The SIMEX Portable Suction / Aspirator / Vacuum system is indicated and used for vacuum suction, extraction, aspiration and removal of surgical fluids, tissue (including bone), bodily fluids or infectious materials from wounds or from patient's airway or respiratory system, either during surgery or at patient's bedside.
Device Story
SIMEX Suction Pump: portable/stationary powered suction device; removes secretions, blood, body fluids, tissue via continuous negative pressure. Input: electrical power (AC or DC battery); Output: negative pressure transmitted through tubing/catheter to collection jar. Operation: maintenance-free electrical piston or diaphragm pump; vacuum level/flow rate manually regulated via secondary air valve and analogue manometer. Used in hospitals, clinics, ambulances, home care; operated by clinicians or patients. Aspirated fluids collected in jar isolated by hydrophobic filter; mechanical overflow protection prevents pump contamination. Benefits: effective fluid/tissue removal from wounds/airways. No electronic process control; all adjustments manual.
Clinical Evidence
Bench testing only. Performance data provided includes suction capacity (l/min), noise levels (db(A)), and operational duration for various models. No clinical data presented.
Technological Characteristics
Powered suction pump; piston or diaphragm vacuum pump; moulded plastic housing; manual control via secondary air valve and analogue manometer; AC or DC battery power; hydrophobic filter and mechanical overflow protection; non-patient contacting; dimensions 130mm x 359mm x 290mm; compliant with IEC 601-1-2/EN 60601-1-2.
Indications for Use
Indicated for vacuum suction, extraction, aspiration, and removal of surgical fluids, tissue (including bone), bodily fluids, or infectious materials from wounds, airways, or respiratory systems in surgical or bedside settings.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{1}------------------------------------------------
# Kolli33
# JUL 1 1 2006
# 510(k) Summary of Safety and Effectiveness
・・・・
# Safe Medical Devices Act of 1990 (SMDA)
### 510(k) Summary
| Name of Firm: | NovaSpine LLC<br>15 Heritage Court<br>Tarrytown, NY 10591<br>Telephone: 914-909-6577<br>Fax: 914-909-6577 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Mr. Hamid Khosrowshahi<br>NovaSpine LLC<br>PO Box 969<br>Elmsford, NY 10523 |
| Manufacture: | SIMEX Medizintechnik, GmbH<br>Mozartstrasse 13<br>78652 Deisslingen, Germany |
| Trade Name: | SIMEX Suction Pumps Models:<br>SIMEX "AC20", SIMEX "AC20K",<br>SIMEX "AC30",<br>SIMEX "DC20", SIMEX "DC30" and<br>SIMEX "DC30S" |
| Common Name: | Suction / Aspiration Pump |
| Classification: | Suction / Aspiration Pump<br>FDA 21 CFR 878.4780<br>Powered Suction Pump<br>Class II |
| Device Product Code | BTA - Pump, Portable, Aspiration,<br>(Manual or Powered) |
| Substantial Equivalency | Code: BTA<br>Precision Medical, Inc.<br>"Easy Go", K 971749 |
{2}------------------------------------------------
#### Device Description:
The SIMEX Suction Pump is a lightweight portable or stationary suction/ aspirator pump for medical suction procedures where secretions, blood and other body fluids must be removed through the application of continuous negative pressure. Applications range from hospital, medical practice, emergency care, ambulance, and in home usage. I
The SIMEX Suction Pumps have two Models. These Models have the same physically size and function but are differentiated in capacity as summarized below:
| 1. The SIMEX AC Model powered<br>with alternating power. This Model<br>has of three versions: | 2. The SIMEX DC Model using an on<br>board battery. This Model has of<br>three versions: |
|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| a. SIMEX "AC20",<br>i. Suction 18 1/min,<br>ii. Noise 56 db (A),<br>iii. Weight 3200 g,<br>iv. Operation Continuous<br>v. Color: Blue | a. SIMEX "DC20",<br>i. Suction 20 1/min,<br>ii. Noise 63 db (A),<br>iii. Weight 3100 g,<br>iv. Operation 20 Min<br>v. Color: Red |
| b. SIMEX "AC20K"<br>i. Suction 18 1/min,<br>ii. Noise 63 db (A),<br>iii. Weight 3200 g,<br>iv. Operation 20 Min<br>v. Color: Blue | b. SIMEX "DC30",<br>i. Suction 28 1/min,<br>ii. Noise 63 db (A),<br>iii. Weight 3500 g,<br>iv. Operation 20 Min<br>v. Color: Yellow |
| c. SIMEX "AC30"<br>i. Suction 28 1/min,<br>ii. Noise 61 db (A),<br>iii. Weight 3900 g,<br>iv. Operation 30 Min<br>v. Color: Green | c. SIMEX "DC30S"<br>i. Suction 28 1/min,<br>ii. Noise 62 db (A),<br>iii. Weight 3500 g,<br>iv. Operation 45 Min<br>v. Color: Yellow |
### Table 1
Models designed for mounting on a panel and packaged with a special "wall holder" are so designated with the letter "W" after the model and version number, e.g., "SIMEX DC 20W". All moving and electrical components are housed in a moulded compact plastic housing (130 mmx359 mmx290 mm). The unit includes attachment for an external power supply and/or a battery charger.
