EPIX IRRIGATION PUMP, EPIX SUCTION IRRIGATION TUBING SET

K122619 · Applied Medical · BTA · Oct 2, 2012 · General, Plastic Surgery

Device Facts

Record IDK122619
Device NameEPIX IRRIGATION PUMP, EPIX SUCTION IRRIGATION TUBING SET
ApplicantApplied Medical
Product CodeBTA · General, Plastic Surgery
Decision DateOct 2, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Epix Suction Irrigation System is intended as a general purpose suction and/or irrigation device for use in laparoscopic and open general surgery, laparoscopic and open gynecological surgery, laparoscopic and open urologic surgery. This device delivers sterile irrigant solution and serves as a conduit for suction.

Device Story

Epix Suction Irrigation System comprises reusable irrigation pump (model C7000) and sterile, single-use suction irrigation tubing set (model C7100). Pump operates outside sterile field; connects to house vacuum and IV fluid bags. Handpiece valves actuate suction and irrigation. Device delivers sterile irrigation fluid to surgical site and evacuates blood, tissue debris, and smoke. Used in laparoscopic and open general, gynecological, and urologic surgeries. Operated by surgical staff. Output allows visualization and site clearance; facilitates surgical workflow.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included irrigation fluid flow, aspiration fluid flow, smoke evacuation flow, fluid leak tests, and tubing collapse testing per ISO 10079-3.

Technological Characteristics

System includes reusable pump and single-use tubing. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993 (biocompatibility). Suction powered by house vacuum; irrigation powered by pump. Flow rates are adjustable via pump settings. Handpiece-actuated valves control fluid delivery and suction.

Indications for Use

Indicated for patients undergoing laparoscopic or open general, gynecological, or urologic surgery requiring suction and/or irrigation.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K122619 Page 1 of 2 # 510(k) Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c). | 510(K) Submitter: | 22872 Avenida Empresa<br>(949) 713-8000 | Applied Medical Resources Corporation<br>Rancho Santa Margarita, CA, 92688 | ОСТ<br>2 2012 | | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------|--| | Contact Person: | Jessica Cho<br>Regulatory Affairs Specialist<br>Applied Medical Resources<br>jcho@appliedmedical.com<br>Tel: (949) 713-8470 ext. 6697<br>Fax: (949) 713-8205 | | | | | Date of Preparation: | August 24, 2012 | | | | | Trade Name: | Epix® Suction Irrigation System | | | | | Common Name: | Suction Irrigator | | | | | Classification: | Table 5-1 - Classifications | | | | | | Product Code | Classification<br>Regulation | | | | | GCX | 21 CFR 880.6740 | Vacuum-powered Body Fluid<br>Suction Apparatus | | | | HET | 21 CFR 884.1720 | Gynecologic Laparoscope and<br>accessories | | | | KQT | 21 CFR 876.4370 | Gastroenterology-urology<br>evacuator | | | Predicate Device: | Stryker® Cassette Pump<br>510(k) #: K042454<br>Product codes: GCX, HET, KQT | | | | | Device Description: | The Epix Irrigation Pump (model C7000) is reusable and operates<br>outside of the sterile field. The Epix Suction Irrigation Tubing Set<br>(model C7100) is a sterile, single use device, used in conjunction with<br>the Epix Irrigation Pump and the house vacuum to deliver sterile<br>irrigation fluids and to evacuate blood, tissue debris, and smoke from<br>the surgical site. | | | | | Intended Use: | The Epix Suction Irrigation System is intended as a general purpose<br>suction and/or irrigation device for use in laparoscopic and open<br>general surgery, laparoscopic and open gynecological surgery,<br>laparoscopic and open urologic surgery. This device delivers sterile<br>irrigant solution and serves as a conduit for suction. | | | | , - 5-1 {2}------------------------------------------------ #### Summary of Technological Characteristics: Page 2 of (2) The Epix® Suction Irrigation System is substantially equivalent in safety and effectiveness to the Stryker® Cassette Pump (K042454). 122619 The subject and predicate systems are technologically similar in that they are both designed to provide controlled suction from and irrigation to the operative site during laparoscopic and open procedures. For each device, the pump is capable of increasing the rate of fluid delivery and has multiple flow rate settings. The tubing set connects with IV fluid bags and the house vacuum. Suction and irrigation are actuated by the valves on the handpiece. Both systems flush blood and tissue debris from the operative site during surgery. #### Discussion of Testing: The Epix Suction Irrigation System is designed to the following safety and performance standards: IEC 60601-1 Medical electrical equipment - General requirements for safety, IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests, and ISO 10993 Biological Evaluation of Medical Devices. Applied Medical created a testing protocol to confirm substantial equivalency between the subject and predicate. The devices were tested side-by-side to evaluate substantial equivalence in performance in a laboratory setting. The bench top tests seen in Table 5-2 were designed to focus on the functional performance of both the suction and irrigation features. | Table 5-2 - Performance Tests | | |-------------------------------|----------------------------| | Irrigation | Suction | | Irrigation Fluid Flow Test | Aspiration Fluid Flow Test | | Irrigation Fluid Leak Test | Smoke Evacuation Flow Test | The subject device was also evaluated by a tubing collapse test based on an applicable section of ISO 10079-3 - Medical suction equipment - Part 3: Suction equipment powered from a vacuum or pressure source. #### Conclusions Drawn from Testing: Testing demonstrates that the subject Epix Suction Irrigation System is substantially equivalent to the predicate Stryker Cassette Pump. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized wing segments. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 ОСТ Applied Medical % Ms. Jessica Cho Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, California 92688 Re: K122619 Trade/Device Name: Epix® Suction Irrigation System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: BTA Dated: August 24, 2012 Received: August 28, 2012 Dear Ms. Cho: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set ` 2 2012 {4}------------------------------------------------ #### Page 2 - Ms. Jessica Cho forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K122 6.19 ## INDICATIONS FOR USE #### 510(k) Number (if known): ### Device Name: Epix® Suction Irrigation System Indications for Use: The Epix® Suction Irrigation System is intended as a general purpose suction and/or irrigation device for use in laparoscopic and open general surgery, laparoscopic and open gynecological surgery, and laparoscopic and open urologic surgery. This device delivers sterile irrigant solution and serves as a conduit for suction. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) Nil Re. Bohn for mxm (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 510(k) Number K122619 #### Section 4 | INDICATIONS FOR USE STATEMENT 4-1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...