis tubing set (part of disposable collection tray). Operation: user-controlled via front panel knob; variable speed control of rollers determines fluid withdrawal rate. Output: controlled suction for fluid removal. Used in clinical settings; operated by healthcare professionals. Safety features include a movable guard door preventing operation while open and one-direction flow clips to prevent incorrect tubing installation. Benefits: provides rapid, convenient fluid drainage from the abdominal cavity.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including simulated use studies to obtain clinical opinions on device performance, flowrate/pressure characterization, life cycle cleaning tests, label durability, and distribution testing. All predetermined acceptance criteria were met.
Technological Characteristics
Reusable peristaltic suction pump; AC-powered. Features variable speed control, grounded enclosure, and movable guard door. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC). Includes firmware for device control.
Indications for Use
Indicated for patients requiring paracentesis drainage procedures for the removal of ascitic fluid from the abdominal cavity.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2016
Cook Incorporated Ms. Erum Nasir Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K161002
Trade/Device Name: Aprima Smartesis Centesis Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: August 31, 2016 Received: September 2, 2016
Dear Ms. Nasir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth
{2}------------------------------------------------
in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K161002
Device Name
Aprima Smartesis TM Centesis Pump
Indications for Use (Describe)
The Aprima Smartesis™ Centesis Pump is a motorized pump to be used in conjunction with the appropriate disposable Aprima Smartesis TM Centesis Collection Tray for paracentesis drainage procedures.
Type of Use (Select one or both, as applicable)
[X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, block letters on a darker red background.
#### 510(K) Summary
## Aprima Smartesis "11 Centesis Pump (21 CFR §878.4780) Date Prepared: August 31, 2016
Bloomington, IN 47404
(812) 335-3575 x102607
Powered suction pump
Aprima Smartesis™ Centesis Pump
Pump, Portable, Aspiration (Manual or Powered) - General &
(812) 332-0281
Plastic Surgery
21 CFR 8878.4780
#### Submitted By: Cook Incorporated Applicant: Contact: Erum B. Nasir Applicant Address: Cook Incorporated 750 Daniels Way
Contact Phone Number: Contact Fax Number:
### Device Information:
Trade Name: Common Name: Classification Name/Panel:
Regulation Product Code
### Predicate Device:
- RenovaRP™ Paracentesis Pump (GI Supply, K970186) o
BTA
### Device Description:
The Aprima Smartesis "11 Centesis Pump is a portable, reusable peristaltic suction pump designed for paracentesis procedure in a manner that is both rapid and convenient. The Aprima Pump is comprised of one pump, one power cord, and operating instructions. The Aprima Pump employs peristaltic pump technology to provide displacement of fluid through the centesis tubing set, a component of the Aprima Smartesis Centesis Trav. Operation of the Aprima Pump is initiated with a control knob on the front panel. The Aprima Pump is equipped with a variable speed control of the rollers, thus allowing the operator to control the rate at which the fluid is withdrawn from the patient.
The Aprima Pump is to be used with the Aprima Smartesis Collection Tray only, which includes the needle for access to the abdominal cavity and the connecting tube with attached drainage bags for removal of the fluid during treatment. The Aprima Smartesis™
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Cook Medical logo. The logo has a red background with the word "COOK" in white, large, sans-serif font. Below the word "COOK" is a smaller red rectangle with the word "MEDICAL" in white, sans-serif font. The Cook logo is a registered trademark, as indicated by the small "®" symbol next to the word "COOK".
Centesis Collection Tray and Bags are not included with the Aprima Pump and are packaged and sold individually. No tools are required to load or to remove the tubing and drainage bags.
## Intended Use:
The Aprima Smartesis™ Centesis Pump is a motorized pump to be used in conjunction with the appropriate disposable Aprima Smartesis Centesis Collection Tray for paracentesis drainage procedures.
### Comparison to Predicates:
The Aprima Pump is substantially equivalent to the predicate device, the RenovaRPTM Paracentesis Pump. The devices have the same intended use, operating mechanism, basic design, portability, and function. The minor differences between the subject and the proposed device are presented in the following table and are listed as dimension, weight, input current/voltage, maximum vacuum, AC power operations, and safety features. These minor differences were appropriately assessed and do not raise new questions regarding safety or effectiveness. In addition, the testing performed demonstrates that the proposed device meets the applicable electrical safety requirements, electromagnetic compatibility requirements, and has comparable performance as compared to the predicate device. The comparison table is listed below,
| Device | RenovaRP™ Paracentesis Pump<br>(K970186) | Aprima Smartesis™ Centesis Pump<br>(Proposed device) |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 CFR §878.4780 | 21 CFR §878.4780 |
| Classification | Class II | Class II |
| Product code | BTA - Pump, Portable, Aspiration<br>(manual or powered) | BTA - Pump, Portable, Aspiration<br>(manual or powered) |
| Intended Use | The GI Supply RenovaRP™<br>Paracentesis Pump is intended as a<br>suction pump to remove ascitic fluid<br>from the abdominal cavity in<br>conjunction with the GI Supply<br>Paracentesis Kit. | The Aprima Smartesis™ Pump is a<br>motorized pump to be used in<br>conjunction with the appropriate<br>disposable Aprima Smartesis™<br>Collection Tray for paracentesis<br>drainage procedures. |
| Duration of use | Reusable | Identical |
| Dimensions | 13 x 9 x 13 in<br>(33 x 23 x 33 cm) | 10 x 6.8 x 6.3 in<br>(25.6 x 17.2 x 16.0 cm) |
| Weight | 8.5 lbs (3.9 kg) | 12.5 lbs (5.7 kg) |
| Input current | 2.5A | 2.0A |
| Input Voltage | 100/230V | 115/230V |
| Input Frequency | 50/60Hz | Identical |
| Power Cord | Hospital Grade power cord | Identical |
| Energy used | AC-powered | Identical |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white, sans-serif letters on a darker red background.