{3}------------------------------------------------
The vacuum is produced with a maintenance-free electrical piston or diaphragm vacuum pump. The vacuum pump transmits negative pressure throughout the tubing-system and the collection jar to a suction catheter that aspirates secretions or liquid from the body. The aspirated fluids are collected in the collection jar that is isolated from the pump by a hydrophobic filter. It also has a mechanical overflow protection in the lid of the collection jar to further prevent the fluids from being sucked into the pump. The filters are single use attachments that can be readily exchanged. The tubing and collection jar can either be replaced after use or sterilized using chemical disinfectants or can be autoclaved in accordance with the instructions supplied. Alternatively the collection jar can be replaced with a holder and a disposable aspiration bag.
A secondary air valve placed in the suction system can regulate the degree of vacuum and flow rate. The actual suction capacity can be estimated from the analogue manometer mounted on the pump. The pump vacuum can be manually adjusted using the regulating valve and the vacuum gage. The device can be adjusted to low vacuum values for use in nasal, pharyngeal and tracheal areas. Note that all controls are manual and the pump does not require any electronic process control.
The pump can be easily moved by hand or (versions that have a "W" designation) rapidly mounted to vehicle sides, room walls or wheel chairs. The SIMEX DC models have integrated electric supply enables charging and operation of the pump with the 12 to 24 Volts DC power supply. Also, if the units are used in home care, hospitals or clinical practices the SIMEX AC or DC models can be directly connected to AC power supplies for operation.
Since neither the pump nor its components come in direct contact with the patient sterilization is not required. The SIMEX Suction Pump is composed of two parts:
- 1. The pump, power unit and power supply all housed in a single unit
- 2. The consumable components such as:
- Collection jar (secretion container), �
- Related hoses (tubing)
- Hose connectors
The operating details of both models, SIMEX AC and SIMEX DC, are given in the operating manuals for the pumps (Product Inserts). These include detail list of all accessories that can be used with each suction pump, the part numbers, and the specifications for their use.
Details regarding the assembly and disassembly of each model are given in Service Instructions. These Service Instruction are for use by trained technicians who are familiar with the product. They are not to be used by untrained end users. These instructions will not be packaged with the pump.
{4}------------------------------------------------
#### Manufacturing:
The SIMEX suction pumps are manufactured in accordance with EEC Directive 93/42/EEC Annex IX and classified as suction unit of Class IIa. They all bear the "CE" marking for CE 0483 and meet the interference requirements for IEC 601-1-2/EN 60601-1-2. All SIMEX pumps are manufactured under recognized medical technical regulations and directions of the laws relating to medical products. The quality management is in accordance with certified international standards.
#### Performance Data:
SIMEX Suction Pumps are classified under FDA 21 CFR 878.4780 "Powered Suction Pump" Class II. A summary performance data is provided in Table 1 above "Device Description".
#### Basis For Substantial Equivalence:
This product is substantially equivalent to similar devices with similar technical specifications currently on the market such as: Precision Medical, Inc. "Easy Go", K 971749 and Medela, Inc. "Vario", K 983552.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Public Health Service
JUL 1 1 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NovaSpine LLC % Mr. Hamid Khosrowshahi President P.O. Box 969 Elmsford, NY 10523
Re: K061133
Trade/Device Name: SIMEX Suction Pump models: SIMEX AC20 pn 10401, SIMEX 20K pn 10400, SIOMEX AC30 pn 10501, SIMEX DC20 pn 10601, SIMEX DC30 pn 10701, SIMEX DC30S pn 10703
Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: II Product Code: BTA Dated: April 20, 2006 Received: April 24, 2006
Dear Mr. Khosrowshahi
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Hamid Khosrowshahi
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Hubert Lenoir
Mark N. Melkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
## 510(k) Number:
### Device Name:
SIMEX Suction Pump models: AC Power
DC Power
| SIMEX AC20 pn 10401 | SIMEX DC20 pn 10601 |
|----------------------|----------------------|
| SIMEX AC20K pn 10400 | SIMEX DC30 pn 10701 |
| SIMEX AC30 pn 10501 | SIMEX DC30S pn 10703 |
## Indication for Use:
The SIMEX Portable Suction / Aspirator / Vacuum system is indicated and used for vacuum suction, extraction, aspiration and removal of surgical fluids, tissue (including bone), bodily fluids or infectious materials from wounds or from patient's airway or respiratory system, either during surgery or at patient's bedside.
Prescription Use v (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------|---------|
| Division of General, Restorative, | |
| and Neurological Devices | |
| 510(k) Number | K061133 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.