| Device | | RenovaRPTM Paracentesis Pump<br>(K970186) | Aprima SmartesisTM Centesis Pump<br>(Proposed device) |
|----------------------------------|------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of pump | | Peristaltic | Identical |
| Maximum measured<br>vacuum (kPa) | | -94.2 (used with 7Fr cannula) | -85.7 (used with 5Fr needle)<br>-88.9 (used with 4Fr needle) |
| Pressure Control | | Not controlled | Identical |
| Speed control | | Dial | Identical |
| Filter | | No | Identical |
| AC power operation | | AC power switch is part of the dial | AC on and off switch |
| Handle for<br>transporting pump | | Yes, portable | Identical |
| Safety<br>features | Usability | Safety guard, but device can be run<br>while open. | Door acts as a movable guard that<br>prevents use while open,<br>One-direction flow (clips in the tubing<br>set prevent incorrect installation). |
| | Electrical | Safety ground | Grounded enclosure |
# Technological Characteristics:
The following tests have been conducted to ensure reliable design and performance under the specified testing parameters. These tests include:
- 1. Electrical Safety Testing Cook Inc. holds the UL certificate for the Aprima Pump. The UL certificate shows that the Aprima Pump complies with the requirements of IEC 60101-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. Tests covered by UL certificate include:
- 0 Leakage Currents Test
- Dielectric Strength Test
- Creepage Distances Test ●
- Spillage Test ●
- Push Test ●
- Impact Test ●
- Drop Test
- Grip/handle Strength Test ●
- Movable Guards Test ●
- Audible Acoustic Energy Test ●
- Maximum Temperature Test during Normal Use ●
- Instability Test ●
- Firmware Verification ●
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The logo is simple and clean, with a focus on the company name.
COOK INCORPORATED OLL FREE: 800.457. WW.COOKMEDICAL.COM
- 2. Electromagnetic Compatibility Testing was performed to show the device conforms to the requirements of IEC 60101-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, including:
- Harmonic Current Emission
- o Voltage Fluctuations and Flicker
- Power Line Conducted Emissions and Radiated Emissions ●
- Electrostatic Discharge Immunity ●
- o Electrical Field Radiated Immunity
- Electrical Fast Transient/Burst Immunity ●
- Surge Immunity
- Conducted Immunity
- Magnetic Field Immunity .
- Voltage Dips, Short Interruptions and Voltage Variations Immunity o
- 3. Performance Testing
- 0 Simulated Use - Testing was conducted to obtain clinical opinions on the use of the Aprima Smartesis™ Centesis Pump and Collection Tray. All users were able to use provided components to successfully perform the centesis procedure in accordance with the intended use statement and the predetermined acceptance criteria for this study were met.
- . Characterization of Flowrate and Pressure at Different Speed Settings for the Aprima Smartesis TM Centesis Pump and the RenovaRP™ Paracentesis Pump -Testing was performed to characterize the flowrate and pressure for the proposed pump and the predicate pump.
- Life Cycle Cleaning Test Testing was performed to show that the Aprima . Smartesis 114 Centesis Pump did not exhibit degradation of electrical safety after the simulation of 1 year life cycle of cleaning. The predetermined acceptance criterion was met.
- Wipe Test of Various Labels Testing was performed to show that the pump . labeling shall be legible after the required number of cleaning cycles. The predetermined acceptance criterion was met.
- Aprima Smartesis™ Centesis Pump Distribution Testing Testing was performed . to show that pump functions correctly without visible damage before and after
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, smaller, bold letters. The logo is simple and clean, with a focus on the company name.
COOK INCORPORATED 750 DANIELS WAY OMINGTON, IN 47404 USA OLL FREE. 800 VW.COOKMEDICAL.COM
distribution cycle testing in packaging. The predetermined acceptance criterion was met.
# Conclusion:
The results of these tests support a conclusion that the Aprima Smartesis™ Centesis Pump met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness. The Aprima Smartesis™ Centesis Pump is substantially equivalent to the RenovaRP™ Paracentesis Pump (GI Supply, K970186).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.